FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.

Slides:



Advertisements
Similar presentations
Mass Spectrometry Chem 3500 Mass Spectrometry Y. Zhao Lecture 2 A Glimpse of Mass Spectrometry Began from 1920s: Aston, Dempster, Thompson MS is often.
Advertisements

Chapter 20 Molecular Mass Spectrometry Mass spectrometry is capable of providing information about (1) the elemental composition of samples of matter.
Mass Spectrometry Mass spectrometry involves ionizing a compound, separating the ions that result on the basis of mass to charge ratio (m/z) and obtaining.
Module 2: Analytical methods and separation techniques.
Mass spectroscopy. In a typical MS procedure:  1- a sample is loaded onto the MS instrument, and undergoes vaporization.  2- the components of the sample.
Mass Spectrometry Dr. Karin Habermehl-Cwirzen. Mass Spectrometry measures individual molecules Definition Mass spectrometry is a method in which a sample.
12-1 Molecular Mass Spectroscopy Molecular structure Composition of mixtures Molecular mass spectra Ion Source Mass Spectrometers Applications.
Mass Spectrometry The substance being analyzed (solid or liquid) is injected into the mass spectrometer and vaporized at elevated temperature and reduced.
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
Chapter 20 Molecular Mass Spectrometry. Introduction... Mass spectroscopy is perhaps one of the most widely applicable of all the analytical tools available.
LC-MS Based Metabolomics. Analysing the METABOLOME 1.Metabolite Extraction 2.Metabolite detection (with or without separation) 3.Data analysis.
Atomic Mass Spectrometry
Atomic Mass Spectrometry
Mass Spectrometry.
Molecular Mass Spectrometry
LC-MS Lecture 7.
Mass Spectroscopy Quantitative Chemical Analysis Harris, 6th Edition
INSTRUMENTAL ANALYSIS CHEM 4811
LC/MS WORKSHOP IOWA STATE UNIVERSITY Kamel Harrata  Instrument Description  Data Acquisition  Data Processing.
Mass Directed Purification from Teledyne Isco
Mass Spectroscopy 1Dr. Nikhat Siddiqi. Mass spectrometry is a powerful analytical technique that is used to identify unknown compounds, to quantify known.
6. Mass Spectrometry Adv. Inst. Techs. How does it work? a very small amount of sample is bombarded by a beam of high energy (usually electron beam) produces.
Russell Rouseff FOS 6355 Summer 2005 What is Mass Spectroscopy Analytical Chemistry Technique Used to identify and quantify unknown compounds Can also.
Proteomics Informatics – Overview of Mass spectrometry (Week 2)
<Instant Notes, D. Kealey & P.J. Haines>
Mass Spectrometry Mass spectrometry (MS) is not true “spectroscopy” because it does not involve the absorption of electromagnetic radiation to form an.
William H. Brown & Christopher S. Foote
Mass Spectrometry Chapter 14 Chapter 14.
Mass spectrometry and the PITZ Sven Lederer Technisches Seminar
Molecular mass spectrometry Chapter 20 The study of “molecular ions” M + e -  M e -
Mass Spectrometry Mass spectrometry involves ionizing a compound, separating the ions that result on the basis of mass to charge ratio (m/z) and obtaining.
Quantitative Chemical Analysis Seventh Edition Quantitative Chemical Analysis Seventh Edition Chapter 22 Mass spectrometry Copyright © 2007 by W. H. Freeman.
1 Chemical Analysis by Mass Spectrometry. 2 All chemical substances are combinations of atoms. Atoms of different elements have different masses (H =
Based on ionization of gas phase molecule followed by analysis of the masses of the ions produced. The Mass Spectrum: Graph of ion intensity versus mass-to-charge.
MS Intro. MS requires gas-phase ions, why? MS uses magnetic and electric fields to control the path of a compound based on mass to charge ratio (m/z)
For all CHEM5161 students: The first day of class for CHEM5161 (Analytical Spectroscopy) will be on TUE Sept 4 (following Labor Day). There will be no.
Ionization energy?. Ionization energy? EI Ionization??
F LORIDA I NTERNATIONAL U NIVERSITY Advanced Mass Spectrometry Piero R. Gardinali/Yong Cai/ Bruce McCord Revised on August 23, 2009.
Separates charged atoms or molecules according to their mass-to-charge ratio Mass Spectrometry Frequently.
Lecture 4b Mass Spectrometry.
Mass Spectrometry Relative atomic masses and the mass of individual isotopes can be determined using a mass spectrometer. The principle behind mass spectrometry.
WELCOME STUDENTS Mobile : Skype: aamarpali.puri.
Chem. 133 – 4/26 Lecture. Announcements Return graded quiz and additional problem Lab – Lab report deadlines (2:4 – Thursday) Today’s Lecture – Mass Spectrometry.
Chapter 29 Mass Spectrometry. 29 A Principles of mass spectrometry In the mass spectrometer, analyte molecules are converted to ions by applying energy.
Chapter 11 Structure Determination: Mass Spectrometry, Infrared Spectroscopy, and Ultraviolet Spectroscopy.
Mass Spectrometry u Chapter 12 Chapter 12.
Yonsei Proteome Research Center Peptide Mass Finger-Printing Part II. MALDI-TOF 2013 생화학 실험 (1) 6 주차 자료 임종선 조교 내선 6625.
Metabolomics Part 2 Mass Spectrometry
CHEM133 Mass Spectrometry: Lecture 1
Mass Spectrometry makes it possible to measure protein/peptide masses (actually mass/charge ratio) with great accuracy Major uses Protein and peptide identification.
Ionization techniques, Mass spectra and MS-Instrumentation
Mass Spectrometry 101 (continued) Hackert - CH 370 / 387D
Components of Mass Spectrometer
Proteomics Informatics – Overview of Mass spectrometry (Week 2)
Chem. 133 – 4/13 Lecture.
Mass Spectrometry 101 Hackert - CH 370 / 387D
Mass Spectrometry Mass spectrometry involves ionizing a compound, separating the ions that result on the basis of mass to charge ratio (m/z) and obtaining.
Mass Spectrometry Obaid M. Shaikh.
Mass Spectrometry Courtesy
Metabolomics Part 2 Mass Spectrometry
Mass Spectroscopy. Mass Spectroscopy Mass Spectrometry Most useful tool for molecular structure determination if you can get it into gas phase Molecular.
Instrumental Chemistry
Instrumental Chemistry
Lecture 22 Introduction to Mass Spectrometry Lecture Problem 7 Due
Mass spectrometry Detection of ionic species to determine the molecular weight of and obtain structural information on a molecule.
Mass Spectrometry THE MAIN USE OF MS IN ORG CHEM IS:
Department of Chemistry St. Joseph’s College (Autonomous)
M.Prasad Naidu MSc Medical Biochemistry, Ph.D,.
Presentation transcript:

FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On

Overview of Mass Spectroscopy: Goals: Terminology Major parts of a mass spectrometer Ionization techniques; advantages & disadvantages Mass analyzers; advantages & disadvantages PurIon Mass Spectrometer overview Why use PurIon? FC-MS overview

Terminology ■ daltons (Da) ■ Unified atomic mass unit (u)- same as Da, based on 12 C ■ amu (atomic mass units)- archaic, technically based on 16 O, but people still use it… ■ m/z- charge-to-mass ratio, may see as m/q (older literature)

Summary of mass spectrometry ■ Analytes are converted to gas-phase ions (source). ■ The ions are separated by their mass-to-charge ratios (m/z, Analyzer) and are detected (Detector). ■ Relative ion current (signal) is plotted versus m/z to produce a mass spectrum.

More Terms that I chuck around casually ■ M: mass of a given molecule ■ [M+H] + : mass of a molecule with a proton; carries a positive charge ■ MeOH: methanol ■ MeCN, ACN: acetonitrile ■ EtOAc: ethyl acetate

Isotopes ■ Same element (determined by number of protons) ■ Different number of neutrons ■ Changes mass of atom (and molecules containing this atom) ■Hydrogen ( 1 H, 2 H deuterium, 3 H tritium) ■Carbon ( 12 C, 13 C, 14 C) ■Chlorine ( 35 Cl, 37 Cl; ~75:25) ■Bromine ( 79 Br, 81 Br; ~50:50)

Mass ■ Masses based on 12 C= ■ Other elements/isotopes do not have integer masses ■ 1 H= ■ 16 O= ■ 14 N= ■ 79 Br=

Mass ■ Nominal mass= integer mass using most abundant isotope; the number we usually state in conversation ■ Monoisotopic mass: Sum of atomic masses using the most common isotope of each element in a molecule ■ Exact Mass: used by some chemistry software; sum of atomic masses of most common isotope when no isotopic species specified. ■ Average mass: sum of average atomic masses in a molecule

Mass Spectrometer block diagram Inlet Source Region Mass AnalyzerDetector Vacuum system

■ Roughing pump ■ Turbo molecular pump ■ Vacuum needed: ■ To avoid further reactions (fragmentation, reactions, etc.) ■ Increase mean-free path (maintain ion energy)

Mass analyzers ■Single quadrupoles (MS) ■Triple quadrupoles (MS/MS) ■Ion traps ■Time-of-flight (TOF) ■Sector analyzers ■Hybrids (ex. QTOF) ■and more hybrids (magnetic sector TOF)

Types of Mass Analyzers: Sector ■ Separate ions by charge-to- mass ratio (m/z; m/q) ■ Generally have a electrical field sector (not shown) to focus ion energies ■ Classical mass spectrometer ■ Good resolution, dynamic range ■ Large, higher cost

Types of Mass Analyzers: Time-Of-Flight (TOF) ■ High mass range ■ Well suited for pulsed ionization methods (MALDI) ■ Requires pulsed ion injection or ion beam switching “Drift” area

Fourier Transform analyzers ■ Ions drift into area of constant magnetic field- ions move in circular motion ■ Use oscillating electric field to “excite” ions ■ Detect ions by their cyclotron radiation, use Fourier transform to obtain masses Highest mass resolution Expensive analyzer

Quadrupoles ■ Less expensive; compact design; low scan times, very common ■ Limited resolution, not suited for pulsed ionization methods ■ Scan the electric fields/ frequency to scan mass range The correct combination of AC and DC electric fields allow resolution of ions by their m/z ratio

Triple Quadrupoles (MS-MS) Select ion of interest Collision induced disassociation Fragment mass analyzer Q1Q2 Q3 Select ion of interest in Q1 Fragment this ion in Q2 Scan Q3 for fragment masses; fragmentation pattern used to deduce original ion structure

Ion traps ■ Trap ions for other purposes ■ Quadrupoles ■ Linear ion traps ■ Orbitraps ■ Some used for analyzers as well as traps

Hybridized techniques ■ Two or more m/z analyzers of different types ■ QTOF: Triple quadrupole, but the last quadrupole replaced by a TOF analyzer (Quadrupole TOF) ■ Provide different information on a molecule ■ Improve signal-to-noise

Ionization Techniques ■ Required to put charge on molecule ■ While ionizing, get molecule into gas phase ■ No ions- no mass spectrometry!!

Ionization Techniques ■ Atmospheric Pressure Ionization (API) ■ Electron and Chemical Ionization (EI/CI) ■ Photo-ionization ■ Matrix-assisted laser desorption (MALDI) ■ Fast atom bombardment (FAB)

Atmospheric Pressure Ionization ■ “Soft” techniques- reduced fragmentation, more molecular ions ■ Useful for liquid chromatography ■ ESI, APCI ■Ionize compounds ■Remove solvents ■Get compounds into analyzer

Electrospray ionization (ESI)- Used on PurIon

ESI Solvents & Additives ■ Solvents ■ Water ■ Acetonitrile ■ Methanol ■ Ethanol ■ Propanol ■ 2-propanol ■Additives ■ Acetic Acid ■ Formic Acid ■ Ammonium hydroxide ■ Ammonium formate * ■ Ammonium acetate * * <= 10 mM What happened here?!

What’s with the solvent additives? ■ Help charge the analyte ■ Acids add protons (positive charge) ■ Bases remove protons (negative charge)

ESI Solvents- Use with care ■ Trifluoroacetic acid ■ Strong ion pair causes neutral molecule? ■ Still, commonly used for LC-MS, some TFA runs seem Ok ■ Triethylamine- may suppress less basic compounds

Atmospheric Pressure Chemical Ionization (APCI)- Used on PurIon ■ Heated probe evaporates solvent ■ Corona discharge places charge on molecules ■ Tetrahydrofuran- very flammable

Compound Ionization Technique Map

Other Ionization- APPI ■ Atmospheric Pressure Photo-Ionization ■ Light “kicks off” an electron, charging molecule

Other ionization- Electron Ionization ■ Formerly known as “electron impact” ■ Molecule charged in vacuum ■ Not compatible with LC ■ Fragments molecules

Other ionization- Chemical Ionization ■ Primary ionization: ■ CH 4 + e -  CH e - ■ Secondary reagent ion ■ CH 4 + CH 3 +  CH CH 3 ■ Product Ion Formation ■ M + CH 5 +  CH 4 + [M+H] + (protonation) ■ M + CH 3 +  CH 4 + [M-H] + (proton abstraction) ■ M + CH 4 +  CH 4 + [M] + (charge exchange)

Other ionization- MALDI ■ Matrix Assisted Laser Desorption/Ionization ■ UV LASER ablates matrix & compound ■ Matrix transfer proton(s) to molecule ■ Commonly used with TOF (pulsed LASER) ■ Used with macromolecules, proteins, bacteria, viruses

Other ionization-DESI, DART ■ Desorption ESI ■ Direct Analysis in Real Time ■ Useful for QC, forensic analysis

Detectors ■ Faraday cup ■ Ions hits cup ■ Enough ions generate a measurable charge ■Electron multiplier ■Used in PurIon ■Amplifies signal by generating electrons

PurIon system

Get Sample into Mass Spectrometer ■ Split/dilute sample ■ Solvent good for ionization (generally methanol) ■ Consistent delivery ■Tubing with restrictions ■“MRA” valve (used in PurIon) ■Both use “make-up” or “carrier solvent” pump

Fluid Interface

Why use FC-MS? Traditional open-access LC/MS workflow Save steps Save time Move right into the next step

Advantages for the chemist ■ Collect only the desired compound(s) ■ Verification the correct compound is being collected ■Ignore previously known compounds (natural products, reverse engineering)

Other uses for PurIon- Flow Injection Analysis Use “Method Development Screen No flow from CombiFlash Useful for reaction monitoring 1 mg/20 mL or less

Chemistry ■ Isotopes ■ Nitrogen Rule ■ Fragmentations/ rearangements ■ Adducts ■ Multiply charged ions ■Note: Many of the rules written for electron ionization Parent peak is M + ■ESI parent is usually [M+1] ■What does this mean?!

Nitrogen rule ■ [M+1] even: odd # of nitrogens ■ [M+1] odd: 0 or even # of nitrogen ■ Applies only to the parent ion!! What is the [M+1] ? Is it even or odd?

Nitrogen rule ■ Synthesized compound has 2 nitrogens ■ See peak at m/z=168 ■ Is this a fragment or a parent ion? ■ Why?

Mass Spectroscopy Chemistry ■ Ions- what we produce ■ Rearrangements- move the charge someplace else ■ Adducts- “share the charge” ■ Fragments- make both charged and uncharged stable pieces ■Lower the energy of the molecule!!!

Ions ■ Multiply charged species ■ What m/z would they appear at? ■ Probably not common on small molecules 93, 94

Fragmentations, rearrangements ■ Generally, fragments occur near heteroatoms (N, O, S) ■ Also can occur with “good” leaving groups, stable ions

Fragmentation case study ■ [M+1] + = 185 ■ Base peak = 168 ■ What drives this reaction?

Rearrangement case Study ■ λ max nm ■ [M+1] = 141 Da expected ■ Solvent system Hexane/EtOAc ■ User advised to use a range of masses, would get weak m/z 123

Rearrangement case Study Not charged- not seen. Major product (loss of another hydrogen between methyls Very minor product- charged, m/z=122 or 123 (depending on loss of H) H + binds to non- bonding electrons on oxygen

Good leaving groups ■ -NH 2 (leaves as ammonia) ■ -OH (leaves as water) ■ COOH (leaves as CO 2 ) ■ Look for increase in conjugation ■ Look for easily formed, stable molecules

Adducts ■ Bond to the molecule- usually detected as [M+H+Adduct] ■ May be more intense than [M+1] ■ May occasionally see dimers [M+H+M] +

Adduct List

Adduct example & sources ■Solvent ■Glassware ■Syringe

Adducts- potential confusion ■ Sample dissolved in methanol- use method development, [M+MeOH+H] observed. Purification in hexane/ethyl acetate- will you see adduct? ■ Sample run on LC-MS mobile phase water/MeCN, [M+MeCN+H] observed. Purification in hexane/ethyl acetate- will you see adduct? ■ Use a range that covers [M] through [M+adduct]

Another Adduct Example ■ Solvent system Hexane/EtOAc ■ m/z=141 expected ■ Carrier = MeOH/0.1% formic acid