1 Detection methods of IgE : a critical view Allergy School 21st-24th September 2007 Laurence Guilloux.

Slides:



Advertisements
Similar presentations
Enzyme Linked Immunosorbent Assay
Advertisements

Radioimmunoassay (RIA)
Analytical Chemistry.
ELISA Enzyme Linked Immunosorbent Assay. Definitions  Antibodies (also known as immunoglobulins abbreviated Ig) are gamma globulin proteins that are.
Clinical Microbiology and Immunology 1 36 Copyright © McGraw-Hill Global Education Holdings, LLC. Permission required for reproduction or display.
Ion Exchange Laboratory. Today’s Schedule Pre-lab discussion Ion Exchange and Spectrophotometer Ion exchange experiment.
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download, adapt, and distribute this.
1 Lecture: Forensic Serology - Immunoassays Antibody/Antigen reaction provides the means of generating a measurable result. “Immuno” refers to an immune.
Part Three Basic Test Methods
World Health Organization
Module Two Labeled Test Methods
Dr Samah Kotb Lecturer of Biochemistry 1 CLS 432 Dr. Samah Kotb Nasr El-deen Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Nasr.
Laboratory Quality Control
Summary 1 l The Analytical Problem l Data Handling.
Measurement in Science Bioassays and Immunoassays Karim Meeran 23 October 2009.
Introduction to Immunoassays
Biomedical Tracers Biology 685 University of Massachusetts at Boston created by Kenneth L. Campbell, PhD.
Immunology vs. Serology Immunology –Study of immune system Serology –Detecting/measuring elements of humoral immune system (I.e., antibodies) –Help diagnose.
Enzyme-linked Immunosorbent Assay
CHEM-3245 Quantitative Analysis
Thoughts on Biomarker Discovery and Validation Karla Ballman, Ph.D. Division of Biostatistics October 29, 2007.
Protein arrays LEAPS technology An array of 110 different antibodies incubated with various levels of the fluorescently labelled cognate antigens in a.
MLAB Coagulation Keri Brophy-Martinez
QUALITY CONTROL OF PHYSICO-Chemical METHODS Introduction :Validation توثيق المصدوقية.
Kyiv, TRAINING WORKSHOP ON PHARMACEUTICAL QUALITY, GOOD MANUFACTURING PRACTICE & BIOEQUIVALENCE Validation of Analytical Methods Used For Bioequivalence.
(Enzyme Linked Immunosorbent Assay)
Application of immunological tests
Quality Assessment 2 Quality Control.
Validation of Analytical Method
Allergen regulation in the future : what will be the place for recombinant allergens ? Jacqueline DAYAN-KENIGSBERG European Academy of Allergology and.
Introduction to Analytical Chemistry
Principles of ELISA. Immunoassay Immunochemical reactions form the basis of a diverse range of sensitive and specific clinical assays. In a typical immunochemical.
Enzyme-Linked Immunosorbent Assay
1 CLINICAL CHEMISTRY CHAPTER 6 IMMUNOASSAYS. 2 Introduction –In the last chapter, we discussed a variety of analytical techniques –In this chapter we’ll.
Analytical considerations
Analytical chemistry MLAB 243 Level 4 Lecture time: every WED 8 -10
Introduction to Analytical Chemistry Dr M. Abd-Elhakeem Faculty of Biotechnology General Chemistry Lecture 7.
Quality Control Lecture 5
Definitions  Antibodies (also known as immunoglobulins abbreviated Ig) are gamma globulin proteins that are found in blood and are used by the immune.
IMMUNOASSAYS Fundamental Questions for an Analytical Chemist How do you rapidly, cheaply and easily detect a single analyte present in a complex heterogeneous.
1 Immunoassay Testing Forensic Toxicology. 2 Introduction Antibody/Antigen reaction provides the means of generating a measurable result. “Immuno” refers.
Students should be able to: 1. Define and differentiate the following terms: Qualitative analysis, Quantitative analysis & Analytes. 2. Define the role.
The Enzyme Linked Immunosorbent Assay (ELISA).. Capture ELISAs Antigen Capture: In this, more specific approach, a capturing Ab is adsorbed onto the solid.
LABORATORY METHODS BASED ON ANTIGEN-ANTIBODY INTERACTIONS I
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
Innovative Paths to Better Medicines Design Considerations in Molecular Biomarker Discovery Studies Doris Damian and Robert McBurney June 6, 2007.
What is an antigen? An antigen is any substance that elicits an immune response and is then capable of binding to the subsequently produced antibodies.
Enzyme linked immunosorbent assay
Developing Real World Tests and Assays Using Gold Particles Arista Biologicals Inc.
QC/QA.
 Routine viral diagnostics: indirect and direct detection of viruses. ◦ Indirect detection: serological tests; ◦ Direct detection:  Viral antigens;
Enzyme Linked Immunosorbent Assay
ELISA Enzyme Linked Immunosorbent Assay Fariba mazrouei.
Chapter 1: The Nature of Analytical Chemistry
Agglutination Part 2.
Establishing by the laboratory of the functional requirements for uncertainty of measurements of each examination procedure Ioannis Sitaras.
이 장 우. 1. Introduction  HPLC-MS/MS methodology achieved its preferred status -Highly selective and effectively eliminated interference -Without.
Analytical Chemistry.
Mass Spectrometry Vs. Immunoassay
Screening library As a well-recognized expert at phage display technology, Creative Biolabs offers our worldwide clients the best phage display library.
Practical clinical chemistry
ANTIGEN-ANTIBODY REACTIONS
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
ANTIGEN-ANTIBODY REACTIONS
And antibody based techniques
Satish Pradhan Dnyanasadhana College, Thane. Department of Chemistry S
RADIOIMMUNOASSAY (RIA)
Enzyme Linked  Immuno sorbent  Assay  
Experimental Systems and Methods
In vitro assays for the diagnosis of IgE-mediated disorders
Presentation transcript:

1 Detection methods of IgE : a critical view Allergy School 21st-24th September 2007 Laurence Guilloux

2 Detection method of IgE  The diagnostic algorithm for human allergic disorders begins with A thorough clinical history And a physical examination  Temporal associations between allergic symptoms and allergen exposures lead to a high degree of suspicion that the patient has an allergic disorder.  Confirmatory testing for IgE antibody (AB) is often performed to strengthen the probability that the working diagnosis is correct. This involve in vivo methods (skin, provocation tests) Laboratory-based in vitro analysis  Cell-based technology  Serum-based technology

3 Serum-based technology History  1960, identification of human « reagin » as IgE  Preparation of IgE Fc fragment and production of polyclonal AB specific for the  heavy chain  1967, first total IgE and allergen-specific IgE RAST were developed (Wide, Lancet)  1974, first generation semi-quantitative IgE AB assay (birch specific IgE calibration curve and results expressed in classes) was first made commercially available by Pharmacia Laboratories.  Over the past decade there have been improvements in all aspects of technology with Second-generation quantitative IgE assays Third-generation automated IgE assays with a detection limit of 0.1 kU/L (Li CCA 2005) Biochip technology, microarrays (Wiltshire, Clin Chem 2000; Kim, Exp Mol Med 2002)

4 The two most representative classes of immunoassays for the measurement of specific IgE are based on the use of  Solid phase allergens  Liquid phase allergens (the fluid phase allergen is bound to the solid phase only after the allergen has complexed with fluid phase IgE AB). Today The assay of specific IgE AB is considerably more complicated than most other immunoassays  Firstly, the concentration of the IgE in the blood is extremely low, (0.004 mg/100ml; IgG: 1200 mg/100ml)  Secondly, each main allergen, contains a large number of different allergenic components,  The assay must be sensitive enough to capture IgE AB to all relevant components but still with a very low level of non-specific binding.

5 « Architecture » of specific IgE immunoassays 1. Allergen-containing reagent and binding chemistry 2. Anti IgE 3. Calibration system 4. And also the tracer and the signal detection system influence the sensitivity and quality of final results The analytical performances of immunoassays depend on the whole « architecture » of the method and particularly on some fundamental variables.

6 1. Allergen-containing reagent and binding chemistry  The solid-phase (allergosorbent) or liquid-phase labelled allergen reagent is the principal component of the assay that confers specificity on the IgE AB assay.  The most complex and highly variable component partly due to: 1-1) the heterogeneity of most allergen extracts 1-2) the different supports 1-3) and the different chemical methods used to insolubilise or label the allergenic material

7 1-1) Heterogeneity of « allergens » Whole extract Heterogeneity has many possible causes, even when the same manufacturer uses; the same extract source (variable stability, inherent biological variation of the raw material, contamination…) the same chemical and manufacturing procedures (extraction process, variable binding of allergen, variable stability….) Purified native and recombinant allergenic molecules The advent of these allergenic molecules offers new perspectives: rAllergens produced in almost unlimited quantities will avoid the variability of natural allergenic extracts and allow standardisation (but different of the native form). Allergenic molecules are essential for the development of innovative techniques using microchip technologies and to improve our understanding of problems related to cross-reactivity of IgE AB to allergens. Spiking of allergen extract with purified native or recombinant allergenic molecules (known to be clinically relevant), can improve the sensitivity of the diagnostic tests.

8 1-2) Different supports Solid phase allergens (allergosorbent)  Insoluble dextran  Cellulose filter paper disc (RAST) Minicolumns (CAP) Thread (CLA)  Mycrocrystalline particles  Nitocellulose (dot blot)  Polystyrene tube (FAST)  Microtiter well activated silica chip (ISAC) Liquid phase allergens  Biotin-labelled allergens (Advia Centaur)  The ligand (biotin) is covalently attached to a backbone (soluble polymer and amino acid copolymer) to wich the allergen molecules are also attached (Immulite 2000)

9 1-3) Different chemical methods Characteristics CNBr CNBr bounds only NH 2 bearing molecules, iso urea N-substituted liaisons are not stable, iso urea group is protonated at neutral to alkaline pH, which implies a positively charged matrix-CNBr-allergen which can bind in a non specific manner negatively charged sIgE. Weak binding capacity (large amount of bound allergen that may increase NSB and IgG interference), Slow kinetics Bethell J Chromat 1981 CDI, BDE Several coupling agents that allow allergen coupling by different groups (NH 2, SH, OH) Neutral matrix-allergen binding Better allergen representation, no masked epitopes, conformation close to their native states.

10 Highly specific for unique determinants on  heavy chains  Used as both capture and detection reagents  Subjected to chemical modification (radio-labelled, enzyme-labelled, chemical and physical immobilisation on solid-phase matrices  Initially polyclonal reagents  Monoclonal reagents are specific and lack batch-to-batch variations, however as they have low affinity and avidity, mixtures of mono and polyclonal antisera are often used 2. Anti - IgE 3. Calibration system (NCCLS 1997) Homologous interpolation methods This is a term that defines a calibration scheme in which the standard or reference (calibration) dose-response curve is constructed using reagents that have the same specificity from those being used to measure the analyte of interest. Heterologous interpolation methods This is a term that defines a calibration scheme in which the standard or reference (calibration) dose-response curve is constructed using reagents that have a different specificity from those being used to measure the analyte of interest. In the contexte of specific IgE assays, this mean that a total serum IgE calibration curve in IU/ml of IgE, which is calibrated back to an IgE primary standard, is used to interpolate unknown allergen-specific IgE response data.

11 Classification of specific IgE antibody immunoassays Qualitative assays (no calibration curve) Positive/negative based on pre-assigned positive threshold Semi-quantitative assays The variations in the positive signal detected are commonly presented in terms of a series of increasing grades (e.g., I to IV; low to high) or classes in arbitrary units (U/ml) determined relative to a supplier-specific, heterologous dose-response curve or an end-point dilution to which the signal becomes negative (e.g., titer), or in comparison to a qualitative grading scheme (e.g., colour chart) Quantitative assay (multipoint calibration curve) A quantitative assay provides an accurate and reproducible estimate of the concentration of IgE and fulfills the analytical criteria for quantification including: Parallelism, recovery, precision and Linearity across the whole assay’s working range Is dependent on the degree with which the assay result accurately reflects the quantity of sIgE AB in the test specimen, and the assay’s precision requirements (according to NCCLS approved guidelines) The most diffuse and typical illustration is the multiallergen screening test in which different allergen specificities are present in a single reagent The results produced are not traceable to any comon reference material.

12 SQA, semi quantitative assay; QA, quantitative assay SP, solid phase; LP, liquid phase; IM, immunometric assay Past and current serologic assays for allergen-specific IgE

13 Conclusion Excellent analytical performances are generally achieved by third generation laboratory systems for sIgE as CAP System, IMMULITE g and ADVIA Centaur in term of quantitative results. On account of differences in term of kU/L results obtained with the different systems and without the avaibality of the « true » value of sIgE we should keep in mind that the different technics are not interchangeable (thresholds using ROC curves, probability curves are not identical). It’s a general rule for immunoassays (ex. tumor markers). Third generation laboratory systems for sIgE demonstrate acceptably high concordance with objective clinical tests (skin, provocation tests). Discrepant results have to be carefully analysed in term of sensitivity and specificity. Immunoassays for the sIgE AB have undergone significant improvements with, the use of more purified extacts, the capacity to quantitate IgE AB in mass units, and total automation of assay system. However, laboratory information should be totally integrated in rational clinical pathways.