MEASUREMENT TO COMPARE THE CORNEAL CENTRAL THICKNESS WITH PENTACAM, VISANTE AND ULTRASONIC PACHYMETRY AT OFTALMOSALUD EYE INSTITUTE Authors: Miranda Mauricio,

Slides:



Advertisements
Similar presentations
Authors have no financial interest
Advertisements

Comparison of Endothelial Cell Count by Manual and Automated Methods in Normal Corneas and in Fuchs' Endothelial Dystrophy Somasheila I Murthy, Debarun.
Corneal biomechanical properties and their association with intraocular pressure (IOP) and central corneal thickness (CCT) Dimitrios P. Bessinis 1, Pandelis.
Comparison of surgically induced astigmatism after phacoemulsification trough 3.2, 2.2 and 1.8 clear corneal incision. Luis Izquierdo Jr MD. PhD. Maria.
Accuracy of Predicted Refractive Error in Resident-Performed Cataract Surgery Using Partial Coherence Interferometry Nickolas P. Katsoulakis, M.D., Paul.
A Prospective Trial Evaluating Scleral Rebound Tonometry Sara Duke, MD, Andrew Logeman, Shuchi Patel, MD Loyola University Chicago, Department of Ophthalmology,
Swept Source Optical Coherence Tomography for Evaluation of Posterior Corneal Changes after Refractive Surgery Dr. Tommy Chung Yan Chan Dr. Vishal Jhanji.
Comparison between Dynamic contour tonometry, Goldmann applanation tonometry and Ehlers-corrected Goldmann applanation tonometry in eyes after Laser In.
Comparison of corneal astigmatism measured with 3 devices Mariko Shirayama, M.D, Li Wang, M.D, PhD, Mitchell P. Weikert, M.D, Douglas D. Koch, M.D. Cullen.
Eugene Ng, Arthur B. Cummings, Patrick P. Collins, Alexander V. Goncharov, Diana Bogusevschi, Chris Dainty, Michael C. Mrochen. The authors of this paper.
Corneal Densitometry and Visual Outcome Comparisons of DSAEK and DMEK Eyes Xiaolin Zhang MD Robert L. Schultze MD Robert A. Eden MD Rahul Raghu.
Anterior Chamber Depth, Iridocorneal Angle Width, and Intraocular Pressure Changes After Phacoemulsification: Narrow vs Open Iridocorneal Angles Huang.
David Zadok MD, Isaac Avni MD
Repeatability of Anterior and Posterior Corneal Higher- Order Aberrations in 4 mm, 6 mm and 8 mm diameters measured by Pentacam System Jadwiga Wojtowicz.
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. David Zadok, MD, Yakov Goldich, MD, Yaniv Barkana,
Dept. of Ophthalmology, Friedrich-Alexander University Erlangen-Nuremberg, Erlangen, Germany. Outcome after big-bubble deep anterior lamellar keratoplasty.
Comparison of Orbscan and Ultrasound Pachymetry Measurements Faik Orucov, MD, Abraham Solomon, MD,Ziv Caspi, David Landau, MD, Eyal Strassman, MD, and.
Effects of IOP Lowering Agents on Myopic Regression after Refractive Surgery Lim, Taehyung M.D., Hong, So Jin M.D., Cho, Beom Jin M.D., Ph.D. Chung Kyu-Hyung.
TRANSPALPEBRAL TONOMETER APPLICATION DURING INTRAOCULAR PRESSURE EVALUATION IN THE PATIENTS WITH REFRACTION ANOMALY BEFORE AND AFTER KERATOPHOTOREFRACTIVE.
Effects of Smoking on Corneal Thickness Candice Giordano, MD February 25, 2011.
Phacoemulsification in eyes with previous anterior chamber phakic IOL surgery Walton Nosé, MD, PhD 1,2 Adriana dos Santos Forseto, MD 1 Mariana Ávila,
Correlation Between Corneal Biomechanics and Wavefront Aberrations Alejandro Rivera, MD Private Practice Mexico City The author has no financial interest.
QUANTITATING THE IMPACT OF CLINICAL AND BASIC SCIENCE RESEARCH POSTER #2469 M. O. GORDON, C. SARLI, E. K. DUBINSKY, E. LONG, C. LA RUE, M. A. KASS, AND.
Universita’ di Catania, Italia Clinica Oculistica Dir.: Prof. A. Reibaldi Purpose: To evaluate the efficacy and safety of Pneumotrabeculoplasty (PNT) in.
Glaucoma Screening With a New Non-Contact Air Impulse Tonometer Pandelis A. Papadopoulos,1,2, Dimitrios P. Bessinis 3, Alexandros A. Papadopoulos 1, Konstantinos.
Relationship between Pachymetry and Ocular Response Analyzer Measurements Alejandro Rivera MD.
The authors have no financial interest in the subject matter of this poster Yinfei Xu, Peter S. Hersh, MD, David S. Chu, MD Institutional Affiliations.
Oliver Findl 1,2, Nino Hirnschall 1,2, Alja Crnej 2 1 VIROS Vienna Institute for Research in Ocular Surgery, Hanusch Hospital, Vienna, Austria 2 Moorfields.
Evaluation of refractive error measurements obtained by three different aberrometers Radha Ram, BA Li Wang, MD, PhD Mitchell P. Weikert, MD, MS Disclosure:
Effect of Corneal Thickness on Selective Laser Trabeculoplasty to Decrease Intraocular Pressure as Primary or Secondary Treatment of Glaucoma ASCRS 2010.
Endothelial Dystrophy Association with Hyperopia Glenn W. Thompson, M.D. Timothy P. Page, M.D. Authors have no financial interest.
Phoebe D. Lenhart, M.D. 1, Amy K. Hutchinson, M.D. 1, Michael J. Lynn, M.S. 2, Scott R. Lambert M.D. 1 1 Department of Ophthalmology, Emory University,
Khairidzan Mohd Kamal MD
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. Yakov Goldich, MD, David Zadok MD, Yaniv Barkana,
ELEVATION AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI ASCRS Boston April Vision Sana Research Group Vision Sana Research Group.
Endothelial Keratoplasty in Patients With an Anterior Chamber Intraocular Lens: A Montreal Experience Georges M. Durr, MD 1,2 Johanna Choremis, MD, FRCSC.
Clinical outcome of DALK in Keratoconus – A one year follow up
Comparison of Central Corneal Thickness and Peripheral Corneal Thickness using Sheimpflug system, Optical Coherence Tomography and Ultrasound Pachymetry.
Prospective, Randomized, Double-Blind Study Comparing Nepafenac 0.1% (Nevanac™) and Bromfenac Sodium 0.09% (Xibrom™) in Post-Op Cataract Patients Hon-Vu.
Partial Coherence Interferometry Failure Rate in a Teaching Hospital Leslie A. Wei 1,2, BA, Nickolaus P. Katsoulakis 2, MD, Theodoros Filippopoulos 3,
Comparison of the Diaton Transpalpebral Tonometer Versus TonoPen- Applination Theodore H. Curtis, M.D.1, Douglas L Mackenzie, M.D.1, Robert J. Noecker.
Naomi R. Goldberg, MD PhD Kenneth J. Wolf, MD Eric J. Wolf, MD FACS The authors have no financial interest in the subject matter of this poster. Comparison.
A Prospective Trial Comparing Pneumotonometry of Cornea and Sclera Usiwoma Abugo BS 1, Sara Duke MD 2, Shuchi Patel, MD 2 1 Loyola University Chicago,
Adriana S. Forseto1, MD Walton Nosé1,2, MD
1 P.O. dei Pellegrini; A.S.L. Napoli 1Centro 2 Dipartimento di Oftalmologia, Seconda Università degli Studi Napoli Scheimpflug topography, Scanning-Slit.
E-Poster Rio de Janeiro Corneal Tomography and Biomechanics Study Group The study of the correlations of the optic disc size with Cornal Geometry and Biomechanical.
Evaluation of Corneal Tomography in Primary Pterygium Sahil Goel, MD (Presenting Author), Murugesan Vanathi MD *The authors have no financial interests.
The effect of corneal edema on intraocular pressure measurements after phacoemulsification surgery Ilker Eser M.D. 1 Arzu Taskiran-Comez M.D. 1 Baris Komur.
Visual and IOP Outcomes after PRK in Pigment Dispersion Syndrome [Poster Number: P190] Kraig S. Bower, Denise A. Sediq, Charles D. Coe, Keith Wroblewski,
Study of Flap Thickness Precision with a Femtosecond Laser Shinagawa LASIK Center Tatsuya Yonekawa,MSc ; Minoru Tomita,MD.PhD ; Youhei Iida,CE ; Yuko Inada,CE.
Comparison of Central Corneal Thickness, Anterior Chamber Depth, and Central Corneal Power Measurements between Two Scheimpflug Imaging Systems Yuichi.
COMPARISON AND EVALUATION OF OCULAR BIOMETRY USING A NEW NON-CONTACT OPTICAL LOW COHERENCE REFLECTOMETER David Goldblum 1,3, Kaspar Rohrer 1, Rudolf Waelti.
Corneal Biomechanical Changes Following Surface Keratorefractive Surgery Teeravee Hongyok, MD, Christopher J. Rapuano, MD, Ajoy Virdi, MBBS, Elisabeth.
Evaluating Post Laser Refractive Surgery Corneal Thickness Using Artemis-2 Very High Frequency Arc-Scanning Ultrasonography Versus Pentacam * † Sharon.
CORNEAL ENDOTHELIAL PROFILE AFTER FERRARA RING IMPLANTATION Leonardo Torquetti, MD, PhD Paulo Ferrara, MD, PhD Paulo Ferrara Eye Clinic Belo Horizonte.
Instituto para Preservación de la Visión
Comparison of Endothelial Cell Loss After Phacoemulsification Performed by Third Year Residents and Anterior Segment Surgeons Alexandra Braunstein, MD.
Measurement of 60 kHz Femtosecond LASIK Flaps with Anterior Segment Ocular Coherence Tomography & Ultrasonic Subtraction Pachymetry Edward E. Manche, MD.
Corneal Pachymetry in Prediction of Refraction After Cataract Surgery
From: Key Developments for Partial Coherence Biometry and Optical Coherence Tomography in the Human Eye Made in Vienna Invest. Ophthalmol. Vis. Sci ;57(9):OCT460-OCT474.
Maria A Henriquez MD Luis Izquierdo MD Jackeline Parraga MD
Comparison of corneal powers obtained from four different devices
Associate Professor of Ophthalmology
Vicente J. Correa-Gomez MD, Guillermo Tapia MD, Oscar Macias Manuel MD, Alejandro Navas Perez MD, Gilberto Islas MD, Tito Ramírez-Luquin MD, Enrique O.
New 3-Dimensional Scheimpflug Analysis of Cataract Density
Bernilla E; Izquierdo Jr. L; Miranda M.
Khairidzan Mohd Kamal MD
“Accommodating IOL Implantation Experience”
The Effect of Corneal Thickness on Ocular Drug Penetration
International Vision Correction Research Centre
Presentation transcript:

MEASUREMENT TO COMPARE THE CORNEAL CENTRAL THICKNESS WITH PENTACAM, VISANTE AND ULTRASONIC PACHYMETRY AT OFTALMOSALUD EYE INSTITUTE Authors: Miranda Mauricio, MD; Gutarra Villavicencio Rosario del Pilar, MD; Izquierdo Jr. Luis, MD; Cañote Flores Renzo, MD; Bernilla F. César, MD. Institution: Instituto de Ojos Oftalmosalud Address: Av. Javier Prado Este 1142 San Isidro. *The authors have no financial interest in the materials being presented

To compare the values of measurement of corneal thickness obtained by Pentacam, Visante and Ultrasonic Pachymetry. PURPOSE

Prospective, comparative simple study of corneal pachymetry, in a sample of 54 eyes of 54 patients taken care of at OFTALMOSALUD Eye Institute during the period, between November and December of the year The values of corneal thickness were obtained by 3 devices (Visante®, Pentacam® and ultrasonic Pachymeter Accutome®). MATERIALS AND METHODS

Inclusion criteria considered: - Age greater than 18 years old - Patient that accepts to participate voluntarily in the study. - Do not have corneal pathology with the exception of keratocono, any ocular surgery or use of contact lenses. -Do not have family history of ocular disease or systemic disease with ocular manifestations Exclusion criteria considered: - Age lower than 18 years old. - Pregnant women - Previous refracting Surgery - Corneal Pathology except keratocono, ocular surgery or use of contact lenses - Family History of ocular disease or systemic disease with ocular manifestations. MATERIALS AND METHODS

All the measurements were made in the afternoon between 4 and 6 p.m., with the purpose of avoiding the normal diurnal variations of the corneal thickness. These studies gave a level of confidence of 95%, during this study 54 subjects showed a probability to have 80% to detect 45µm, of difference in averages. MATERIALS AND METHODS

RESULTS The age average of patients was of 28, DS The patient with smaller age was 18 years old and the one of greater age was 58 years old. The 85,2% (46 patients) were of female sex and 14.8% (8 patients) male.

RESULTS We obtained a medium correlation between the VISANTE and U.S. of 0,647 (64,7%) and between U.S. and PENTACAM of 0,690 (69%).

RESULTS

There were variations in the readings of corneal central thickness between the 3 devices. The measures of central corneal thickness in normal eyes with Pentacam revealed an excellent correlation with the results obtained with ultrasonic pachymetry. The measures of corneal thickness with the Visante System turned out to be smaller than with the Pentacam and ultrasonic pachymetry. The results suggest that the Pentacam and the U.S. is equally precise, giving similar readings. CONCLUSIONS

REFERENCES 1.Miglior S, Albe E, Guareschi M, Mandelli G, Gomarasca S, Orzalesi N (2004) Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness. Br J Ophthalmol; 88: Rainer G, Petternel V, Findl O, Schmetterer L, Skorpik C, Luksch A, Drexler W (2002) Comparison of ultrasound pachymetry and partial coherence interferometry in the measurement of central corneal thickness. J Cataract Refract Surg; 28: Gordon A, Boggess EA, Molinari JF (1990) Variability of ultrasonic pachometry. Optom Vis Sci; 67: Marsich MW, Bullimore MA (2000) The repeatability of corneal thickness measures. Cornea; 19: Barkana Y, Gerber Y, Elbaz U, Schwartz S, Ken-Dror G, Avni I, Zadok D (2006) Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg; 31: Hitzenberger CK, Baumgartner A, Drexler W, Fercher AF (1994) Interferometric measurement of corneal thickness with micrometer precision. Am J Ophthalmol; 118: Drexler W, Baumgartner A, Findl O, Hitzenberger CK, Sattmann H, Fercher AF (1997) Submicrometer precision biometry of the anterior segment of the human eye. Invest Ophthalmol Vis Sci; 38: Bohnke M, Chavanne P, Gianotti R, Salathe RP (1998) Continuous non-contact corneal pachymetry with a high speed reflectometer. J Refract Surg; 14: Wong AC, Wong CC, Yuen NS, Hui SP (2002) Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry. Eye; 16: Muscat S, McKay N, Parks S, Kemp E, Keating D (2002) Repeatability and reproducibility of corneal thickness measurements by optical coherence tomography. Invest Ophthalmol Vis Sci; 43: Bechmann M, Thiel MJ, Neubauer AS, Ullrich S, Ludwig K, Kenyon KR, Ulbig MW (2001) Central corneal thickness measurement with a retinal optical coherence tomography device versus standard ultrasonic pachymetry. Cornea; 20: Modis L, Jr., Langenbucher A, Seitz B (2001) Scanning-slit and specular microscopic pachymetry in comparison with ultrasonic determination of corneal thickness. Cornea; 20: Chakrabarti HS, Craig JP, Brahma A, Malik TY, McGhee CN (2001) Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes. J Cataract Refract Surg; 27: Suzuki S, Oshika T, Oki K, Sakabe I, Iwase A, Amano S, Araie M (2003) Corneal thickness measurements: scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetry. J Cataract Refract Surg; 29: Rabsilber TM, Khoramnia R, Auffarth GU (2006) Anterior chamber measurements using Pentacam rotating Scheimpflug camera. J Cataract Refract Surg; 32: Price FW, Jr., Koller DL, Price MO (1999) Central corneal pachymetry in patients undergoing laser in situ keratomileusis. Ophthalmology; 106: Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol; 44: Stodtmeister R (1998) Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand; 76: Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK, 2nd, Wilson MR, Kass MA (2002) The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol; 120: ; discussion