Digital Image Manipulation I By Professor Stelmark.

Slides:



Advertisements
Similar presentations
DIGITAL RADIOGRAPHY.
Advertisements

Technique Guidance Systems
IMAGE ARTIFACTS.
Image Quality Bushong Chapter 7.
The Field of Digital Radiography
Primary Exposure Factors IV
Quality Assurance and Digital Radiography
Digital Radiography.
Dr Mohamed El Safwany, MD. 2 The student should be able to recognize technological principles of radiographic Abdominal imaging.
Control of Scatter Radiation Beam Restricting Devices
Chapter 8 Planar Scintigaraphy
Dr. Mustafa Zuhair Mahmoud Mr.Ali B Alhailiy بسم الله الرحمن الرحيم 1.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
QC/PACS Artifact Identification. Artifact Causes Image receptor/Hardware –Dirty –Foreign material –Failures Software –Algorithm –Enhancements/manipulations.
FILM ARTIFACTS.
Optical Density and Brightness
Unsharpness Calculations and Resolution By Professor Stelmark.
3D CT Image Data Visualize Whole lung tissues Using VTK 8 mm
Radiography Introduction. Objectives To describe  Properties of x-rays  Production of x-rays  Formation of radiographic image  Components of an x-ray.
Basic quality control tests in medical radiography
Quality Assurance/Quality Control Chapter 10. Quality Assurance Definition A system of activities whose purpose is to provide assurance that the overall.
Digital Radiography Chapter 22. History of Digital Radiography Slower process of conversion because no pressing need to convert to digital radiography.
CR Image Acquisition By Professor Stelmark.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Digital Image Characteristic
CT Quality Control for CT Scanners. Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends.
Unit III Creating the Image. Unit III Creating the Image.
Chapter 6: Digital Radiographic Imaging
RT Wk 14 Digital Artifacts & Imaging ERRORS
Computers in Imaging Q & A Robert Metzger, Ph.D..
Quality Control.
The Digital Image Dr. John Ryan.
Optimisation Strategies Ali B alhailiy
Factors affecting CT image RAD
Computed Radiography (CR), & Digital Radiography (DR) Images
Computed Radiography By Prof. Stelmark. Presently, an acceleration in the conversion from screen-film radiography (analog) to digital radiography (DR)
Digital Radiographic Image Processing and Manipulation Chapter 7.
Elsevier items and derived items © 2009 by Mosby, Inc., an affiliate of Elsevier Inc. Digital Image Artifacts Chapter 31.
Preparing Variable kVp Technique Charts By Prof. Stelmark.
COMP322/S2000/L171 Robot Vision System Major Phases in Robot Vision Systems: A. Data (image) acquisition –Illumination, i.e. lighting consideration –Lenses,
Sensitometry By Prof. Stelmark. The study of the relationship between the intensity of exposure of the film and the blackness after processing is called.
Radiographic Quality Visibility and Sharpness
Image Quality Q&A. RAPHEX Diagnostic Question 2001 D9: The imaging system which is best for visualizing small high contrast objects is: A. Computed tomography.
More digital reading explaining LUT RT 244 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of.
Quality Assurance in the Dental Office
RADIOGRAPHIC TECHNIQUE - I
Objectives Identify the image quality characteristics that apply to all medical imaging modalities Understand the concept of image optimization Review.
More digital 244 wk 12 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of Medicine Atlanta, GA,
CR Image Acquisition By Professor Stelmark.
Grid Performance By Prof. Stelmark.
Digital Artifacts Edge Enhancement High contrast and edge definition –Lose fine detail on the edges of high contrast areas.
DIGITAL IMAGING.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Digital Radiographic Image Processing and Manipulation Chapter 7.
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Thursday Case of the Day What is the source of the artifact in the upper right-hand quadrant of the image? History: An AP view of the pelvis was acquired.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Physics Case of the Day History : Patient presented for a follow-up mammogram. The cranio-caudal view was suboptimal due to very low contrast and was repeated.
History of digital radiology 1980 Kinos L.T. ET all developed a portable radiographic X-ray camera made of tantalum and aluminum. Digital radiography.
B ASIC P HYSICS OF D IGITAL R ADIOGRAPHY By : Maisa Alhassoun Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS &
{ Digital Image Processing By Professor Stelmark.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Technique Guidance Systems By Prof. Stelmark. Anatomic Programming Anatomic programming, or anatomically programmed radiography (APR), refers to a radiographic.
Control of Scatter Radiation Beam Restricting Devices By Prof. Stelmark.
Chapter 10 Digital Signal and Image Processing
Pediatric Radiology By Dr. Wambani, J.S. Chief Radiologist
Digital Artifacts.
Mobile Fluoroscopy Based on:
Digital Radiographic Image Processing and Manipulation
Digital radiography.
Presentation transcript:

Digital Image Manipulation I By Professor Stelmark

Image Histogram Digital radiographic image histograms are very important for digital image production. However, they can be the source of bothersome digital radiographic image artifacts if they are not properly understood and manipulated. All digital radiographic imaging systems have the ability to evaluate the original image data through histogram analysis. A histogram is a plot of the frequency of appearance of a given object characteristic.

Simulated chest radiograph shows areas of lung and tissue that are unexposed (collimated) or fully exposed (raw x-ray beam)

PA chest image histograms and averaging the value of each frequency interval, a representative histogram is produced for each image receptor. The histogram can be regularly updated from newer images. This places an additional responsibility on the radiographer. In addition to selecting technique, the radiographer must engage the appropriate histogram before examination so as to apply the appropriate reconstruction algorithm to the final image.

Collimation and Partition If the x-ray exposure field is not properly collimated, sized, and positioned, exposure field recognition errors may occur. These can lead to histogram analysis errors because signal outside the exposure field is included in the histogram.

Collimation of the projected area x-ray beam is important for patient radiation dose reduction and for improved image contrast in screen-film radiography. In DR, proper collimation has the added value of defining the image histogram. If improperly collimated, the histogram can be improperly analyzed.

Digital image receptors normally can recognize even-numbered (i.e., two or four) x-ray exposure fields that are centered and cleanly collimated. Three on one and four on one are not recommended unless the unexposed portion is shielded.

For the image histogram to be properly analyzed, each collimated field should consist of four distinct collimated margins.. The use of three collimated margins usually works, but when fewer than three are used, artifacts may result. If images are not collimated and centered, image receptor exposure will not be accurate and cannot be used for image quality evaluation.

If multiple fields are projected onto a single IP, each must have clear, collimated edges and margins between each field. This process, called partitioning,

Alignment Alignment of the exposure field on the IP is important in the same way and for the same reason as collimation.

Preprocessing Before an image is prepared “for processing,” several manipulations of the output of an image receptor may be necessary to correct for potential artifacts. Such artifacts can occur because of dead pixels or dead rows or columns of pixels Preprocessing of digital images is largely automatic.

Preprocessing is designed to produce artifact-free digital images. In this regard, preprocessing provides electronic calibration to reduce pixel-to-pixel, row-to-row, and column-to-column response differences. The processes of pixel interpolation and noise correction are automatically applied with most systems.

A single pixel or a single row or column normally will not interfere with diagnosis. However, many of these defects must be corrected. Correction algorithms specific to each type of digital image receptor use interpolation techniques to assign digital values to each dead pixel, row, or column. Interpolation is the mathematical process of assigning a value to a dead pixel based on the recorded values of adjacent pixels.

Flatfielding is a software correction that is performed to equalize the response of each pixel to a uniform x-ray beam. Exposure to a raw x-ray beam shows the heel effect on the image. B, Flatfielding corrects this defect and makes the image receptor response uniform.

Computed radiography cassettes are highly sensitive to background radiation and scatter. If a CR cassette has not been used for several days, it should be inserted into the reader for re-erasure. The practice of leaving cassettes in a supposedly “radiation-safe” area in an x-ray room during an examination must be discouraged.