Focused ion beam (FIB) Overview. Ion source and optics.

Slides:



Advertisements
Similar presentations
Secondary Ion Mass Spectrometry
Advertisements

Basic Electron Microscopy
Chapter 12 Additional Analytical Methods. Analytical Methods Technique Type Technique application Subdivisions Specific application DescriptionDestruction.
Focused ion beam (FIB) 1.Overview. 2.Ion source and optics. 3.Ion-solid interaction, damage. 4.Scanning ion beam imaging. ECE 730: Fabrication in the nanoscale:
SEM & TEM in Polymer Characterization
Ion Beam Analysis techniques:
Electron-Specimen Interactions
Scanning Electron Microscope (SEM)
1 Introduction to Plasma Immersion Ion Implantation Technologies Emmanuel Wirth.
Chapter 8 Ion implantation
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
Lecture 5.1 Scanning Electron Microscopy (SEM)
XPS and SIMS MSN 506 Notes.
Secondary Ion Mass Spectrometry Professor Paul K Chu.
Transmission Electron Microscopy (TEM) By Austin Avery.
Secondary Ion Mass Spectrometry Professor Paul K Chu.
Introduction to scanning electron microscopy
BY SANTANU PRAMANIK(09369) HITESH KUMAR GUPTA(09320) CHANDAN SINGH(09260) SCANNING ELECTRON MICROSCOPE MATERIAL SCIENCE ASSIGNMENT.
Thin Film Deposition Prof. Dr. Ir. Djoko Hartanto MSc
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
Radiation therapy is based on the exposure of malign tumor cells to significant but well localized doses of radiation to destroy the tumor cells. The.
Secondary Ion Mass Spectrometry (SIMS)
M.H.Nemati Sabanci University
Chapter 8 Ion Implantation Instructor: Prof. Masoud Agah
ECE/ChE 4752: Microelectronics Processing Laboratory
Ion Implantation Topics: Deposition methods Implant
III. Analytical Aspects Summary Cheetham & Day, Chapters 2, 3 Chemical Characterization of Solid-State Materials Chemical Composition: Bulk, Surface, …
Chapter 8 Ion Implantation
NANOFABRICATION -3 NOVEL PROCESSES EEE5425 Introduction to Nanotechnology1.
PC4250 Secondary Ion Mass Spectrometry (SIMS). What is SIMS? SIMS is a surface analysis technique used to characterize the surface and sub-surface region.
B. Spatial coherancy & source size Spetial coherancy is related to the size of the source. Source size governs spatial coherancy and maller source sizes.
EDS Energy Dispersive Spectroscopy
Molecular mass spectrometry Chapter 20 The study of “molecular ions” M + e -  M e -
1. In a Liquid metal ion source (LMIS), a metal (typically gallium) is heated to the liquid state and provided at the end of a capillary or a needle.
Scanning Electron Microscope (SEM)
Post Anneal Solid State Regrowth
Other modes associated with SEM: EBIC
Chemistry Topic: Atomic theory Subtopic : Mass Spectrometer.
TEM charcaterization Basic modes – Bright field microscopy – Dark field Microscopy –STEM – EDAX – EELS.
NANO 225 Micro/NanoFabrication Electron Microscopes 1.
Sputter deposition.
Mass spectrometry (Test) Mass spectrometry (MS) is an analytical technique that measures masses of particles and for determining the elemental composition.
NANO 230 Micro/Nano Characterization
Scanning capacitance microscopy
Daniela Adriana LĂCĂTUŞ1 Supervisor: Alexander Pavlovich KOBZEV
Characterization of Nanomaterials…
08/03/09 SEM signal generation
SEM Scanning Electron Microscope
Secondary Ion Mass Spectrometry A look at SIMS and Surface Analysis.
X-ray absorption spectroscopy (XAS)
Questions/Problems on SEM microcharacterization Explain why Field Emission Gun (FEG) SEM is preferred in SEM? How is a contrast generated in an SEM? What.
© 2004 Dieter Ast, Edwin Kan This material has been edited for class presentation. Ion Implantation: The most controlled way to introduce dopants into.
Electrons Thermionic Emission
Lecture 5: Secondary ion mass spectroscopy (SIMS) Assoc. Prof. Dr. Zainovia Lockman, PPKBSM, USM EBB 245. Materials Characterisation.
SARDAR PATEL INSTITUTE OF TECHNOLOGY E.NO : Guide By:- V.N.Thakkar.
Scanning Transmission Electron Microscope
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
BY GAJENDRA KUMAR ID- 2011uit1721. WHAT IS SEM  It is a microscope that uses a focused electron probe to extract structural and chemical information.
CHAPTER 5 ION IMPLANTATION In ion implantation, ionized impurity atoms are accelerated through an electrostatic field and strike the surface of the wafer.
Introduction to scanning electron microscopy
Presentation on SEM (Scanning of Electron Microscope) Represented by:-Ravi Kumar Roll:- (BT/ME/1601/006)
Introduction to scanning electron microscopy
NANO 230 Micro/NanoFabrication
TEM and SEM.
Chapter 8 Ion Implantation
Do it with electrons !.
1.6 Glow Discharges and Plasma
Shukui Zhang, Matt Poelker, Marcy Stutzman
High resolution transmission electron microscopy (HRTEM) investigations of defect clusters produced in silicon by electron and neutron irradiations Leona.
Presentation transcript:

Focused ion beam (FIB) Overview. Ion source and optics. Ion-solid interaction, damage. Scanning ion beam imaging. ECE 730: Fabrication in the nanoscale: principles, technology and applications Instructor: Bo Cui, ECE, University of Waterloo; http://ece.uwaterloo.ca/~bcui/ Textbook: Nanofabrication: principles, capabilities and limits, by Zheng Cui

Focused ion beam overview Mainly developed in 1970’s and 80’s (Escovitz, Levi-Setti, Orloff, Swanson…) Ion column structure similar to that of SEM. Source: Liquid Metal Ion Source (LMIS). Ga, Au, Be, Si, Pd, B, P, As, Ni, Sb, alloy … Principle: a strong electromagnetic field causes the emission of positively charged ions. Secondary Ion Mass Spectrometry (SIMS) is a analytical tool with high spatial resolution and high sensitivity. It uses a highly focused ion beam (generally oxygen or cesium ions for inorganic samples) which “sputters” material from a selected area on sample surface. The ejected “secondary ions” pass through a mass spectrometer, which separates the ions according to their mass/charge ratio, in effect providing chemical analysis of a very small sampling volume. FIB is based on similar technology as SIMS Primary ion beam Secondary ions

Comparison of electrons and ions

Comparison of electrons and ions Energy E=½mv2 momentum=mv=(2mE)1/2m1/2 for same E.

Gun with LMIS ion source Focused ion beam (FIB) Gun with LMIS ion source Ion source Extraction electrodes Condensor lens Ion current selection aperture Wien filter Mass selection aperture Ion optics Blanking plates Faraday cup Scanning and Stigmation octupole Objective lens Chamber Sample stage

Focused ion beam (FIB) Overview. Ion source and optics. Ion-solid interaction, damage. Scanning ion beam imaging.

Ion sources Early ion sources were developed for mass spectrometry and nuclear physics research, then for ion implantation for semiconductor manufacturing. Four types. Electron bombardment ion sources. Gas molecules bombarded by electrons become ionized, producing a stream of ions. Small ion current and small ion energy spread, used in mass spectrometry. Gas discharge ion source. High current, used in nuclear physics instrumentation such as high energy accelerators and ion implanters. Field ionization source. Gas molecules absorbed on the surface of a fine needle tip can be directly ionized at extremely high electrical field near the tip apex. Used as field ion microscope to study atomic structure of a material. Liquid metal ion source (LMIS). A field emission source from liquid metal under a strong electrical field.

Liquid metal ion source (LIMS) W wire electrochemically etched into a needle with tip radius 5-10m. The needle wetted by molten metal. The electrical field at the liquid apex can reach 108V/cm. At this field, metal atoms at the apex become ionized and escape in the form of field evaporation. Experiment has shown the following. There exists a threshold voltage (2kV) for ion emission. The emission angle is large, around 30o. The angle distribution of emission current is rather uniform. Energy spread of emitted ions is large, 15V, leading to large chromatic aberration in an ion optical system. At current <10A, almost 100% ions are single charged.

TEM image of operating LIMS The W wire is not sharp at all! Electric field AuGe W LMIS emitter substrate The W wire is not sharp at all! But the “Taylor” cone of the liquid metal induced by electric field is very sharp. As a result, electric field at cone apex is very high for field emission. LMIS emitter substrate with AuGe Taylor cone Driesel W, Dietzsch C, Muhle R, J. Vac. Sci. Technol. B, 14, 3367(1996)

Liquid metal ion source Heating current 10 mm

Liquid metal ion source: why Ga Melting point at 30°C → liquid around room temperature. Low vapor pressure → applicable in high vacuum. [Ga2+]/[Ga+]  10‐4 at 10μA → narrow energy distribution. Long life (up to 1500 hr sources). Heavy ion for sputtering (Z=31). Blunt W with grated surface for Ga transport Ga forms a Taylor‐ Gilbert cone

Ion optics: overview Similar to electron optics. But use only electrostatic lens and deflectors to focus and deflect ion beam, because for magnetic lens (though they have superior optics): It must be fabricated impractically large to focus 30kV Ga ions. The focusing plane depends on mass/charge ratio. There is not much room for improvement in electrostatic lens performance given that working distances must remain ≥ 10 mm to allow room for gas injectors etc. Higher space charge effect, as the repulsion of particles of same charge is inversely proportional to their speed (ion is a lot slower than electrons). This leads to: Large energy spread (5-15eV) and thus large chromatic aberration. Enlarged focused ion beam. State-of-the-art FIB has focal spot size below 5nm at current of few pA. At medium current (60pA, 1mrad), chromatic aberration (=CcdE/E) dominates. At large current (5nA, 10mrad), spherical aberration (=0.5Cs3) dominates. Force independent of m, same for electron and ion. v is speed, much smaller for ions than for electrons; so much smaller force, need impractically large B to focus. V is acceleration voltage.

Ion source and optics

Beam size vs. current Spot size 20nm at 100pA. For electron, spot size < 10nm at 1nA. Source spot size is about 50nm (fixed). Small aperture  small beam current (slow) and narrow beam (high resolution). Need a tradeoff. Beam tails can extend up to some m, is one limiting factor when milling deep high aspect ratio (depth/width) trenches/holes (the other factor is re-deposition of sputtered material).

Focused ion beam (FIB) Overview. Ion source and optics. Ion-solid interaction, damage. Scanning ion beam imaging.

Ion-solid interactions ? Imaging (secondary electron image, 2-3 SE per ion), milling and deposition simultaneously. Leads to lattice defects (vacancies, interstitials, dislocations). Leads to damages, amorphization, re-crystallization.

Ion-ion and ion-electron scattering: the equations Nuclear differential scattering cross-section, used to calculate scattering angle distribution. En: energy loss to solid atom (nucleus), lead to sputtering, dislocation, recoiling… Ee: energy loss to electrons, leads to secondary electron emission, ionization… L: mean free path between two collisions. N: atomic density of solid material.

Ion-solid interactions Three regimes of ion-solid interactions Regime I (knock-on regime) M1<<M2 or E0 is low, minimum sputtering Regime II (linear cascade regime), where FIB operates. M1M2, E0 is moderate, governed by nuclear effects Regime III (spike-on regime) M1>M2 and/or E0 is high, majority of atoms move in collision cascade Ion range (penetration/attenuation depth) Ion range varies with ion energy and substrate material, order 1nm/keV (for electron, a few m at 30keV)

Trajectories for Ga+ bombardment 30 keV Ga+ on Si sample 30 keV Ga+ on W/Si sample Penetrate deeper inside Si than W Grazing angle, less damage, fast milling rate T. Ishitani, et al., J. Vac. Sci. Technol. B16, 1907 (1998)

Damage by Ga+ bombardment Ion Implantation - Ga atoms remain in the sample target and may reach critical composition for second phase formation. Amorphization of surface – loss of crystallinity Lattice defects Vacancies – displaced or “missing” atoms from their equilibrium lattice positions Interstitials – atoms which are positioned in between equilibrium lattice positions Dislocations – a missing “half-plane” of atoms Local heating due to large displacement of atoms that may occur within the collision cascade (10’s of nanometers from surface) Concentration of primary radiation defects (knock-outs from lattice sites) can be evaluated by Kinchin-Pease formula: nD = kE/2Ed, where k ≈ 0.8 is a coefficient, E is ion energy, Ed is displacement energy. Average 1000 defects per ion. Ga in most semiconductors is acceptor, affecting electronic, optical, magnetic and thermal properties. Concentration of Ga in the irradiated zone can be given by: CGa = 1/(1+γ), where γ is sputter yield.

25keV Ga+ depth distribution for silicon (implantation) Ga+ flux SIMS data, concentration is relative fraction of Ga H. Gnaser et al., J. Vac. Sci. Technol. B13, 19 (1995)

Damage by Ga+ bombardment InP, 40 nm damage layer thickness Amorphization depth 10 keV 30 keV Si 6 nm 28 nm GaAs 4 nm 24 nm InP 15 nm 40 nm J. F. Walker and R. F. Broom, Inst. Phys. Conf. Ser. 157, 473 (1996)

Use low keV FIB milling to reduce ion damage Reduced amorphous layer with reduced FIB energy What material? GaAs? 30keV 5keV 2keV 21nm 2nm 0.5-1.5nm

Focused ion beam (FIB) Overview. Ion source and optics. Ion-solid interaction, damage. Scanning ion beam imaging.

Ion beam imaging Scanning ion microscope (SIM) FIB imaging is destructive, yet may be used to remove (insulating) oxide layer. Topographic contrast. Secondary electron and secondary ion images. Ion channeling contrast for grain size measurements. Material contrast - local compositional differences. Voltage contrast - electrical state differences - passive and active voltage contrast. Scanning ion microscope (SIM) Secondary electron mode Detector biased positive Emitted from top 5-10 nm Typically 30kV 40pA for optimal resolution and signal Only charging up a few volts to go dark (electrons cannot escape) Grounded metals very bright, oxides dark Secondary ion mode Detector biased negative Emitted from top 0.5-1nm (very surface sensitive) No voltage-contrast Oxides brighter Lower yield, so images noisier.

Imaging resolution vs. sputtering yield Problem with FIB imaging resolution: if an object is small enough, it will be sputtered away before sufficient signal (secondary electrons) can be collected to resolve it. Before imaging Sputtering rate (m3/nC) Image resolution (nm) After 1min imaging Orloff et al., JVST (1996)

Materials contrast Scanning ion microscopy (SIM) SEM (secondary electron image) Scanning ion microscopy (SIM) Secondary electron image SIM image has higher material contrast than SEM image, and with fine structure (peaks and valleys in the curve). Unlike SEM, the signal brightness is lower for heavier materials (larger Z). For SEM, heavier materials have higher SE yield since there are more electrons for high Z materials. Relative secondary electron (not ion) intensity “Origin of material contrast in scanning ion microscope”, J Electron Microscopy, 2002.

Passive voltage contrast SE cannot escape, dark region Positively charged Bridging open contact with metal deposition indicates remaining contacts are good. Open contact in chain shown by charging.

Ion channeling contrast

Ion channeling contrast Ions penetrate deeper in crystalline material for certain grain orientations. Those channeled ions have lower sputtering yield (slower milling rate). Secondary electron (SE) yields are also lower for areas that channel better. Polycrystalline materials have grains with different orientations. Grain size can be determined by images at different incidence angles. Typical SE intensity with respect to tilt angle for Fe single crystal sample under 30keV Ga-FIB bombardment.

Ion channeling contrast Left: not aligned with crystal direction Right: channeled (aligned) Ion channeling contrast

Channeling contrast for grain size analysis