Chapter 16 Infrared Absorption Spectroscopy An IR spectrum contains information about the functional groups in a molecule, and this is used to uniquely.

Slides:



Advertisements
Similar presentations
FT-IR Instrument 1. 2 i)Dispersive spectrometers or ii)Fourier transform spectrometers. Most commercial instruments separate and measure IR radiation.
Advertisements

Applications of IR spectroscopy
Raman Spectroscopy Laser 4880 Å. Raman Spectroscopy.
FOURIER TRANSFORM INFRARED SPECTROSCOPY
Mass Spectrometry Introduction:
Infrared Spectroscopy
Lecture 6. FT-IR and Raman Spectroscopy. FT-IR Analytical infrared studies are based on the absorption or reflection of the electromagnetic radiation.
Light Source: Globar Silicon Carbide Rod (5mm diameter, 50 mm long) Heated electrically to 1300 – 1500 K Positive temperature coefficient of resistance.
Structural Information
Infrared Radiation 780 nm m Near, Mid and Far
FT-IR Instrument. Components Source Michelson Interferometer Sample Detector.
Rotational Spectra Simplest Case: Diatomic or Linear Polyatomic molecule Rigid Rotor Model: Two nuclei joined by a weightless rod J = Rotational quantum.
FT-IR Instrument. Components Source Michelson Interferometer Sample Detector.
Vibrational Spectroscopy HH O Bend. Diatomic Molecules So far we have studied vibrational spectroscopy in the form of harmonic and anharmonic oscillators.
Introduction to Infrared Spectrometry Chap 16. Quantum Mechanical Treatment of Vibrations Required to include quantized nature of E From solving the wave.
Fourier Transform IR Spectroscopy Saptarshi Basu.
Lecture 3 INFRARED SPECTROMETRY
Introduction and Applications of Infrared Spectrometry
Infrared Spectroscopy. Theory of Infrared Absorption Spectroscopy IR photons have low energy. The only transitions that have comparable energy differences.
Spectroscopy (continued) Last time we discussed what spectroscopy was, and how we could use the interaction of light with atoms and molecules to measure.
IR – Infrared Spectroscopy. What is IR Spectroscopy? Infrared spectroscopy is the analysis of infrared light interacting with a molecule. IR spectroscopy.
10/11/ ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY.
 PART Requirements for Spectroscopic Techniques for Polymers 1. High resolution 2. High sensitivity (>1%) 3. High selectivity between molecular.
Chemical Analysis. Analytical Techniques When chemical evidence is collected at a crime scene, it must be run through an instrument. These instruments.
1 University of Petra Faculty of Science & Arts Department of Chemistry Seminar I.R Spectroscopy By Firas Al-ouzeh Supervisor : Nuha I. Swidan Summer 2007.
Pharmacy 325 Infrared (IR) Spectroscopy Dr. David Wishart Rm Ph Hours: anytime after 4 pm.
1.1 Introduction Energy of IR photon Near IR 12, cm -1 Mid IR cm -1 Far IR cm -1 Energy of IR photons insufficient to cause.
Advanced Spectroscopy 3. Infrared Spectroscopy. Revision 1.What molecular or structural features give rise to absorption of infrared (IR) radiation? covalent.
Infrared Spectroscopy Chapters Chapter 7- I.
Molecular Vibrations
TOPIC D: SPECTROMETRY AND SPECTROSCOPY. Mass spectrometry is used to detect isotopes. mass spectrometer uses an ionizing beam of electrons to analyze.
Infrared Spectroscopy
 a mathematical procedure developed by a French mathematician by the name of Fourier  converts complex waveforms into a combination of sine waves, which.
Chapter 12 Infrared Spectroscopy Jo Blackburn Richland College, Dallas, TX Dallas County Community College District  2006,  Prentice Hall Organic Chemistry,
Chapter 2: IR Spectroscopy Paras Shah
5.3 Notes Light & Spectrometry Pg Theory of Light  Color is a visual indication of the fact that objects absorb certain portions of visible.
(IR spectroscopy) is the subset of spectroscopy that deals with the infrared region of the electromagnetic spectrum. spectroscopy infraredelectromagnetic.
Introduction to Spectrochemical Methods
INFRARED SPECTROSCOPY
Infrared Spectroscopy
1 Spectroscopy  Atomic emission spectra  UV/Vis spectra  Infrared (IR)
Infrared Spectroscopy
1.1 What’s electromagnetic radiation
For 3 rd year students- Biophysics Department – Faculty of Science – Cairo University Prepared by: Dr. Heba Mohamed Fahmy and Dr. Alaa Mamdouh.
Fourier Transform IR Spectroscopy. Absorption peaks in an infrared absorption spectrum arise from molecular vibrations Absorbed energy causes molecular.
Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)
1 Increasing frequency CH 2 =CH-CH=CH 2 Absorption spectrum for 1,3-butadiene.
1 Instrumental Analysis Infrared Spectroscopy Dr. Nermin Salah Lecture 5 09 October 2012.
PRESENTED BY :- ARPIT S. PANDYA I-M.PHARM INDUSTRIAL PHARMACY SRINIVAS COLLEGE OF PHARMACY.
Dr. S. M. Condren Chapter 16 An Introduction to Infrared Spectroscopy.
FTIR: Fourier Transform Infrared Spectroscopy
 FT-IR stands for Fourier Transform Infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR radiation is passed through.
Fourier Transform IR (FTIR)
Infrared Spectroscopy: Comparison of Transmission and ATR Techniques Matt Herring Lewis University.
LAB. S4 Identification of Drugs Using Attenuated Total Reflectance Fourier Transform Infrared Spectroscopy (ATR-FT-IR Spectroscopy)
Optical Characterization Techniques
Lecture 10 IR Theory This Week In Lab: Ch 6 PreLab Due
IR-Spectroscopy IR region Interaction of IR with molecules
Fourier Transformation Infra-red
Analytical methods Prepared By Dr. Biswajit Saha.
IR-Spectroscopy IR region Interaction of IR with molecules
Raman Spectroscopy A) Introduction IR Raman
Infrared Spectroscopy
Infrared Sampling Methods
Presentation On INFRARED SPECTROSCOPY
Infrared Spectroscopy
Lab Equipment.
Harmonic Oscillator.
Infrared Spectroscopy
Presentation transcript:

Chapter 16 Infrared Absorption Spectroscopy An IR spectrum contains information about the functional groups in a molecule, and this is used to uniquely identify the compound.

Wavenumbers (cm -1 ) are used since they are directly proportional to energy – e.g. convert 2.5 µm to wavenumbers (cm -1 ) 400 nm780 nm2500 nm or 4000 cm -1 50,000 nm or 200 cm -1 UVVisNear IRMid IRFar IR

Typical Infrared Spectrum Higher energy vibrations C-H “overtones” C=O C-C

Mechanical Model of a Stretching Vibration in a Diatomic Molecule Treats the vibrating bond like a spring with a given “stiffness” -- = frequency of vibration k = force constant (spring “stiffness”) µ = “reduced mass”

Quantum Mechanical Treatment of Normal Modes 1.Only certain vibrations are “allowed” 2.The vibrational quantum number v = 0, 1, 2, 3…… 3.Only transitions between adjacent energy levels are possible, i.e.  v =  1 4.The frequency of the photon has to equal the frequency of the vibration (see next slide) 5.The molecule must have a change in “dipole moment” as a result of the vibration Dipole moment = charge difference X separation distance Homonuclear diatomics never have an Infrared spectrum. Why? ++ --

++ -- -- ++ IR Absorption if the vibration results in a change in dipole moment Frequency of absorbed light  frequency of vibration SoSo S1S1  v =  1

Common Types of Vibrations (“Normal Modes”) Higher energy modes Lower energy modes

Example – predict the normal modes of CO 2 The number of normal modes = 3N-6 For linear molecules it’s 3N-5

Which normal modes are Infrared active?

Example 16-1, p.436 Calculate the approximate wavenumber and wavelength of the fundamental absorption peak due to the stretching vibration of a C=O group (k = 1.0 x 10 3 N/m)

Force Constants increase with Bond Strength Bond Type* Force Constant, k (N/m) Wave- number (cm -1 ) Bond Energy (kJ/mol) C-C5 x C=C10 x CΞCCΞC15 x Bond Type* Reduced mass (kg) Wave- number (cm -1 ) Bond Energy (kJ/mol) C-H1.55 x C-N1.07 x C-O1.14 x * all bonds are consider these as being approx. equal (a). (b) (a).as  then frequency  (b) When masses approx. equal, then see peak at same wavenumbers

IR Sources 1.Nichrome – Ni/Cr alloy, resistively heats up and emits IR, 1100 K 2.Globar – SiC rod, 1500 K 3.Nernst Glower – electrically heated rare earth-oxide ceramic, 2000 K Output from a Nernst Glower cm -1

Optics – IR spectra are measured in the mid-IR, so have to use halides such as NaCl and KBr Never use aqueous samples, or water to clean salt plates 400 nm780 nm2500 nm or 4000 cm -1 50,000 nm or 200 cm -1 UVVisNear IRMid IRFar IR

Sample Handling – a few drops of “neat” sample between “salt plates”

Solutions

Detectors for the Mid-IR – cm -1 1.Pyroelectric – crystal of DTGS (Deuterated Triglyceine Sulfate) Nice link: 1.DTGS maintains polarization when heated to just below the Curie Point 2.Above the Curie Point, the permanent polarization of the DTGS crystal disappears. 3.The closer to the Curie Point, the more responsive the detector 4.Much faster response so can take a spectrum much more quickly (FTIR)

Detectors for the Mid-IR – cm -1 2.Photoconduction - Mercury-Cadmium-Tellurium (MCT) Liquid N 77K 1.Semiconductor-based 2.Cooled to 77 K using liquid N 2 3.Much faster response so can take a spectrum much more quickly (FTIR)

Definition of the Fourier Transform (  ) - The Inverse Fourier Transform - The functions f(t) and  (  ) are called "transform pairs" f(t)   (  )

time (t)  frequency (s -1, Hz)

e.g. FTIR distance (cm)  wavenumber (cm -1 ) background interferogram P/P o = IR spectrum background spectrum = P o sample spectrum = P sample interferogram

e.g. FT-NMR time (s)  ppm (Hz) Free Induction Decay (FID) Signal (time) NMR Spectrum (ppm, Hz)

Fourier Transforms "invert dimensionality" – signal domainFT domainInstrument time (t)frequency (s -1, Hz)oscilloscope mirrorwavenumber (cm -1 )FTIR distance (cm) free induction ppm (from Hertz)FT-NMR decay (t) Fast Fourier Transform (FFT) on computers - Cooley-Tukey algorithm very efficient on a PC

movable mirror → D fixed mirror 50/50 beamsplitter IR source mirror distance, x (cm) → ¼ interferogram   /42 /43 /44 /4

If polychromatic radiation enters the interferometer, then each wavelength produces a separate interferogram. The output from the interferometer will therefore be a superposition of all wavelengths - 1  1 mirror distance (+x, cm) → 2  2 3  3 ←mirror distance (-x, cm)

signal at detector S(x) this term contains the IR spectrum Taking the Fourier Transform of S(x) results in the IR spectrum, A( ) -

Depends on the distance the mirror moves in the Michelson Interferometer – the further it moves, the greater the resolution. minimum detectable difference in wavenumber x = distance mirror moves e.g. What length of mirror drive in a Michelson Interferometer is required to separate and  m? "retardation“ = 2x

Fourier Transform Infrared (FTIR) Spectrometers The Multiplex (Fellgett's) Advantage - entire spectrum obtained virtually instantaneously; results in a higher S/N because during the time a scanning instrument is slowly obtaining the spectrum, a multiplex instrument such as an FTIR can acquire 100’s pf spectra. Averaging these extra spectra causes the increased S/N (see next slide) and allows quantitative IR. S/N increases as where N = number of "resolution elements" 4000 cm cm -1 e.g. scan range cm -1 at a resolution of 2 cm -1 2 cm -1 "resolution elements" so the S/N increases by IF the FTIR can obtain the entire spectrum for the same amount of time a slower, scanning instrument requires to acquire only one resolution element.

Signal-to-Noise Ratio (S/N)

Single-Beam – run background first (see next slide) and store in memory (P o ) Higher energy throughput (Jacquinot’s Advantage) and no stray light problems

Typical IR background spectrum (P o )

inexpensive (~ $25K) range = 350 – 7800 cm -1 (29 to 1.3  m) max resolution = 4 cm -1 scan time = as fast as 1 sec detector = DTGS expensive (~ $100K) range = 10 – 50,000 cm -1 (1000  m to 200 nm) max resolution < 0.01 cm -1 scan time = can be minutes at high-res detector(s) = DTGS, MCT

Advantages over Scanning Instruments 1.fast scan times 2.higher S/N because of signal averaging 3.higher resolution 4.because there are no slits, there’s a higher energy throughput (Jacquinot’s Advantage); means larger signals for the same concentration (therefore lower LODs) 5.no stray light problems Disadvantages 1.higher cost than scanning instruments 2.difficult to align the interferometer (automation helps) 3.IR optics water soluble (beamsplitter made of KBr)

Chapter 17 Applications of IR Spectrometry

1. Qualitative Analysis

"fingerprint region""group frequency region"

2. Quantitative Analysis 1.lower source powers than in UV-Vis (in FTIR Jacquinot's Advantage partially offsets) 2.detectors suffer from thermal noise (i.e. larger backgrounds or  blank ) 3.salt plates have a very short path length (A =  bc) Signal averaging with FTIR's has improved the S/N enough to allow more sensitive quantitative work in the IR. Much less sensitive (i.e. higher LODs) in the IR compared to the UV-Vis because of -

3. Diffuse Reflectance Accessory

Typically used for powdered samples, i.e. forensic drug analyses

Kubelka-Munk Units = converts the reflectance spectrum to the equivalent of an absorbance spectrum. R’  = sample refectivity/KBr reflectivity f(R’  ) = (1 - R’  )2/2 R’  = k/S k =  C  = the molar absorptivity C is the sample concentration S = "scattering coefficient" For a sample to follow a linear relationship between f(R’  ) and concentration, the following criteria must be met: 1.the sample must be diluted in a non-absorbing matrix such as KBr or KCl. 2.the scattering coefficient S must remain constant over the entire spectrum. 3.there must be no specular, or regular reflectance off the surface of the sample.

4. Attenuated Total Internal Reflectance (ATR) Accessory Total Internal Reflection and the "evanescent wave" - Samples - conventional (solutions, liquids, etc) as well as powders, pastes, suspensions, colloids

5. Infrared Microscopy