DESY drift chambers efficiency and track reconstruction

Slides:



Advertisements
Similar presentations
Analysis for J2 chamber Yousuke Kataoka (University of Tokyo) Atsuhiko Ochi, Yuki Edo (Kobe University) 11 / 12 / 2012 Micromegas weekly meeting 1.
Advertisements

E/π identification and position resolution of high granularity single sided TRD prototype M. Târzilă, V. Aprodu, D. Bartoş, A. Bercuci, V. Cătănescu, F.
The performance of Strip-Fiber EM Calorimeter response uniformity, spatial resolution The 7th ACFA Workshop on Physics and Detector at Future Linear Collider.
TRACK DICTIONARY (UPDATE) RESOLUTION, EFFICIENCY AND L – R AMBIGUITY SOLUTION Claudio Chiri MEG meeting, 21 Jan 2004.
1 Calice Meeting 20/9/06David Ward What did we learn from DESY 2005 run? DESY run May CERN run August Data/MC comparisons for ECAL.
Drift Chambers at DESY and CERN Mike Green, Fabrizio Salvatore, Michele Faucci Giannelli.
Status of Tracking at DESY Paul Dauncey, Michele Faucci Giannelli, Mike Green, Hakan Yilmaz, Anne Marie Magnan, George Mavromanolakis, Fabrizio Salvatore.
Drift velocity and tracking Michele Faucci Giannelli, Mike Green, Fabrizio Salvatore.
March 6th, 2006CALICE meeting, UCL1 Position and angular resolution studies with ECAL TB prototype Introduction Linear fit method Results with 1, 2, 3,
© Imperial College LondonPage 1 Preliminary Studies of the Tracking Resolution at DESY Hakan Yilmaz.
Beam profile vs time Analyzed Vx vs Vy distributions vs time for –Run 72 (13:26:02 – 28:25) –Run 73 (13:36:10 – 13:37:30) –Run 74 (43:12 – 44:21) Binned.
Octal ASD Certification Tests at Michigan J. Chapman, Tiesheng Dai, & Tuan Bui August 30, CERN.
Tracking System at CERN 06 and 07 test beams Michele Faucci Giannelli.
Veto Wall Test Hyupwoo Lee MINERvA/Jupiter Group Meeting Apr, 16, 2008.
29 Mar 2007Tracking - Paul Dauncey1 Tracking/Alignment Status Paul Dauncey, Michele Faucci Giannelli, Mike Green, Anne-Marie Magnan, George Mavromanolakis,
Rate and Gain Measurements of the 1-m long GEM detector Aiwu Zhang EIC tracking R&D weekly meeting.
Pion test beam from KEK: momentum studies Data provided by Toho group: 2512 beam tracks D. Duchesneau April 27 th 2011 Track  x Track  y Base track positions.
Preliminary comparison of ATLAS Combined test-beam data with G4: pions in calorimetric system Andrea Dotti, Per Johansson Physics Validation of LHC Simulation.
PERFORMANCE OF THE MACRO LIMITED STREAMER TUBES IN DRIFT MODE FOR MEASUREMENTS OF MUON ENERGY - Use of the MACRO limited streamer tubes in drift mode -Use.
19/07/20061 Nectarios Ch. Benekos 1, Rosy Nicolaidou 2, Stathes Paganis 3, Kirill Prokofiev 3 for the collaboration among: 1 Max-Planck-Institut für Physik,
Measuring the Position Resolution of a COMPASS Drift Chamber Prototype Rojae Wright 1, Ihnjea Choi 2, Caroline Riedl 2, Matthias Grosse Perdekamp 2 (1)
Muon-raying the ATLAS Detector
Performance of a Large-Area GEM Detector Prototype for the Upgrade of the CMS Muon Endcap System Vallary Bhopatkar M. Hohlmann, M. Phipps, J. Twigger,
1 A.Andronic 1, H.Appelshäuser 1, V.Babkin 2, P.Braun-Munzinger 1, S.Chernenko 2, D.Emschernmann 3, C.Garabatos 1, V.Golovatyuk 2, J.Hehner 1, M.Hoppe.
Track extrapolation to TOF with Kalman filter F. Pierella for the TOF-Offline Group INFN & Bologna University PPR Meeting, January 2003.
1 Energy loss correction for a crystal calorimeter He Miao Institute of High Energy Physics Beijing, P.R.China.
Mitglied der Helmholtz-Gemeinschaft Calibration of the COSY-TOF STT & pp Elastic Analysis Sedigheh Jowzaee IKP Group Talk 11 July 2013.
21 Jun 2010Paul Dauncey1 First look at FNAL tracking chamber alignment Paul Dauncey, with lots of help from Daniel and Angela.
CEA DSM Irfu Reconstruction and analysis of ANTARES 5 line data Niccolò Cottini on behalf of the ANTARES Collaboration XX th Rencontres de Blois 21 / 05.
© Imperial College LondonPage 1 Tracking & Ecal Positional/Angular Resolution Hakan Yilmaz.
Progress on the beam tracking instrumentation Position measurement device Tests performed and their resolution Decision on electronics Summary.
RPC Timing Results with Final Splitters Gianpaolo Carlino INFN Napoli The Napoli RPC Group: M.Alviggi, V. Canale, M. Caprio, G.C., R. de Asmundis, M. Della.
Test beam preliminary results D. Di Filippo, P. Massarotti, T. Spadaro.
CALICE Tungsten HCAL Prototype status Erika Garutti Wolfgang Klempt Erik van der Kraaij CERN LCD International Workshop on Linear Colliders 2010, October.
Active polarimeter simulation Suguru Shimizu Osaka University Sep. 1, 2007 JPARC TREK Collaboration meeting at Saskatchewan.
(s)T3B Update – Calibration and Temperature Corrections AHCAL meeting– December 13 th 2011 – Hamburg Christian Soldner Max-Planck-Institute for Physics.
Semi Digital HCAL Data analysis results Robert Kieffer Institut de Physique Nucléaire de Lyon 10/02/
Beam Test of a Large-Area GEM Detector Prototype for the Upgrade of the CMS Muon Endcap System Vallary Bhopatkar M. Hohlmann, M. Phipps, J. Twigger, A.
Construction and beam test analysis of GE1/1 prototype III gaseous electron multiplier (GEM) detector V. BHOPATKAR, E. HANSEN, M. HOHLMANN, M. PHIPPS,
N. Saoulidou & G. Tzanakos1 Analysis for new Period 3 Events N. Saoulidou and G. Tzanakos University of Athens, Department of Physics’ Div. Of Nuclear.
Giuseppe Ruggiero CERN Straw Chamber WG meeting 07/02/2011 Spectrometer Reconstruction: Pattern recognition and Efficiency 07/02/ G.Ruggiero - Spectrometer.
A New Upper Limit for the Tau-Neutrino Magnetic Moment Reinhard Schwienhorst      ee ee
Paolo Massarotti Kaon meeting March 2007  ±  X    X  Time measurement use neutral vertex only in order to obtain a completely independent.
Plots of RPC performance G. Cattani, University of Rome “Tor Vergata” & INFN Roma 2 on behalf of ATLAS Muon Collaboration.
Pattern recognition with the triplet method Fabrizio Cei INFN & University of Pisa MEG Meeting, Hakata October /10/20131 Fabrizio Cei.
SHIP calorimeters at test beam I. KorolkoFebruary 2016.
ΜTPC Reconstruction Status Report (…mostly a late night / weekend project … ) D. A. Jensen March 1, 2007.
ICARUS T600: low energy electrons
Michele Faucci Giannelli
Analysis of LumiCal data from the 2010 testbeam
The GOOFIE detector as a Ternary Mixture Monitor
Activities on straw tube simulation
EZDC spectra reconstruction and calibration
Simulation of Properties of COMPASS Drift-Chamber Prototypes
First results from prototype measurements
LXe Alignment with X-rays
Tracking results from Au+Au test Beam
TOF CALIBRATION DATABASE
Tracking System at CERN 06 and 07 test beams
The Silicon Track Trigger (STT) at DØ
Han Jifeng BESIII MUON GROUP IHEP, CAS
Fake Jet Rejection: Sum Jt
Performance of MDC during New Data Taking
MEG実験アップグレードに向けたSiPMを用いた ポジトロン時間測定器の研究開発
Cosmic ray test of RPC for the ATLAS experiment
Michele Faucci Giannelli
Calibration of DT-MTCC data
ZHH Analysis preliminary results on different detector models
Test Beam models in Mokka
Beam properties for run
Presentation transcript:

DESY drift chambers efficiency and track reconstruction Michele Faucci Giannelli

Michele Faucci Giannelli Outline The gas mixture (Ar, Ethane) in the DCs changed since last year, from 50%-50% to Ar 96% Ethane 4%. The gas is now NON flammable. (no night shift!) Need to proof the DCs continue to work with the new gas. Calculate the track resolution to look at the minimum requirement of hits. 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Setup wire gas Bubbler DC2 DC1 electrons DC3 DC4 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Efficiency method 1 Sum of two parallel wire give this spectrum Fit with a gaus + ax^2 + bx +c 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Efficiency method 1 The area of the Gaussian distribution over the total number of events is the product of the efficiencies of the two wire Last year values ( at 2800 V) X12 Y12 X34 Y34 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Efficiency method 2 But we can control only the voltages so we developed a second method. Now we are looking at the scatter plot of times in the chambers. This is an example chamber 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Beam profile 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Efficiency method 2 The events in the chamber over the total is our efficiency. Last year values ( at 2800 V) DC1 DC2 DC3 DC4 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Drift Velocity The drift velocity is calculated using the formula L is the dimension of the chamber (72 mm) The time sum is obtained from the previous plot This year values is: 0.027 mm/ns 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli DCDigi Processor (1) I developed a Processor for the digitization of the DC signal in MC. This processor is now available, if not ask me. It’s possible to smear a signal according to a chosen sigma. The sigma of each chamber is: where is the sigma of the previous distribution. 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli DCDigi Processor DATA MC 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Track reconstruction A linear fit is performed to reconstruct the track. The fitted point at the front face of the calorimeter is compared with the real point of impact. 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Track resolution Multiple Scattering only (σDC=0) Y X Sigma = 0.66±0.02 Sigma = 0.64±0.02 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Track resolution Smearing of DC hits Y X Sigma = 1.70±0.05 Sigma = 1.69±0.05 17/01/2019 Michele Faucci Giannelli

Track resolution All effects on (MS, DC smearing, misalignment 0.2 μm) Y X Sigma = 1.61±0.05 Sigma = 1.65±0.05 17/01/2019 Michele Faucci Giannelli

Track resolution 3 hits (last chamber missing) Y X Sigma = 2.83±0.07 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Track resolution σall = 1.64±0.05 mm σ4 = 1.96±0.05 mm σ3 = 1.64±0.05 mm σ1 = 1.80±0.05 mm σ2 = 2.89±0.07 mm The number indicate the missing chamber 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Summary The new gas mixture in DC caused a reduction of the efficiency of 15%, still the efficiency was high enough. This has been proved using two methods. The new drift velocity has been calculated. A MC digitization processor has been developed. Now available. The track resolution is 1.64±0.05 mm using 4 points. The resolution is still under 2 mm if you require 3 hits with the last chamber as one of them. 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Backup slides 17/01/2019 Michele Faucci Giannelli

Track resolution 3 hits (1°chamber missing) Y X Sigma = 1.93±0.05 17/01/2019 Michele Faucci Giannelli

Track resolution 3 hits (2°chamber missing) Y X Sigma = 1.61±0.05 17/01/2019 Michele Faucci Giannelli

Track resolution 3 hits (3°chamber missing) Y X Sigma = 1.81±0.05 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Problems Bubbler: the old bubbler didn’t worked, now we are much more confident with the gas flux “34”: some time the TDC read a fake leading edge at 34 ns, to be understand (see next slide). Wire Y of DC3 is not working, readout electronics will be analyzed. The triggered events read by the DAQ are fewer (1/5) than those really triggered. This happens only at high rate (100 Hz). 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Mean vs Voltage Time is increasing, Drift velocity is decreasing. 17/01/2019 Michele Faucci Giannelli

Michele Faucci Giannelli Summary Still some problems to be solved… …but the chambers are working! The average efficiency is lower so we will need more time to collect enough data. This is a good tradeoff for not having night shift! To do: calculate the new drift velocity. 17/01/2019 Michele Faucci Giannelli