Nuclear Magnetic Resonance (NMR) Magnetic Resonance Imaging (MRI)

Slides:



Advertisements
Similar presentations
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Advertisements

The analytical measurement principle
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Ari Borthakur, PhD Associate Director, Center for Magnetic Resonance & Optical Imaging Department of Radiology Perelman school of Medicine, University.
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Imaging Sequences part I
MRI Phillip W Patton, Ph.D..
BE 581 Lecture 3- Intro to MRI.
PHYSICS OF MAGNETIC RESONANCE
Nuclei With Spin Align in Magnetic Fields HoHo anti-parallel parallel Alignment Energy  E = h  H o Efficiency factor- nucleus ConstantsStrength of magnet.
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Parameters and Trade-offs
Magnetic Resonance Imaging
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
Principles of MRI: Image Formation
Chapter 9 Basic MRI I Mark D. Herbst, MD, PhD. Notice This lecture contained many drawings on the whiteboard, so get these from one of the other students.
Magnetic Resonance Imaging
Principles of MRI. Some terms: –Nuclear Magnetic Resonance (NMR) quantum property of protons energy absorbed when precession frequency matches radio frequency.
Lecture 2 1 H Nuclear Magnetic Resonance. Gas Chromatograph of Molecular Hydrogen at –100 °C Thermoconductivity Detector 12.
NMR SPECTROSCOPY.
ELEG 479 Lecture #9 Magnetic Resonance (MR) Imaging
Nuclear Magnetic Resonance Spectrometry Chap 19
Nuclear Magnetic Resonance Spectrometry Chap 19. Absorption in CW Experiments Energy of precessing particle E = -μ z B o = -μ B o cos θ When an RF photon.
Basics of Magnetic Resonance Imaging
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
Magnetic Resonance MSN 506 Notes. Overview Essential magnetic resonance Measurement of magnetic resonance Spectroscopic information obtained by magnetic.
MRI: an Introduction By Mohammad Ali Ahmadi Pajouh Amirkabir University of Technology Biomedical Eng. Dep.
Medical Imaging Systems: MRI Image Formation
Magnetic Resonance Imaging 4
Physics of Magnetic Resonance Chapter 12
대구가톨릭대학병원 영상의학과 이 영 환 M M R R Basic Physics. MR Signal T1-, T2-weighted TR, TE MR Signal T1-, T2-weighted TR, TE.
Principles of MRI Physics and Engineering
Principles of Magnetic Resonance
Magnetic Resonance Imaging
Prof. Dr. Moustafa. M. Mohamed Dr. Eng. Safa Ahmed El-Askary
Medical Imaging Systems: MRI Image Formation
Basic Physical Principles of MRI
Basic Concept of MRI Chun Yuan. Magnetic Moment Magnetic dipole and magnetic moment Nuclei with an odd number of protons or neutrons have a net magnetic.
Basic of Magnetic Resonance Imaging Seong-Gi Kim Paul C. Lauterbur Chair in Imaging Research Professor of Radiology, Neurobiology and Bioengineering University.
fMRI Methods Lecture2 – MRI Physics
NMR in Medicine and Biology
NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY Basics of …….. NMR phenomenonNMR phenomenon Chemical shiftChemical shift Spin-spin splittingSpin-spin splitting.
Variable Rate Selective Excitation Radio Frequency Pulse in Magnetic Resonance Imaging Stephen Stoyan.
Nuclear Magnetic Resonance I Magnetization properties Generation and detection of signals.
Allen W. Song, PhD Brain Imaging and Analysis Center Duke University MRI: Image Formation.
Magnetic Resonance Imaging – Basic Principles –
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
MRI. Vector Review x y z Vector Review (2) The Dot Product The Cross Product (a scalar) (a vector) (a scalar)
BME1450 Intro to MRI February 2002 The Basics The Details – Physics The Details – Imaging.
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
Anna Beaumont FRCR Part I Physics
V.G.Wimalasena Principal School of Radiography
Introducing cross- sectional imaging Dr.Ahmed S. Tawfeek FIBMS(Rad.)
MRI Physics: Spatial Encoding Anna Beaumont FRCR Part I Physics.
RT 4912 Review (C) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
Magnetic Resonance Imaging Glenn Pierce, King’s College London, Department of Physics Introduction Edward Purcell and Felix Bloch were both awarded the.
Spinning Nucleus Produces Magnetic Moment
Magnetic Resonance Learning Objectives
Charged particle. Moving charge = current Associated magnetic field - B.
Lecture 1: Magnetic Resonance
Lecture 3: The Bloch Equations Solved & the MR Signal Equation Solve the Bloch equation –Precession –T2 relaxation –T1 relaxation MR signal equation –Understand.
 This depends on a property of nuclei called spin.  Gyroscope: Principle: As long as its disc remains spinning rapidly the direction of the spin axis.
NUCLEAR MAGNETIC RESONANCE
بسم الله الرحمن الرحيم Dr. Maged Ali Hegazy Assistant Professor Alazhar University.
MRI Physics in a Nutshell Christian Schwarzbauer
Basic MRI I Chapter 3 Notes.
MRI: From Protons to Pixels
How MRI Works By Wesley Eastridge, adapted from and with illustrations from The Basics of MRI by Joseph P. Hornak, Ph.D.
(4)ELECTRONIC SUPPORT SYSTEM
T2 Relaxation Time T2 relaxation time is defined as the time needed to dephase up to 37% of the original value. T2 relaxation refers to the progressive.
Presentation transcript:

Nuclear Magnetic Resonance (NMR) Magnetic Resonance Imaging (MRI)

What is the source and how is it working? How is contrast created? How is direction created? How is the image constructed?

The source

Nuclear Magnetism Nucleons (protons, neutrons) have a quantum property known as spin (1/2 Bohr magneton, spin=1/2)

Why are protons important? They are plenty in the body Positively charged They spin about a central axis A spinning charge creates a magnetic field The arrow indicates the direction of the magnetic field

Nuclear Magnetism Normally the nuclear spins of the protons cancel out and there is no net magnetic field to observe.

Nuclear Magnetism In a homogeneous magnetic field, the proton spin will orient themselves. Most spin will be oriented along the external field giving a net magnetic field.

Mo y x z Bo

Energy and populations An external magnetic field create an energy difference between nuclei aligned and against Bo: Each level has a different population (N), and the difference between the two is related to the energy difference by the Boltzman distribution: Na / Nb = e DE / kT The DE for 1H at 400 MHz (Bo = 9.5 T) is 3.8 x 10-5 Kcal /mol b Bo > 0 DE = h n a Bo = 0 Na / Nb = 1.000064

Mo z x i B1 = C * cos (wot) B1 Transmitter coil (y) y + = +wo -wo Bo

Once again Mo is composed by individual atomic spin with the Larmour frequency wo. If a RF field of the same frequency is applied a magnetic vector B1 with the same frequency wo is created that interact with Mo under a resonant condition. The alternating magnetic field generates a torque on Mo, and the system absorbs energy : Since the system absorbed energy, the equilibrium of the system is altered. We modified the populations of the Na and Nb energy levels. Again, keep in mind that individual spins flipped up or down (a single quanta), but Mo can have a continuous variation. B1 off… (or off-resonance) Mo z x B1 Mxy y wo

ENERGY ABSORPTION MRASUREMENT OF SIGNAL RELAXATION 2 2 2

Energy Absorption no change M0=x B RF in 0 equilibrium M0=x 16 18

Energy Absorption 0 B M0=y M0=x RF in energy absorption equilibrium M0=x 17 19

Energy Absorption tip B M0=x M0=y ““ RF pulse 0 18 20

Energy Absorption 900 tip B M0=x M0=z 900 RF pulse 0 19 21

Energy Absorption 1800 tip B M0=x M0=-x 1800 RF pulse 0 20 22

What happens to the released energy? released as heat exchanged and absorbed by other protons released as radio waves 5 5

Measuring the MR Signal B z y x cannot easily record ML the signal is measured in the transverse (x, y) plane 34

Rotation in the xy-plane z y x z y x z y x 900 RF t=t0 t=t0+ t=t0++ 35 motion in the xy-plane =   >> T1 relaxation, MHz vs seconds

Mxy and ML the magnitude of Mxy depends on the size of ML immediately prior to the 900 RF pulse z y x z y x fat protons - short T1 water protons - long T1 900 RF

Measuring the MR Signal z y x RF signal from precessing protons 43 RF antenna

Relaxation relaxation is the release of energy from an excited state to a lower energy state protons absorb and release energy in MRI 4 4

CONTRAST

Basic Principle The signal in MR images vary due to proton density T1 relaxation T2 relaxation flow contrast 3 3

T1 Relaxation synonyms longitudinal relaxation thermal relaxation spin-lattice relaxation

T2 Relaxation immediately after the absorption of the RF energy and the initiation of precession, the process of T2 relaxation begins results from loss of phase coherence among groups of protons rotating in the transverse plane

Spin Dephasing after excitation y x y Bo x y x y Spin Dephasing after excitation

180o 180o 180o TE Time Echo returned RF Signal

T2 and T2* Relaxation T2* is the observed T2 relaxation time T2 is the natural relaxation time T2* < T2

FID=Free Induction Decay TE=Time Echo TR=Time Repetition

26

SPATIAL ENCODING

Spatial encoding is accomplished by superimposing gradient fields. There are three gradient fields in the x, y, and z directions. Gradients alter the magnetic field resulting in a change in resonance frequency or a change in phase. 29

MRI The main magnetic field is oriented in the z direction. Gradient fields are located in the x, y, and z directions. 30

29

MRI The three magnetic gradients work together to encode the NMR signal with spatial information. Remember: the resonance frequency depends on the magnetic field strength. Small alterations in the magnetic field by the gradient coils will change the resonance frequency. 31

Gradients The gradients have different names Z gradient: slice select X gradient: frequency encode (readout) Y gradient: phase encode 32

Slice Selection For axial imaging, slice selection occurs along the long axis of the magnet. Superposition of the slice selection gradient causes non-resonance of tissues that are located above and below the plane of interest. 33

Slice Selection  y z z gradient x    2 2 2 imaging plane

Slice Selection     t1 t2 t3 slice thickness is determined by gradient strength     RF bandwidth     t1 t2 t3

Slice Selection      Selection of an axial slice is accomplished by the z gradient. z gradient direction    z-axis graph of the z magnetic gradient 

Slice Selection     slice location is determined by the null point of the z gradient slice 1 slice 2 slice 3   RF bandwidth      

The shape of the RF-pulse relates to the thickness and shape of the slice.

In the gradient field protons will precise with slightly different frequency. They will rapidly be out of face and the signal strength will decrease. An opposite gradient will compensate for this. The protons will spin with the same face which increases the strength of the signal.

Frequency Encoding y higher frequency x z x gradient lower frequency R

Frequency Encoding Within the imaging plane, a small gradient is applied left to right to allow for spatial encoding in the x direction. Tissues on the left will have a slightly higher resonance frequency than tissues on the right. The superposition of an x gradient on the patient is called frequency encoding. Frequency encoding enables spatial localization in the L-R direction only.

In the chosen slice the protons will spin with different frequency in the x-gradient. The antenna will pick up a wider frequency spectrum. A Fourier analysis gives the number of spinning protons as a function of frequency which is x-coordinate dependant.

Fourier analysis of the signal give frequency and amplitude Frequency and x-coordinate is coupled You get a projection of spin-amplitude as a function of x Collect projections from all directions Apply FBP

Phase Encoding An additional gradient is applied in the y direction to encode the image in the remaining direction. Because the x gradient alters the frequencies in the received signal according to spatial location, the y gradient must alter the phase of the signal.

Waves and Frequencies simplest wave is a cosine wave properties frequency (f) phase () amplitude (A)

After picking out the slice the protons are having the same frequency and the same phase. For a short moment a gradient is applied (y-gradient). It gives a phase shift which is related to the y-coordinate.

Then we apply the read-out gradient while listening Then we apply the read-out gradient while listening. Each x-position have a variation in phase that contains information about the y-coordinate

Phase Encoding The technique of phase encoding the second dimension in the imaging plane is sometimes referred to as spin warping. The phase encoding gradient is “stepped” during the acquisition of image data for a single slice. Each step provides a unique phase encoding. For a 256 x 256 square image matrix, 256 unique phase encodings must be performed for each image slice. The second 256 points in the x direction are obtained by A to D conversion of the received signal.

Frequency Encoding RF signal from entire slice 1 line of A/D conversion, 256 points 1 line of k-space RF signal from entire slice

Phase Encoding y x z y gradient, phase step #64 y gradient,

RF signal FT A/D conversion image space k-space

What is k-space? a mathematical concept not a real “space” in the patient nor in the MR scanner key to understanding spatial encoding of MR images 4

k-space is the temporary image space in which data from digitized MR signals are stored during data acquisition. When k-space is full (at the end of the scan), the data are mathematically processed to produce a final image. Thus k-space holds raw data before reconstruction.

k-space is in spatial frequency domain. We define kFE and kPE such that                   and where FE refers to frequency encoding, PE to phase encoding, Δt is the sampling time (the reciprocal of sampling frequency), τ is the duration of GPE,    (gamma bar) is the gyromagnetic ratio, m is the sample number in the FE direction and n is the sample number in the PE direction (also known as partition number), The 2D-Fourier Transform of this encoded signal results in a representation of the spin density distribution in two dimensions. k-space has the same number of rows and columns as the final image. During the scan, k-space is filled with raw data one line per TR (Repetition Time).

k-space and the MR Image x y f(x,y) ky  kx F(kx,ky) Image-space K-space 5

k-space and the MR Image each individual point in the MR image is reconstructed from every point in the k-space representation of the image like a card shuffling trick: you must have all of the cards (k-space) to pick the single correct card from the deck all points of k-space must be collected for a faithful reconstruction of the image

Discrete Fourier Transform F(kx,ky) is the 2D discrete Fourier transform of the image f(x,y) x y f(x,y) ky  kx F(kx,ky) 7 image-space K-space

k-space and the MR Image If the image is a 256 x 256 matrix size, then k-space is also 256 x 256 points. The individual points in k-space represent spatial frequencies in the image. Image contrast is represented by low spatial frequencies; detail is represented by high spatial frequencies. 8

low spatial frequencies high spatial frequencies all frequencies

Spatial Frequencies low frequency = contrast high frequency = detail The most abrupt change occurs at an edge. Images of edges contain the highest spatial frequencies. 12

Properties of k-space k-space is symmetrical all of the points in k-space must be known to reconstruct the signal faithfully truncation of k-space results in loss of detail, particularly at edges most important information centered around the middle of k-space k-space is the Fourier representation of the waveform 26