Chapter 5 Quality Assurance and Calibration Methods

Slides:



Advertisements
Similar presentations
Quality is a Lousy Idea-
Advertisements

Calibration Techniques
Analytical Method Development and Validation
Instrumental Analysis
Errors in Chemical Analyses: Assessing the Quality of Results
CHEMISTRY ANALYTICAL CHEMISTRY Fall
6-1 Introduction To Empirical Models 6-1 Introduction To Empirical Models.
Result validation. Exercise 1 You’ve done an analysis to the best of your ability using the correct procedure. Is your answer correct? possibly, hopefully.
Interpreting Your Lab Report & Quality Control Results
World Health Organization
Basic Questions Regarding All Analytical & Instrumental Methods (p 17-18) What accuracy and precision are required? How much sample do I have available,
Instrumental Methods: Intro  Types of Instrumental Methods  Fundamental Components of an Instrument  Instruments Measure Voltages and Currents! (“Machines”
Data Handling l Classification of Errors v Systematic v Random.
Quality Assurance Chapter 29. Quantitative Chemical Analysis, Daniel C. Harris, 6 th Edition, New to this edition and a very important topic in industry.
Types of experimental error
QUALITY CONTROL OF PHYSICO-Chemical METHODS Introduction :Validation توثيق المصدوقية.
CALIBRATION METHODS.
Chemometrics Method comparison
Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay.
Quality Assurance.
Quality Assessment 2 Quality Control.
Validation of Analytical Method
Introduction to Linear Regression and Correlation Analysis
Quality assurance of sampling and analytical instruments
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Analytical considerations
Introduction to Analytical Chemistry Dr M. Abd-Elhakeem Faculty of Biotechnology General Chemistry Lecture 7.
HD 2007 Rule Diesel Fuel Sulfur Testing and Sampling Methods and Requirements US EPA Office of Transportation and Air Quality November 20, 2002.
Quality WHAT IS QUALITY
Chapter 5 Errors In Chemical Analyses Mean, arithmetic mean, and average (x) are synonyms for the quantity obtained by dividing the sum of replicate measurements.
Quality Control Lecture 5
Quality Assurance How do you know your results are correct? How confident are you?
5. Quality Assurance and Calibration Quality assurance is We do to get the right answer for our purpose. Have Sufficient accuracy and precision to support.
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
Data Analysis: Quantitative Statements about Instrument and Method Performance.
Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function.
Instrumental Methods: Intro
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
BME 353 – BIOMEDICAL MEASUREMENTS AND INSTRUMENTATION MEASUREMENT PRINCIPLES.
Wenclawiak, B.: Fit for Purpose – A Customers View© Springer-Verlag Berlin Heidelberg 2003 In: Wenclawiak, Koch, Hadjicostas (eds.) Quality Assurance in.
Industrial Technology Institute Test Method Validation & Verification H.P.P.S.Somasiri Principal Research Scientist / SDD-QAD /QM Industrial Technology.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
Chapter 5 Quality Assurance and
EQUIPMENT and METHOD VALIDATION
means to “TO CHECK OR PROVE THE VALIDITY OF” According to FDA – “ The goal of validation is to establish a documented evidence which provides a high degree.
Art PowerPoints Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 05: QUALITY ASSURANCE AND CALIBRATION METHODS.
Demand Estimation & Forecasting
Chapter 7 Demand Estimation and Forecasting
Quality is a Lousy Idea-
World Health Organization
Instrumental Methods: Intro
Quality is a Lousy Idea-
Quality Assurance Chapter 29. Quantitative Chemical Analysis, Daniel C. Harris, 6th Edition, New to this edition and a very important topic in.
What it Means, Why it Works, and How to Comply
Analytical Method Validation
Introduction to Instrumentation Engineering
Why Use Them? By: Marcy Bolek – Alloway
Chapter 1: The Nature of Analytical Chemistry
ANALYTICAL METHOD VALIDATION
World Health Organization
Chapter 7: Demand Estimation and Forecasting
Introduction To Medical Technology
Quality Control Lecture 3
▪Internal quality control:
Chapter 7: Demand Estimation and Forecasting
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Chapter 7 Demand Estimation & Forecasting
Demand Estimation & Forecasting
MGS 3100 Business Analysis Regression Feb 18, 2016
Presentation transcript:

Chapter 5 Quality Assurance and Calibration Methods

Overview Quality assurance Method validation External standards calibration Standard additions calibration Internal standards calibration

5-1: Quality Assurance Panel a shows results for the determination of lead in the same sample of river water. Of 181 labs, 18 reported results more than 50% above and four reported results more than 50% below the certified level of 62.3 ± 1.3 nM. Even though most labs in the study employed recognized quality management procedures, a large fraction of results did not include the certified range.

5-1: Quality Assurance Quality assurance is what we do to get the right answer for our purpose. Use objectives, from which specifications for data quality can be derived. Specifications could include requirements for sampling, accuracy, precision, specificity, detection limit, standards, and blank values. Assessment is the process of (1) collecting data to show that analytical procedures are operating within specified limits, and (2) verifying that final results meet use objectives.

5-1: Specifications Specifications could include: sampling requirements accuracy and precision rate of false results selectivity sensitivity acceptable blank values recovery of fortification (spike recovery) calibration checks quality control samples

5-1: Specifications Sampling requirements: Must collect representative samples, and analyte must be preserved after sample is collected. Otherwise, even the most accurate analysis is meaningless. Quality assurance begins with sampling. Accuracy and precision: Within practical restraints, (cost, time, and limited amounts of material), what level of accuracy and precision will satisfy the use objectives? What rate of false results is acceptable? A false positive implies that the concentration exceeds the legal limit when, in fact, the concentration is below the limit. A false negative implies that the concentration is below the limit when it is actually above the limit.

5-1: Specifications Selectivity (specificity): Sensitivity: distinguish analyte from other species in the sample (avoiding interference). Sensitivity: Capability of responding reliably and measurably to changes in analyte concentration The detection limit of an analytical method must be lower than the concentrations to be measured. Slope of the calibrations curve Example: method below is more sensitive for perchlorate in reagent water than in groundwater.

5-1: Specifications Acceptable blank values Blanks account for: Interference by other species in the sample Traces of analyte found in reagents used for sample preservation, preparation, and analysis Frequent measurements of blanks detect whether analyte from previous samples is carried into subsequent analyses by adhering to vessels or instruments. Three types – method, reagent, and field blanks.

5-1: Blanks Method blank: Reagent blank: Field blank: All components except analyte Taken through all steps of the analytical procedure Subtract the response of the method blank from the response of sample before calculating the quantity of analyte Reagent blank: Similar to a method blank, but it has not been subjected to all sample preparation procedures Field blank: Indicates if analyte is inadvertently picked up by exposure to field conditions

5-1: Matrix Interferences Sometimes, response to analyte can be decreased or increased by something else in the sample. Matrix refers to everything in the sample other than analyte.

5-1: Spike Recovery A spike, also called a fortification, is a known quantity of analyte added to a sample to test whether the response to the spike is the same as that expected from a calibration curve. Spiked samples are analyzed in the same manner as unknowns. If drinking water is found to contain 10.0 mg/L of nitrate, a spike of 5.0 mg/L could be added. Ideally, the concentration in the spiked portion found by analysis will be 15.0 mg/L. If a number other than 15.0 mg/L is found, then the matrix could be interfering with the analysis.

5-1: Spike Recovery

5-1: Specifications Calibration check: Quality control: Analyze solutions with known concentrations of analyte. A specification might, for example, call for one calibration check for every 10 samples. Calibration check solutions should be different from the ones used to prepare the original calibration curve. Helps verify that the initial calibration standards were made properly. Quality control: Samples of known composition are provided to the analyst as unknowns. Helps eliminate bias introduced by an analyst who knows the concentration of the calibration check sample.

5-1: Accuracy and Precision Accuracy can be assessed by: analyzing certified standards calibration checks performed by the analyst spikes made by the analyst analyzing blind quality control samples Gauge precision: replicate samples replicate portions of same sample

5-1: Standard Operating Procedures Written standard operating procedures (SOP) must be followed rigorously to avoid inadvertent changes in procedure that could affect the outcome. It is implicit that everyone follows the standard operating procedures. Adhering to these procedures guards against the normal human desire to take shortcuts based on assumptions that could be false.

5-1: Control Charts A control chart is a visual representation of confidence intervals for measurements having a Gaussian distribution. Warn us when a property being monitored strays dangerously far from an intended target value. Can be used to monitor accuracy, precision, or instrument performance as a function of time.

5-2: Method Validation Method validation is the process of proving that an analytical method is acceptable for its intended purpose. In validating a method, we typically demonstrate that requirements are met for specificity, linearity, accuracy, precision, range, limit of detection, limit of quantitation, and robustness.

5-2: Method Validation Specificity is the ability to distinguish analyte from anything else. Linearity is usually measured by the square of the correlation coefficient for the calibration curve. Types of precision include instrument precision, intra-assay precision, intermediate precision, and, most generally, interlaboratory precision. The “Horwitz trumpet” is an empirical statement that precision becomes poorer as analyte concentration decreases. Range is the concentration interval over which linearity, accuracy, and precision are acceptable.

5-2: Detection Limit (DL) The detection limit (LOD) is usually taken as three times the standard deviation of the blank. The lower limit of quantitation (LOQ) is 10 times the standard deviation of the blank. The reporting limit is the concentration below which regulations say that analyte is reported as “not detected,” even when it is observed. Robustness is the ability of an analytical method to be unaffected by small changes in operating parameters.

5-2: Detection Limit (DL)

5-3: Standard Addition A standard addition is a known quantity of analyte added to an unknown to increase the concentration of analyte. Standard additions are especially useful when matrix effects are important. Use Equation 5-7 to compute the quantity of analyte [X]i after a single standard addition. The initial concentration [X]i, increases to [X]f while the initial signal, Ix increase to Ix+s. [s]f is the concentration of standard added. Equation 5-7

5-3: Standard Addition

5-3: Multiple Standard Additions and Uncertainty For multiple standard additions to a single solution, use Equation 5-9 to construct the graph in Figure 5-6, in which the x-intercept gives us the concentration of analyte. For multiple solutions made up to the same final volume, the slightly different graph in Figure 5-7 is used. Equation 5-10 gives the x-intercept uncertainty in either graph.

5-3: Multiple Solutions Made up to the Same Final Volume For multiple solutions made up to the same final volume, the slightly different graph in Figure 5-7 is used.

5-3: Multiple Standard Addition to a Constant Volume

5-3: Uncertainty in the x-Intercept (ux) m = slope k = number of replicate measurements for unknown n = number of data points for calibration line 𝑦 = mean value of measured y for unknown x Sy = error of the regression

5-4: Internal Standard An internal standard is a known amount of a compound, different from analyte, that is added to the unknown. Signal from analyte is compared with signal from the internal standard to find out how much analyte is present. Internal standards are useful when the quantity of sample analyzed is not reproducible, when instrument response varies from run to run, or when sample losses occur in sample preparation. The response factor F is the relative response to analyte and standard.

5-4: Internal Standard Compensates for changes in experimental conditions, when Ix varies for same concentration [X] Internal Standard: Added standard is different from analyte

5-4: Internal Standards Calibration

5-4: Internal Standard For the sake of accuracy, it is best to prepare a series of standard mixtures of analyte plus internal standard and prepare a graph such as Figure 5-10. The response factor is the slope of the graph, and the intercept should be within statistical limits of zero. The graph should confirm linear response over the desired analytical range.

5-4: Internal Standard