Capillary Electrophoresis

Slides:



Advertisements
Similar presentations
Isofocusing Chromatography Prepared by- Shirin Akhter.
Advertisements

Jace James & Lauren Erland.  Separation based on charge & frictional force Capillary Electrophoresis (CE)
Cruces-Blanco, C., Gamiz-Gracia, L., Garcia-Campana A.M., Applications of Capillary Electrophoresis in Forensic Analytical Chemistry Trends in Analytical.
Gas Chromatography There an be many parts to a gas chromatography system but the basic components include: An injection system. A column (controllable.
Capillary Electrophoresis and Capillary Electrochromatography
Capillary Electrophoresis and the Human Genome
Capillary electrophoresis. Electrophoretic migration Z i : charge # of component i e 0 : elemental charge [ 1.602× C ] E=V/L ( V : applied voltage.
Qualification of Capillary Electrophoresis Instrument M. Koupparis.
Chromatography for Protein purification 1
ANALYTIC METHODS II. PART Jana Švarcová.  Chromatography  Electrophoresis  Potentiometry  Titration  Spectrophotometry.
Chapter 26 Other Methods. Ion-Exchange Chromatography The mechanism of separation will be the exchange of ions from the column to the solution. Water.
PRINCIPLE AND INSTRUMENTATION OF CAPILLARY ELECTROPHORESIS
Capillary Electrophoresis
SFC and SF Extraction Intermediate between HPLC and GC SF are substances above their critical pressure and temperature (critical point SFs has great solvating.
Ch. 30 Capillary Electrophoresis, Capillary Electrochromatography
Instant Notes Analytical Chemistry
Salting out is a method of separating proteins based on the principle that proteins are less soluble at high salt concentrations. The salt concentration.
Polyacrylamide Gel Electrophoresis
By Dan Janiak and Mark Hanna September Electrokinetics Electroosmosis- Mark Electrophoresis- Dan.
Biochemical instrumental analysis-7 Dr. Maha Al-Sedik.
Qualitative Analysis of Product
Wilkes University -CHM 342 Electrochromatography - A Hybrid Separation Technique Gel Filtration Chromatography + Capillary Electrophoresis = Electrochromatography.
Chapter 6 - Chromatography
Gel Electrophoresis.
High Performance Liquid Chromatography
Chromatography Separates components in mixture: Based on - polarity
Magnet Analytical Chemistry Unit 4
Capillary Electrokinetic Separations
INSTRUMENTAL ANALYSIS CHEM 4811
Types of Liquid Chromatography I. Ion Exchange Chromatography A. Factors influencing retention B. Suppressed ion exchange II. Partitioning Chromatography.
Factors affecting electrophoresis The electric field ·       Voltage. If the separation of the electrodes is d (cm) and the potential difference.
History of Chromatography n Early LC carried out in glass columns n diameters: 1-5 cm n lengths: cm n Size of solid stationary phase n diameters:
Supercritical Fluid Chromatography Introduction : Supercritical Fluid Chromatography (SFC) is a hybrid of gas and liquid chromatography that combines some.
Advanced Analytical Chemistry – CHM 6157® Y. CAIFlorida International University Updated on 10/17/2006Chapter 9Capillary Electrophoresis Chapter 9Capillary.
Aquious Halide Concentration Determination via Leucigenin Quenching and Indirect CE Christopher Hampton, Dr. Eamonn F. Healy, Chemistry Department, St.
Supercritical Fluid Chromatography Introduction : Supercritical Fluid Chromatography (SFC) is a hybrid of gas and liquid chromatography that combines some.
Simultaneous determination of inorganic anions and cations by capillary electrophoresis with indirect UV detection I. Haumann, J. Boden, A. Mainka, U.
Chromatographic Methods & Capillary Electrophoresis
CHAPTER 21 SUPERCRITICAL-FLUID CHROMATOGRAPHY, CAPILLARY ELECTROPHORESIS, AND CAPILLARY ELECTROCHROMATOGRAPHY Introduction to Analytical Chemistry.
High Throughput Enantiospecific Separations Using Multiplexed Capillary Electrophoresis with Absorbance Detection Jeremy Kenseth, Andrea Bastin, and Brett.
Capillary Electrophoresis (CE) PHAR Lecture Objectives By the end of the lecture, students should be able to: 1.Illustrate the CE instrumental.
Chapter 23 Chromatographic Methods & Capillary Electrophoresis.
Created with MindGenius Business 2005® Capillary Electrophoresis Capillary Electrophoresis Advantages Only needs nL sample High speed and resolution, virtually.
High Performance Liquid Chromatography. The chromatogram is a record of detector output Vs time as the analyte passes through the chromatography.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 23 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Principles Separation is carried out by applying a high potential (10-30 kV) to a narrow (25-75 pm) fused silica capillary filled with a mobile phase.
19-1 Supercritical Fluid Chromatography Theory Instrumentation Properties of supercritical fluid §Critical temperature àAbove temperature liquid cannot.
HPLC.
Chem. 230 – 11/25 Lecture.
CAPILLARY ELECTROPHORESIS Prepared by Dr: Sherif A.Abdel-Gawad.
Analytical methods Houdková Chocholáčová. Analytical methods צ Absolute analytical methods (direct evaluation of the analytical signals with regard to.
General Tests for Chemistry 101
Electrophoresis Chapter 3-Lecture 7
Chapter 34 Miscellaneous Separation Methods. 34 A Supercritical fluid separations Supercritical fluids are an important class of solvents that have unique.
Welcome To my presentation.
Lecture 10 Modes and Methods in Capillary Electrophoresis
Principles of chromatography
High Performance Liquid Chromatography. What is HPLC ? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube.
Advanced Analytical Chemistry – CHM 6157® Y. CAIFlorida International University Updated on 10/17/2006Chapter 9Capillary Electrophoresis Chapter 9Capillary.
Lecture 9 Introduction to Capillary Electrophoresis Lecture 9- PHCM662-SS2016 Dr. Rasha Hanafi 1© Dr. Rasha Hanafi, GUC.
Capillary electrophoresis. Principle of electrophoresis Electrophoresis is the process of moving charged molecule in solution by applying an electric.
ELECTROPHORETIC METHODS
Tutorial 11 Modes and Methods in Capillary Electrophoresis
Lecture 2   ELECTROPHORESIS Electrophoresis is a separation technique that is based on the mobility of the ions in an electric field.. it is a Greek word.
High Performance Liquid Chromatography HPLC
Chapter: Chromatography
Chapter: Chromatography
Capillary Electrophoresis and Capillary Electrochromatography
High-Performance Liquid Chromatography
High-Performance Liquid Chromatography
Presentation transcript:

Capillary Electrophoresis

History Its origin can be traced back to the 1880s it got major recognition in 1937, when Tiselius reported the separation of different serum proteins by a method called moving boundary electrophoresis the moving boundary method was enhanced further with the development of techniques such as the paper electrophoresis (obsolete) and gel electrophoresis (joule heating).

History in 1967, Hjerten used glass tubes with an internal diameter (I.D.) around 3 mm (tube improves the dissipation of heat). In 1979, Mikkers provided a theoretical basis for migration dispersion in free zone electrophoresis in 1981, Jorgenson and Lukacs was introduced the term "capillary electrophoresis (CE)“ fused silica-100um -30kV major challenge toward practical applications of CE Coupling with mass spectrometry (MS) .

Electrophoresis Electro = flow of electricity, phoresis, from the Greek = to carry across A separation technique based on a solute’s ability to move through a conductive medium under the influence of an electric field. The medium is usually a buffered aqueous solution In the absence of other effects, cations migrate toward the cathode, and anions migrate toward the anode.

Electrophoresis is a separation technique that is based on the differential migration of charged compounds in a semi-conductive medium under the influence of an electric field.

Principle of Capillary Electrophoresis A semiconducting medium and an electric field are the basic needs electrophoresis. In the case of CE the semi conducting medium is composed of a capillary filled with an electrolyte or a gel. An electric field is generated by applying a voltage difference across the capillary. As a result components in the capillary are affected by physical forces coming from electro osmosis and electrophoresis (Fig. 1).

Electrophoretic Mobility As a result components in the capillary are affected by physical forces coming from electro osmosis and electrophoresis Electrophoretic Mobility The movement of ions solely due to the electric field, potential difference Cations migrate toward cathode Anions migrate toward anode Neutral molecules do not favor either

Electrophoretic Mobility v=Eq/f E electric field strength f vep = μepE μ = q/(6πηr) q net ionic charge η is buffer viscosity r is solute radius Properties that effect mobility Voltage applied Size and charge of the solute Viscosity of the buffer

Electroosmotic Flow As the buffer sweeps toward the anode due to the electric field, osmotic flow dictates the direction and magnitude of solute ion flow within the buffer All ions are then swept toward the anode. Negative ions will lead the neutral ions toward the anode Positive ions will trail the neutral ions as the cathode pulls them

Electroosmotic Mobility veof = μeofE μeof = ɛζ / (4πη) ɛ = buffer dielectric constant ζ = zeta potential Zeta Potential The change in potential across a double layer Proportional to the charge on the capillary walls and to the thickness of the double layer. Both pH and ion strength affect the mobility

Total Mobility vtot = vep + veof Migration times t = lL/((μep + μeof)V vtot = l/t l = distance between injection and detection t = migration time to travel distance l t = lL/((μep + μeof)V L = length of capillary V = voltage

Electrophoretic Migration The overall migration in CE is determined by the combined effect of the effective and the electro osmotic mobility. In CE using fused silica capillaries the EOF is directed toward the cathode; therefore, the apparent migration velocity of cations is positively affected, while the migration of anions is negatively affected. Neutral compounds are also transported through the capillary toward the cathode because of the EOF. When the electroendosmotic mobility is sufficiently high it is ever possible to separate both cations and anions in one single run (Fig. 5). When µeof is greater than µeff anions that originally migrate toward the anode are still carried toward the cathode: due to a positive apparent velocity. Neutral compounds migrate with the velocity of the EOF, but are unresolved under one peak in the electropherogram.  Migration of cations, anions, and neutral compounds in capillary zone electrophoresis in an ordinary fused silica capillary

As a result, the EOF has a flat plug-like flow profile, compared to the parabolic profile of hydrodynamic flows (Fig. 4). Flat profiles in capillaries are expected when the radius of the capillary is greater than seven times the double layer thickness (Schwer and Kenndler, 1990) and are favorable to avoid peak dispersion. Therefore, the flat profile of the EOF has a major contribution to the high separation efficiency of CE.

Capillary Electrophoresis Instrument

electropherogram

Instrumentation Power supply Anode compartment Cathod compartment narrow-bore fused-silica capillary tube; injection system; detector; Recorder Both with buffer reservoir

Capillary tube Varied length but normally 25-100 cm Small bore and thickness of the silica play a role Using a smaller internal diameter and thicker walls help prevent Joule Heating, heating due to voltage

Joule Heating Joule heating is a consequence of the resistance of the solution to the flow of current if heat is not sufficiently dissipated from the system the resulting temperature and density gradients can reduce separation efficiency Heat dissipation is key to CE operation: Power per unit capillary P/L  r2 For smaller capillaries heat is dissipated due to the large surface area to volume ratio capillary internal volume =  r2 L -capillary internal surface area = 2 r L End result: high potentials can be applied for extremely fast separations (30kV)

• Pressure • Vacuum • Siphoning • Electrokinetic Injection: • Pressure • Vacuum • Siphoning • Electrokinetic

Detector UV/Visible absorption Fluorescence Radiometric (for radioactive substances) Mass Spec.

Different Modes in Capillary Electrophoresis Moving boundary CE (outdated) (2) Steady-state CE  Isotachophoresis. (ITP) Isoelectric focusing (IEF) (3) Zone CE Capillary gel electrophoresis (CGE) Capillary zone electrophoresis (CZE) Micellar electrokinetic capillary chromatography (MEKC) Chiral Capillary Electrophoresis (CCE) Capillary electrochromatography (CEC). Free solution CE In a steady state process, the composition of the background electrolyte is not constant. Both the electric field and the effective mobilities may change along the migration path. The most common practical realization of this type of separation process is to form a pH gradient along the migration path.

Capillary isotachophoresis

Capillary isoelectric focusing Separation due to differences in isoelectric point (pI). Coated column to avoide electroosmosis

Capillary gel electrophoresis Separation mainly due to differences in shape and size.

Capillary zone electrophoresis Separation due to differences in charge, shape and size.

Micellar electrokinetic chromatography Separation due to difference in hydrophobicity.

Separation parameters To achieve a good separation: Narrow bands narrow peaks efficiency:

Resolution:

Capillary Temperature Electrode Polarity Applied Voltage Capillary Temperature Capillary Dimensions Buffers Separation Optimization Parameters Length Internal Diameter

The effect of separation factor ion each other The effect of separation factors The effect of separation factor ion each other

Characteristics -1 Electrophoresis in narrow-bore(25-150 μm id), fused silica capillaries High voltages (10-30 kV) and high electric fields applied across the capillary High resistance of the capillary limits current generation and internal heating High efficiency (N>105-106) Short analysis time(5-20 min) Detection performed on-capillary (no external detection cell)

Characteristics -2 Small sample volume required (1-50 nlinjected) Limited quantities of chemicals and reagents required (financial and environmental benifits) Operates in aqueous media Simple instrumentation and method development Automated instrumentation Numerous modes to vary selectivity and wide application range Applicable to wider selection of analytes compared to other techniques (LC, TLC, SFC, cGC) Applicable to macro-and micromolecules Applicable to charged and neutral solutes Modern detector technology used (DAD, MS)

Why we need chiral separation? Nature is chiral because it mainly uses one of the two enantiomers of a chiral compound.

most biological processes have a high degree of enantioselectivity: each enantiomer may have a different biological activity. drug is administered as a racemic mixture, one enantiomer may have pharmacological effects while the other could have antagonist effect or it could show some undesired side effects.

All of this shows that there are many reasons to discriminate between the enantiomers of a chiral compound and to study them separately. CE has been applied extensively for the separation of chiral compounds in chemical and pharmaceutical analysis.

principle Chiral separation by capillary electrophoresis Not based on an electrophoretic mechanism because the electrophoretic mobilities of the enantiomers of a chiral compound are equal and nonselective. This separation principle relies on the different partition of enantiomers between the bulk solution and the chiral pseudophase ( chiral selector), Electrokinetic Chromatography

Mechanism of chiral sepatation using cyclodextrine as chiral selector Anode Cathode Detector μCD(-) μEOF k2 K 1 Inclusion R S

chiral selector Number of papers published using the different chiral selectors described in EKC.

Types of CDs Natural Cyclodextrins Charged Cyclodextrins Anionic Cyclodextrins Ex. Highly sulfated CDs Ex.Carboxymethylated CDs Cationic Cyclodextrins Dual Cyclodextrin System

Part II   Enantioselective capillary electrophoresis method for determination of tertatolol in plasma and dosage forms using highly sulphated gama cyclodextrin as chiral selector: mechanistic and molecular modeling studies* *This part has been submitted to J. Chromatogr. A

Experimental work Optimization one-variable Type of chiral selectore Capillary Dimensions Effect of BGE Concentration Effect of pH HS-γ-CD Concentration. Neutral CDs such as α-CD, β-CD, γ-CD and their derivatives hydroxy propyl-CD (HP-CD) and dimethyl-β-CD (DM-β-CD) were examined. The charged CDs investigated were sulfated- β-CD (S-β-CD), carboxy methyl-β-CD (CM-β-CD), highly sulfated-α-CD (HS-α-CD), highly sulfated-β-CD(HS-β-CD) and highly sulfated-γ-CD (HS-γ-CD). Tow concentration (3 and 5 mM) from each bile salt, taurocholic sodium (STDC), taurodeoxycholate (STC), deoxycholate (SDC) and cholate (SC) were also studied under normal and reverse polarity at pH (2.5 and 8) of phosphate and acetate buffer, respectively. No separations were obtained using bile salts and CDs except for HS-γ-CD in 25mM triethylammonium phosphate (TEAP) buffer at pH 2.5 where base line separations was achieved within 20 min Application in human plasma & pharmaceutical preparation.

Electrophoretic Condition Reverse polarity 7 kV voltage 25 mM triethylammonium phosphate (pH 2.5 ) 5% HS-γ-CD.

Results and Discussion

Electropherograms of spiked human plasma with 100 ng/ml of Rs =1.23 Rs =17.12 Electropherograms of spiked human plasma with 100 ng/ml of (-)-tertatolol (1), (+)- tertatolol (2) and 400 ng/ml tolterodine L- tartarate (3).

Study the molecular mechanics for both chiral drugs

Schematic representation of the two most probable inclusion models

(+)-Model-A (wide ring) (-)-Model-A (wide ring) (-)-Model-B (wide ring) (+)-Model-B (wide ring) Inclusion complex of (+)- & (-)-tertatolol with HS-γ-CD showed Model-A (upper panel) and Model-B (lower panel) from wide rings views.

The method was linear in the range of 100‑2000 ng/ ml (r = 0.999) for each enantiomer LOD = 50 ng/ml. LOQ = 100 ng/ml The mean RSD of the results within-day and intra- day precision was ≤ 5% Accuracy of the drug were E% ≤ 2.5 % . The method was highly specific, where the co formulated compounds did not interfere.

500 ng/ml of (-)-tertatolol (1), (+)-tertatolol (2) % recovery = 98.32% RSD = 0.85 % (+)-Tertatolol % recovery = 100.96% RSD = 0.99 % Electropherograms of 500 ng/ml of (-)-tertatolol (1), (+)-tertatolol (2) and 500 ng/ml tolterodine L- tartarate (3) recovered from tertatolol tablets.

Stability study The two drugs were subjected to thermal, photolytic, hydrolytic, and oxidative stress conditions and the stressed samples were analyzed by the proposed method.

mAU Minute Minute

Degradation products (UK-55-410), (PD 0162910-00) for AM and AT respectively produced as a result of stress studies did not interfere with the detection of AM and AT and the assay can thus be considered stability indicating.