Lecture Three: "MALDI-TOF MS, Up Close, and Personal Review of principles Delayed Extraction Reflector / reflectron Post-source Decay (PSD)

Slides:



Advertisements
Similar presentations
Module 2: Analytical methods and separation techniques.
Advertisements

Mass spectroscopy. In a typical MS procedure:  1- a sample is loaded onto the MS instrument, and undergoes vaporization.  2- the components of the sample.
Introduction to MALDI-TOF MS
MALDI-TOF Mass Spectrometry and Introduction to Proteomics Dr. Steve Hartson Oklahoma State University Dept. Biochemistry and Molecular Biology Recombinant.
Lecture 9. MALDI/TOF. Introduction Matrix-assisted laser desorption/ionization (MALDI) is a soft ionization technique used in mass spectrometry, allowing.
Mass Spectrometry in Life Science: Technology and Data-Evaluation H. Thiele Bruker Daltonik, Germany.
High Performance MALDI-TOF Mass Spectrometry Marvin Vestal and Kevin Hayden Virgin Instruments Corp. Sudbury, MA.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
Mass Spectrometry in the Biosciences: Introduction to Mass Spectrometry and Its Uses in a Company Like Decode. Sigurður V. Smárason, Ph.D. New Technologies.
PROTEIN IDENTIFICATION BY MASS SPECTROMETRY. OBJECTIVES To become familiar with matrix assisted laser desorption ionization-time of flight mass spectrometry.
Mass Labels While Mass Spectroscopy is not covered in our texts, other mass/density dependent technologies including X-ray Spectroscopy, Gamma Spectroscopy,
Mass Spectrometry.
Molecular Mass Spectrometry
Mass Spectroscopy Quantitative Chemical Analysis Harris, 6th Edition
LC/MS WORKSHOP IOWA STATE UNIVERSITY Kamel Harrata  Instrument Description  Data Acquisition  Data Processing.
Basics of 2-DE and MALDI-ToF MS
Proteomics Informatics Workshop Part I: Protein Identification
Mass Spectroscopy 1Dr. Nikhat Siddiqi. Mass spectrometry is a powerful analytical technique that is used to identify unknown compounds, to quantify known.
My contact details and information about submitting samples for MS
MALDI-TOF-TOF with High Resolution Precursor Selection and Multiplexed MS-MS Poster Number ThP 618 Kevin Hayden, Stephen C. Gabeler, Mark Dahl and Marvin.
Proteomics Informatics – Overview of Mass spectrometry (Week 2)
TOF Mass Spectrometer &
Chapter 8 – Mass Spectrometry. Mass Spectrometry The mass spectrometer can be used for: – Quantitative analysis – as a sophisticated and very sensitive.
Comparison of chicken light and dark meat using LC MALDI-TOF mass spectrometry as a model system for biomarker discovery WP 651 Jie Du; Stephen J. Hattan.
Gaucher’s Disease – A glycolipid storage disorder characterized by the accumulation of glucosylceramide in the spleen, liver, lungs, bone marrow and brain.
Molecular mass spectrometry Chapter 20 The study of “molecular ions” M + e -  M e -
Mass Spectrometry I Basic Data Processing. Mass spectrometry A mass spectrometer measures molecular masses. The mass unit is called dalton, which is 1/12.
UPDATE! In-Class Wed Oct 6 Latil de Ros, Derek Buns, John.
INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable.
Common parameters At the beginning one need to set up the parameters.
1 Chemical Analysis by Mass Spectrometry. 2 All chemical substances are combinations of atoms. Atoms of different elements have different masses (H =
Mass Spectroscopy 1 Mass Spectroscopy (Mass Spec) Applying Atomic Structure Knowledge to Chemical Analysis.
Marvin Vestal and Kevin Hayden SimulTOF Systems
MS Calibration for Protein Profiles We need calibration for –Accurate mass value Mass error: (Measured Mass – Theoretical Mass) X 10 6 ppm Theoretical.
MS Intro. MS requires gas-phase ions, why? MS uses magnetic and electric fields to control the path of a compound based on mass to charge ratio (m/z)
For all CHEM5161 students: The first day of class for CHEM5161 (Analytical Spectroscopy) will be on TUE Sept 4 (following Labor Day). There will be no.
INF380 - Proteomics-51 INF380 – Proteomics Chapter 5 – Fundamentals of Mass Spectrometry Mass spectrometry (MS) is used for measuring the mass-to-charge.
Chemistry 367L/392N Macromolecular Chemistry Lecture 8.
Starter: Spec links 1.06–1.10. Green pens out! Spec links 1.06–1.10.
Mass Analyzers : Time-of-flight Mass Spectrometry CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Joel Kimmel Fall 2007.
F LORIDA I NTERNATIONAL U NIVERSITY Advanced Mass Spectrometry Piero R. Gardinali/Yong Cai/ Bruce McCord Revised on August 23, 2009.
1 CH908 Structural Analysis by Mass Spectrometry revision lecture. Prof. Peter O’Connor.
Separates charged atoms or molecules according to their mass-to-charge ratio Mass Spectrometry Frequently.
F Don Lincoln, Fermilab f Fermilab/Boeing Test Results for HiSTE-VI Don Lincoln Fermi National Accelerator Laboratory.
Metabolomics MS and Data Analysis PCB 5530 Tom Niehaus Fall 2015.
Chem. 133 – 2/18 Lecture. Announcements Homework Set 1.2 (bold problems) – due today Quiz 2 Today Electronics Lab Reports – due Tuesday Seminar this Friday.
WELCOME STUDENTS Mobile : Skype: aamarpali.puri.
2014 생화학 실험 (1) 6주차 실험조교 : 류 지 연 Yonsei Proteome Research Center 산학협동관 421호
Chapter 29 Mass Spectrometry. 29 A Principles of mass spectrometry In the mass spectrometer, analyte molecules are converted to ions by applying energy.
How do we know? Spectroscopy: Experimental Evidence.
What is Mass Spectrometry? Mass spectrometry could be considered as an analytical technique that involves the study in the gas phase of ionized molecules.
Chapter 11 Structure Determination: Mass Spectrometry, Infrared Spectroscopy, and Ultraviolet Spectroscopy.
Yonsei Proteome Research Center Peptide Mass Finger-Printing Part II. MALDI-TOF 2013 생화학 실험 (1) 6 주차 자료 임종선 조교 내선 6625.
An introduction to Mass spectrometry. What is mass spectrometry? Analytical tool measuring molecular weight (MW) of sample Only picomolar concentrations.
Post translational modification n- acetylation Peptide Mass Fingerprinting (PMF) is an analytical technique for identifying unknown protein. Proteins to.
Peptide Mass Finger-Printing Part II. MALDI-TOF
Mass Spectrometry 101 (continued) Hackert - CH 370 / 387D
Review of Basic Atomic Structure, Mass Spectrometry
Acknowledgements Slides and animations were made by Dr. Jon Karty Mass Spectrometry Facility Indiana University, Bloomington.
Mass Spectrometry 101 Hackert - CH 370 / 387D
Tandem MS.
Time of Flight Analyzers
Mass Spectrometry Obaid M. Shaikh.
Starter Complete the quiz and hand in..
Mass Spectrometry Courtesy
Chem. 133 – 2/16 Lecture.
Mass Spectrometry.
Schematic of MS1 filtering.
Mass Spectrometry THE MAIN USE OF MS IN ORG CHEM IS:
M.Prasad Naidu MSc Medical Biochemistry, Ph.D,.
Presentation transcript:

Lecture Three: "MALDI-TOF MS, Up Close, and Personal Review of principles Delayed Extraction Reflector / reflectron Post-source Decay (PSD)

Mass Spec vocabulary Generic mass spectrometer Ions & isotopes Mass: –m/z –monoisotopic mass –average mass –peak centroid Resolution

Ions Only ions are detected in MS For ionization techniques that are typically used for biological molecules, ions are generated via the ejection or capture of a proton. The mass of a proton is ~1 AMU; the charge of a proton is +1 Ionic mass [MH] +1 versus molecular mass (subtract ~1 from raw data)

Peaks A peak represents a packet of peptide ions hitting the detector. The distribution of ions’ flight times creates a rapid rise, and then fall, of current from the detector (y-axis). This current is captured digitally. Base Peak (BP): the strongest peak in the spectrumBase Peak (BP): the strongest peak in the spectrum

Resolution: the MS gold standard Limits mass accuracy R=mass/peak width R=1000 R=3000 R=10,000 R=30,000

Measuring Resolution Analyte mass divided by Full peak Width, as measured at one-Half peak's Maximum height (FWHM) Industry standard, despite being somewhat arbitrary Alternatively, the DE-PRO can resolve analytes with a difference of 1 part per thousand (linear mode)… …or a difference of 1 part per 6 thousand (reflector mode).

monoisotopic mass, average mass, peak centroid high-resolution spectrum with isotopes resolved low-resolution spectrum with unresolved “isotopic envelope”

Mass versus m / z Mass divided by charge +1 ions, +2 ions, +3 ions,… +20…+30. "twice the charge" behaves like "half the size"

Isotopes For peptide-sized molecules, most mass spec’s can resolve (n) versus (n +1 AMU). The result is that a single peptide actually yields a series peaks differing by one AMU.

Calibration and mass error MALDI-TOF’s must be rigorously calibrated, due to TOF variance across the face of the probe plate. Other MS’s need less frequent calibration. You will always have error –error as a pitfall –error as a tool

Drift time allows mass determination because: drift time~velocity~ acceleration~mass. The measurement is calibrated by co-analysis of standards whose masses are known.

TOF ~ m / z

Calibration drift time~velocity~ acceleration~mass. The relationship between TOF and mass can be calibrated using standards with known masses. …or "default" estimates.

Calibrations must follow the laws of physics drift time~velocity~ acceleration~mass. This relationship is linear, and major departures are not physically possible. NONSENSE 

Calibration drift time~velocity~ acceleration~mass. In a calibration effort and in a database search result, errors between data and theory must be systematic and/or within instrument tolerances.

Calibration and Mass Error In PMF d-base search results, differences between theoretical mass and experimental mass arise from two causes: –Calibration error –An invalid, coincidental match between your data and the theoretical protein

Mass Error

Calibration drift time~velocity~ acceleration~mass. In a calibration effort and in a database search result, errors between data and theory must be systematic and/or within instrument tolerances.

Fractional Mass as a tool Although 12 C is AMU, other atoms have a decimal component which is not zero-a fractional mass. This fractional mass contributes to peptide mass in a consistent manner: ~ 0.5 Da per kDa. This consistent trend can be used to assess calibrations: if the FM is wrong, the calibration is suspect. This trend is a powerful way to ID artifacts in peak lists (matrix, de-isotoping errors, noise). powefu

Lecture Three: "MALDI-TOF MS, Up Close, and Personal Review of principles Delayed Extraction Reflector / reflectron Post-source Decay (PSD)

Practical MALDI considerations You need crystals of peptide:matrix. You’ll have matrix noise, especially: –When signal is low; –In the low-mass range Ionization is a competitive process: – minimal matrix, salt, trypsin

The problem: The desorption process imparts intitial velocities to analyte molecules (independent of, and prior to, the accelerating voltage). These initial velocities are not uniform; they have a significantly wide range. These non-uniform initial velocities are significant and affect TOF. The result is broad TOF peaks of analytes with identical masses (poor resolution).

The solution: Exploit the initial velocities by delaying the “extraction” (the application of the accelerating voltage). Delay allows the initial velocities to be translated into distance from the plate. When the plate is charged, this distance will impact the time spent in the accelerating field. Time spent in the accelerating field will impact the magnitude of acceleration (and thus, velocity/TOF).

The skate park analogy: Identical twins on skate boards Have the same mass...

…but the green skater cheats; he hits the board running, and thus has a greater initial velocity... “Go!”

…which will allow him to pull ahead of his twin, despite gravity’s equal acceleration of both skaters... “And they’re off…”

…thus, despite their equal masses, the skaters will not hit the finish line (the “detector”) at the same time.

The solution... (Again, our identical twins have the same mass but different initial velocities…) “Go!”

Use a slight delay to translate the initial velocity into distance before applying the acceleration... And they’re off! (No “slope” = no electrostatic field applied to accelerate the ions.)

…then apply the accelerating field. The purple skater will experience more acceleration than the green skater.

This greater acceleration of the formerly slow ion results in a greater velocity during the time of flight...

…allowing the purple skater to catch up...

…and the two skaters of equal mass reach the finish line (the detector) simultaneously, despite having different initial velocities. We have focused the arrival of our twin analytes at the detector.

One last twist: grid voltage

We break the accelerating voltage into two slopes using the grid. This gives us more control in our manipulation of time lag focusing. Delay time and Grid voltage are interdependent parameters.

Lecture Three: "MALDI-TOF MS, Up Close, and Personal Review of principles Delayed Extraction Reflector / reflectron Post-source Decay (PSD)

The reflector is an “ion mirror” that redirects the vector of ion flight.

Reflectors dramatically increase resolution... …and by creating a slightly longer flight path (greater separation between peaks)… …by focusing the arrival of ions having the same mass, but slightly different velocities (sharper, narrower peaks.

MALDI-TOF Theory-overview Our instrument General theory of MALDI-TOF Delayed Extraction Reflector Post-source decay (PSD)

Post Source Decay PSD is a non-specific cleavage tool that can be used to generate fingerprints of individual peptides. Referred to as an “MS/MS” technique because ions are re-accelerated via the reflector.) CID and CAF can be used to facilitate decay.

Peptides decay (break) during flight.

Breakage generates characteristic products (and nomenclature).

The masses of daughter ions can be determined. TOF ~acceleration~mass First acceleration: from the source Second acceleration: reflection in the mirror.

Only ions with a specific TOF (the parent ion and its daughters) are allowed to enter the ion mirror. Too many analytes! A specific peak mass range is chosen by the operator. “Timed ion selector” (Bradbury-Neilson gate)

Problem: daughters don’t arrive on time No single reflector voltage (“mirror ratio” relative to accelerating voltage gradient) will properly focus all daughter ions. Although all fragments from a single parent have the same velocity... light daughter ions have less energy, and will be reflected too easily. heavy daughter ions have more energy, and may pass through the ion mirror.

The solution: Use a slightly lower “mirror ratio” (reflector voltage) for each daughter size range. “Mirror ratio” is the reflector’s voltage relative to the accelerating field.

At mirror ratio Parent MH+ (1000 Da ) is properly focused. Daughter AH+ (700 Da) is poorly focused. Daughter BH+ (300 Da) is poorly focused.

At mirror ratio 0.7 (a shallower voltage gradient)... Parent MH+ (1000 Da) is not reflected. Daughter AH+ (700 Da) is properly focused. Daughter BH+ (300 Da) is poorly focused.

At mirror ratio 0.3 (an even shallower voltage gradient)... Parent MH+ (1000 Da) is not reflected. Daughter AH+ (700 Da) is not reflected. Daughter BH+ (300 Da) is properly focused.

The result is several “mini-spectra” which are “stitched” together to yield a composite spectrum.

In the composite spectrum... The heaviest peak represents the parent (this intense peak may need to be cropped off for display purposes). Other peaks represent decay products.

Uses of PSD Confirmation tool Can be used for searching. Analysis of labile post-translational modifications. CAF Fragmentation of organic.

Limitations Efficiency of breakage Mass accuracy Our manufacturer: –no automation=tedious –Their corporate investments focused on MS/MS

End of Lecture Three: Questions? Next up: "Bigger, Better, Faster, Stronger: the Cutting Edge"