Atomic Absorption & Atomic Emission Spectroscopy

Slides:



Advertisements
Similar presentations
Lecture 6 ATOMIC SPECTROSCOPY
Advertisements

Atomic Absorption & Emission Spectroscopy
1.1 Atomic Absorption Spectrometry (AAS) determination of elements not compounds needs radiation source high temperature for atomization Atomization a.
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Atomic spectroscopy  It’s a class of spectroscopic method in which the species examined in the spectrometer are in the form of atoms.
Spectroscopy Lecture 4 Ahmad Razali Bin Ishak Department of Environmental Health Faculty of Health Sciences UiTM Puncak Alam.
1 Atomic Absorption Spectroscopy Atomic Emission Spectroscopy Lecture 18.
Atomic Emission Spectroscopy
Chapter 10 ATOMIC EMISSION SPECTROMETRY
AAS and FES (Ch 10, 7th e, WMDS)
Atomic Emission Spectroscopy
Atomic Absorption Spectroscopy Prof Dr Hisham E Abdellatef 2011.
ATOMIC SPECROSCOPY (AS)
AAS and FES (Ch 9, 7th e, WMDS)
AA and Atomic Fluorescence Spectroscopy Chapter 9
Atomic Spectroscopy Atomic Spectroscopic Methods Covered in Ch 313: Optical Atomic Spectrometry (Ch 8-10) Atomic X-ray Spectrometry (Ch 12) Atomic Mass.
INTRODUCTION TO OPTICAL METHODS
Atomic Absorption Spectroscopy AAS Comparatively easy to use Low maintenance Low consumables Good for measuring one element at a time. Comparatively easy.
Atomic Absorption Spectrometry Dr AKM Shafiqul Islam University Malaysia Perlis.
HL Chemistry - Option A: Modern Analytical Chemistry ATOMIC ABSORPTION SPECTROSCOPY.
1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission.
Flame photometry.
Molecular Fluorescence Spectroscopy
Atomic Emission Spectroscopy
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Atomic Emission - AES M* → M + hn Thermal excitation M → M*
Fluorescence. Topics Sensitivity (nM-pM) –Contrast UV-vis measurements Derivatization Laser-Induced Fluorescence.
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy Elemental analysis Dissolved metals Source –Hallow Cathode Lamp –Cathode of specific element Sample –Acetylene-air flame,
ATOMIC ABSORPTION SPECTROSCOPY
BC ILN Atomic Absorption Spectroscopy (AAS) 1 Thompson Rivers University.
Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh
Chapter 8 An Introduction to Optical Atomic Spectroscopy
Biochemical instrumental analysis-2
Atomic Emission Spectroscopy
Atomic Emission Spectrometry
Atomic Absorption Spectrophotometer
Chapter 10 Atomic Emission Spectrometry
1.1 Inductively coupled plasma (ICP) Three Argon flow 1.Plasma gas (10-20 L/min) 2.Nebulizer gas (~1L/min) 3.Optional auxiliary gas (~0.5L/min) Radio-frequency.
ATOMIC SPECROSCOPY (AS)
Atomic Absorption Spectroscopy
1 Atomic Emission Spectroscopy Lecture Advantages of Plasma Sources 1.No oxide formation as a result of two factors including Very high temperature.
1 Atomic Absorption Spectroscopy Lecture Performance Characteristics of Electrothermal Atomizers Electrothermal atomization is the technique of.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 20
Atomic spectroscopy Elemental composition Atoms have a number of excited energy levels accessible by visible-UV optical methods ä Must have atoms (break.
1 Introduction to Atomic Spectroscopy Atomic Absorption Spectroscopy Lecture 12.
Adv. Inst. Techs.  flame emission (eg flame photometer) known as low temperature emission ( K) ◦ first form of spectroscopy ◦ used in commercial.
Atomic Fluorescence Spectroscopy. Background l First significant research by Wineforder and Vickers in 1964 as an analytical technique l Used for element.
Atomic-absorption spectroscopy
ATOMIC ABSORPTION SPECTROSCOPY (AAS) Atomization: It is the conversion of molecules to their component atoms in gaseous state using a source of heat (flame).
1 Atomic Emission Spectroscopy Molecular Absorption Spectroscopy Lecture 21.
Plasma A Plasma consists of a collection of free-moving electrons and ions and is very hot. Energy must be continually applied to sustain the plasma.
Chapter 10 Atomic Emission Spectrometry
A TOMIC - ABSORPTION SPECTROSCOPY Lab no. 3 Done by : Iman Al Ajeyan.
Chemistry 4631 Instrumental Analysis Lecture 18 Chem 4631.
ATOMIC ABSORPTION SPECTROSCOPY SUPERVISOR: PRESENTED BY: Dr. SANGEETA OBROI SURITI SHARMA M.Sc. Chem.-3 rd Sem. M.Sc. Chem.-3 rd Sem
Chem. 133 – 4/4 Lecture.
Satish Pradhan Dnyanasadhana College Department Of Chemistry
Flame Emission Spectrometry
Atomic Absorption and Atomic Fluorescence Spectrometry
The ratio of excited state to ground state atoms as a function of temperature is determined by the Maxwell-Boltzmann expression. Fig Excited state.
Elemental composition
Dnyanasadhana College, Thane. Department of Chemistry T. Y. B. Sc
Chapter 8 An Introduction to Optical Atomic Spectroscopy
Пламена Спектроскопија
Satish Pradhan Dnyanasadhana College Department Of Chemistry
Atomic Absorption Spectroscopy. Atomic absorption spectroscopy is based on the same principle as the flame test used in qualitative analysis.
SPECTROPHOTOMETRY Applied Chemistry.
Atomic Absorption & Atomic Emission Spectroscopy
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Presentation transcript:

Atomic Absorption & Atomic Emission Spectroscopy Dr.Syed Muzzammil Masaud Mphill.Pharmaceutical Chemistry

BASIC PRINCIPLE ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light by these elements in order to measure their concentration .

- Atomic-absorption spectroscopy quantifies the absorption of ground state atoms in the gaseous state . The atoms absorb ultraviolet or visible light and make transitions to higher electronic energy levels . The analyte concentration is determined from the amount of absorption.

Concentration measurements are usually determined from a working curve after calibrating the instrument with standards of known concentration. - Atomic absorption is a very common technique for detecting metals and metalloids in environmental samples.

Elements detectable by atomic absorption are highlighted in pink in this periodic table                                                                                              

The Atomic Absorption Spectrometer Atomic absorption spectrometers have 4 principal components 1 - A light source ( usually a hollow cathode lamp ) 2 – An atom cell ( atomizer ) 3 - A monochromator 4 - A detector , and read out device .

Schematic Diagram of an Atomic Absorption Spectrometer Detector and readout device Light source (hollow cathode Lamp ) atomizer monochromator

Atomic Absorption Spectrophotometer

1 – Light Source The light source is usually a hollow cathode lamp of the element that is being measured . It contains a tungsten anode and a hollow cylindrical cathode made of the element to be determined. These are sealed in a glass tube filled with an inert gas (neon or argon ) . Each element has its own unique lamp which must be used for that analysis .

Hollow Cathode Lamp cathode Anode Quartz window Pyrex body Anode

How it works Applying a potential difference between the anode and the cathode leads to the ionization of some gas atoms . These gaseous ions bombard the cathode and eject metal atoms from the cathode in a process called sputtering. Some sputtered atoms are in excited states and emit radiation characteristic of the metal as they fall back to the ground state .

Scheme of a hollow cathode lamp

The shape of the cathode which is hollow cylindrical concentrates the emitted radiation into a beam which passes through a quartz window all the way to the vaporized sample. Since atoms of different elements absorb characteristic wavelengths of light. Analyzing a sample to see if it contains a particular element means using light from that element .

For example with lead, a lamp containing lead emits light from excited lead atoms that produce the right mix of wavelengths to be absorbed by any lead atoms from the sample . A beam of the electromagnetic radiation emitted from excited lead atoms is passed through the vaporized sample. Some of the radiation is absorbed by the lead atoms in the sample. The greater the number of atoms there is in the vapor , the more radiation is absorbed .

2 – Atomizer Elements to be analyzed needs to be in atomic sate Atomization is separation of particles into individual molecules and breaking molecules into atoms .This is done by exposing the analyte to high temperatures in a flame or graphite furnace .

The role of the atom cell is to primarily dissolvate a liquid sample and then the solid particles are vaporized into their free gaseous ground state form . In this form atoms will be available to absorb radiation emitted from the light source and thus generate a measurable signal proportional to concentration . There are two types of atomization : Flame and Graphite furnace atomization .

Flame Flame AA can only analyze solutions , where it uses a slot type burner to increase the path length, and therefore to increase the total absorbance . Sample solutions are usually introduced into a nebuliser by being sucked up a capillary tube .In the nebuliser the sample is dispersed into tiny droplets , which can be readily broken down in the flame.

FLAME ATOMIZERS Used in all Atomic Spectroscopic techniques Converts analyte into free atoms in the form of vapor phase free atoms Heat is required Routes for sample introduction

Various flame atomization techniques

Types of Flames Used in Atomic Spectroscopy

Processes that take place in flame

Effect of flame temperature on excited state population # atoms in Excited state Boltzmann constant Temperature # atoms in Ground state Energy difference Statistical factor

For Zn: N*/No = 10-15%

Thus 99.998% of Na atoms are in the ground state Atomic emission uses Excited atoms Atomic absorption uses Ground state atoms

Effect of flame temperature on excited state population

ATOMIZATION DEVICES ATOMIZATION : A process of forming free atoms by heat Atomizers are devices that carry out atomization: Continuous Non-continuous Continuous: (Constant temperature with time) Flame Plasma Non-Continuous: (temperature varies with time) Electrothermal Spark discharge

SAMPLE INTRODUCTION SYSTEMS In continuous atomizers sample is constantly introduced in form of droplets, dry aerosol, vapor Nebulizer : A device for converting the solution into fine spray or droplets Continuous sample introduction is used with continuous nebulizers in which a steady state atomic population is produced. Sample is introduced in fixed or discrete amounts. Discontinuous samplers are used with continuous atomizers

1- Discrete samples are introduced into atomizers in many ways: Electrothermal atomizers a syringe is used a transient signal is produced as temperature changes with time and sample is consumed 2- Indirect insertion (Probe) sample is introduced into a probe (carbon rod) and mechanically moved into the atomization region vapor cloud is transient because sample introduced is limited

3- Flow Injection 4- Hydride Generation The analyte is introduced into the carrier stream into a nebulizer as mist 4- Hydride Generation the volatile sample is stripped from the analyte solution and carried out by a gas into the atomizer. This strip is followed by chemically converting the analyte to hydride vapor form.

5- With Arc Spark Solids are employed 6- Laser Microbe Technique A beam of laser is directed onto a small solid sample, gets vaporized, atomized by relative heating. Either sample is probed by encoding system or vapor produced is swept into a second absorption or fluorescence

Nebulization gas is always compressed, usually acts as the oxidant; it is oxygen (O2) in flame and argon (Ar) in plasma Nebulization chambers produce smaller droplets and remove or drain larger droplets called aerosol modifiers Aspiration rate is proportional to compressed gas pressure. The pressure drops through capillary, here 1/4 capillary diameters are recommended. This is inversely proportional to viscocity of the solution Peristaltic and/or syringe pumps could be used

Oxidant and fuel are usually brought into the nebulization chamber through a separate port. They mix and pass the burner head called premixed burner system. Add organic solvents to reduce the size of the drop

The Atomic Absorption Spectrometer Sample Introduction System Nebuliser Capillary Solution

The fine mist of droplets is mixed with fuel ( acetylene ) , and oxidant ( nitrous oxide) and burned. The flame temperature is important because it influences the distribution of atoms. It can be manipulated by oxidant and fuel ratio.

Graphite Furnace The graphite furnace has several advantages over a flame. First it accept solutions, slurries, or solid samples. Second it is a much more efficient atomizer than a flame and it can directly accept very small absolute quantities of sample. It also provides a reducing environment for easily oxidized elements. Samples are placed directly in the graphite furnace and the furnace is electrically heated in several steps to dry the sample, ash organic matter, and vaporize the analyte atoms. It accommodates smaller samples but it’s a difficult operation, because the high energy that is provided to atomize the sample particles into ground state atoms might excite the atomized particles into a higher energy level and thus lowering the precision .

3- Monochromators This is a very important part in an AA spectrometer. It is used to separate out all of the thousands of lines. Without a good monochromator, detection limits are severely compromised. A monochromator is used to select the specific wavelength of light which is absorbed by the sample, and to exclude other wavelengths. The selection of the specific light allows the determination of the selected element in the presence of others.

4 - Detector and Read out Device The light selected by the monochromator is directed onto a detector that is typically a photomultiplier tube , whose function is to convert the light signal into an electrical signal proportional to the light intensity. The processing of electrical signal is fulfilled by a signal amplifier . The signal could be displayed for readout , or further fed into a data station for printout by the requested format.

Calibration Curve A calibration curve is used to determine the unknown concentration of an element in a solution. The instrument is calibrated using several solutions of known concentrations. The absorbance of each known solution is measured and then a calibration curve of concentration vs absorbance is plotted. The sample solution is fed into the instrument, and the absorbance of the element in this solution is measured .The unknown concentration of the element is then calculated from the calibration curve

Calibration Curve A 1.0 - b 0.9 - S 0.8 - . o 0.7 - . r 0.6 - . 10 20 30 40 50 60 70 80 90 100 Concentration ( g/ml )

Determining concentration from Calibration Curve A 1.0 - absorbance measured b 0.9 - S 0.8 - . o 0.7 - . r 0.6 - . b 0.5 - . . a 0.4 - . n 0.3 - . concentration calculated c 0.2 - e 0.1 - 10 20 30 40 50 60 70 80 90 100 Concentration ( mg/l )

Interferences The concentration of the analyte element is considered to be proportional to the ground state atom population in the flame ,any factor that affects the ground state atom population can be classified as an interference . Factors that may affect the ability of the instrument to read this parameter can also be classified as an interference .

The different interferences that are encountered in atomic absorption spectroscopy are : - Absorption of Source Radiation : Element other than the one of interest may absorb the wavelength being used. - Ionization Interference : the formation of ions rather than atoms causes lower absorption of radiation .This problem is overcome by adding ionization suppressors. - Self Absorption : the atoms of the same kind that are absorbing radiation will absorb more at the center of the line than at the wings ,and thus resulting in the change of shape of the line as well as its intensity .

- Back ground Absorption of Source Radiation : This is caused by the presence of a particle from incomplete atomization .This problem is overcome by increasing the flame temperature . - Transport Interference : Rate of aspiration, nebulization, or transport of the sample ( e g viscosity, surface tension, vapor pressure , and density ) .

2 Atomic Emission Spectroscopy Atomic emission spectroscopy is also an analytical technique that is used to measure the concentrations of elements in samples . It uses quantitative measurement of the emission from excited atoms to determine analyte concentration .

The analyte atoms are promoted to a higher energy level by the sufficient energy that is provided by the high temperature of the atomization sources . The excited atoms decay back to lower levels by emitting light . Emissions are passed through monochromators or filters prior to detection by photomultiplier tubes.

The instrumentation of atomic emission spectroscopy is the same as that of atomic absorption ,but without the presence of a radiation source . In atomic Emission the sample is atomized and the analyte atoms are excited to higher energy levels all in the atomizer .

Schematic Diagram of an Atomic Emission spectrometer

Introduction to AES Atomization Emission Sources Advantages of plasma Flame – still used for metal atoms Electric Spark and Arc Direct current Plasmas Microwave Induced Plasma Inductively Coupled Plasma – the most important technique Advantages of plasma Simultaneous multi-element Analysis – saves sample amount Some non-metal determination (Cl, Br, I, and S) Concentration range of several decades (105 – 106) Disadvantages of plasma very complex Spectra - hundreds to thousands of lines High resolution and expensive optical components Expensive instruments, highly trained personnel required

10A Plasam Source AES Plasma Three main types an electrically conducting gaseous mixture containing significant concentrations of cations and electrons. Three main types Inductively Coupled Plasma (ICP) Direct Current Plasma (DCP) Microwave Induced Plasma (MIP)

ICP Inductively Coupled Plasma (ICP) Plasma generated in a device called a Torch Torch up to 1" diameter Ar cools outer tube, defines plasma shape Rapid tangential flow of argon cools outer quartz and centers plasma Rate of Argon Consumption 5 - 20 L/Min Radio frequency (RF) generator 27 or 41 MHz up to 2 kW Telsa coil produces initiation spark Ions and e- interact with magnetic field and begin to flow in a circular motion. Resistance to movement (collisions of e- and cations with ambient gas) leads to ohmic heating. Sample introduction is analogous to atomic absorption.

Sample introduction Nebulizer Electrothermal vaporizer Table 8-2 methods of sample introducton

Nebulizer convert solution to fine spray or aerosol Ultrasonic nebulizer uses ultrasound waves to "boil" solution flowing across disc Pneumatic nebulizer uses high pressure gas to entrain solution

Electro-thermal vaporizer ETV electric current rapidly heats crucible containing sample sample carried to atomizer by gas (Ar, He) only for introduction, not atomization

Plasma structure Brilliant white core Flame-like tail Ar continuum and lines Flame-like tail up to 2 cm Transparent region where measurements are made (no continuum)

Plasma characteristics Hotter than flame (10,000 K) - more complete atomization/ excitation Atomized in "inert" atmosphere Ionization interference small due to high density of e- Sample atoms reside in plasma for ~2 msec and Plasma chemically inert, little oxide formation Temperature profile quite stable and uniform.

DC plasma First reported in 1920s DC current (10-15 A) flows between C anodes and W cathode Plasma core at 10,000 K, viewing region at ~5,000 K Simpler, less Ar than ICP - less expensive Less sensitive than ICP Should replace the carbon anodes in several hours

Atomic Emission Spectrometer May be >1,000 visible lines (<1 Å) on continuum Need higher resolution (<0.1 Å) higher throughput low stray light wide dynamic range (>1,000,000) precise and accurate wavelength calibration/intensities stability computer controlled Three instrument types: sequential (scanning and slew-scanning) Multichannel - Measure intensities of a large number of elements (50-60) simultaneously Fourier transform FT-AES

Desirable properties of an AE spectrometer

Sequential vs. multichannel Sequential instrument PMT moved behind aperture plate, or grating + prism moved to focus new l on exit slit Pre-configured exit slits to detect up to 20 lines, slew scan characteristics Cheaper Slower Multichannel instrument Polychromators (not monochromator) - multiple PMT's Array-based system charge-injection device/charge coupled device Expensive ( > $80,000) Faster

Sequential vs. multichannel

Sequential monochromator Slew-scan spectrometers even with many lines, much spectrum contains no information rapidly scanned (slewed) across blank regions (between atomic emission lines) From 165 nm to 800 nm in 20 msec slowly scanned across lines 0.01 to 0.001 nm increment computer control/pre-selected lines to scan

Slew scan spectrometer Two slew-scan gratings Two PMTs for VIS and UV Most use holographic grating

Scanning echelle spectrometer PMT is moved to monitor signal from slotted aperture. About 300 photo-etched slits 1 second for moving one slit Can be used as multi channel spectrometer Mostly with DC plasma source

AES instrument types Three instrument types: sequential (scanning and slew-scanning) Multichannel - Measure intensities of a large number of elements (50-60) simultaneously Fourier transform FT-AES

Multichannel polychromator AES Rowland circle Quantitative det. 20 more elements Within 5 minutes In 10 minutes

Applications of AES AES relatively insensitive small excited state population at moderate temperature AAS still used more than AES less expensive/less complex instrumentation lower operating costs greater precision In practice ~60 elements detectable 10 ppb range most metals Li, K, Rb, Cs strongest lines in IR Large # of lines, increase chance of overlap

Detection power of ICP-AES

ICP/OES INTERFERENCES Spectral interferences: caused by background emission from continuous or recombination phenomena, stray light from the line emission of high concentration elements, overlap of a spectral line from another element, or unresolved overlap of molecular band spectra. Corrections Background emission and stray light compensated for by subtracting background emission determined by measurements adjacent to the analyte wavelength peak. Correction factors can be applied if interference is well characterized Inter-element corrections will vary for the same emission line among instruments because of differences in resolution, as determined by the grating, the entrance and exit slit widths, and by the order of dispersion.

Physical interferences of ICP cause effects associated with the sample nebulization and transport processes. Changes in viscosity and surface tension can cause significant inaccuracies, especially in samples containing high dissolved solids or high acid concentrations. Salt buildup at the tip of the nebulizer, affecting aerosol flow rate and nebulization. Reduction by diluting the sample or by using a peristaltic pump, by using an internal standard or by using a high solids nebulizer.

Interferences of ICP Chemical interferences: include molecular compound formation, ionization effects, and solute vaporization effects. Normally, these effects are not significant with the ICP technique. Chemical interferences are highly dependent on matrix type and the specific analyte element.

Memory interferences: When analytes in a previous sample contribute to the signals measured in a new sample. Memory effects can result from sample deposition on the uptake tubing to the nebulizer from the build up of sample material in the plasma torch and spray chamber. The site where these effects occur is dependent on the element and can be minimized by flushing the system with a rinse blank between samples. High salt concentrations can cause analyte signal suppressions and confuse interference tests.

Typical Calibration ICP curves

Calibration curves of ICP-AES

10B. Arc and Spark AES Arc and Spark Excitation Sources: Limited to semi-quantitative/qualitative analysis (arc flicker) Usually performed on solids Largely displaced by plasma-AES Electric current flowing between two C electrodes

Carbon electrodes Sample pressed into electrode or mixed with Cu powder and pressed - Briquetting (pelleting) Cyanogen bands (CN) 350-420 nm occur with C electrodes in air -He, Ar atmosphere Arc/spark unstable each line measured >20 s needs multichannel detection

Arc and Spark spectrograph

spectrograph Beginning 1930s photographic film Cheap Long integration times Difficult to develop/analyze Non-linearity of line "darkness“ Gamma function Plate calibration

Multichannel photoelectric spectrometer multichannel PMT instruments for rapid determinations (<20 lines) but not versatile For routine analysis of solids metals, alloys, ores, rocks, soils portable instruments Multichannel charge transfer devices Recently on the market Orignally developed for plasma sources

Comparison Between Atomic Absorption and Emission Spectroscopy - Measure trace metal concentrations in complex matrices . - Atomic absorption depends upon the number of ground state atoms . Emission - Measure trace metal concentrations in complex matrices . - Atomic emission depends upon the number of excited atoms .

- It measures the radiation absorbed by the ground state atoms. - Presence of a light source ( HCL ) . - The temperature in the atomizer is adjusted to atomize the analyte atoms in the ground state only. - It measures the radiation emitted by the excited atoms . - Absence of the light source . - The temperature in the atomizer is big enough to atomize the analyte atoms and excite them to a higher energy level.