The transmission electron microscope. Additional web resources – Eric Stach (2008), ”MSE 528 Lecture 4: The instrument,

Slides:



Advertisements
Similar presentations
Basic Electron Microscopy
Advertisements

Electron Optics Basic Introduction Bob Ashley
A.E. GunnæsMENA3100 V10 Transmissions electron microscopy Sample preparation Basic principles Imaging aberrations (Spherical, Chromatic, Astigmatism) contrast.
Electron Optics.
Do it with electrons ! II.
Scanning Electron Microscope (SEM)
Groups: WA 2,4,5,7. History  The electron microscope was first invented by a team of German engineers headed by Max Knoll and physicist Ernst Ruska in.
Foundations of Physics
Lab 10: Lenses 1.Focal Length 2.Magnification 3.Lens Equation 4.Depth of Field 5.Spherical Aberrations and Curved Focal Plane 6.Chromatic Aberration 7.Astigmatism.
Chapter 31 Images.
Components of Initial Eprobe Startup Electron Gun –turning on beam, setting beam current, acceleration, saturation level Focusing lenses –Permit focussed.
First of all, do you know any methods to check chemical composition? Or how you know what is what? First of all, do you know any methods to check chemical.
Chapter 23 Mirrors and Lenses.
Chapter 23 Mirrors and Lenses. Notation for Mirrors and Lenses The object distance is the distance from the object to the mirror or lens Denoted by p.
Electron Optics Two essential components: 1)Electron source (gun) 2)Focusing system (lenses) Add scanning apparatus for imaging Electron gun Cathode Anode.
Electron Microscope. Light Resolution  The resolution of a microscope is limited by the diffraction of light. Single diffractionSingle diffraction 
The TEM system and components:
Transmission Electron Microscopy (TEM) By Austin Avery.
Physics 52 - Heat and Optics Dr. Joseph F. Becker Physics Department San Jose State University © 2005 J. F. Becker San Jose State University Physics 52.
Chapter 23 Mirrors and Lenses.
Introduction to scanning electron microscopy
Transmission Electron Microscopy
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
Diffraction methods and electron microscopy
Microscope.
A.E. GunnæsMENA3100 V13 Analytical Transmissions Electron Microscopy (TEM) Part I: Basic principles Operational modes Diffraction Part II: Imaging Sample.
Electron Microscopy Jim Atherton. Development of Light Microscope 1590 Hans Zacharias Janssen 1660 Robert Hook o Onserved cells (cork) 1 Klein, Aaron.
Scanning Electron Microscopy
Visual Angle How large an object appears, and how much detail we can see on it, depends on the size of the image it makes on the retina. This, in turns,
Chapter 23 Mirrors and Lenses.
Fundamental of Optical Engineering Lecture 3.  Aberration happens when the rays do not converge to a point where it should be. We may distinguish the.
Introduction to transmission electron microscopy
B. Spatial coherancy & source size Spetial coherancy is related to the size of the source. Source size governs spatial coherancy and maller source sizes.
Chapter 34 Lecture Eight: Images: II. Image Formed by a Thin Lens A thin lens is one whose thickness is small compared to the radii of curvature For a.
Scanning Electron Microscope (SEM)
PHYS 430/603 material Laszlo Takacs UMBC Department of Physics
Lesson 25 Lenses Eleanor Roosevelt High School Chin-Sung Lin.
Scanning Electron Microscope (SEM)
A.E. GunnæsMENA3100 V08 Transmissions electron microscopy Basic principles Sample preparation Imaging aberrations (Spherical, Chromatic, Astigmatism) contrast.
The Hong Kong Polytechnic University Optics 2----by Dr.H.Huang, Department of Applied Physics1 Diffraction Introduction: Diffraction is often distinguished.
NANO 225 Micro/NanoFabrication Electron Microscopes 1.
Transmission Electron Microscope Basic premise of a TEM is to project a magnified image of the specimen onto a fluorescent screen where it can be viewed.
Chapter 38: Diffraction and Polarization  For a single opening in a barrier, we might expect that a plane wave (light beam) would produce a bright spot.
The law of reflection: The law of refraction: Image formation
Optics (Lecture 2) Book Chapter 34,35.
Transmission Electron Microscope
A.E. GunnæsMENA3100 V08 Electron Diffraction (ED) in the transmissions electron microscope.
GEOMETRICAL OPTICS. Laws of Reflection Laws of Refraction.
Signature (unit, name, etc.) Transmissions electron microscopy Basic principles Sample preparation Imaging aberrations (Spherical, Chromatic, Astigmatism)
Part 10 Optics --Mirrors and Lenses Chapter 24 Geometric Optics.
EMALEMAL U of M The Transmission Electron Microscope: A Tool for Diffraction & Imaging in Materials Science & Engineering John Mansfield North Campus Electron.
Scanning Transmission Electron Microscope
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
Microscopy.
Refraction & Lenses. Refraction of Light When a ray of light traveling through a transparent medium encounters a boundary leading into another transparent.
Analytical Transmissions Electron Microscopy (TEM)
Presentation on.  There are many methods for measuring of fiber structure. Such as:  The absorption of infrared radiation  Raman scattering of light.
Electron sources and guns
Introduction to scanning electron microscopy
Presentation on SEM (Scanning of Electron Microscope) Represented by:-Ravi Kumar Roll:- (BT/ME/1601/006)
Introduction to scanning electron microscopy
FYS4340/FYS9340 Diffraction methods and electron microscopy
TRANSMISSION ELECTRON MICROSCOPy (TEM).
Laboratory equipment Lecture (3).
Ch. 30 Lenses.
TEM and SEM.
Do it with electrons !.
SEM Imaging - Magnification and Deflection System
Chapter 16: Electron Diffraction
Presentation transcript:

The transmission electron microscope

Additional web resources – Eric Stach (2008), ”MSE 528 Lecture 4: The instrument, Part 1,

MENA3100 V08 Objective lense Diffraction plane (back focal plane) Image plane Sample Parallel incoming electron beam Si 1,1 nm 3,8 Å Objective aperture Selected area aperture Simplified ray diagram

JEOL 2000FX Wehnelt cylinder Filament Anode Electron gun 1. and 2. beam deflectors 1.and 2. condenser lens Condenser aperture Condenser lens stigmator coils Condenser lens 1. and 2. beam deflector Condenser mini-lens Objective lens pole piece Objective aperture Objective lens pole piece Objective lens stigmators 1.Image shift coils Objective mini-lens coils (low mag) 2. Image shift coils 1., 2.and 3. Intermediate lens Projector lens beam deflectors Projector lens Screen Mini-lens screws Specimen Intermediate lens shifting screws Projector lens shifting screws

Eric Stach (2008), ”MSE 528 Lecture 4: The instrument, Part 1,

Eric Stach (2008), ”MSE 528 Lecture 4: The instrument, Part 1,

The requirements of the illumination system High electron intensity – Image visible at high magnifications Small energy spread – Reduce chromatic aberrations effect in obj. lens Adequate working space between the illumination system and the specimen High brightness of the electron beam – Reduce spherical aberration effects in the obj. lens

The electron gun The performance of the gun is characterised by: – Beam diameter, d cr – Divergence angle, α cr – Beam current, I cr – Beam brightness, β cr at the cross over Cross over α d Image of source

Brightness Brightness is the current density per unit solid angle of the source β = i e /(πd c α c ) 2 Cross over α d Image of source

The electron source Two types of emission sources – Thermionic emission W or LaB6 – Field emission W ZnO/W Cold FEG Schottky FEG

The electron gun Bias -200 V Ground potential -200 kV Anode Wehnelt cylinder Cathode d cr Cross over α cr Equipotential lines Thermionic gun FEG

Thermionic guns Filament heated to give Thermionic emission -Directly (W) or indirectly (LaB 6 ) Filament negative potential to ground Wehnelt produces a small negative bias -Brings electrons to cross over

Thermionic guns

Thermionic emission Current density: – Ac: Richardson’s constant, material dependent – T: Operating temperature (K) – φ: Work function (natural barrier that prevents electrons from leaving the solid) – k: Boltzmann’s constant J c = A c T 2 exp(-φ c /kT) Richardson-Dushman Maximum usable temperature T is determined by the onset of the melting/evaporation of material.

Field emission Current density: Fowler-Norheim Maxwell-Boltzmann energy distribution for all sources

Field emission The principle: – The strength of an electric field E is considerably increased at sharp points. E=V/r r W < 0.1 µm, V=1 kV → E = V/m – Lowers the work-function barrier so that electrons can tunnel out of the tungsten. Surface has to be pristine (no contamination or oxide) – Ultra high vacuum condition (Cold FEG) or poorer vacuum if tip is heated (”thermal” FE; ZrO surface tratments → Schottky emitters).

Characteristics of principal electron sources at 200 kV WLaB 6 FEG Schottky (ZrO/W) FEG cold (W) Current density J c (A/m 2 )2-3* *10 4 1*10 7 Electron source size (µm) Emission current (µA) ~100 Brightness B (A/m 2 sr)5*10 9 5* *10 12 Energy spread ΔE (eV) ~0.80.3~0.7 Vacuum pressure (Pa)* Gun temperature (K) * Might be one order lower

Advantages and disadvantages of the different electron sources W Advantages:LaB 6 advantages:FEG advantages: Rugged and easy to handleHigh brightnessExtremely high brightness Requires only moderate vacuum High total beam currentLong life time, more than 1000 h. Good long time stabilityLong life time ( h) High total beam current W disadvantages: LaB 6 disadvantages:FEG disadvantages: Low brightnessFragile and delicate to handle Very fragile Limited life time (100 h)Requires better vacuumCurrent instabilities Long time instabilitiesUltra high vacuum to remain stable

Electron lenses Electrostatic – Require high voltage - insulation problems – Not used as imaging lenses, but are used in modern monochromators or deflectors Magnetic – Can be made more accurately – Shorter focal length F= -eE F= -e(v x B) Any axially symmetrical electric or magnetic field has the properties of an ideal lens for paraxial rays of charged particles.

General features of magnetic lenses Focuses near-axis electron rays with the same accuracy as a glass lens focuses near axis light rays. Same aberrations as glass lenses. Converging lenses. The bore of the pole pieces in an objective lens is about 4 mm or less. A single magnetic lens rotates the image relative to the object. Focal length can be varied by changing the field between the pole pieces (changing magnification).

Electromagnetic lens Bore Soft Fe pole piece gap Cu coil Current in the coil creates A magnetic field in the bore. The magnetic field has axial symmetry, but is inhomogenious along the length of the lens. The soft iron core can increase the field by several thousand times.

Electron ray paths through magnetic fields B See fig 6.9 r θ v v2 v1 The electron spirals through the lens field: A helical trajectory. For electrons with higher keV, we must use stronger lenses (larger B) to get similar ray paths.

Simple ray diagrams Electron lenses act like a convex glas lens Thin lens β: variable giving the fraction of rays collected by the lens ~ 10 m rad ~0.57 o β α Point obj Point image Never a perfect image

Changing the strength of the lens The further away rays are from the optical axis the stronger they are bent by a convex lens. What happens to the focal and image plane when the strength of the lens is changed? What happens to the image?

Under conditions normally found in the TEM, strong lenses magnify less and demagnify more (not in VLM). When do we want to demagnify an object? The strength of the lens

Spherical aberration d s = 0.5MC s α 3 (disk diameter, plane of least confusion) d s = 2MC s α 3 (disk diameter, Gaussian image plane) M: magnification C s :Spherical aberration coefficient α: angular aperture/ angular deviation from optical axis r1r1 r2r2 Plane of least confusion α Gaussian image plane 2000FX: C s = 2.3 mm 2010F: C s = 0.5 nm Highest intensity in the Gaussian image plane

Chromatic aberration v v - Δ v Diameter for disk of least confusion: d c = C c α ((ΔU/U) 2 + (2ΔI/I) 2 + (ΔE/E) 2 ) 0.5 C c : Chromatic aberration coefficient α: angular divergence of the beam U: acceleration voltage I: Current in the windings of the objective lens E: Energy of the electrons 2000FX: C c = 2.2 mm 2010F: C c = 1.0 mm Thermally emitted electrons: ΔE/E=kT/eU, LaB 6 : ~1 eV Disk of least confusion The specimen will introduce chromatic aberration. The thinner the specimen the better!! Correcting for Cc effects only makes sense if you are dealing with specimens that are thin enough.

Lens astigmatism Loss of axial symmetry y-focus x-focus y x This astigmatism can not be prevented, but it can be corrected! Disk of least confusion Diameter of disk of least confusion: d a : Δfα Due to non-uniform magnetic field as in the case of non-cylindrical lenses. Apertures may affect the beam if not precisely centered around the axis.

Depth of focus and depth of field (image) Imperfections in the lenses limit the resolution but give a better depth of focus and depth of image. – Use of small apertures to minimize aberrations. The depth of field (Δb or D ob ) is measured at, and refers to, the object. – Distance along the axis on both sides of the object plane within which the object can move without detectable loss of focus in the image. The depth of focus (Δa, or D im ), is measured in, and referes to, the image plane. – Distance along the axis on both sides of the image plane within which the image appears focused.

α im D ob D im d ob d im Depth of focus and depth of field (image) β ob Ray 1 and 2 represent the extremes of the ray paths that remain in focus when emerging ± D ob /2 either side of a plane of the specimen. α im ≈ tan α im = (d im /2)/(D ob /2) β ob ≈ tan β ob = (d ob /2)/(D im /2) Angular magnification: M A = α im / β ob Transvers magnification: M T = d im / d ob M T = 1/M A Depth of focus: D im =(d ob / β ob )M T 2 Depth of field: D ob = d ob / β ob α im ≈ tan α im = (d im /2)/(D ob /2) β ob ≈ tan β ob = (d ob /2)/(D im /2) Angular magnification: M A = α im / β ob Transvers magnification: M T = d im / d ob M T = 1/M A Depth of focus: D im =(d ob / β ob )M T 2 Depth of field: D ob = d ob / β ob

Depth of field: D ob = d ob / β ob Carefull selection of β ob Thin sample: β ob ~10 -4 rad Thicker, more strongly scattering specimen: β ob (defined by obj. aperture) ~10 -2 rad Depth of field Example: d ob / β ob = 0.2 nm/10 mrad = 20 nm Example: d ob / β ob = 2 nm/10 mrad = 200 nm D ob = thickness of sample all in focus D ob = thickness of sample all in focus

Depth of focus Depth of focus: D im =(d ob / β ob )M T 2 Example: To see a feature of 0.2 nm you would use a magnification of ~ x (d ob / β ob )M 2 = 20 nm *(5*10 5 ) 2 = 5 km Example: To see a feature of 0.2 nm you would use a magnification of ~ x (d ob / β ob )M 2 = 20 nm *(5*10 5 ) 2 = 5 km Example: To see a feature of 2 nm you would use a magnification of ~ x (d ob / β ob )M 2 = 200 nm *(5*10 4 ) 2 = 500 m Example: To see a feature of 2 nm you would use a magnification of ~ x (d ob / β ob )M 2 = 200 nm *(5*10 4 ) 2 = 500 m Focus on the wieving screen and far below! Focus on the wieving screen and far below!

Fraunhofer and Fresnel diffraction Fraunhofer diffraction: far-field diffraction – The electron source and the screen are at infinite distance from the diffracting specimen. Flat wavefront Fresnel diffraction: near-field diffraction – Either one or both (electron source and screen) distances are finite. Electron diffraction patterns correspond closely to the Fraunhofer case while we ”see” the effect of Fresnel diffraction in our images.

Airy discs (rings) Fraunhofer diffraction from a circular aperture will give a series of concentric rings with intensity I given by: I(u)=I o (J I (πu)/ πu) 2

Strengths of lenses and focused image of the source If you turn up one lens (i.e. make it stronger, or ‘over- focus’ then you must turn the other lens down (i.e. make it weaker, or ‘under-focus’ it, or turn its knob anti-clockwise) to keep the image in focus.

Magnification of image, Rays from different parts of the object If the strengths (excitations) of the two lenses are changed, the magnification of the image changes

The Objective lens Often a double or twin lens The most important lens – Determines the resolving power of the TEM All the aberations of the objective lens are magnified by the intermediate and projector lens. The most important aberrations – Astigmatism – Spherical – Chromatic

Astigmatism Can be corrected for with stigmators

Cs can be calculated from information about the shape of the magnetic field – Cs has ~ the same value as the focal length (see table 2.3) The objective lens is made as strong as possible – Limitation on the strength of a magnetic lens with an iron core (saturation of the magnetization Ms) – Superconductiong lenses (give a fixed field, but need liquid helium cooling) The objective lens

Apertures

We use apertures in the lenses to control the divergence or convergence of electron paths through the lens which, in turn, affects the lens aberrations and controls the current in the beam hitting the sample.

A.E. GunnæsMENA3100 V08 Use of apertures Condenser apertures: Limit the beam divergence (reducing the diameter of the discs in the convergent electron diffraction pattern). Limit the number of electrons hitting the sample (reducing the intensity). Objective apertures: Control the contrast in the image. Allow certain reflections to contribute to the image. Bright field imaging (central beam, 000), Dark field imaging (one reflection, g), High resolution Images (several reflections from a zone axis).

BF image Objective aperture Objective aperture: Contrast enhancement All electrons contribute to the image. Si Ag and Pb glue (light elements) hole Only central beam contributes to the image. Bright field (BF)

Small objective aperture Bright field (BF), dark field (DF) and weak-beam (WB) BF image Objective aperture DF imageWeak-beam Dissociation of pure screw dislocation In Ni 3 Al, Meng and Preston, J. Mater. Scicence, 35, p , Diffraction contrast

Large objective aperture High Resolution Electron Microscopy (HREM) HREM image Phase contrast

Use of apertures Condenser aperture: It limits the beam divergence (reducing the diameter of the discs in the convergent electron diffraction pattern). It limits the number of electrons hitting the sample (reducing the intensity). Objective aperture: It controls the contrast in the image. It allows certain reflections to contribute to the image. Bright field imaging (central beam, 000), Dark field imaging (one reflection, g), high resolution images (several reflections from a zone axis). Selected area aperture: It selects diffraction patterns from small (> 1µm) areas of the specimen. It allows only electrons going through an area on the sample that is limited by the SAD aperture to contribute to the diffraction pattern (SAD pattern).

Selected area diffraction Objective lense Diffraction pattern Image plane Specimen with two crystals (red and blue) Parallel incoming electron beam Selected area aperture Pattern on the screen

Diffraction with no apertures Convergent beam and Micro diffraction (CBED and µ-diffraction) Convergent beam Focused beam Convergent beam Illuminated area less than the SAD aperture size. CBED pattern µ-diffraction pattern C2 lens Diffraction information from an area with ~ same thickness and crystal orientation Small probe

Shadow imaging (diffraction mode) Objective lense Diffraction plane (back focal plane) Image plane Sample Parallel incoming electron beam

Magnification and calibration MicroscopeLensModeMagnification JEM-2010ObjectiveMAG LOW MAG Philips CM30TwinTEM Super twinTEM TwinSA Super twinSA Resolution of the photographic emulsion: µm Magnification depends on specimen position in the objective lens Magnification higher than x can be calibrated by using lattice images. Rotation of images in the TEM.