Getting the YSI Sonde Ready For Use

Slides:



Advertisements
Similar presentations
Analyze-Immediately Parameters pH Analyze Immediately Within 15 minutes of collection: pH Residual Chlorine Temperature Chlorine Dioxide Dissolved Oxygen.
Advertisements

PREPARATION OF INFANT FEEDS Chapter 27 Doris Corkin and Andrea McDougall.
SECTION 11 LEVEL B SUIT OUT. SECTION 12 AIR MONITORING.
Sonde Maintenance and Care March 2014 Sonde Maintenance and Care 10 th Annual Red River Basin Water Quality Monitoring Training March 5, 2014 Bruce Paakh,
Salk Institute Mobile Lab Gel Electrophoresis Teacher Instructions Edvotek Set Up 12 groups.
IC Controls 855 & 802 ppm Dissolved Oxygen System pH / ORP Conductivity Dissolved Oxygen Chlorine Standards R1.0 © 2004 IC CONTROLS.
Salk Institute Mobile Lab Gel Electrophoresis Teacher Instructions VWR Set Up 12 groups Mira Costa kit.
Use and Maintenance of Micro-pipettes
Multi-parameter Water-Quality Probes (YSI/Hydrolab) Lecture 4.
Cleaning of the Optical Block, Critical Orifice, and Focusing Nozzle of the Brechtel Model 1710 Mixing CPC.
Chapter 10 Control Loop Troubleshooting. Overall Course Objectives Develop the skills necessary to function as an industrial process control engineer.
Tool removed during cycle Fault #2 Conditions for setting Tool cocked prox switch goes open during cycle AND force on load cell drops below limit in fault.
Abaxis Piccolo xpress The piccolo xpress™ chemistry system is compact and easy to transport. The system consists of a portable analyzer and single-use.
ThermexWELD Exothermic Welding Training Guide
Salk Institute Mobile Lab Gel Electrophoresis Teacher Instructions BioRad Set Up 8 groups Grossmont Kit.
IC Controls 865 ppb Dissolved Oxygen System pH / ORP Conductivity Dissolved Oxygen Chlorine Standards ppb Galvanic Dissolved Oxygen.
WORKING PRINCIPLE OF DO AND pH PROBE Prepared by: Tan See Yean Yuganesh a/l Murugiah Jelita Bungin Nazzira Md Zin.
AquaGlo Water Detector Kit For Detecting Water in Jet Fuel
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering NRP 3: Let’s get started.
Standard Operating Procedures l At l Call for help at Orlando Office Tallahassee.
Water Quality Supplies Waste bottle 1413 solution pH 10 solution pH 4 solution Solution cups Water sample cup Hanna Meter.
AccuMax Multi-Point Injection Mechanics
Soil Water Potential Peter Cull ICT International.
Kuskokwim Watershed Council Water Quality Training Workshop
Acids and Bases Biotechnology I. Life Chemistry  Based on water  Cells contain 80-90% water  Proper pH essential to ALL living systems Plants cannot.
IC Controls Quality Water Solutions for pH pH / ORP Conductivity Dissolved Oxygen Chlorine Standards R1.0 © 2004 IC CONTROLS Concerns.
Tel: +44 (0) Turbidity Application and Product Data.
Overview of Continuous Water-Quality Monitoring. Purpose of Monitoring Define the objectives of the water quality monitoring project 1. Environmental.
Maintenance of Machine tools
1 pH observed with Honeywell pH electrodes Current Plan for Calibration and Cleaning Gel replacement Checking with Buffers pH: SATURN-03; SATURN-04.
Gauge Operation and Software by Scott A. Ager. Computer Recommendations 750 MHz Pentium III 64 Meg SRAM 40 Gig Hard Drive 1024 x 768 graphics CD Writer.
Getting Ready Gather tools, oil and filter. If you plan to change your oil regularly, consider investing in jack stands, a socket set and an oil drain.
Horticulture II - Landscape UNIT C LANDSCAPE INSTALLATION AND MAINTENANCE.
The Larsen Autoblocker Calibrating Your System for Maximum Productivity and Quality.
Free Residual Chlorine Measurement
Spring Hill Assembly1 InnoSys Park Brake Tool Master Control Panel Screens 1 3-Mar-2008.
Dye Solar Cell Assembly Glass handling Titania deposition Sintering process Sensitizer impregnation Platinum deposition Sealing electrodes Electrolyte.
QC - THE OUT-OF-CONTROL PROBLEM 謝 昆 穎 Assistant Product Manager, CDG BIO-RAD Laboratories, Taiwan Branch.
Preventative Maintenance For Water Treatment Systems Presented by:
Field Equipment Calibration PH, Temperature, Conductivity, Turbidity, Dissolved Oxygen, Free & Total Chlorine Gregory Langland CCSF.
Basics of Medical Equipment Lecture 5
BioTector COD TOC TN analyzer
Let’s watch a DVD… DVD Instructor Notes
Freestyle Precision Pro
Let’s watch a DVD… DVD Instructor Notes
Quick Start Guide for Flex Flow Pump
EMSA 22 Lab Module 1 Field Equipment Calibration
The Step by Step Charging Techniques
Cleaners Only use cleaners that are: Stable Noncorrosive Safe to use
Conductivity Lecture.
Instrument Maintenance
Fault detection Lecture (1).
StatSensor® Creatinine Operator Training Program
Gel Electrophoresis Teacher Instructions BioRad Set Up 12 groups
How to: Walchem W600 Calibration Introduction Calibration Overview
Use and Maintenance of Micro-pipettes
 .
Instructor Notes Play the “Cleaning and Sanitizing” section from the Facilities, Cleaning and Sanitizing, and Pest Management DVD
How to: Walchem W600 Calibration Introduction Calibration Overview
Multi-Ion Kit Product Demonstration
Electrode Training HORIBA Instruments, Inc. Pamela Millett Scientific
Overview Menus & Settings
Let’s watch a DVD… DVD Instructor Notes
Instructor Notes Play the “Cleaning and Sanitizing” section from the Facilities, Cleaning and Sanitizing, and Pest Management DVD
Let’s watch a DVD… DVD Instructor Notes
Bedside glucose XceedPro Meter
AquaGlo Water Detector Kit For Detecting Water in Jet Fuel
Bedside glucose XceedPro Meter
URINALYSIS DIPSTICK TESTING
School Based Enterprise: Service
Presentation transcript:

Getting the YSI Sonde Ready For Use

Quality Data Proper Maintenance & Calibration Keep all electrical contacts dry Proper removal / installation of probes Good DO membrane installation Use Good Laboratory Practice Proper Rinse method between buffers/standards Pre-rinse…using small amount of calibration solution Allowing time for Equilibration Quality Calibration Understanding Sensor Diagnostics

Probe Installation/Removal Review Handout DISSOLVED OXYGEN ISE 4 ISE 3 4 3 5 CHLOROPHYLL TURBIDITY C T ISE1 / ISE2 (pH / ORP) COND/TEMP ISE 5

Reconditioning DO Probe for Maintenance and/or High DO Charge Remove membrane, rinse probe in clean water, and dry off probe surface Use the YSI 6035 kit to service electrodes (note that the sanding disc must be stroke parallel with the electrodes) (Care & Maintenance Section) Rinse probe surface area with clean water and wipe away residue Re-membrane probe and check DO Charge (after 5 minutes of probe burn-in). NOTE: Install new O-RING If reconditioning DO Probe does not reduce High Charge, repeat conditioning. If it does not help, then replace probe or contact Technical Support at YSI for further assistance.

DO Probe Service 2 Dirty Anodes 1 Dirty Anode Serviced Probe

Membrane Installation / Quality Data Prepare the Probe with KCl

Prepare to Install Membrane

Finish Installation of Membrane

Keep Membrane Tight to Install O-Ring

Cut Excess Material Away

DO Probe Ready to Proceed

Precision Certified Thermometer Temperature No calibration or service available But check accuracy with a Precision Certified Thermometer

Conductivity Service and Calibration Always service conductivity cell before calibration (Care and Maintenance Section) Always use a standard that is a 1000 uS/cm or greater Pre-rinse before calibration Always use fresh standards, and allow time to equilibrate Ensure that conductivity cell is totally submerged in solution and lightly shake sonde to remove any air bubbles Never accept an “Out of Range”

Conductivity Probe Service

Conductivity Calibration Precautions & Diagnostics Calibrate in SpCond (Specific Conductance) Enter correct value e.g.: 1000 uS/cm Calibration value has to be entered as a mS/cm) 1000 uS/cm = 1 mS/cm After calibration “ALWAYS” check “AND” record Cell Constant in “Advanced / Cal Constants Menu” Range 5.0 ± 0.5 (should be very close with each cal.) If Cell Constant is out of range clean and recalibrate SEE CALIBRATION WORK SHEET

pH/ORP Maintenance Tips During calibration; if mV readings are out of specs, or if there is slow response, do cleaning procedure to probe (Care and Maintenance Section) Soak probe for 24 - 48 hrs. in 2M KCl note: Proper Storage will extend the life of the probe If all this fails, replace electrode (Reference first)

pH Probe Service Clean pH Probe Dirty pH/ORP Probe

pH / ORP Diagnostics and Calibration In “Report Menu” enable pH mV Place into a pH 7.0 buffer, let Equilibrate and record mV value. Considered acceptable range: 0 +/- 50 ?? 4.0 buffer should be +177 (from 7 buffer mV value) range  165 to 180 mV 10.0 buffer should be -177 (from 7 buffer mV value) range  -165 to -180 mV

pH The potential is related to the pH by a form of the Nernst Equation pH 4.0 +177 mV pH 7.0 0 mV pH 10.0 -177 mV

pH Calibration Tips Calibrate using at least a 2 Point Cal. (7 buffer always being the first point) Make sure Temperature Probe is in the pH Buffer and allowing time for Equilibration Pre-rinse between buffers Record mV values in all buffers and check probe response between buffers Never accept an “Out of Range” SEE CALIBRATION WORK SHEET

ORP / REDOX Calibration CALIBRATION CHART Temperature Celsius Zobell Solution Value, MV -5 270.0 0 263.5 5 257.0 10 250.5 15 244.0 20 237.5 25 231.0 30 224.5 35 218.0 40 211.5 45 205.0 50 198.5 NOTE: All ISE sensors, requires the pH and temperature probe to be in solution during calibration. Never accept an “Out of Range” SEE CALIBRATION WORK SHEET

Depth Sensor Maintenance Use syringe from Maintenance Kit; this is the only tool to be used on the depth sensor If further cleaning is required, soak in warm soapy tap water, then try syringe again

Depth Sensor Calibration Calibrate in air to zero For more accurate reading in reference to WQ sensors; apply offset to calibration Never accept an “Out of Range” SEE CALIBRATION WORK SHEET

DO Sensor Diagnostics In Report menu; Enable DO Charge confirm that DO charge is 50 +/- 25, if out of specs recondition probe Put into RUN MODE for 15 - 20 minutes before calibration (Probe Burn In) Confirm DO Warm-up Pattern (PROBE HAS TO BE COLD BEFORE STARTING THIS TEST) Check DO Gain “Advance/Cal Const. Menu” (After Calibration) should be 1 within a range of .7 to 1.5

Dissolved Oxygen Calibration Calibrate with probe cold OR fully warmed up ?? (RS232 auto sleep enabled) ; UNATTENDED (RS232 auto sleep disabled) ; SPOT SAMPLING Membrane and temperature sensor must be dry Probe must be in a water saturated air, or air saturated water Enter current barometer reading (mm/Hg) Never accept an “Out of Range” SEE CALIBRATION WORK SHEET

Optic Probes Service and Calibration Turbidity Chlorophyll Rhodamine BGA-PC…Fresh to Brackish Water BGA-PE…Brackish to Marine Water DO

Dirty Optic Wiper Pads

Clean Optic Wiper Pads

Optic Probe Wiper Wiper Motor Wiper Sponge Wiper

Proper Spacing for Wiper Block

Turbidity Probe test and Calibration Finger Test to confirm probe response and wiper parking Calibration process more critical in certain applications: (review handout) Standards to use / value to enter (review handout) For 6600EDS Sonde; remove brush and install clean wiper block with clean pad

Chlorophyll, Rhodamine and BGA Calibration Accept default calibration and use for screening and/or monitoring trends Check for drift and/or slightly enhance accuracy using dye solution to simulate field data Use extractive analysis results from field samples to post calibrate all readings (Chlorophyll, BGA)

Post-calibration of In Vivo Chl Readings

Calibration of Optical DO Calibration method the same as for Rapid Pulse, e.g. air saturated water / water saturated air, entering correct barometric pressure value, etc. (RS232 can be enabled for either application) Can do a 2 point calibration; 1st. Point being in zero oxygen environment; 2nd. Point in saturated environment Diagnostics is DO gain; acceptable range is .75 to 1.25 MIKE Testing has shown the sensor drift is very minimal with the ROX system. Drift has been so minimal that when a drift is observed the user is strongly advised to double check their setup.

Membrane Removal The wiper set screw is loosened with the Allen tool provided, then the wiper is removed.

Membrane Removal The membrane can now be removed. The three membrane screws are removed using the Allen tool provided. MIKE

Membrane Replacement Remove the new sensor membrane assembly from its hydrated container and dry the underside completely. Ensure both sides of the membrane cavity are dry. Be sure to enter the new calibration code values into sonde menu. MIKE Topside Underside

Final test of all Sensors after Calibration Place sensors in tap water (aerated for 20 minutes or more) Check for response time to stabilize on all sensors (this will vary from sensor to sensor), and look for realistic values Review handout on pH response test

REVIEW CALIBRATION WORK SHEET “GLP” FILE TRANSPORT METHODS OF SONDE

Sensor Storage Recommendations “General Rule” Short term (4 weeks or less): Saturated Air Long term (Seasonal): Varies with Sonde Model 2. Sonde Care and Maintenance Review Handout

Questions….