1 May 6, 2009 C. Lu Continuously Monitoring the RPC Gas Mixture with a Gas Chromatograph C. Lu Princeton University.

Slides:



Advertisements
Similar presentations
Lecture 8b Gas Chromatography.
Advertisements

1 Monitor of Isobutane Content in the RPC Gas Mixture Changguo Lu, Kirk McDonald Princeton University (10/22/2008)
The RPC Closed loop Gas System. Schedule for completion - update Status Supplier: ok Mixer: ok Humidifier: ok Purifiers: installed, but not completely.
Results of ageing tests C. Garabatos, GSI Experimental set-up TEDLAR ARALDIT 2013 Copenhagen,
Agenda Review homework Lecture/discussion Week 10 assignment
Daya Bay RPC Gas Safety System FDR, June 19, Daya Bay RPC Gas Safety System Design Changguo Lu, Kirk McDonald Princeton University (June 19/20,
C. Lu, K.T. McDonald RPC Gas System Final Design Review April 11, Daya Bay Reactor Antineutrino Experiment Resistive Plate Chamber Gas System Final.
Daya Bay RPC Gas System: Safety Subsystem Changguo Lu, Princeton University (Jan. 12, 2008)
1 Humidity Effect on the RPC Changguo Lu Princeton University 10/1/2007.
E.H. #1 RPC Gas System Commissioning  “Leaky” modules’ study 1 Changguo Lu, Princeton University Qingmin Zhang, IHEP Shih-Kai Lin, University of Houston.
1 Daya Bay Muon system Installation Workshop, IHEP, July 14, Updated RPC Gas System Installation Plan Changguo Lu Princeton University.
1 April 7, 2009 C. Lu Monitoring the RPC Gas Mixture with a Gas Chromatograph C. Lu Princeton University.
C. Lu, RPC System Installation Workshop, 6/11/2009, CUHK 1 1 RPC gas system installation (step by step plan) (CUHK, 6/11/2009) Changguo Lu, Kirk McDonald.
1 March 27, 2007 IHEP RPC Bakelite Resistivity in-Situ Test Results and Strip Plane Considerations W. Sands, C. Lu and K. McDonald Princeton University.
Daya Bay RPC Gas System Installation at EH#1 Lu Changguo 1Daya Bay Muon SYstem Installation Workshop, Daya Bay, 10/18/2010 Our original gas system installation.
1 December 19, 2007 RPC Gas Flammability C. Lu Princeton University.
1 Daya Bay collaboration meeting, IHEP, Beijing, Aug. 1-3, 2007 Test OPERA gas mixture on IHEP RPC w/ 8m long strip C. Lu Princeton University.
1 May 6, 2009 C. Lu Continuously Monitoring the RPC Gas Mixture with a Gas Chromatograph C. Lu Princeton University.
Daya Bay Muon Subsystem Review, 7/28-29/2007, IHEP, Beijing Daya Bay RPC Gas System: Design Report & Budget Estimate Changguo Lu, Kirk McDonald Princeton.
C. Lu, RPC Gas Slow Control Workshop, 7/15/2008, Dongguan, China 1 1 Some thoughts on RPC gas system slow control (Dongguan, 7/15/2008) Changguo Lu Princeton.
1 Daya Bay Muon system Installation Workshop, Daya Bay, March 22, Daya Bay RPC Gas System Installation Plan Changguo Lu Princeton University.
C. Lu, K.T. McDonald RPC Gas System Final Design Review April 11, 2008 Daya Bay RPC gas system fire safety issue Slides from Gas System FDR + a few more.
C. Lu, RPC Gas Slow Control Workshop, 6/10/2009, CUHK 1 1 RPC gas system slow control (CUHK, 6/10/2009) Changguo Lu Princeton University.
1 Daya Bay Collaboration Meeting, IHEP, Beijing, 8/1-3/2007 RPC aging concern Changguo Lu Princeton University Daya Bay Collaboration Meeting IHEP, Beijing,
Introduction to Gas Chromatography Written by Bette Kreuz Produced by Ruth Dusenbery University of Michigan-Dearborn 2000.
2011 HV scan SF6 flow-meter accident 2011 Results comparison RPC HV efficiency scan Pigi Paolucci on behalf of RPC collaboration.
1 Alhussein Abuhoza, S. Biswas, U. Frankenfeld, J. Hehner, Ch. Schmidt GSI, Darmstadt, Germany and H.R. Schmidt University of Tübingen, Germany Study of.
Thomas Jefferson National Accelerator Facility Page 1 Hall B Drift Chambers Review, March 2007 George Jacobs Jefferson Lab Hall B 12 GeV Upgrade Drift.
C. Lu, K.T. McDonald RPC Flammable Gas Monitor April 15, RPC Flammable Gas Monitor Based on Gas Chromatography Changguo Lu, Kirk T. McDonald Princeton.
Design of a Computer Controlled Test System for Micro-Electro-Mechanical-Resonator (MEMR) Based Gas Sensor Presenter: Joshua Ward Co-Author: Robert MacKinnon.
Contact person: E. Swietlicki, A. Massling NMR – Nordic Aerosol Project DMPS/SMPS Intercomparison Division of Nuclear Physics, Lund University,
PORTABLE HIGH-SPEED GAS CHROMATOGRAPH
Electronic Detector status and run organization Logistics and organization Updates on the electronic detector since last year run Conclusions.
GIF++ gas systems 21/01/2014Ernesto, Roberto PH-DT-DI gas service1.
Summer Updates- Summer Student Report Emine Altuntas Stephen Fee Supervisors: Mar Capeans, Ivan Glushkov, Roberto Guida Gamma Irradiation Facility.
Thomas Jefferson National Accelerator Facility Page 1 Hall B Drift Chambers Review, March 2007 George Jacobs Jefferson Lab Hall B 12 GeV Upgrade Drift.
EH#1 RPC Gas System Repair Report - Zhijian Zhang; Y. Lei; Qiang Xiao & Changguo Lu 11/5/2013.
First results of an aging test of a full scale MWPC prototype for the LHCb muon system Workshop on “Aging Phenomena in Gaseous Detectors”, October 2-5.
Advanced Method for Renewable Ethanol by Direct Synthesis from Syngas for Renewable Fuel Applications Julia Fisher, Chemical Engineering, Arizona State.
Active gas system status. S.Konovalov, K.Zhukov. Active gas system operation. S.Konovalov "Active gas system..."TRT Overview  Good design.
RPC detection rate capability ≈ 1/ρ Float-Glass ( Ω-cm) RPC rate capability < Bakelite( Ω-cm) one. New kind of doped glass was developed.
5. Quality Assurance and Calibration Quality assurance is We do to get the right answer for our purpose. Have Sufficient accuracy and precision to support.
The RPC Closed loop Gas System Slides from Bianco and Guida.
Andrea Di Simone Andrea Di Simone – INFN Roma2 Andrea Di Simone CERN PH/ATC and INFN-CNAF On behalf of ATLAS RPC groups: Lecce, Napoli, Protvino, Roma2.
1 Daya Bay RPC Gas Safety System - Gas Cabinet Test Report Changguo Lu, Kirk McDonald Princeton University (October 11, 2009)
GIF++ gas systems 11/07/2014Ernesto, Roberto PH-DT-DI gas service1.
Progress on the beam tracking instrumentation Position measurement device Tests performed and their resolution Decision on electronics Summary.
RPC gas purity During 2/18 – 2/26/2013 EH#1 RPC gas flow rate showed anomaly, the total gas and branch #1 flow rate displayed a bump, but all four gases.
Results from the CMS-RPC tests at CERN Gamma Irradiation Facility Roberto Guida CMS-RPC (Bari, Beijing, Napoli, Pavia, Sofia) Seoul, October 2005.
Gas Quality Checks Maria Luisa Porto Oriented by: Burkhard Schmidt 23/02/2009.
RPC Operation Jilei, Qingmin, Changguo, Ningzhe 2013/4/3.
GIF ++ ambient condition The Prep. area outside Tent and its gradient CSC area Geneva meteo station Ian Crotty6 Oct 2015.
Active gas system status & plans. S.Konovalov, K.Zhukov, K.Vorobev.
CMS-GEM gas system GEMs for CMS High Eta Muons - Workshop III, CERN April R.Guida PH-DT-DI.
1 Analysis of Small RPC DHCAL Prototype Data (noise and cosmic ray) LCWA09, Albuquerque, New Mexico Friday, October 02, 2009 Qingmin Zhang HEP Division,
S. Colafranceschi - Ischia CMS RPC Italia1 CMS RPC studies at ISR Stefano Colafranceschi Laboratori Nazionali Frascati Università di Roma – La Sapienza.
Siena 2002 Siena, 24/10/2002 G. Passaleva New results from an extensive aging test on bakelite RPC G. Passaleva INFN Firenze.
GIF++ gas system Outline:  Beam and cosmic trigger  Gas distribution panels  Gas detection (present and final layout)  Conclusions 14/1/2015R. Guida,
S.Colafranceschi 20 June Stefano Colafranceschi LNF Frascati CERN Universita’ di Roma – La sapienza An overview on gas studies, purifier materials.
Performance and Aging Studies of BaBar RPC’s Henry R. Band University of Wisconsin for the BaBar IFR-RPC group RPC 2005 VIII Workshop.
Final results from an extensive aging test on bakelite Resistive Plate Chambers Stefano de Capua University of Rome II “Tor Vergata” and INFN.
Control System for a four component Gas Mixing Unit Suvendu Nath Bose, Satyajit Saha, Sudeb Bhattacharya, Saha Institute of Nuclear Physics, Kolkata
Long-term environmental effects on the ARGO-YBJ RPC array studied with the Detector Control System P. CAMARRI 1, R. CARDARELLI 1, L. PALUMMO 1, C. VIGORITO.
R & D Status report on INO Naba K Mondal Tata Institute of Fundamental Research Tata Institute of Fundamental Research Mumbai, India.
PORTABLE HIGH-SPEED GAS CHROMATOGRAPH
R. Guida PH-DT-DI Boston Students Program CERN, 16/1/2013
Agenda Review homework Lecture/discussion Week 10 assignment
Design of a Computer Controlled Test System for
Evaluative practice What is Evaluation? The consideration of:
Evaluative practice What is Evaluation? The consideration of:
Presentation transcript:

1 May 6, 2009 C. Lu Continuously Monitoring the RPC Gas Mixture with a Gas Chromatograph C. Lu Princeton University

2 May 6, 2009 C. Lu Run GC in sequence mode We created a sequence, which repeatedly execute a GC method file (program) and collect a gas chromatogram every 21’. The corrected gas mixing ratio is directly wrote into a data base. We have collected data continuously for more than 2 weeks. Ar SF6 Isobutane R134A See the interpretation of the plot on next slide.

3 May 6, 2009 C. Lu Interpretation of the results The GC tested mixing ratio for each gas component is not so stable. Is this due to gas mixing panel or GC? Most likely it is due to the gas mixing panel. The repeatability of the MKS flow controller is 0.2% of full range. The full range for the mixing panel we are using is for the Daya Bay near hall RPC, Ar full range is ~2800 sccm, thus the precision would be 2800*0.2% = 5.6 sccm. Now we only use 65 sccm for Ar controller, so we should expect the fluctuation at the level of 5.6/65 = 8.6%. The other three controllers are having similar limitation. To confirm if GC can distinguish small increase of Isobutane percentage, we adjusted the gas mixing ratio at two points. At the first point we decreased the flow rate for Ar and R134A, at the second point we change their flow rates back to original. The small percentage increase of Isobutane can be clearly identified. It gives us a confidence that the GC will be able to carry out the task as a safety device as well.

4 May 6, 2009 C. Lu Dark current and ambient parameters We recorded an RPC chamber’s current with a Co-60 source: HV = 7600V, Daya Bay gas mixture. We also recorded the ambient parameters. Looks to me the most sensitive ambient parameter to the current is relative humidity. We can see that the current follows R.H. nicely. temperature pressure humidity Current

5 May 6, 2009 C. Lu Conclusion The Varian 3900GC system is a reliable, high-quality tool to check Daya Bay RPC gas system, including the concentration of isobutane. 0.5% of Isobutane mixing ratio change can be reliably detected by GC. Ambient parameters are sensitive to RPC operating current. Since the humidity can change wildly, its effect on the dark current is very important.