Chapter 7 Components of Optical Instruments. Typical spectroscopic instruments contain five components: (1) a stable source of radiant energy, (2) a transparent.

Slides:



Advertisements
Similar presentations
Chapter 25 Instruments for Optical Spectroscopy
Advertisements

Raman Spectroscopy A) Introduction IR Raman
Chapter 7 Components of Optical Instruments
11 Components of Optical Instruments Lecture Spectroscopic methods are based on either: 1. Absorption 2. Emission 3. Scattering.
 PART Absorption Spectrometer Dr. S. M. Condren SourceWavelength SelectorDetector Signal Processor Readout Sample.
Components of Optical Instruments or What’s inside that spectrometer?
Molecular Fluorescence Spectroscopy
1 Molecular Luminescence Spectroscopy Lecture 31.
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Chem. 133 – 4/1 Lecture. Announcements I Lab –Returning Term Project Proposals (I expect to be getting boxes of supplies ready) –Returning Set 2 Period.
Fluorescence. Topics Sensitivity (nM-pM) –Contrast UV-vis measurements Derivatization Laser-Induced Fluorescence.
06/02/2008CCDs1 Charge Coupled Device M.Umar Javed M.Umar Javed.
Instrumental Chemistry
Lecture 5 An Introduction to Spectroscopy Electromagnetic radiation, electromagnetic wave Emission, absorption, fluorescence.
Atomic Absorption Spectroscopy Elemental analysis Dissolved metals Source –Hallow Cathode Lamp –Cathode of specific element Sample –Acetylene-air flame,
Lecture 3 INFRARED SPECTROMETRY
4-1 Chap. 7 (Optical Instruments), Chap. 8 (Optical Atomic Spectroscopy) General design of optical instruments Sources of radiation Selection of wavelength.
What Are Some Types of Spectroscopy ?
Atomic Emission Spectroscopy
Components of Optical Instruments, Cont… Lecture 8.
Chapter 25 Instruments for Optical Spectrometry
What is spectroscopy? It is the study of how matter interacts with electromagnetic radiation (gamma rays down to radio waves) Matter can interact with.
5 Components Common to All Optical Spectrometers Source Transparent Sample Holder Wavelength Selector Radiation Detector Signal Processor and Readout.
Principles of instrumentation Prepared by: Ibtisam H. AlAswad Reham S. Hammad.
Chapter 10 Atomic Emission Spectrometry
Atomic Emission Spectrometry By: Alexa Kunch
RamanRaman. Scattering Tyndall scattering – if small particles are present During Rayleigh scattering (interaction of light with relatively small molecules)
Dispersive spectrometers measure transmission as a function of frequency (wavelength) - sequentially--same as typical CD Interferometric spectrometers.
1 Components of Optical Instruments Lecture Silicon Diode Transducers A semiconductor material like silicon can be doped by an element of group.
Spectrophotometer Prof.Dr. Moustafa M. Mohamed Vice Dean Faculty of Allied Medical Science Pharos University in Alexandria, EGYPT.
1 Atomic Emission Spectroscopy Lecture Advantages of Plasma Sources 1.No oxide formation as a result of two factors including Very high temperature.
Optical Characterization methods Rayleigh scattering Raman scattering transmission photoluminescence excitation photons At a glance  Transmission: “untouched”
Fluorometric determination of riboflavin. Introduction.
Fluorometric Analysis
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 19 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Chapter 13 An Introduction to Ultraviolet- Visible Molecular Absorption Spectrometry.
Chapter 12 Infrared Spectroscopy Jo Blackburn Richland College, Dallas, TX Dallas County Community College District  2006,  Prentice Hall Organic Chemistry,
Spectrophotometry.
Introduction to Spectrochemical Methods
Atomic Fluorescence Spectroscopy. Background l First significant research by Wineforder and Vickers in 1964 as an analytical technique l Used for element.
Colorimeters or photometers
Chapter 3 Components of optical instruments
Chapter 9 Atomic Absorption and Atomic Fluorescence Spectrometry.
Spectrophotometer.
CLS 332 CLINICAL INSTRUMENTAL ANALYSIS. A VISIBLE ABSORPTION SPECTROMETER.
Molecular Luminescence Spectroscopy
Chapter 10 Atomic Emission Spectrometry
Instruments for Optical Spectrometry
hn1 hn2 optical l selector source sample detector signal processor
Lab1 A VISIBLE ABSORPTION SPECTROMETER. -One of the simplest and most widely used methods to determine concentration of a substance in solution -Measures.
Principles of instrumentation. Photometry - Photometry means “the measurement of light” If a substance can be converted to a soluble, colored material,
GammaX-rayUVVisIRMicroRadio Short wavelength High energy High frequency Long wavelength Low energy Low frequency Solvent effects: Neat: has fine structure.
Chem. 133 – 3/16 Lecture.
Lab1 A VISIBLE ABSORPTION SPECTROMETER
Luminescence spectroscopy
Components of Optical Instruments
Chem. 133 – 3/14 Lecture.
Optical Characterization Techniques
Chapter 5: Analytic Techniques
Colorimeters or photometers
Spectrophotometer Dr . S. Jayakumar.
Chapter 3. Components of Optical Instruments
Measurements involving light –A Basic Principles and Instrumentation
Intro to UV-Vis Molecular Absorption Spectrometry
Atomic Absorption and Atomic Fluorescence Spectrometry
Satish Pradhan Dnyanasadhana College, Thane
INTRODUCTION TO SPECTROSCOPIC METHODS OF ANALYSIS
Colorimeters OR photometers
Raman Spectroscopy A) Introduction IR Raman
Clinical instrumental analysis
Presentation transcript:

Chapter 7 Components of Optical Instruments

Typical spectroscopic instruments contain five components: (1) a stable source of radiant energy, (2) a transparent container for holding the sample, (3) a device that isolates a restricted region of the spectrum for measurement, (4) a radiation detector that converts radiant energy into a signal detector, (5) a signal processor and readout.

Components of Optical Instruments  Absorption  Fluorescence  Phosphorescence  Scattering  Emission  Chemiluminescence

Rayleigh & Raman Scattering  Occur when the dimensions of the particles that cause the scattering are small in comparison to the wavelength of the incident radiation. Dissolved particles can result in Rayleigh and Raman scattering.  A type of Raman Spectrometer: FRA 106/S Spectrometer

Sources of Radiation Requirements  Sufficient power  Stability over long periods of time  Voltage regulation required as radiant power varies exponentially with voltage

Lasers

 Pumping  Spontaneous emission (fluorescence)  Stimulated emission  Absorption

Wavelength Selectors Filters  Interference Filters  Interference Wedges  Absorption Filters

Wavelength Selectors Monochromators- one color - pass a narrow band of wavelengths The SURE_SPECTRUM is an imaging spectrograph and scanning monochromator that features dual exit ports for maximum flexibility.

Wavelength Selectors View of inside of monochromator

Wavelength Selectors

Radiation Transducers Photomultiplier

Radiation Transducers Photodiode Arrays (PDA)

Radiation Transducers Charge Transfer Devices  Charge Injection Devices (CID)  Charge-Coupled Devices (CCD)

Signal Processors & Readouts Photon Counting Advantages:  Improved signal-to-noise ratio  Sensitivity to low radiation levels  Improved precision for a given measurement time  Lowered sensitivity to photomultiplier tube voltage and temperature fluctuations  Detection method of choice in fluorescence, chemiluminescence, and Raman spectrometry

Signal Processors & Readouts Photon Counting Disadvantages:  Required equipment is more complex and expensive  Technique has not been widely applied for routine molecular absorption measurements in ultraviolet and visible regions

Principles of Fourier Transform Optical Measurements  Transforms data set from time domain to frequency domain  Advantages  Throughput  High resolving power  Interferometers  Michelson  Mach-Zender  Fabry-Perot