Lecture 9 Yuri Kazakevich Seton Hall University

Slides:



Advertisements
Similar presentations
Chem. 231 – 3/4 Lecture. Announcements I Set 1 Lab Reports –Both Lab Reports due 3/13 –Only one of the lab reports needs to be detailed, the other can.
Advertisements

Chromatography Integration and Trouble-shooting
Compact & Perfect for Your Satisfaction
1 HPLC Columns and Stationary Phases Lecture 2 Yuri Kazakevich Seton Hall University.
Practical HPLC. 2 In This Section, We Will Discuss: How to set up an HPLC System for a sample injection including:  Solvent Handling  Mobile Phase preparation.
KROMATOGRAFİ Sedat Türe. HPLC Liquid Chromatography.
Reversed Phase HPLC Mechanisms Nicholas H. Snow Department of Chemistry Seton Hall University South Orange, NJ 07079
High Performance Liquid Chromatography. HPLC originally refered to: High Pressure Liquid Chromatography currently refers to: High Precision Liquid Chromatography.
Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.
1 Chapter 25 HPLC High-Performance Liquid Chromatography.
Chromatography.
Analyte Ionization Lecture 3 Yuri Kazakevich Seton Hall University
Chem. 31 – 4/8 Lecture. Announcements I Exam 2 – Monday –Covering Ch. 6 (topics since exam 1), 7, 8-1, 17, and parts of 22 (up to and including retention.
Chem. 133 – 5/5 Lecture. Announcements Lab Report 2.4 due Thursday – can turn in today for reduction of late penalties Term Project Progress Report –
1 Salt Effects Lecture 5 Yuri Kazakevich Seton Hall University.
1 Introduction HPLC Process Lecture 1 Yuri Kazakevich Seton Hall University.
Mobile Phase pH Analyte pK a Shift Lecture 4 Yuri Kazakevich Seton Hall University.
High Performance Liquid Chromatography. HPLC originally refered to: High Pressure Liquid Chromatography currently refers to: High Precision Liquid Chromatography.
CHAPTER 29 Supercritical Fluid Chromatography The mobile phase is a supercritical fluid (a fluid above its critical T and critical pressure) Supercritical.
Analytical Chemistry Section D Separation Techniques.
Chapter 6 - Chromatography
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC). When particles of small diameter (microns) are used as a stationery phase support, the technique is called.
Created with MindGenius Business 2005® Performance parameters (Advanced) Theoretical Plate Model Performance parameters (Advanced) Theoretical Plate Model.
Gas Chromatography Carrier Gas Flow Control Injector Port Column Column Oven Detector Recorder The GC system consists of gas supplies for the mobile phase.
High Performance Liquid Chromatography
Chromatography Chapter 6.
1.1 General description - Sample dissolved in and transported by a mobile phase - Some components in sample interact more strongly with stationary phase.
High Performance Liquid Chromatography (HPLC) Jeannette Comeau CHEE April 2004.
GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5.
High Performance Liquid Chromatography Instrumentation.
History of Chromatography n Early LC carried out in glass columns n diameters: 1-5 cm n lengths: cm n Size of solid stationary phase n diameters:
900-LC P. van Poppel – Field Support Specialist October 2009.
DEVELOPMENT OF A RP-HPLC METHOD FOR THE DETERMINATION OF METFORMIN IN HUMAN PLASMA.
HPLC – High Performance Liquid Chromatography
Introduction  High-performance liquid chromatography (HPLC) is a form of liquid chromatography.liquid chromatography  The main purpose is to separate.
1 HPLC Lecture Displacement pumps Displacement pumps, on the other hand, is composed of a one directional motor driven plunger that pushes the mobile.
Best Broken into four categories
Created with MindGenius Business 2005® Instrumentation (1) Instrumentation (1) Origins Preparative column chromatography (Glass columns) Internal diameter:
Chromatography High Performance Liquid Chromatography HPLC Chapter Dr Gihan Gawish.
HPLC.
Waters HPLC Competitive Review
High Performance Liquid Chromatography
Chapter 28 High Performance Liquid Chromatography.
HPLC.
Why HPLC? Almost universal applicability Remarkable precision Highly commercially available (competition)
HPLC 1. Introduction 1.Introduction  INSTUMENTAL ANALYSIS  PRACTICAL 213 PHC  HPLC.
Comparison of the Varian 920-LC with Agilent 1200 Series HPLC.
Lecture 8 Peak Parameters and Quantitative chromatography
Chem. 133 – 5/3 Lecture. Announcements Lab – Term Project Progress Report Due Today – Last Assignments: Term Project Poster and Peer Review Grading (Friday,
Instrumental Analysis (I) HPLC Tutorial 8. Graded presentation Students in groups of 4-5 individuals are asked to prepare a presentation (weight=5% of.
HPLC TROUBLESHOOTING. There are five major of symptoms to help you quickly identify the source of the problem(s): I.Pressure abnormalities II.Leaks III.Problems.
Instrumental Analysis (I)  HPLC Tutorial #7 PHCMt561 – 5 th Sem. Pharm.
Experiments in Analytical Chemistry
1.1 General description - Sample dissolved in and transported by a mobile phase - Some components in sample interact more strongly with stationary phase.
High Performance Liquid Chromatography. What is HPLC ? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube.
High Performance Liquid Chromatography Presented by Dr. Kamal Modi 2 nd Year Resident.
Gel Permeation Chromatography Size Exclusion Chromatography
Chromatography- TLC & HPLC
High Performance Liquid Chromatography (HPLC)
High Performance Liquid Chromatography HPLC
Size Exclusion Chromatography (SEC)
Horizon Chromatography
Kromasil Eternity™ Designed for long life
Best Broken into four categories
High Performance Liquid Chromatography
HPLC TROUBLE SHOOTING Dr. Mirza Shahed Baig.
High Performance Liquid Chromatography
Liquid Chromatography - Method Development and Validation
33A-2 Pumping Systems The requirements for liquid chromatographic pumps include the generation of pressures of up to 6000 psi (lb/in2), (2) pulse-free.
Transferring LC-UV to LC-MS.
Presentation transcript:

Lecture 9 Yuri Kazakevich Seton Hall University System and Method Troubleshooting Yuri Kazakevich Seton Hall University

Troubleshooting Locate the problem by ranking possible causes. There is no standard troubleshooting procedure. General Pattern: Locate the problem by ranking possible causes. Verify the presence of the most probable cause. If present – fix the problem, otherwise verify the existence of the next possible cause. First try to distinguish System problem or Method problem

Method vs. System Troubleshooting System Parameters Flow stability Backpressure Clogging Detector problems Injection siutability Injection volume Temperature Method Parameters Flow rate Eluent type Eluent composition pH pH modifier (type) Injection volume Temperature Gradient profile

System Parameters Solvent Degasser Pump Autosampler Column Detector Simple preliminary verification of system setup can save time. Solvent Degasser Pump Autosampler Column Detector Bottle fill-in Inlet filter date Flush if solvent change >15 mL Backpressure Flow stability Check-valves Vial fill-in connections cross-contamination Column type connections Wavelength Cell volume multiple detectors connection sequence Sampling rate Critical connections. Minimize tubing length

Available HPLC system set margins for column selection. System Suitability Available HPLC system set margins for column selection. 20 ml detector flow-cell incompatible with <3 mm I.D. columns 10 ml sample loop incompatible with <1 mm I.D. columns. 0.2 ml micro-injector is useless for conventional columns. Suitability Rule Injection volume < Cell volume Column Dead Volume » 0.65 of the empty column volume

System Suitability (Injection Volume) Column: 150 x 4.6 mm (C18), Vo = 1.7 ml Efficiency: 10,000 t.p. Eluent: MeCN/Water 70/30 VR(benzene)=2.2 ml; VR(benz-a-pyrene)=14.6 ml wbenzene=88 ml; w(b-a-p) = 584 ml 100 ml injection 8 ml injection

Effect of flow-cell volume and sampling rate Response time effect Flow cell volume Cell volume

HPLC System set up Minimize the volume and connections between autosampler, column, and detector. No guard, no prefilter

Tubing & connections

Unions

Critical Connections Injector - Column, Column - Detector No unions, in-line filters, guard columns Single piece tubing (0.007” and smaller I.D.)

Eluent Composition Effects on the Column Back Pressure

Guard Columns Guard column always decreases system efficiency Purpose - trapping retentive impurities Disadvantage - introduces extra-connections in critical zone Sample has 1% impurity. How many injections will kill 1% of column surface with 1% sample solution and 10 ml injection volume? 1% column surface ~ 2-3 m2, it could adsorb ~ 0.1 mMole 300 injections will reach this level. Guard column always decreases system efficiency

Autosampler – Column/Pump Connections Wrong connection Correct connection

Waters system (Injection, Drawing Sample) Lecture 9 Waters system (Injection, Drawing Sample) Mobile phase flow with sample drawn into loop

Waters system (Injection, Injecting Sample) Lecture 9 Waters system (Injection, Injecting Sample) Sample is swept from the loop and into the flowing stream.

Sample Diluent Effect 50/50 Buffer/MeOH Buffer: 20 mM Citrate, pH=4.6 1- 50/50 MeOH/Water 2 - 80/20 MeOH/Water 3 - 90/10 MeOH/Water 4 - 95/5 MeOH/Water 5 - 100 MeOH 50/50 Buffer/MeOH Buffer: 20 mM Citrate, pH=4.6 1 2 3 4 5 Incompatible solvents may cause sample precipitation and column clogging Different eluent pH and composition may cause peak splitting

Column Length Column length is a compromise between the efficiency and backpressure Column efficiency is proportional to the column length Specific efficiency (# of particles per one plate) decreases with length increase.

Column Overloading 1 ml 5 ml

Effect of pH on Aniline UV absorbance Lecture 9 Effect of pH on Aniline UV absorbance The mobile phase pH at a constant organic composition may have an effect on an ionizable analyte’s UV response. At 232 nm there is a decrease in aniline’s absorbance as this analyte becomes progressively more ionized. A plot of the UV absorbance at a particular wavelength versus the sspH of the mobile phase will lead to a sigmoidal dependence. The inflection point corresponds to the analyte pKa. Abs. sspH 5.2-9.2 sspH 4.2 sspH 3.2 sspH 1.2-2.2 Chromatographic Conditions Column: 15 cm x 0.46 cm Luna C18(2) Eluent: 90% Aqueous:10% MeCN Aqueous: 15 mM K2HPO4•7H2O adj. to wwpH 1 - 9 with H3PO4 Flow rate: 1 ml/min Temp: 25oC Wavelength (nm)

Effect of pH on Aniline Retention and UV response (220 nm) Lecture 9 Effect of pH on Aniline Retention and UV response (220 nm) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Detector Response (mV) wwpH 2 wwpH 4 wwpH 5 wwpH 6 wwpH 9 Time (min.) Chromatographic Conditions Column: 15 cm x 0.46 cm Luna C18(2) Eluent: 90% Aqueous: 10% MeCN Aqueous: 15 mM K2HPO4•7H2O adj. to wwpH 1.5 - 9 with H3PO4 Flow rate: 1 ml/min Temp: 25oC Enhanced sensitivity is obtained by analyzing aniline in its neutral state

Column Equilibration Column equilibrates within 30 min in normal eluent composition range. Check retention time stability by injecting standard mixture 3 - 4 times. Very high organic (>98%) or very high aqueous (>80%) need ~1 - 2 h equilibration at 1 ml/min. In pure water after ~20 h equilibration all analytes elute with void volume. “Chain collapse”? - No. After 20 h of water pumping all organic removed from adsorbent pores. Water is not wetting the alkylated hydrophobic surface. There is no flow through adsorbent particles, only around.

Solvent Purity n (moles) = How much solvent (0.1 ppm total impurity) will contaminate 10% of adsorbent surface? Average column - 200 m2/g Assume molecular area of 100 Å2 n (moles) = Assume average 100 g/mole - 3 mg total accumulation this comes from 30 L of solvent with 0.1 ppm total purity Column has to be cleaned at least once a week

Gradient High pressure vs. low pressure mixing System dwell volume effect J.Dolan, LC-GC V.16 #1, 16

Column Cleaning Solvent front disturbs phase equilibrium Release of trapped impurities

Method troubleshooting Problems are usually related to one of the following: System Column Sample Mobile Phase

System System-to-system compatibility Differences in configuration (detector sequence, etc.) Different dwell volume Detector sensitivity always different Wavelength accuracy Bandwidth Environment effects

Sample Avoid particulate in the sample Filter Centrifuge Sample vials Lecture 9 Sample Avoid particulate in the sample Typical cause of inlet filter clogging Filter Centrifuge MD or Qualitative analyses. (Sample filtration can change composition) Usually cumbersome Sample vials Type of the vial cap and septa affect contamination and carry-over Waters systems require 75% filling of 2 mL vial

Troubleshooting sequence Pump Any reciprocal pattern on chromatogram Pressure fluctuations Baseline drift (possible contamination of the solvent) Autosampler Injection marks (baseline disturbance) Cross-contamination Vial fill-in (sample level) Detector Response (baseline noise, drift, etc.) Wavelength (bandwidth, accuracy, etc.)

Troubleshooting sequence First check is always the plumbing (leak, flow rate, pressure) Output (chromatogram) evaluation 1 2 3 1- flow or detection problem 2 – possible injection problem 3 – correct chromatogram

Troubleshooting sequence Analysis of chromatogram Compare with previous results Peak tailing Retention shift Reverse elution