Point Spread Function, Spectral Calibration & Spectral Separation: Quality Assurance Testing Light Microscopy Research Group Richard W. Cole Wadsworth.

Slides:



Advertisements
Similar presentations
Today’s Topic (Slight ) correction to last week’s lecture
Advertisements

Bio-optics & Microscopy (MEDS 6450) 11/16/2010 Presented by: Mathilde Bonnemasison Leia Shuhaibar Steve Pirnie Ronghua (Ronnie) Yang Neil MAA, Juskaitis.
ERT107 MICROBIOLOGY FOR BIOPROCESS ENGINEERING Pn Syazni Zainul Kamal PPK Bioprocess.
Nick Beavers Project Manager Deconvolution from Andy Molnar Software Engineer.
February 19 th – 22 nd Technologies to Enable Personalized Medicine San Antonio, TX Point Spread Function, Spectral Calibration and beyond: Quality Assurance.
Microscope Objective Parameters. What do the Numbers on the Objective Mean ?
Blizard Advanced Light Microscopy club Making friends with your microscope.
Spectral Resolution and Spectrometers
Dial in: (319) Access: # Unmute/mute: #6 Deconvolution of Microscopy Images Issues to Consider for Widefield and Confocal Microscopy Jonathan.
Deconvolution of Laser Scanning Confocal Microscope Volumes: Empirical determination of the point spread function Eyal Bar-Kochba ENGN2500: Medical Imaging.
ABRF meeting 09 Light Microscopy Research Group. Why are there no standards? Imaging was largely an ultrastructure tool Digital imaging only common in.
Optical Microscopy Lecture 1.
R. Hui Photonics for bio-imaging and bio- sensing Rongqing Hui Dept. Electrical Engineering & Computer Science, The University of Kansas, Lawrence Kansas.
Image-Pro Premier Basic Training Course
Lasers and Confocal. Laser Acronym: Light Amplification by Stimulated Emission of Radiation Ordinary light emission: Comes from spontaneous decay of excited.
Comparison and Implications of Three Optical Microscopy Data Acquisition Modalities James Butler Ph.D. Nikon Instruments, Inc. Widefield Fluorescence Confocal.
1 Laser Beam Coherence Purpose: To determine the frequency separation between the axial modes of a He-Ne Laser All sources of light, including lasers,
Fluorescence microscopy – Principle and practical consideration Hiro Ohkura.
Optics. Thomas Young ( ) Wave Optics Diffraction & Interference Christiaan Huygens ( )
P. Moghe, 125:583 1 Microscopy Techniques for Biomaterials and Cell Based Interfaces Professor Prabhas V. Moghe October 26, :583 Fall 2006.
Media Cybernetics Deconvolution Approaches and Challenges Jonathan Girroir
Microscopy Techniques for Biomaterial Characterization: A Primer Prabhas V. Moghe Lecture 3 September 21, 1999 RU CBE 533 or BME 553; NJIT BME 698.
Study of Protein Association by Fluorescence-based Methods Kristin Michalski UWM RET Intern In association with Professor Vali Raicu.
1 Basics of Digital Imaging Digital Image Capture and Display Kevin L. Lorick, Ph.D. FDA, CDRH, OIVD, DIHD.
Microscope.
A microscope is a Complicated Lens n The microscope can be represented by a single convex lens.
Microscopy.
Illumination and Filters Foundations of Microscopy Series Amanda Combs Advanced Instrumentation and Physics.
Biophotonics lecture 23. November Last week: -Fluorescence microscopy in general, labeling, etc… -How to do optical sectioning and fill the missing.
Confocal Reflectance Theta Line-Scanner for Intra-operative Imaging Confocal Reflectance Theta Line-Scanner for Intra-operative Imaging Peter J. Dwyer*,
TIRF Total Internal Reflection Fluorescence Microscopy specialized fluorescence microscopy technique specifically images a very thin optical section (50-250nm)
Page 1 AO Microscopy of Biological Systems Slides thanks to Prof. Joel Kubby, EE at UCSC.
IPC Friedrich-Schiller-Universität Jena 1 Reduction of out of focus light  Excitation light excites fluorescence more or less within the whole sample.
- 1 - LSM 510 META AIM Carl Zeiss - PMLSM 510 META EF-Guided Tour April 2002 Guided tour through:  Configuration  Defining spectral range  Scan control.
Agenda How to make the specimen visible – Definition of Contrast
High Throughput Microscopy
1)Adaptive optics: optimization and wavefront sensing 2)Novel microscope enhancements.
BIO 211 Human Anatomy and Physiology I
Basics of Deconvolution
DECam Daily Flatfield Calibration DECam calibration workshop, TAMU April 20 th, 2009 Jean-Philippe Rheault, Texas A&M University.
- 1 - LSM 510 META AIM Carl Zeiss - PMLSM 510 META EF-Guided Tour April 2002 Guided tour through:  Configuration  Defining spectral range  Scan control.
Molecular Cell Biology Light Microscopy in Cell Biology Cooper Modified from a 2010 lecture by Richard McIntosh, University of Colorado.
Use Snell’s law to determine the angle of total internal reflection in the coverslip (without oil). Oil immersion objectives More light (no total internal.
Designing a Microscopy Experiment Kurt Thorn, PhD Director, Image from Susanne Rafelski, Marshall lab.
(Image: T. Wittman, Scripps) Principles & Practice of Light Microscopy: 2.
Date of download: 5/28/2016 Copyright © 2016 SPIE. All rights reserved. Confocal to multiphoton conversion. (a) Schematic of system adaptation. The near.
Confocal Microscopy Kurt Thorn NIC.
FROM IMAGES TO ANSWERS Deconvolution of Widefield and Confocal images The growing role of deconvolution NE 2007.
1 Setting Laser Power and PMT Gain Begin with laser power at 20% Set PMT gain to 6.00.
Principles & Practice of
Date of download: 6/22/2016 Copyright © 2016 SPIE. All rights reserved. Index-matching effect. Matching the index of refraction of the bead with the solution.
Reporter: Chien-Chung Tsai
Date of download: 7/5/2016 Copyright © 2016 SPIE. All rights reserved. Spatial light modulator-two-photon microscope (SLM-2PM) scheme: (1) Ti:Sa laser.
Biology 227: Methods in Modern Microscopy Andres Collazo, Director Biological Imaging Facility Yonil Jung, Graduate Student, TA Week 10 Multispectral Imaging.
Light Microscopy Research Group Robert F. Stack, Richard W. Cole Wadsworth Center / NYSDOH Albany, N.Y.
Light Microscope Terms and Practices.
Characterisation of Bessel and Gaussian beam illumination modes on LSM imaging quality of early chick embryo development Ricardo F Bango Da Cunha Correia1,
Microscope.
Lasers and Confocal.
Oil immersion objectives
Microscope Calibration
Volume 104, Issue 4, Pages (February 2013)
LIGHT MICROSCOPY basic
LIGHT MICROSCOPY variations
Supraresolution Imaging in Brain Slices using Stimulated-Emission Depletion Two- Photon Laser Scanning Microscopy  Jun B. Ding, Kevin T. Takasaki, Bernardo.
Francesca Pennacchietti, Travis J. Gould, Samuel T. Hess 
Harvard Medical School Neurobiology Imaging Facility
Microscopy.
Harvard Medical School Neurobiology Imaging Facility
OL 750 Spectroradiometer & Linear Spectral Transmission Measurements
Presentation transcript:

Point Spread Function, Spectral Calibration & Spectral Separation: Quality Assurance Testing Light Microscopy Research Group Richard W. Cole Wadsworth Center / NYSDOH Albany, New York

Why ? – until the last 5-10 yrs, simply observing a specimen was sufficient; advances in light microscopes necessitates traceable standards & procedures Overall Goal – the creation of a range of imaging parameters traceable to standard references  NIH – Realizes the need for and supports the “core” model – 40% of S10 grants funded were imaging in general; 13% confocal  NIST goal of moving medical imaging & lab testing from an art to a science  FDA –ensure manufacturers systems are reliable, guaranteeing that the drugs will be safe & efficacious  Congress – provide the financial support for comparable standards of research Quality and standards: Making bioimaging ‘measure up’ Susan M. Reiss BioOptics World, Jan/Feb 2010, Vol.3 No.1, p Access sparks action Lila Guterman NCRR Reporter, Winter 2010, p.4-8

 Phase One a worldwide research study to ascertain the current state of light microscope performance using simple, efficient & robust tests for LASER stability, field illumination & coregistration  define & improve cross-platform standards--assist core managers & users maintaining microscopes for optimal operation with the ultimate goal of improving the validity of quantitative measurements in light microscopy  the results of this study were accepted for publication in late 2010 in Microscopy and Microanalysis, one of the highest rated imaging journals  throughout 2011, the LMRG tested and defined additional areas of instrument performance (Phase Two) and refined the methodology for determining a system’s Spectral calibration, Spectral separation ability and finally the Point Spread Function of an imaging system What we have done Stack, R., Bayles, C., Girard, A., Martin, K., Opansky, C., Schulz, K., and Cole, R. (2011) Quality Assurance Testing for Modern Optical Imaging Systems. Microscopy & Microanalysis 17(4): Cole, R.W., Jinadasa, T., and Brown, C.M. (2011) Resolution and Quality Control of Confocal Microscopy Optics. Nature Prot. 6 (12): 1929–1941.

Spectral calibration Purpose: Measure spectral calibration of the detection system. MIDL lamp / mirror slide protocol:  Use 10x lens or no lens (system dependent)  Set up the MIDL lamp as the illumination source or use laser(s) and mirror slide (remove blocking)  Set the PMT gains to be ‘equivalent’  Perform a lambda scan and measure the signal-to-noise  Compare acquired spectra with published spectra Analysis: 1.Determine if your PMT(s) show significant spectral variation (sliders) or signs of aging reference:

PARISS Spectral Calibration Lamp, Lightform,Inc. Asheville, NC Overlay of 5 PMT responses and MIDL lamp calibrated output / before repair

PARISS Spectral Calibration Lamp, Lightform,Inc. Asheville, NC Overlay of 5 PMT responses and MIDL lamp calibrated output / after repair

Quality of Spectral un-mixing: Purpose: Measure the spectral un-mixing capability of an imaging system. Protocol: Bead slide: 6.0 µm FocalCheck Double Orange fluorescent microspheres (excitation/emission maxima: core = 532/552 & shell = 545/565) o use same optical settings/components (i.e. laser line/excitation, dichroic filter) to acquire reference and experimental spectra o set detection to maximize S/N without any pixel saturation o select a detection bandwidth wide enough to encompass full emission range (e.g. DoubleOrange beads nm) o if available, choose detection set-up (i.e. parallel vs. lambda) o split detection into smallest discreet ‘bins’ if using lambda scanning mode o select an area of the reference spectra (via ROI) with the highest S/N and store in database Analysis: Select the most appropriate unmixing data-processing algorithm available:  automatic mode (1 st pass / not generally adequate)  parallel mode (simultaneous data acquisition across multiple PMTs)  lambda mode (lambda scanning utilizing one PMT)

Spectral separation FocalCheck™ fluorescence microscope test slide core = 532/552 & shell = 545/565 ) Image of a bead where the core and ring have a small spectral separation Ring and core are pseudo-colored for illustration purposes

Linear Unmixing Algorithms  The measured spectra of a ‘mixed pixel’ is broken down into a collection of component spectra (endmembers) and a set of subsequent fractions (abundances) that indicate the ratio of each endmember in the pixel  Three distinct stages of spectral unmixing : - dimension reduction (i.e. data reduction) - endmember determination (i.e. # of distinct spectra) - inversion (i.e. abundance estimation)  Employs a linear mixing model A Survey of Spectral Unmixing Algorithms Nirmal Keshava Lincoln Laboratory Journal, Vol.14 No.1, 2003, p.55-78

Brain tissue (5 different labels)

blue = cell nuclei, green = Nissl-specific for neurons, yellow = reactive astrocytes, red = microglia, purple = endothelial cells representing blood vessels.

Resolution  point at which two objects are perceived as separate and distinct from one another  Resolution obtained from an imaging system is affected by: the specific wavelength of light in use the diffraction of light (Rayleigh, Abbe & Sparrow limits) lens aberrations sample prep (coverslip thickness, mounting media, RI matching)  Lens imperfections such as coma, astigmatism and spherical aberrations will result in a loss of resolution  microscope resolution and the extent of image ‘blur’ is typically described in terms of its Point Spread Function (PSF)  an ‘ideal’ PSF demonstrates symmetric balance and proportion Limit of resolution: d = (λ) / N.A.

What is a Point Spread Function & why is it so important ?  a measure of the degree of blurring of an object & any potential aberrations  speaks directly to the quality/resolution of an imaging system  Image = convolution of an object and the point spread function  an object plane light wave refocused by a lens produces a blurred focal plane point commonly referred to as an ‘airy disc / airy pattern’  sub-resolutional beads are typically used

Point Spread Function: Purpose: Measure the point spread function of an imaging system. Protocol: Bead slide: 175 nm PS-Speck beads (mixture of blue, green, orange & deep red single-color beads) o test multiple lens: i.e. 20x, 40x, 63x & 100x (all objectives routinely used for imaging in your lab) o collect a Z series or scan in XZY mode o if needed, suitably rotate image to obtain a ‘side view’ o if your system is filter based (non-AOBS), check various dichroic filters Analysis:  use the MetroloJ plug-in (Fiji / ImageJ) to determine the FWHM lateral & axial resolution  compare the experimental vs. theoretical resolution values  check the curve fits for all three

MetroloJ PSF report

‘Idealized’ PSF images courtesy of Zeiss

Theoretical PSF images / Confocal vs. Widefield courtesy of Media Cybernetics

Widefield PSF of thick specimen coverslip  increasing depth & worsening PSFs | V

‘3D’ Widefield PSF

20x / Refractive Index mismatch collar incorrectly set to water // RI(water)=1.33, RI(Leica imm.oil)=1.518

40x oil NA 1.25 / pinhole = 0.5 & 5 airy units

63x N.A. 1.4 oil immersion lens / Brownian motion

Corrective Actions Spectral Calibration – a.Service call Spectral Unmixing – a.Try a different unmixing algorithm - avoid using ‘automatic’ - try various linear algorithms - try non-linear (e.g. SWCCA) algorithms b.Try a different detector set-up - use (5) PMTs with simultaneous scanning OR - use (1) PMT with lambda scanning c. Improve the signal-to-noise Point Spread Function – a.Clean the lens and optics / remove all air bubbles b.Check for any possible refractive index mismatches c.Try a different lens d.Open pinhole aperture to mimic widefield conditions e.Check for optical misalignment ** It is important to note that the above suggestions DO NOT encompass all possible solutions to these issues **

The test specimens proposed for both phases of this study were decided upon by the members of the LMRG for their applicability, robustness, ease-of-use and relative cost. While the phase I & II tests utilize materials from specific vendors who offer excellent products for these purposes, neither the members of the LMRG nor the ABRF endorse the use of these specific vendors, and fully acknowledge the use of legitimate alternatives for the purposes of instrument performance testing.

Acknowledgements Light Microscopy Research Group Carol BaylesCornell University Claire Brown (chair)McGill University Richard ColeWadsworth Center / NYSDOH Brady Eason McGill University Anne-Marie GirardOregon State University Jay JeromeVanderbilt University Tushare Jinadasa McGill University Karen Jonscher(EB Liaison)University of Colorado Cynthia OpanskyBlood Center of Wisconsin George McNamaraUniversity of Miami Katherine SchulzBlood Center of Wisconsin Marc Thibault Ecole Polytechnique * We would also like to thank the ABRF for their financial support and commitment to this project *