Dr. Marc Madou, UCI, Winter 2012 Class V Potentiometric and Amperometric Sensors (I) Electrochemistry MAE-295.

Slides:



Advertisements
Similar presentations
SVCC Respiratory Care Programs Analysis of ABG Samples.
Advertisements

Review of Analytical Methods Part 2: Electrochemistry
Potentiometric Methods A.) Introduction:
Chapter 20 Electrochemistry
EEE 529 Microsystems Amperometric Biosensors
Fig. 22-1a (p.629) A galvanic electrochemical cell at open circuit
A galvanic cell is made from two half cells. In the first, a platinum electrode is immersed in a solution at pH = 2.00 that is M in both MnO 4 -
Summary of Potentiometry:
Introduction to Analytical Chemistry
Electrode measuring principle
Electricity from Chemical Reactions
IC Controls 855 & 802 ppm Dissolved Oxygen System pH / ORP Conductivity Dissolved Oxygen Chlorine Standards R1.0 © 2004 IC CONTROLS.
Electrochemistry Chapter 20.
Chemical equilibrium: electrochemistry 자연과학대학 화학과 박영동 교수.
MLAB 2401: C LINICAL C HEMISTRY K ERI B ROPHY -M ARTINEZ Designs in Instrumentation Electrochemistry.
Ch. 21 Honors Chem. Electrochemistry
3-D reconstructions based on electron micrographs of isolated mitochondria taken with a large depth of field, at different tilt angles have indicated.
Tracking Reaction Progress via Electrochemical Methods.
Electroanalytical – Introduction Ch 21, 7th e, WMDS)
Measuring concentration using electrodes Indicator electrodes used with reference electrode to measure potential of unknown solution E cell = E indicator.
Dr. Marc Madou Class IV. Microfabrication of electrochemical sensors Winter 2011 BIOMEMS.
Electrodes and Potentiometry
Section 11 Potentiometry. Potentiometric Electrodes Potentiometric electrodes measure: Activity not concentration Concepts to review: Activity and affect.
14-6 Cells as chemical probes
Potentiometry. pH is a Unit of Measurement  pH = Power of Hydrogen (H + )  Defined as the Negative Logarithm of Hydrogen Ion Activity  pH = log (1/H.
Chapter 17 Electrochemistry
ELECTROCHEMISTRY. During electrolysis positive ions (cations) move to negatively charged electrode (catode) and negative ions (anions) to positively charged.
Fall 2014, Prof. JB Lee Ion Sensitive FET (ISFET) - What and Why?
Chemistry 1011 Slot 51 Chemistry 1011 TOPIC Electrochemistry TEXT REFERENCE Masterton and Hurley Chapter 18.
Potentiometric Methods
IC Controls 865 ppb Dissolved Oxygen System pH / ORP Conductivity Dissolved Oxygen Chlorine Standards ppb Galvanic Dissolved Oxygen.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 15 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Lecture 14 CM1001. Oxidation Numbers Positive or negative numbers decided using agreed rules. Help us work out whether a substance is oxidized or reduced.
Electroanalytical Chemistry
Dr. Marc Madou Class II. Electrochemistry Background (I) Winter 2009 BIOMEMS.
PH probe
File Name: Dissolved Oxygen.pptFeb 2001 Dissolved Oxygen Overview.
Chapter 19 Electrochemistry
ELECTROLYSIS. Compare and contrast voltaic (galvanic) and electrolytic cells Explain the operation of an electrolytic cell at the visual, particulate.
POTENTIOMETRY 8th lecture
South Fork Instruments Advances in pH/ORP Measurement – Polymeric Sensors Presented by John Daly South Fork Instruments.
Potentiometry and potentiometric measurements. potentiometer A device for measuring the potential of an electrochemical cell without drawing a current.
1. Blood or Fluid Analysis Using carried out in central labs with problems of time delay costs and potential misslabeling Drive toward more bedside or.
An Introduction to Electroanalytical Chemistry Electrochemistry: The study of the interchange of chemical and electrical energy Oxidation is the loss of.
Chapter 15 Electrode Measurements “Chem 7” test in hospital lab. Na +, K +, Cl -, CO 2, glucose, urea, & creatinine ion-selective electrodes.
Electrochemistry.
 Deals with the relation of the flow of electric current to chemical changes and the conversion of chemical to electrical energy (Electrochemical Cell)
Topic 19 Oxidation and reduction
Electrochemical Methods Dr M.Afroz Bakht. Potentiometry Potentiometry is a method of analysis used in the determination of concentration of ions or substances.
Chapter 17 Corrosion and Degradation of Materials.
A.) Introduction : 1.) Coulometry: electrochemical method based on the quantitative oxidation or reduction of analyte - measure amount of analyte by measuring.
Electrochemical Cells in Actions Batteries and Fuel Cells Chapter 15.
Electrodes and Potentiometry Introduction 1.)Potentiometry  Use of Electrodes to Measure Voltages that Provide Chemical Information - Various electrodes.
PH THEORY What is it pH?What is it pH?/How is pH Measured?/What Equipment is Required to Measure pH?/How is a pH Measurement Device Calibrated?/Why is.
POTENTIOMETRY.
Home Reading Skoog et al. Fundamental of Analytical Chemistry. Chapters 18, 19, 21.
Ch. 21 Potentiometry 1. General principles
Electrochemical Cells. Electrochemical Electrochemical cells are a way of storing chemical potential energy. When batteries operate, electrons in high.
POTENTIALS AND THERMODYNAMICS OF CELLS (1) POTENTIALS AND THERMODYNAMICS OF CELLS (1)
SCI3023 ELECTROCHEMISTRY Chapter 8e: Potentiometry
Inovace bakalářského studijního oboru Aplikovaná chemie Reg. č.: CZ.1.07/2.2.00/
Electrochemistry MAE-212
Ion Selective Electrode
Potentiometry and potentiometric measurements
Ion Selective Electrode
Rapid Lab Systems measuring principle
Analytical Chemistry II Spring 2017
Electrode Training HORIBA Instruments, Inc. Pamela Millett Scientific
Redox in Electrochemistry
Presentation transcript:

Dr. Marc Madou, UCI, Winter 2012 Class V Potentiometric and Amperometric Sensors (I) Electrochemistry MAE-295

Table of content Potentiometric Sensors Amperometric Sensors Nanosensors as electrochemical sensors [Potentiometric and Amperometric Sensors (II)]

Potentiometric Sensors Potentiometric techniques are the most widely used electroanalytical method: Direct potentiometry – pH and ions (pH sensors and ion selective probes) Indirect potentiometry: Enzyme sensors, Gas sensors Miniaturization of Potentiometric Sensors

Direct Potentiometric Sensors Best know example is the pH sensor. Combination electrodes (indicator+reference) for convenience (tube within a tube) pH sensing component of the indicator electrode is the glass bulb, which is a thin glass membrane ~ 0.03 – 0.1 mm thick When immersed, H + ions from the solution enter the Si-O lattice structure of the glass membrane in exchange for Na + Inner tube: pH indicator electrode (pH sensing membrane, Ag/AgCl reference electrode and HCl Outer tube: reference electrode (Ag/AgCl) and salt bridge (KCl)

Direct Potentiometric Sensors A traditional pH measurement with a glass electrode is the best known potentiometric ion selective electrode (ISE) (e.g. a thin glass layer with this composition 22% Na 2 O, 6% CaO, 72% SiO 2 ) There is no change in the inner solution and there is no actual contact between inner and outer solution for any potentiometric probe or sensor How to construct a combination electrode?

Direct Potentiometric pH Sensors The glass bulb creates an electric ‘boundary’ potential across the membrane w.r.t. the internal Ag/AgCl reference electrode. This is called the Donnan potential: Where a H+ = activity of H + (= concentration in very dilute solutions). Slope factor (2.303RT/F) is temperature dependent, pH meter must be adjusted for changes in temperature All modern pH meters record potential (mV) and transform the voltage caused by H + into pH units Standard buffers (4.0, 7.0, 10.0) are used for calibration Automatically recognize standard buffers and adjust for temperature

Electrochemical Methods Applications in Environmental Analysis Direct Potentiometric pH Sensors

Direct Potentiometric Sensors Measurement of Ions by Ion Selective Electrodes (ISEs) Uses direct potentiometry to measure ion concentration Membrane responds selectively to a given ion mV reading between sensing and reference electrode

Direct Potentiometric Sensors There are many other types of potentiometric ion sensors or ISE’s. The so-called Donnan potential is established on both sides of any ion selective membrane-the potential on one side is kept constant through the internal reference solution while the other side is determined by the analyte solution For other ions than protons (cations and anions) other membranes are available (see e.g. LaF 3 for F - and a wide variety of polymeric membranes

Direct Potentiometric Sensors An ion selective polymeric membrane is often made by mixing an ionophore (e.g. valinomycin, a natural occuring antibiotic) with PVC and a plasticizer (to make the rigid plastic more flexible) In these types of ISE’s one sometimes does not use an internal reference solution at all or one incorporates a hydrogel to replace the aqueous solution. This makes the electrode easier to handle and store. Especially with no internal reference electrode drift tends to be larger! The polymeric ISE’s lend themselves well to miniaturization and cost reduction (it is much more difficult to miniaturize a glass pH electrode)

Indirect Potentiometric Sensors: Enzyme Base Potentiometric Sensor A potentiometric urea sensor may consist of two pH sensors one with the enzyme coated on its surface and one without (the reference electrode) The electrode with the urease will sense a local pH change The pH difference bewteen the two electrodes is proportional to the urea concentration As an example two IrO x electrodes may be used

Indirect Potentiometric Sensors: Carbon Dioxide Sensor

Indirect Potentiometric Sensors: Carbon Dioxide Sensor (3D)

Indirect Potentiometric Sensors: Carbon dioxide sensor (MEMS version) A pH, CO 2 and oxygen electrochemical sensor array for in-vivo blood measurements was made using MEMS techniques The pH and CO 2 sensors are potentiometric and the oxygen sensor is amperometric (see further in this class) The pH sensor is an ISE based on a pH sensitive polymer membrane. The CO 2 sensor is based on an IrOx pH sensor and a Ag/AgCl reference electrode..

Miniaturization of Potentiometric Sensors By making ISE’s planar (e.g. on a polyimide sheet) many sensors can be made in parallel (i.e. batch fabnrication). From 3D structures to 2D ! Mass production can make them very small (e.g. 2 by 3 mm), cheap (perhaps disposable), reproducible and even electronics might be integrated (see below under ISFETs)

Miniaturziation of Potentiometric Sensors Potentiometric sensors have been made the size of a transistor in ISFETs (almost).

Amperometric Sensors Our first example of an amperometric sensors involves a "Fuel cell" oxygen sensors consisting of a diffusion barrier, a sensing electrode (cathode) made of a noble metal such as gold or silver, and a working electrode made of a metal such as lead or zinc immersed in a basic electrolyt (such as a solution of potassium hydroxide). Oxygen diffusing into the sensor is reduced to hydroxyl ions at the cathode: O 2 + 2H 2 O + 4e OH- Hydroxyl ions in turn oxidize the lead (or zinc) anode: 2Pb + 4OH PbO + 2H 2 O + 4e- 2Pb + O PbO Fuel cell oxygen sensors are current generators. The amount of current generated is proportional to the amount of oxygen consumed (Faraday's Law).

Amperometric Sensors A second example of an amperometric sensors is a simple (first generation) glucose sensor. This sensor is based on the enzyme Glucose Oxidase (GO). Enzymes are high-molecular weight biocatalysts (proteins) that increase the rate of numerous reactions critical to life itself Enzyme electrodes are devices in which the analyte is either a substrate (also called reactant) or a product of the enzyme reaction, detected potentiometrically or amperometrically Here we consider an amperometric glucose sensor where the substrate (glucose) diffuses through a membrane to the enzyme layer where glucose is converted and H 2 O 2 is produced and electrochemically detected.

Amperometric Sensors Amperometric glucose sensor based on peroxide oxidation, The lateau of the limiting current is proportional to the peroxide concentration which in turn is proportional to glucose typical 0.6 to 0.8 V vs Ag cathode Glucose oxidase is an oxidase type enzyme, urease is a hydrolytic type enzyme. Other sensors can be constructed based on those enzymes. -  i l Anodic Cathodic +i -i +  V

Amperometric Sensors Measurement of Dissolved Oxygen e.g. Polarographic Clark cell

Amperometric Sensors Measurement of Dissolved Oxygen e.g. Polarographic Clark cell O 2 + 2H 2 O + 4e - ⇌ 4OH - (O 2 reduced at gold cathode) 4Ag(s) + 4Cl - (aq) ⇌ 4AgCl(s) + 4e - (oxidation of silver at anode) Membrane is susceptible to degradation, must be replaced if it dries out Calibrated in air (O 2 ), air saturated water (aerated water) or by Winkler method

Amperometric Sensors Measurement of Dissolved Oxygen Calibrate the probe (in air) Place the probe below the surface of the water Set the meter to measure temperature and allow the temperature reading to stabilize Switch the meter to 'dissolved oxygen‘ For saline waters, measure electrical conductivity level or use correction feature Re-test water to obtain a field replicate result NOTE: The probe needs to be gently stirred to aid water movement across the membrane