Goldmann Applanation Tonometry

Slides:



Advertisements
Similar presentations
Fitting the New Gas Perm Contacts
Advertisements

AT 900 / AT 870 General Goldmann applanation tonometry is the gold standard for accurate and reproducible measurement of the intra ocular pressure. The.
A Wireless, Implantable Intra- Ocular Pressure Sensor for the Management of Glaucoma Gabriel Simon, M.D. Ph.D. Sept 16, 2008 ESCRS.
Scanning Tonometer for Eye Pressure (STEP™) Problem Measurement of eye pressure by Goldmann applanation tonometry is a technique first embodied in the.
Week 4 Soft lens fitting techniques and care
Technician role in fitting lenses
Goldmann Tonometer.
8.6 Frictional Forces on Collar Bearings, Pivot Bearings and Disks
Advanced Basic Procedure. Handheld Autorefractor Welch-Allyn SureSight Retinomax.
STATICALLY DETERMINATE STRESS SYSTEMS
Measurement of lens power by Lensometer
Physiological optics 5th lecture
Hemoglobin Testing Using the HemoCue. Instrument calibration At the beginning of each survey day, check the instrument accuracy using the calibrator cuvette.
Canadian Ophthalmological Society Evidence-based Clinical Practice Guidelines for the Management of Glaucoma in the Adult Eye.
Schiotz tonometer.
Corneal topography orbscan
OPTOMETRIC EQUIPMENTS IN REFRACTION ROOM
Fluorescein Patterns. General Information Hydrostatic forces from the tear film holds the CL to the eyeHydrostatic forces from the tear film holds the.
Chapter 23 Mirrors and Lenses.
Chapter 23 Mirrors and Lenses. Notation for Mirrors and Lenses The object distance is the distance from the object to the mirror or lens Denoted by p.
Applanation tonometry:
Recent Advances in IOP measurements
The Hong Kong Polytechnic University Optics II----by Dr.H.Huang, Department of Applied Physics1 Interference Conditions for Interference: (i) (  2 
Newton’s Rings Another method for observing interference in light waves is to place a planoconvex lens on top of a flat glass surface, as in Figure 24.8a.
Rigid lens verification and evaluation
Blood pressure. Blood pressure Preparation for measurement.
Chapter 21 & 22 Electric Charge Coulomb’s Law This force of repulsion or attraction due to the charge properties of objects is called an electrostatic.
Lab 3 - Centering. Centering; or the smart way to align centered optical elements and systems This lab will make use of concepts used in the previous.
1.1 Intro to the microscope and Calculating cell size.
© 2010 Pearson Education, Inc. Conceptual Physics 11 th Edition Chapter 28: REFLECTION & REFRACTION Reflection Principle of Least Time Law of Reflection.
3-D Vision One person holds test tube at arms length Other holds pencil in arm upright Try to swing down lower arm to place pencil directly in test tube.
Basic Contact Lenses COT/COMT Review.
INTRAOCULAR PRESSURE LECTURE
Keratometry Walter Huang, OD Yuanpei University Department of Optometry.
Chapter 23 Mirrors and Lenses.
Refraction is the change of direction of a light wave caused by a change in speed as the wave crosses a boundary between materials.
To many people, accuracy and precision mean the same thing: to someone involved in measurement, the two terms should have very different meanings. Accuracy.
Methods of measuring intra ocular pressure. What is intra ocular pressure ? Definition Intra ocular pressure refers to the pressure exerted by the intra.
Optical Manufacturing Solutions 1 FOLLOW THE DIRECTIONS BELOW AFTER INSTALLING A NEW CUTTER BODY WITH POLISHING WHEELS INTO THE 7E TYPE MACHINES. IT IS.
 When light strikes the surface of an object  Some light is reflected  The rest is absorbed (and transferred into thermal energy)  Shiny objects,
Halliday/Resnick/Walker Fundamentals of Physics
Sphere Standards and Standard Spheres Dr. Richard Young Optronic Laboratories, Inc.
Getting the Markings Right
Presents: Diaton Tonometer Tonometry through the Eyelid! September 2006.
Diffraction is the bending of waves around obstacles or the edges of an opening. Huygen’s Principle - Every point on a wave front acts as a source of tiny.
Oliver Findl 1,2, Nino Hirnschall 1,2, Alja Crnej 2 1 VIROS Vienna Institute for Research in Ocular Surgery, Hanusch Hospital, Vienna, Austria 2 Moorfields.
Subject 7: The lens-meter: Measuring Posterior Vertex Power. Description of the lens-meter Calibrating the instrument Measuring Posterior Vertex Power.
Newton’s Rings. AB P Q O Air film Plano convex lens Glass plate Thickness of air film is zero at point of contact O.
Ch 16 Interference. Diffraction is the bending of waves around obstacles or the edges of an opening. Huygen’s Principle - Every point on a wave front.
Surface Rescue Swimmer School Mask, Fins, and Snorkel Mask, Fins, and Snorkel LT 4.2.
1© Manhattan Press (H.K.) Ltd. Young’s double slit experiment Young’s double slit experiment 9.10 Interference of light waves Relationship between x,,
To many people, accuracy and precision mean the same thing: to someone involved in measurement, the two terms should have very different meanings. Accuracy.
Tonometry Objective measurement of IOP based, most commonly, on the forces required to flatten the cornea,or the degree of corneal indention produced.
Direct ophthalmoscopy
TONOMETER.
Lecture 8: Measurement Errors 1. Objectives List some sources of measurement errors. Classify measurement errors into systematic and random errors. Study.
Examination Techniques for Accuracy and Efficiency Astigmatism Detection and Management Options A VOSH-Florida Presentation.
Seeing Things Through Stuff 1)Stare straight ahead at the word “Refraction” on the blackboard. 2)Close your right eye. 3)Hold the index finger of your.
Refractive error introduced in the application of a refractive surgical treatment and in the subsequent response of the ocular tissue Charles Campbell.
Keratometry and Corneal Topography Instructions. Generally the keratometer & corneal topographer are combined into one instrument that does each reading.
Lab 3 - Centering.
Glaucoma.
Ch 16 Interference.
55th ASMR National Scientific Conference
Interference of EM Waves
Keratometry: Main Points so Far
OPTO 4101: Contact lens1- lab L01: Slit lamp microscope
Circles in Perspective - Elliptical Shapes
2/24/2013.
Yakov Goldich MD,Yaniv Barkana MD, Isaac Avni MD, David Zadok MD,
Presentation transcript:

Goldmann Applanation Tonometry Ted Barnett

Introduction Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann, Perkins, Draeger, MacKay-Marg, and Tono-Pen and Pneumatonometer): area of the cornea being applanated held constant, variable for applied.

Principles based on Imbert-Fick law: pressure within a sphere (P) is roughly equal to the external force (f) needed to flatten a portion of the sphere divided by the area (A) of trhe sphere which is flattened: P = f / A applies to surfaces which are perfectly spherical, dry, flexible, elastic and infinitely thin

Principles (cont.) include force of cornea which pushes applanating surface away from eye (N), subtract surface tension of tear film toward the eye (M) since cornea has thickness, consider only flattening of inner corneal area (A1) P = f / A1 + M - N when A1 = 7.35, M and N cancel out so: P = f / 7.35 mm2

Principles (cont.) this internal area achieved when diameter of external area of corneal applanation is 3.06mm at this external diameter, grams of force applied multiplied by 10 is directly converted to mmHg measured pressure is 3% greater than IOP before applanation (not corrected) minimal displacement (0.5ul) of fluid or increase in IOP with applanation, thus unaffected by ocular rigidity

Technique of measurement plastic biprism which contacts cornea creates two semicircles edge of corneal contact is visible after placing fluorescein into tear film & viewing with cobalt blue light manually rotate the dial calibrated in grams, force is adjusted by changing the length of a spring within the device. inner margins of semicircles touch when 3.06 mm of cornea is applanated.

Instructions to patient press head firmly against chin and forehead rest. look straight ahead and fixate on a target (e.g. examiners opposite ear) breathe normally, do not hold your breath blink immediately prior to measurement to moisten cornea.

Measurement (cont.) position patient’s head with forehead rest well above eyebrows, allowing raising of eyebrows. anesthetic & fluorescein (0.25%) ,separately, or as mixture (preserved) placed in inferior cul-de-sac. with maximal illumination of biprism the lamp is moved toward the eye until the tip of biprism contacts the apex of the cornea stop moving forward when limbus shines with light, best observed with naked eye

Measurement (cont.) After contact, semicircles visible through left (or right) ocular. Center in field of view. Adjust vertically until semicircles equal in size. Tension dial adjusted so that inner edge of upper and lower semicircles are aligned. Multiply dial reading (grams of force) by 10 to obtain IOP (mmHg) Read at median over which arcs glide to control for excursions due to ocular pulsations.

Measurement (cont.) If slit-lamp moved too far toward patient the pressure arm will push against a spring which will press against the eye with a low inoffensive force. Mires (flattened area) too large, moving dial doen’t alter appearance. Solution: Draw back until regular pulsation noted and appearance of mires normalizes.

Measurement (cont.) Blue central area represents applanated cornea, green semicircles are fluorescein-stained tears, inner border of ring is demarcation between flattened and non-flattened cornea. Without staining of tears, bright reflection from air-cornea interface is seen; leads to underestimation of IOP. Mires should be approximately 10% of circle width.

Errors in Measurement The fluorescein ring is too wide or too narrow: Too wide: occurs if prism not dried after cleaning or lids touch prism. Overestimates IOP. Solution: dry prism Too narrow: inadequte fluorescein concentration may cause hypofluorescence. Underestimates IOP. Solution: patient blinks or additional fluorescein added.

Errors (cont.) thin corneas produces underestimate thick cornea d/t increased collagen gives overestimate, if d/t edema gives underestimate. inadequate vertical alignment of semicircles leads to overstimate of IOP. distortion d/t irregular cornea influences accuracy, less useful with corneal scarring.

Errors (cont.) squeezing of eyelids, breath holding or other Valsalva maneuvers, pressure on globe, excessive EOM force applied to restricted globe, vertical gaze, tight collars, retreating patient, inaccurately calibrated tonometer. repeated tonometry may induce decline in estimated IOP.

Error d/t corneal curvature increase of 1 mmHg for every 3D increase in corneal power. more fluid displaced under steep cornea, increases contribution of ocular rigidity in overestimating IOP. the steeper the cornea, the more cornea must be indented to produce standard area of contact. >3D astigmatism produces elliptical rather than circular area

Correction for astigmatism With semicircles displaced horizontally, IOP underestimated by 1 mmHg for every 4D of WTR astigmatism, vice versa for ATR astigmatism. To minimize, prisms should be rotated so that axis of least corneal curvature is opposite red line on prism holder (i.e. align negative cylinder axis). Can average reading with vertical and horizontal alignment of prism.

Sterilization CDC recommendation (HIV, HSV, and adenovirus): wipe tip clean and disinfect tip only with bleach (1:10 dilution x 5”, changed once daily). Alternative is 3% H2O2, changed at least twice daily (affects tip less than bleach or ETOH). Alternative #2: wiping tip with 70% ETOH

Reliability Goldmann applanation is standard against which others measured. Good accuracy in gas-filled eyes. Inter- and intraobserver variability (>30% varied by 2-3 mmHg), due to subjective nature of optical endpoint. Assume error of 2 mmHg.

Calibration: Wessels & Oh (1990) Tested tonometers in ophthalmologists offices. 19% outside range of manufacturers specifications (1mmHg of calibration), 4.5% > 2mmHg error. Annual recalibration in 86% of instruments. Practitioners who themselves performed calibration had the most accurate instruments. Less than 15% knew how to perform calibration check. Calibration here done 4 times/year