MRI Physics II: Gradients, Imaging Douglas C. Noll, Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor.

Slides:



Advertisements
Similar presentations
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Advertisements

Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Imaging Sequences part I
MRI Phillip W Patton, Ph.D..
In Chan Song, Ph.D. Seoul National University Hospital
Richard Wise FMRI Director +44(0)
BE 581 Lecture 3- Intro to MRI.
PHYSICS OF MAGNETIC RESONANCE
MR Sequences and Techniques
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Statistical Parametric Mapping
Parameters and Trade-offs
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
Principles of MRI: Image Formation
Chapter 9 Basic MRI I Mark D. Herbst, MD, PhD. Notice This lecture contained many drawings on the whiteboard, so get these from one of the other students.
Basic Principles MRI related to Neuroimaging Xiaoping Hu Department of Biomedical Engineering Emory University/Georgia Tech
Magnetic Resonance Imaging
Noll An Overview of Echo Planar Imaging (EPI) Douglas C. Noll, Ph.D. Depts. of Biomedical Engineering and Radiology University of Michigan, Ann Arbor.
Wald, fMRI MR Physics Massachusetts General Hospital Athinoula A. Martinos Center MR physics and safety for fMRI Lawrence L. Wald, Ph.D.
Noll Spin-Warp Imaging For each RF pulse:For each RF pulse: –Frequency encoding is performed in one direction –A single phase encoding value is obtained.
Terry M. Button, Ph.D. Principals of Magnetic Resonance Image Formation.
Relaxation Exponential time constants T1 T2 T2*
Steady-state free precession and other 3D methods for high-resolution FMRI Steady-state free precession and other 3D methods for high-resolution FMRI Karla.
EPI – Echo Planar Imaging Joakim Rydell
Basics of Magnetic Resonance Imaging
Psy 8960, Fall ‘06 Fourier transforms1 –1D: square wave –2D: k x and k y Spatial encoding with gradients Common artifacts Phase map of pineapple slice.
FMRI: Biological Basis and Experiment Design Lecture 7: Gradients and k-space FFT examples –Sampling and aliasing Gradient Gradient echo K-space
MRI Physics I: Spins, Excitation, Relaxation
MRI Physics 3: Hardware Douglas C. Noll Biomedical Engineering University of Michigan.
Psy 8960, Fall ‘06 EPI, Part 2: variants1 Segmentation Partial Fourier Spin echo vs. gradient echo Inversion recovery Long vs. short TE.
MRI Image Formation Karla Miller FMRIB Physics Group.
Medical Imaging Systems: MRI Image Formation
Principles of MRI Physics and Engineering
Imaging Sequences part II
ELEG 479 Lecture #12 Magnetic Resonance (MR) Imaging
Gradients (Continued), Signal Acquisition and K-Space Sampling
Parallel Imaging Reconstruction
Medical Imaging Systems: MRI Image Formation
Partial Parallel imaging (PPI) in MR for faster imaging IMA Compressed Sensing June, 2007 Acknowledgement: NIH Grants 5RO1CA and 5P41RR008079, Pierre-Francois.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
fMRI Methods Lecture2 – MRI Physics
G Practical MRI 1 Gradients.
Allen W. Song, PhD Brain Imaging and Analysis Center Duke University MRI: Image Formation.
Contrast Mechanism and Pulse Sequences
FMRI – Week 4 – Contrast Scott Huettel, Duke University MR Contrast FMRI Graduate Course (NBIO 381, PSY 362) Dr. Scott Huettel, Course Director.
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
MR Image Formation FMRI Graduate Course (NBIO 381, PSY 362)
MRI Physics: Spatial Encoding Anna Beaumont FRCR Part I Physics.
MRI: Contrast Mechanisms and Pulse Sequences
Spinning Nucleus Produces Magnetic Moment
Magnetic Resonance Learning Objectives
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Charged particle. Moving charge = current Associated magnetic field - B.
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Lecture 1: Magnetic Resonance
Receive Coil Arrays and Parallel Imaging for fMRI of the Human Brain
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
Parameters which can be “optimized” Functional Contrast Image signal to noise Hemodynamic Specificity Image quality (warping, dropout) Speed Resolution.
FMRI data acquisition.
MRI Physics in a Nutshell Christian Schwarzbauer
Magnetic Resonance Imaging: Physical Principles
Assume object does not vary in y
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
Basic MRI I Mark D. Herbst, MD, PhD
Magnetic Resonance Imaging
Introduction to MR Robert Frost
(4)ELECTRONIC SUPPORT SYSTEM
The echo time (TE) The echo time (TE) refers to the time between the application of the radiofrequency excitation pulse and the peak of the signal induced.
Introduction to MR Robert Frost
Presentation transcript:

MRI Physics II: Gradients, Imaging Douglas C. Noll, Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor

Noll Magnetic Fields in MRI B 0 – The main magnetic field.B 0 – The main magnetic field. –Always on (0.5-7 T) –Magnetizes the object to be imaged –After excitation, the magnetization precesses around B 0 at  0 =  B 0 B 1 – The rotating RF magnetic field.B 1 – The rotating RF magnetic field. –Tips magnetization into transverse plane –Performs “excitation” –On for brief periods, then off

Noll Gradient Fields The last magnetic field to be used in MRI are the gradient fieldsThe last magnetic field to be used in MRI are the gradient fields –3 of them: G x, G y, G z –These are for localization –Make the magnetic field different in different parts of the body, e.g. for the x-gradient: B(x) = B 0 + G. x B(x) = B 0 + G. x –Observe the field points in the same direction as B 0 so it adds to B 0.

Noll Gradients x-gradient (G x ) y-gradient (G y ) z-gradient (G z )

Noll Imaging Basics To understand 2D and 3D localization, we will start at the beginning with one- dimensional localization.To understand 2D and 3D localization, we will start at the beginning with one- dimensional localization. –Here we “image” in 1D - the x-direction. (e.g. the L-R direction) We start with the simplest form of localization called “frequency encoding.”We start with the simplest form of localization called “frequency encoding.”

Noll 1D Localization We acquire data while the x-gradient (G x ) is turned on and has a constant strength.We acquire data while the x-gradient (G x ) is turned on and has a constant strength. Recall that a gradient makes the magnetic field vary in a particular direction.Recall that a gradient makes the magnetic field vary in a particular direction. In this case, having a positive x-gradient implies that the farther we move along in the x-direction (e.g. the farther right we move) the magnetic field will increase.In this case, having a positive x-gradient implies that the farther we move along in the x-direction (e.g. the farther right we move) the magnetic field will increase. B(x) = B 0 + G. x

Noll Frequency Encoding A fundamental property of nuclear spins says that the frequency at which they precess (or emit signals) is proportional to the magnetic field strength:A fundamental property of nuclear spins says that the frequency at which they precess (or emit signals) is proportional to the magnetic field strength: This says that precession frequency now increases as we move along the x- direction (e.g. as we move rightwards).This says that precession frequency now increases as we move along the x- direction (e.g. as we move rightwards).  (x) =  (B 0 + G. x).  =  B - The Larmor Relationship

Noll Frequency Encoding

Noll Spins in a Magnetic Field GxGxGxGx  (x) x

Noll The Fourier Transform The last part of this story is the Fourier transform.The last part of this story is the Fourier transform. A function of time is made up of a sum of sines and cosines of different frequencies.A function of time is made up of a sum of sines and cosines of different frequencies. We can break it down into those frequency componentsWe can break it down into those frequency components Recall that e i2πft = cos(2πft) + i sin(2πft) Courtesy Luis Hernandez

Noll Fourier Transforms In short, the Fourier transform is the mathematical operation (computer program) that breaks down each MR signal into its frequency components.In short, the Fourier transform is the mathematical operation (computer program) that breaks down each MR signal into its frequency components. If we plot the strength of each frequency, it will form a representation (or image) of the object in one-dimension.If we plot the strength of each frequency, it will form a representation (or image) of the object in one-dimension.

Noll Fourier Transforms

Noll 2D Imaging - 2D Fourier Transform 2D FFT Acquired Data Resultant Image In MRI, we are acquiring Fourier components – works in two dimensions as well

Noll Fourier Representation of Images Decomposition of images into frequency components, e.g. into sines and cosines.Decomposition of images into frequency components, e.g. into sines and cosines. 1D Object Fourier Data 1D Object Fourier Data

Noll 1D Fourier Transform New Components Cumulative Sum New Components Cumulative Sum of Components of Components 0 th Frequency Component

Noll 1D Fourier Transform New Components Cumulative Sum New Components Cumulative Sum of Components of Components 1 st Frequency Component

Noll 1D Fourier Transform New Components Cumulative Sum New Components Cumulative Sum of Components of Components 2 nd Frequency Component

Noll 1D Fourier Transform New Components Cumulative Sum New Components Cumulative Sum of Components of Components 3 rd Frequency Component

Noll 1D Fourier Transform New Components Cumulative Sum New Components Cumulative Sum of Components of Components 5 th Frequency Component

Noll 1D Fourier Transform New Components Cumulative Sum New Components Cumulative Sum of Components of Components 20 th Frequency Component

Noll 1D Fourier Transform New Components Cumulative Sum New Components Cumulative Sum of Components of Components 63 rd Frequency Component

Noll Fourier Acquisition In MRI, we are acquiring Fourier componentsIn MRI, we are acquiring Fourier components –Remember, we take the FT of the acquired data to create an image The more Fourier components we acquire, the better the representationThe more Fourier components we acquire, the better the representation

Noll Spatial Frequencies in 2D FourierDataImageData Full Low Freq High Freq Low Res Edges Low Res Edges (contrast) (contrast)

Noll Resolution and Field of View Field of view is determined by spacing of samples: FOV = 1 /  kField of view is determined by spacing of samples: FOV = 1 /  k Resolution is determined by size of the area acquired:  x = 1 / WResolution is determined by size of the area acquired:  x = 1 / W W kxkxkxkx kykykyky kykykyky xy yyyy FOV Courtesy Luis Hernandez

Noll Aliasing Courtesy Luis Hernandez

Noll Aliasing K space Courtesy Luis Hernandez

Noll Aliasing K space If you don’t sample often enough, higher frequencies look like lower frequencies (Collected only every other line) Courtesy Luis Hernandez

Noll Resolution and Field of View kxkxkxkx kykykyky Resolution is determined by size of the area acquired:  x = 1 / W Field of view is determined by spacing of samples: FOV = 1 /  k W kkkk

Noll Goals of Image Acquisition Acquire 2D Fourier dataAcquire 2D Fourier data Acquire samples finely enough to prevent aliasing (FOV)Acquire samples finely enough to prevent aliasing (FOV) Acquire enough samples for the desired spatial resolution (  x)Acquire enough samples for the desired spatial resolution (  x) Acquire images with the right contrastAcquire images with the right contrast Do it fast as possibleDo it fast as possible Do it without distortions and other artifactsDo it without distortions and other artifacts

Noll Some Common Imaging Methods Conventional (spin-warp) ImagingConventional (spin-warp) Imaging Echo Planar Imaging (EPI)Echo Planar Imaging (EPI) Spiral ImagingSpiral Imaging

Noll Conventional (Spin-Warp) Imaging kxkxkxkx kykykyky 128x128 FLASH/SPGR TR/TE/flip = 50ms/30ms/30º 0.2 slices per sec, single slice One Line at a Time

Noll Conventional (Spin-Warp) Imaging Known as:Known as: –GRE, FLASH, SPGR Typically matrix sizes for fMRITypically matrix sizes for fMRI –128x64, 128x128 Acquisition ratesAcquisition rates –3-10 sec/image –1-4 slices Usually best for structural imagingUsually best for structural imaging kxkxkxkx kykykyky One Line at a Time

Noll Echo Planar Imaging (EPI) kxkxkxkx kykykyky Single-shot EPI, TE = 40 ms, TR = 2 s, 20 slices Zig-Zag Pattern

Noll Echo Planar Imaging (EPI) kxkxkxkx kykykyky Zig-Zag Pattern Single-shot acquisitionSingle-shot acquisition Typically matrix sizes for fMRITypically matrix sizes for fMRI –64x64, 96x96 –128x128 interleaved Acquisition ratesAcquisition rates –TR = 1-2 sec –20-30 slices Suffers some artifactsSuffers some artifacts –Distortion, ghosts

Noll EPI Geometric Distortions Jezzard and Balaban, MRM 34: field maphigh res image warped epi image unwarped epi image Courtesy of P. Jezzard

Noll EPI Nyquist Ghost Courtesy of P. Jezzard

Noll Spiral Imaging kxkxkxkx kykykyky Single-shot spiral, TE = 25 ms, TR = 2 s, 32 slices Spiral Pattern

Noll Spiral Imaging kxkxkxkx kykykyky Spiral Pattern Single-shot acquisitionSingle-shot acquisition Typically matrix sizes for fMRITypically matrix sizes for fMRI –64x64, 96x96 –128x128 interleaved Acquisition ratesAcquisition rates –TR = 1-2 sec –20-40 slices Suffers some artifactsSuffers some artifacts –Blurring

Noll Spiral Off-Resonance Distortions perfect shimpoor shim Courtesy of P. Jezzard

Noll Single-shot Imaging Single-shot imaging is an extremely rapid and useful class of imaging methods.Single-shot imaging is an extremely rapid and useful class of imaging methods. It does, however, require some special, high performance hardware. Why?It does, however, require some special, high performance hardware. Why? –In spin-warp, we acquire a small piece of data for an image with each RF pulse. –However in EPI and spiral, we try to acquire all of the data for an image with a single RF pulse.

Noll Single-shot Imaging Need powerful gradient ampsNeed powerful gradient amps Limitations:Limitations: –Peripheral nerve stimulation –Acoustic noise –Increased noise –Heating and power consumption in gradient subsystem Other issues:Other issues: –Limited spatial resolution –Image distortions –Some limits on available contrast

Noll Pulse Sequences Two Major AspectsTwo Major Aspects –Contrast (Spin Preparation) What kind of contrast does the image have? What is the TR, TE, Flip Angle, etc.? Gradient echo/spin echo/etc. –Localization (Image Acquisition) How is the image acquired? How is “k-space” sampled? Spatial Resolution?

Noll Pulse Sequences Spin Preparation (contrast)Spin Preparation (contrast) –Spin Echo (T1, T2, Density) –Gradient Echo –Inversion Recovery –Diffusion –Velocity Encoding Image Acquisition Method (localization, k-space sampling)Image Acquisition Method (localization, k-space sampling) –Spin-Warp –EPI, Spiral –RARE, FSE, etc.

Noll Localization vs. Contrast In many cases, the localization method and the contrast weighting are independent.In many cases, the localization method and the contrast weighting are independent. –For example, the spin-warp method can be used for T1, T2, or nearly any other kind of contrast. –T2-weighted images can be acquired with spin-warp, EPI, spiral and RARE pulse sequences.

Noll Localization vs. Contrast But, some localization methods are better than others at some kinds of contrast.But, some localization methods are better than others at some kinds of contrast. –For example, RARE (FSE) is not very good at generating short-TR, T1-weighted images. In general, however, we can think about localization methods and contrast separately.In general, however, we can think about localization methods and contrast separately.

Noll The 3 rd Dimension We’ve talked about 1D and 2D imaging, but the head is 3D.We’ve talked about 1D and 2D imaging, but the head is 3D. Solution #1 – 3D ImagingSolution #1 – 3D Imaging –Acquire data in a 3D Fourier domain –Image is created by using the 3D Fourier transform –E.g. 3D spin-warp pulse sequence Solution #2 – Slice SelectionSolution #2 – Slice Selection –Excite a 2D plane and do 2D imaging –Most common approach

Noll Slice Selection The 3 rd dimension is localized during excitationThe 3 rd dimension is localized during excitation –“Slice selective excitation” Makes use of the resonance phenomenonMakes use of the resonance phenomenon –Only “on-resonant” spins are excited

Noll Slice Selection Slice 1 Slice 2 Slice 3 GzGzGzGz B(z) z With the z-gradient on, slices at different z positions have a different magnetic fields and therefore different frequencies :  (z 1 ) <  (z 2 ) <  (z 3 )

Noll Slice Selection Slice 1 Slice 2 Slice 3 GzGzGzGz B(z) z Slice 1 is excited by setting the excitation frequency to  (z 1 ) Slice 2 is excited by setting the excitation frequency to  (z 2 ) Interesting note: Exciting a slice does not perturb relaxation processes that are occurring in the other slices.

Noll Slice Thickness Slice thickness is adjusted by changing the “bandwidth” of the RF pulseSlice thickness is adjusted by changing the “bandwidth” of the RF pulse Bandwidth ~ 1 / (duration of RF pulse)Bandwidth ~ 1 / (duration of RF pulse) –E.g., for duration = 1 ms, BW = 1 kHz GzGzGzGz  (z) z  zzzz

Noll Multi-Slice Imaging Since T1’s are long, we often would like to have long TR’s ( ms)Since T1’s are long, we often would like to have long TR’s ( ms) While one slice is recovering (T1), we can image other slices without perturbing the recovery processWhile one slice is recovering (T1), we can image other slices without perturbing the recovery process

Noll Multi-Slice Imaging RF pulses Data acquisition RF pulses Data acquisition RF pulses Data acquisition Slice 1 Slice 2 Slice 3  (z 1 )  (z 2 )  (z 3 ) TR

Noll Advanced Image Acquisition Topics T2 vs. T2*T2 vs. T2* –Spin-echo vs. Gradient-echo Parallel Imaging (GRAPPA/SENSE/iPAT/ASSET)Parallel Imaging (GRAPPA/SENSE/iPAT/ASSET) Simultaneous Multi-slice ImagingSimultaneous Multi-slice Imaging

Noll What is T2*? T2* has two parts:T2* has two parts: –Inter-molecular interactions leading to dephasing, a.k.a. T2 decay –Macroscopic or mesoscopic static magnetic field inhomogeneity leading to dephasing, a.k.a. T2’ Pulse sequence issues:Pulse sequence issues: –Spin echoes are sensitive to T2 –Gradient echoes are sensitive to T2*

Noll Spin-Echo Pulse Sequence RF pulses Data acquisition 90 o 180 o 180 o pulse “pancake flipper” Gradient Echo Spin Echo

Noll Parallel Imaging Basic idea: combining reduced Fourier encoding with coarse coil localization to produce artifact free imagesBasic idea: combining reduced Fourier encoding with coarse coil localization to produce artifact free images –Artifacts from reduced Fourier encoding are spatially distinct in manner similar to separation of the coil sensitivity patterns Goes by many names. Most common:Goes by many names. Most common: SENSE (SENSitivity Encoding)SENSE (SENSitivity Encoding) –Pruessmann, et al. Magn. Reson. Med. 1999; 42: GRAPPA (GeneRalized Autocalibrating Partially Parallel Acquisitions)GRAPPA (GeneRalized Autocalibrating Partially Parallel Acquisitions) –Griswold, et al. Magn. Reson. Med. 2002; 47:

Noll Localization in MR by Coil Sensitivity Coarse localization from parallel receiver channels attached to an array coilCoarse localization from parallel receiver channels attached to an array coil

Noll SENSE Imaging – An Example Full Fourier Encoding Volume Coil Pixel A Pixel B Full Fourier Encoding Array Coil S 1A A S 1B B S 3A A S 3B B S 2A A S 2B B S 4A A S 4B B Unknown Pixel Values A & B Known Sensitivity Info S 1A, S 1B,…

Noll SENSE Imaging – An Example Reduced Fourier – Speed-Up R=2 Array Coil S 1A A+S 1B B S 2A A+S 2B B S 3A A+S 3B B S 4A A+S 4B B Reduced Fourier – Speed-Up R=2 Volume Coil A+B Insufficient Data To Determine A & B Extra Coil Measurements Allow Determination of A & B

Noll SENSE Imaging – An Example y1y1y1y1 y2y2y2y2 y4y4y4y4 y3y3y3y3 Solving this matrix equation leads to A & B and the unaliased image A B

Noll Parallel Imaging In the lecturer’s opinion, parallel imaging is only moderately useful for fMRI.In the lecturer’s opinion, parallel imaging is only moderately useful for fMRI. Pros:Pros: –Higher spatial resolution –Some reduction of distortions Cons:Cons: –But lower SNR –Minimal increase in temporal resolution

Noll Simultaneous Multislice Basic Idea: Use coil localization information to separate two overlapping slices – similar to parallel imagingBasic Idea: Use coil localization information to separate two overlapping slices – similar to parallel imaging –Larkman, et al. J. Magn. Reson. Imaging 2001; 13: Pros:Pros: –Increase in temporal resolution (2x-3x) –Allows for thinner slices –Reduce effects of physio noise Cons:Cons: –Small increase in noise, artifact from imperfect decoding of slices