An Introduction to Flame Atomic Absorption Spectrometry (FAAS) Steve Badger and Charity Wessel.

Slides:



Advertisements
Similar presentations
Atomic Absorption Spectrometer
Advertisements

By: Mae Temporal TOMIC BSORPTION PECTROSCOPY. ATOMS Nucleus- protons (+ve) and neutrons (neutral). Electrons- (-ve) charged particle. Shells- consists.
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
427 PHC. Direct-Current Plasma  A direct-current plasma (DCP) is created by an electrical discharge between two electrodes. A plasma support gas is necessary,
Spectroscopy Lecture 4 Ahmad Razali Bin Ishak Department of Environmental Health Faculty of Health Sciences UiTM Puncak Alam.
Electron Configuration and Energy Levels: Ground state: All the electrons in an atom have the lowest possible energies Most stable electron configuration.
Atomic Absorption and Atomic Fluorescence Spectrometry Wang-yingte Department of Chemistry
AAS and FES (Ch 9, 7th e, WMDS)
Flame Test Lab.
INTRODUCTION TO OPTICAL METHODS
Atomic Absorption Spectrometry Dr AKM Shafiqul Islam University Malaysia Perlis.
HL Chemistry - Option A: Modern Analytical Chemistry ATOMIC ABSORPTION SPECTROSCOPY.
Flame photometry.
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Atomic Emission Spectroscopy (AES)
Atomic Absorption Spectroscopy (AAS)
Spectroscopy Chapter 7.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering Atomic-Absorption (AA) Spectroscopy 
Advanced Higher Chemistry Unit 1 Spectroscopy. Spectroscopy  Spectroscopy is used to give information regarding the structure of atoms or molecules.
Safety Before starting work, students must familiarise themselves with the safety and handling information provided with the procedures. Laboratory coat.
A TOMIC A BSORPTION S PECTROSCOPY By Hisham E Abdellatef 1.
BC ILN Atomic Absorption Spectroscopy (AAS) 1 Thompson Rivers University.
Biochemical instrumental analysis-2
Atomic Emission Spectroscopy
Atomic Emission Spectrometry
Flame Photometry Flame atomic emission spectrometry
Honors Forensic Science.  Introduction  Organic substances constitute a substantial portion of physical evidence submitted to crime labs  Carbon does.
Electrons in Atoms Chapter 5.
Atomic Absorption Spectroscopy
6- PRENTICE HALL ©2007 Pearson Education, Inc. Upper Saddle River, NJ CRIMINALISTICS An Introduction to Forensic Science, 9/E By Richard Saferstein.
Chapter 10.2 Radiation Tells Us the Temperature, Size, and Composition of Stars.
Emission Spectra Bohr Model of the Atom.
Atomic Absorption Spectrometry
Lab 12 Atomic spectra and atomic structure
Fluorometric determination of riboflavin
Fluorometric determination of riboflavin. Introduction.
1 Atomic Absorption Spectroscopy Lecture Performance Characteristics of Electrothermal Atomizers Electrothermal atomization is the technique of.
Atomic Fluorescence Spectroscopy. Background l First significant research by Wineforder and Vickers in 1964 as an analytical technique l Used for element.
Electrons Negative charge e- Located in the electron cloud far from the nucleus Have mass, but it is negligible Also have wave-like properties.
Topic #9 : Ground State vs. Excited State What is the difference between ground state and excited state electron configurations? How can we identify unknown.
Ground State vs. Excited State
Atomic-absorption spectroscopy
ATOMIC ABSORPTION SPECTROSCOPY (AAS) Atomization: It is the conversion of molecules to their component atoms in gaseous state using a source of heat (flame).
A TOMIC - ABSORPTION SPECTROSCOPY. Atomic absorption spectroscopy (AAS) is a technique for determining the concentration of a particular metal element.
Temperature profiles for a nature gas/air flame.
UNIT 3 ELECTRON CONFIGURATION AND MODERN ATOMIC THEORY.
Atoms & Nuclei The Structure of the Atom. Atoms are made up of a dense, positively charged nucleus and negatively charged electrons that orbit in specific.
Atomic-absorption spectroscopy Lab3 Atomic-absorption spectroscopy.
Fluorometric determination of riboflavin
Atomic Absorption Spectroscopy.  Qualitative test for metals in solution  Can also be used for coloured lights  This emission of light is called an.
A TOMIC - ABSORPTION SPECTROSCOPY Lab no. 3 Done by : Iman Al Ajeyan.
By Maryum Atique M.Phill chemistry University of Agriculture.
Atomic Absorption Spectrometer
Satish Pradhan Dnyanasadhana College Department Of Chemistry
Flame Emission Spectrometry
Topic: Atomic Concepts Aim: What is the difference between ground state and excited state electron configurations? How can we identify unknown elements?
Spectroscopy Chapter 7.
General Physics (PHY 2140) Lecture 35 Modern Physics Atomic Physics
Flame Test Lab.
Atomic Absorption and Emission
Dnyanasadhana College, Thane. Department of Chemistry T. Y. B. Sc
Aim: What happens when electrons get excited?
Electronic Structure of the Atom
Satish Pradhan Dnyanasadhana College Department Of Chemistry
Atomic Absorption Spectroscopy. Atomic absorption spectroscopy is based on the same principle as the flame test used in qualitative analysis.
Atomic Emission Spectra
Flame Tests.
Atomic absorption spectroscopy
Spectroscopy Uses emission and absorption of light by electrons moving between ground and excited state configuration, hence electronic configuration.
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Presentation transcript:

An Introduction to Flame Atomic Absorption Spectrometry (FAAS) Steve Badger and Charity Wessel

Introduction FAAS, developed in the 1950s, is a common method of quantitative analysis of many elements FAAS is extensively used to determine trace quantities of elements in biological, environmental, clinical, geological, and edible samples. –Elemental health hazards can be detected with % error.

Theory of Operation Structure of an atom –Atoms consist of a nucleus and electrons. The nucleus contains protons and neutrons. The negatively charged electrons are located in orbitals around the nucleus. –When electrons occupy the innermost available orbital, their energy is at a minimum (ground state).

Theory of Operation, continued When atoms are subjected to heat or some form of EMR, one or more electrons jump to a higher energy level, leaving a vacancy in the inner shell We say the electron is excited As this happens, energy is absorbed

Theory of Operation, continued When the excited electron in the outer orbital returns to the lower energy level of the inner, vacant orbital, energy is released in the form of a photon

Theory of Operation, continued Because the atoms of each element have a different electronic structure, each one emits light of different wavelengths In FAAS, the concentration of an element present in a sample can be measured by noting the absorbance caused by the excited sample

Instrument Design FAAS components: –Fuel (acetylene) –Oxidizer (air, N 2 O) –Hollow Cathode Lamp (HCL) –Nebulizer –Burner Head –Flame –Exhaust hood

Instrument Design: Acetylene Acetylene is a flammable, compressed gas that is used as a fuel for FAAS. During FAAS operation, the valve connected to the acetylene cylinder is opened completely. Storing the tank in a lab can be extremely dangerous. That is why the tanks are supported upright with a metal chain.

Instrument Design: HCL The HCL is a glass tube filled with an inert gas (neon, argon, or helium) and the pure element to be subjected to FAAS. The atoms of the element become ionized and excited when the FAAS is on.

Instrument Design: HCL The spectrum emitted by the HCL corresponds to the element in the cathode For example, if a Cd HCL is used, the characteristic wavelengths of Cd are emitted

Do you recall emission spectra from general chemistry?

7.3

Instrument Design: HCL Single-element HCLs generally achieve greater sensitivity On the other hand, multi-element HCLs save money

Instrument Design: Burner Head The burner head is made of titanium This is where the sample atoms are excited

Instrument Design: Nebulizer The nebulizer changes liquid sample to a mist This arrow is pointing to the nebulizer

Instrument Design: Flame The flame is ignited by pushing the red burner button on the instrument The maximum temperature of the air-acetylene flame is 3,095°C

Typical Units of Concentration Parts Per Million (ppm) A unit of concentration often used when measuring levels of pollutants in air, water, body fluids, etc ppm is equal to 1.00 mg/liter And 1.00  g/mL

Typical Units of Concentration Parts Per Billion (ppb) Another typical concentration unit used when measuring trace levels of pollutants in air, water, body fluids, etc ppb is equal to 1.00  g/liter

Preparation of Standards Typically 4-5 standards and a blank are prepared to construct a calibration curve (a/k/a a standard curve) –When preparing your solutions, think about the need for precision, low contamination, and minimal waste –Standard solutions should be prepared using appropriate glassware and distilled water Consult a reference to determine appropriate concentrations of standards for a given element

Constructing a Calibration Curve As with other analytical techniques, FAAS requires careful calibration The absorbance of the standards are plotted versus concentration The plot often deviates from a straight line

Preparation of Samples A standard reference should be consulted to determine appropriate methods of preparing various samples for FAAS

Operation Open the valve on top of the acetylene cylinder completely. Turn on the exhaust vent using the switch on the hood. Turn on the workstation connected to the FAAS. Load the FAAS software.

Operation: Optimization Only a few steps are required for optimization –The software steps you through the optimization, because a method has already been created for doing so After optimization, you are now ready to start the analysis

Analysis Again, the software leads you through the procedure Sequentially aspirate the blank (distilled water), the calibration standards, and the samples

Evaluation of Results After the analyses are complete, look at the calibration curve printed out by the FAAS system’s printer

Limitations of FAAS The chemical form of the analyzed element is not detected –For example, if copper was being analyzed, all that would be known is how much copper is present. –It would not be known if it is copper(II) cyanide, copper(III) sulfate, etc. The preparation of the standards and samples can be time consuming The whole FAAS procedure requires detailed work Some elements cannot be examined by FAAS The FAAS procedure destroys the sample

Conclusion FAAS is an accurate method of quantitative analysis for many elements