Techniques of Proton Radiotherapy Bernard Gottschalk Harvard University

Slides:



Advertisements
Similar presentations
Heuristic Search techniques
Advertisements

Isoparametric Elements Element Stiffness Matrices
(See also arXiv v2 [physics.med-ph] 29 Apr 2012.)
5/3/2015J-PARC1 Transverse Matching Using Transverse Profiles and Longitudinal Beam arrival Times.
Lateral penumbra is a measure of the sharpness of the shadow cast by a patient aperture. It depends on depth in patient, air gap and beam line design,
Sampling Distributions
Gauss’ Law Besides adding up the electric fields due to all the individual electric charges, we can use something called Gauss’ Law. Its idea is similar.
McGraw-Hill Ryerson Copyright © 2011 McGraw-Hill Ryerson Limited. Adapted by Peter Au, George Brown College.
With Curved Mirrors Created by Derek J. Wells. Under the expressed written consent of Derek J. Wells in accordance with the rules and by-laws of Derek.
Chapter 23 Mirrors and Lenses. Notation for Mirrors and Lenses The object distance is the distance from the object to the mirror or lens Denoted by p.
Introduction The cubic spline is a standard mathematical function commonly used to interpolate a data set. (The spline is a pre-CAD draftsman’s tool which.
Instructor: Mircea Nicolescu Lecture 13 CS 485 / 685 Computer Vision.
Planning under Uncertainty
Could CKOV1 become RICH? 1. Characteristics of Cherenkov light at low momenta (180 < p < 280 MeV/c) 2. Layout and characterization of the neutron beam.
Evaluating Hypotheses
Definitions Uniform Distribution is a probability distribution in which the continuous random variable values are spread evenly over the range of possibilities;
Physics of fusion power Lecture 8 : The tokamak continued.
CSC 160 Computer Programming for Non-Majors Lecture #7: Variables Revisited Prof. Adam M. Wittenstein
Value of Information for Complex Economic Models Jeremy Oakley Department of Probability and Statistics, University of Sheffield. Paper available from.
Maximum likelihood (ML)
lecture 2, linear imaging systems Linear Imaging Systems Example: The Pinhole camera Outline  General goals, definitions  Linear Imaging Systems.
Binary Degraders Often we want to control both scattering and energy loss in a beam line element. For instance, we might want a contoured scatterer with.
Ch 8.1 Numerical Methods: The Euler or Tangent Line Method
Chapter 5 Sampling Distributions
UNIVERSITI MALAYSIA PERLIS
Frame by Frame Bit Allocation for Motion-Compensated Video Michael Ringenburg May 9, 2003.
880.P20 Winter 2006 Richard Kass 1 Confidence Intervals and Upper Limits Confidence intervals (CI) are related to confidence limits (CL). To calculate.
Copyright © 2010 Pearson Education, Inc. Chapter 6 The Standard Deviation as a Ruler and the Normal Model.
Ch 9 pages Lecture 23 – The Hydrogen Atom.
R 2R2R a a Today… More on Electric Field: –Continuous Charge Distributions Electric Flux: –Definition –How to think about flux.
Module 1: Statistical Issues in Micro simulation Paul Sousa.
ECE 8443 – Pattern Recognition ECE 8423 – Adaptive Signal Processing Objectives: Deterministic vs. Random Maximum A Posteriori Maximum Likelihood Minimum.
1 Lesson 8: Basic Monte Carlo integration We begin the 2 nd phase of our course: Study of general mathematics of MC We begin the 2 nd phase of our course:
Copyright © 2010, 2007, 2004 Pearson Education, Inc. Chapter 6 The Standard Deviation as a Ruler and the Normal Model.
Copyright © 2009 Pearson Education, Inc. Chapter 6 The Standard Deviation as a Ruler and the Normal Model.
Lesson 10 The Fields of Accelerating Charges. Class 29 Today we will: learn about threads and stubs of accelerating point charges. learn that accelerating.
We often draw two or more E field vectors at one point, when there are two or more charges causing the field. Each vector then represents the field that.
CEE 262A H YDRODYNAMICS Lecture 15 Unsteady solutions to the Navier-Stokes equation.
Laser-Based Finger Tracking System Suitable for MOEMS Integration Stéphane Perrin, Alvaro Cassinelli and Masatoshi Ishikawa Ishikawa Hashimoto Laboratory.
D. M. J. Tax and R. P. W. Duin. Presented by Mihajlo Grbovic Support Vector Data Description.
Numerical Methods.
© 2005 Pearson Prentice Hall This work is protected by United States copyright laws and is provided solely for the use of instructors in teaching their.
Design Studies 20 ‘Show Off’ Project How to make a computer monitor In Google Sketchup By: Liam Jack.
1 8. One Function of Two Random Variables Given two random variables X and Y and a function g(x,y), we form a new random variable Z as Given the joint.
Background Subtraction and Likelihood Method of Analysis: First Attempt Jose Benitez 6/26/2006.
21 Jun 2010Paul Dauncey1 First look at FNAL tracking chamber alignment Paul Dauncey, with lots of help from Daniel and Angela.
Single scattering is the simplest way to spread out a proton beam. It is relatively immune to beam alignment. On the other hand, it is inefficient and.
Modulators We have seen how to flatten the dose transversely by single or double scattering. We now flatten the dose longitudinally (in depth) by using.
Computer simulation Sep. 9, QUIZ 2 Determine whether the following experiments have discrete or continuous out comes A fair die is tossed and the.
Improvement of the Monte Carlo Simulation Efficiency of a Proton Therapy Treatment Head Based on Proton Tracking Analysis and Geometry Simplifications.
1 Introduction to Statistics − Day 4 Glen Cowan Lecture 1 Probability Random variables, probability densities, etc. Lecture 2 Brief catalogue of probability.
Lecture 8 Source detection NASSP Masters 5003S - Computational Astronomy
Bragg Peak The Bragg peak, as we use the term, is the on-axis depth-dose distribution, in water, of a broad quasi-monoenergetic proton beam. A carefully.
One Function of Two Random Variables
Copyright © 2008 Pearson Education, Inc. Publishing as Pearson Addison-Wesley Chapter 6 The Standard Deviation as a Ruler and the Normal Model.
If information seems to be missing, make any reasonable assumptions. 1.A target has an areal density of 2.3 g/cm 2 and a thickness of 0.8 inch. What is.
Copyright © Cengage Learning. All rights reserved. 5 Joint Probability Distributions and Random Samples.
G. Cowan Lectures on Statistical Data Analysis Lecture 5 page 1 Statistical Data Analysis: Lecture 5 1Probability, Bayes’ theorem 2Random variables and.
Lecture 9: PID Controller.
The accuracy of averages We learned how to make inference from the sample to the population: Counting the percentages. Here we begin to learn how to make.
Review on Coulomb’s Law and the electric field definition
ITK 9.3. Level Set Segmentation Shape Detection Segmentation
Virtual University of Pakistan
Lesson 8: Basic Monte Carlo integration
Hidden Markov Models Part 2: Algorithms
Algorithm An algorithm is a finite set of steps required to solve a problem. An algorithm must have following properties: Input: An algorithm must have.
Slope measurements from test-beam irradiations
8. One Function of Two Random Variables
8. One Function of Two Random Variables
IntroductionLecture 1: Basic Ideas & Terminology
Presentation transcript:

Techniques of Proton Radiotherapy Bernard Gottschalk Harvard University

Harvard University, the Physics Department, and the Lab for Particle Physics and Cosmology (LPPC) made this course possible by their support.

Double Scattering Double scattering wastes less energy and fewer protons than single scattering. It is used for large fields or maximum penetration. We’ll survey various techniques, emphasizing the compensated contoured scatterer. We’ll outline design and compensation methods, using projection, scaling, and the idea of a universal or generic shape. Although some pretty good shapes are known (45% efficiency), global optimization is still elusive and better shapes may exist. Double scattering is sensitive to beam centering at the second scatterer S2. Active steering (feedback) is usually needed. On the other hand, it is relatively insensitive to beam emittance (shape and divergence) and small problems can usually be corrected by adjusting S1.

Outline methods: annulus, dual ring, contoured design procedure: projection, scaling, useful radius de-scaling: converting back to physical quantities optimizing the generic shape beam steering: skewness and all that summary

The first double scattering system was described by Koehler, Schneider and Sisterson (Med. Phys. 4 (1977) ). S2 consists of a beam stopper or ‘occluding ring’ in front of a uniform scatterer. They also described using a rotating permanent magnet for S1 with a uniform S2, but attributed the first use of magnetic pencil beam scanning to Uppsala: ‘highly efficient, if complex’. The method can be generalized to more occluding rings. Occluding Ring (Annulus) Method

The annulus method shares with all double-scattering methods a sensitivity to beam mis-steering on S2, which causes the Gaussian there to be off center causing a tilted (not displaced!) dose distribution downstream. It is the centering of the projected beam at S2 that matters. An angle error at S2 merely displaces the ‘good’ field by a small amount at isocenter. Sensitivity to Beam Mis-Steering on S2

Takada (Japan J. Appl. Phys. 33 (1994) ) described a dual-ring system. The inner ring is of material that scatters more strongly. The outer one has less scattering but the same energy loss. Sensitivity to beam problems was studied in a later paper. This method, which can be generalized to more rings, can be regarded as a two step discretization of the compensated contoured scatterer to be described next. Takada’s Dual-Ring System

The contoured scatterer was invented independently at Uppsala and HCL around 1988 and a detailed design procedure was published later (Grusell et al. Phys. Med. Biol. 39 (1994) ). S1 forms a Gaussian on S2, which is shaped so that the central part scatters more strongly. The shape can be optimized so the dose distribution in the measuring plane is exactly flat out to some ‘design radius’. Of course, central protons lose more energy, which must be compensated by a complementary plate of plastic. Contoured Scatterer

This profile (J.V. Siebers and D.W. Miller, AAPM Annual Meeting (1992) abstract V5) shows that the contour need not be bell-shaped. Flattening it yields a slightly smaller energy loss. This scatterer from the HCL neurosurgery beam is now used in the Burr Center ‘outside user’ beam. Note the resemblance to the Takada dual-ring scheme. The Siebers Profile

Outline methods: annulus, dual ring, contoured design procedure: projection, scaling, useful radius de-scaling: converting back to physical quantities optimizing the generic shape beam steering: skewness and all that summary

Designing a Contoured Scatterer Today, the compensated contoured scatterer is the preferred technique for passive beam spreading. The Uppsala report (Grusell et al.) reads quite differently from the method we will describe but the differences are superficial. The same basic techniques are used: projection onto the measuring plane; their ‘universal radial profile’ which is our ‘generic solution’; scaling (though the choice of scaling variable is different). Grusell’s universal profile is a nine-step deformed cosine whereas ours is a cubic spline. A major difference is that Grusell’s derivation cites generalized Fermi-Eyges theory whereas ours is based directly on the Preston and Koehler construction. Even here, the difference is more one of style and derivation. The final computations are quite similar.

If nothing else, pictures like this are hard to draw (maybe easier now with Google SketchUp). Therefore, following Preston and Koehler, we’ll focus on the projection of the double scattering event on a ‘measuring plane’ (MP), which can be anywhere along the beam line. Beam Line Geometry

t is the displacement vector of the beam (0 for now). u is the projected first scatter and q the projected radius at which the proton hits S2. v is the second scatter which takes the proton to radius r in the MP. R is the maximum radius of the contoured scatterer (collimator radius) projected onto the MP. All quantities of interest in the MP are distances. Double Scattering in Projection u q r v t S R P 0

Projected first and second scatters u, v and their characteristic (rms) values u 0, v 0 in the measuring plane at z m are related to the corresponding angles via the geometry (distances) of the nozzle : Projected radii are related to radii at S2 by a magnification factor as though projected by a point light source at z 0 : The fluence/proton at the field point is a double integral. f is either the Gaussian or the Molière 2D probability density : Double Scattering Formalism

Scaling So far we have discussed projection. To flatten the dose, we need to find some dependence of scattering on projected radius v 0 (q) that will render Φ(r) constant out to some design radius. Suppose this has been done. It is obvious from the form of the integral that an optimum solution will remain optimum if all quantities in the MP are multiplied by the same constant. That is equivalent to saying that any one quantity can be taken as unity and the others scaled accordingly. We choose for this quantity the design radius and define the corresponding solution as the generic solution. In other words the generic solution is any optimum solution for a design radius of 1 cm. Therefore all projected quantities in the generic solution are distances of the order of 1 cm, since either much larger or much smaller distances yield tiny scattering probabilities. All authors use scaling in one form or another, but they differ in their choice of the magic variable.

The Useful Radius We need to define one more scaled quantity in the MP. If optimization succeeds, we will have a dose which is perfectly constant out to 1 cm. However, there is usually a finite tolerance for dose uniformity which means that the dose is acceptable out to a larger scaled radius, say 1.3 cm, which we call the useful radius. It varies from one design to another and must be found numerically. If we have somehow found a generic solution and its useful radius, we can adapt it to any reasonable real-world situation by ‘unscaling’ it. Any efficient beam line design program works that way.

Outline methods: annulus, dual ring, contoured design procedure: projection, scaling, useful radius de-scaling: converting back to physical quantities optimizing the generic shape beam steering: skewness and all that summary

Model System for the Next Slide Assume we have a satisfactory generic double scattering solution, and that we have actual positions for S1 and S2 relative to the patient. The problem is to find the actual thicknesses of Pb and plastic to realize that solution, if possible.

Back to the Real World 1.Multiply all generic quantities by (radius desired/generic useful radius) 2.Assuming a reasonable z 0, covert the new u 0 and v 0 back to angles by multiplying by the appropriate distance. Convert new r i to radii at S2 by multiplying by the magnification factor. Let j =1 (first mod step). 3.From depth of penetration for that modulator step, find T into the water tank (out of S2). 4.If j = 1 : find the thickness of lead at center required for that output energy and the desired scattering at center. Then find lead and Lexan at each radius to r max for desired scattering at that radius and same energy loss as center. S2 is designed. Now (any j) compute energy loss in S2. 5.Knowing energy in and out of S1 as well as the desired scattering angle. Solve the binary degrader problem to find Pb and Lexan. S1 j is designed. 6.Unless this is the last mod step, increment j and go to 3. 7.Iterate the whole computation over z 0 (converges quickly). Given a generic solution, energy in, actual distances and a desired SOBP (depth and modulation), design a physical modulator and contoured scatterer.

Outline methods: annulus, dual ring, contoured design procedure: projection, scaling, useful radius de-scaling: converting back to physical quantities optimizing the generic shape beam steering: skewness and all that summary

Optimization First we need to choose a functional form for the scattering profile that we wish to optimize. We use a cubic spline at fixed radii: only the y values are adjusted during optimization. This form can describe either a bell-shaped function (top) or the Siebers variant (bottom). The program reads starting values for the spline from the input file. If instructed to optimize, it uses either a grid search or the Marquardt algorithm (see Numerical Recipes).

Optimization Goals 1.dose uniformity: fluence constant out to some radius 2.efficiency: fluence within useful radius as large as possible 3.energy loss: total thickness of scatterers as small as possible 4.compensation: generic solution can be energy compensated 5.practicality: the physical scatterer is easy to fabricate Present day techniques (including NEU) only guarantee #1 and even that depends on starting conditions. Other goals must be attained ‘by hand’. For #4, we need a generic solution with max/min ≤ 4/1. Our best generic solution so far (next slide) has 45% efficiency. Placement of S2 is another variable. Too far upstream: hard to fabricate and sensitive to beam. Too far downstream: larger energy loss, transverse penumbra and unwanted neutron dose to the patient.. Comprehensive optimization would consider at least 5 goals:

A physical scatterer derived from one of our best generic solutions (parameters given in NEU User Guide). Efficiency is 45%, energy loss the lowest we have found so far, but better profiles may exist. Our Personal Best

Outline methods: annulus, dual ring, contoured design procedure: projection, scaling, useful radius de-scaling: converting back to physical quantities optimizing the generic shape beam steering: skewness and all that summary

Active Beam Steering We noted that dose flatness in double scattering is very sensitive to the centering on S2 of the Gaussian from S1. In other words, the beam centroid after scattering in S1 must project to the center of S2. The beam direction at S2 matters much less. In a gantry it is impractical to reduce mechanical and magnetic errors to the required level. In addition, magnets may drift slightly with time. Therefore active beam steering is needed and the feedback loop should remain active during irradiation. The error signal is obtained from a large segmented IC. At HCL this IC was near the patient where the transverse dose was already flat, so the measured slope of the flat part of the distribution (ideally 0) was used directly. In the IBA gantry (Burr Center) the IC is near S2 where the dose is not yet flat, so an indirect method is required.

Roughly to scale but transverse dimensions ×10. IC2 has 32 strips in x and y which measure projected transverse dose fairly close to S2. Before the nozzle was built we did a Monte Carlo study of beam steering possibilities. IBA Nozzle Layout

(right) Transverse dose at the patient with a deliberate steering error of just 1 mm at S2, resulting in the overall tilt. At the same time S1 is 20% too strong resulting in the ‘dished’ appearance which we quantify as curv. (left) Transverse dose at the monitor IC. There is an obvious asymmetry as well as a slight distortion in the curvature. Statistical analysis of this distribution shows its skewness is proportional to tilt and kurtosis is proportional to curv. Error Signal

In case you forgot about skewness and kurtosis, here are the first five moments of a distribution as conventionally defined. Crucially rms, skewness and kurtosis are defined relative to the mean so the IC’s transverse position is relatively unimportant. Moments of a Distribution

The MC study predicted skewness would be proportional to tilt over a wide range and therefore is a good error signal. A subsequent study (D. Prieels, ‘Clinical performance in double scattering,’ IBA technical memo 1999) confirmed this. Kurtosis is less useful because of quantization ‘noise’, but S1 errors are ruled out by the hardware. Since the skewness must be held to ≤1% or so, extremely stable current-measuring electronics is needed for the strip IC. Skewness Proportional to Tilt

Given an error signal, the feedback system itself requires some attention. At HCL the loop was closed except when the detected dose was below some threshold. Thus each treatment started with the final values from the last one. The bars shows the HCL system response to a 12% (huge) tilt error. Even this is corrected in 3 seconds. The feedback is digital with short time constants so it is easily modeled. Open squares show the predicted response. Settling Time (HCL)

Monte Carlo simulations can also be used to study the effect of other beam parameters (size and divergence at S1). If the beam is divergent the transverse dose will be slightly ‘dished’. That can be fixed by decreasing S1 or moving it nearer S2. You should always be prepared to make small adjustments to the computed S1. Effect of Beam Emittance

Outline methods: annulus, dual ring, contoured design procedure: projection, scaling, useful radius de-scaling: converting back to physical quantities optimizing the generic shape beam steering: skewness and all that summary

Summary and a Comment We have described a variety of double scattering techniques which are more efficient and waste less energy than single scattering. In our opinion, the compensated contoured scatterer is best. Projection and scaling allow one to define a generic solution which, once found, can be rescaled to any reasonable physical requirement. The process of finding the ‘best’ generic solution still leaves something to be desired. Some good solutions are known, but better ones may exist. The sensitivity of double scattering to beam misalignment necessitates an active beam steering system. The Burr Center system has been described here and the HCL system is detailed in the lecture on ionization chambers. The position of S2 along the beam is a design choice that must be made by hand, with tradeoffs. A given S2 may be moved along the beam line (‘zoom’) to create larger or smaller fields. Combining scattering and magnetic scanning in the same nozzle forces undesirable tradeoffs. For instance, in the Burr Center nozzle S2 is far downstream, to make room for the scanning magnets. If it were further upstream the lateral penumbra would be sharper, the depth of penetration greater, and unwanted neutron dose to the patient somewhat smaller.