524.3 Purge Flow Study Anne Jurek – Sr. Applications Chemist

Slides:



Advertisements
Similar presentations
Hit “esc” to quit hit any key to continue Effects of Aging on Surface Inertness and Analyte Recoveries in SUMMA Passivated and Fused Silica Lined Canisters.
Advertisements

Hit “esc” to quit hit any key to continue Detection of Ignitable Liquids from Smoke Residues by Large Volume Static Headspace.
PURGE & TRAP Training Commercial July, 22th 2010 Saint-Antoine - France.
Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS.
The Foundations: Classical Split and Splitless Injection
Week 3 Collection and Analysis of Biodiesel
Identification of unsaturated fatty acid from discarded Goya seeds. Okinawa National College of Technology Bioresources engineering * Momoko Mashiyama,
I n t e g r i t y - S e r v i c e - E x c e l l e n c e Headquarters U.S. Air Force 1 Direct Sampling Ion Trap Mass Spectrometry Rob R. Smith Oak Ridge.
Gradient Manipulation and 1.8 µm LC Columns for High Resolution Analysis of Herbal Supplements John W Henderson Jr, Thierry Faye, Ulrik Witteg, Maureen.
July 29, Bench Top Tests for Surfactant Selection Ayantayo Ajani The University of Tulsa.
Results Initial chromatographic conditions [Figure 2a caption] for the separation of the degradation products of aspirin were problematic due to the long.
11 Determination of the Levels of Fungicides on Citrus Fruits By Miseung Borgers Chem 4101, Fall 2008.
Detection of Copper in Wastewater Seth Holm Chem4101 December 2009.
Preparing Solutions LG: I can use formulas for concentration to prepare standard solutions from a solid.
Gaseous Pollutants mini-course TAMS Center February 2009 Creating a Test Atmosphere.
Extraction Methods Static Headspace (HS) Dynamic Headspace
Gas Chromatography Carrier Gas Flow Control Injector Port Column Column Oven Detector Recorder The GC system consists of gas supplies for the mobile phase.
ASTM WK Standard Test Method for Dissolved Gases Anne Jurek – Applications Chemist.
High Performance Liquid Chromatography
A Single Calibration for Waters and Soil Samples Performing EPA Method 8260 Anne Jurek – Applications Chemist.
Utilization of Accuracy Profiles as a Tool for the Validation of Analytical Methods A. Ceccato, P. Jacobs, A. Flament, M. Gibella and W. Dewé Lilly Development.
Overbrook Scientific.
Polybrominated diphenyl ethers by HRGC/HRMS Coreen Hamilton Brian Fowler Louis Haviland Axys Analytical Services, Ltd.
Section 15.2 Describing Solution Composition 1. To understand mass percent and how to calculate it 2. To understand and use molarity 3. To learn to calculate.
J. A. Mock a, A. I. Bolozdynya a, C. E. Dahl b, T. Shutt a a Department of Physics, Case Western Reserve University, Cleveland, OH USA b Department.
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
Waters HPLC Competitive Review
Micro Analytical GC System for Tea Aroma in Product Quality Control Fred Lewis, Principal Chemist, Analytical Science Laboratory, Unilever John Crandall,
Providing comprehensive scientific resources to environmental clients worldwide. Dissolved Gas Analysis: A Review and Comparison.
FAME in AVTUR using GC Heart-Cut and refocusing – IP PM EL Piet Koppen – AC Analytical Controls EI Fame in Jet workshop March 1, 2012.
The world leader in serving science For Research Use Only. Not for use in diagnostic procedures Quantitative Analysis of 4 Immunosuppressant Drugs in Whole.
Introduction to Gas Chromatography
Flavor Delivery via Lipid Encapsulation Flavor Retention in cake and Loaf Volume in Bread.
Different Ways of Expressing Concentration Grade 10 ST.
John D. Vargo 1 ; Michael Schueller 1 ; Mary Skopec 2 1 State Hygienic Laboratory at the University of Iowa, Coralville, IA; 2 Iowa Department of Natural.
Evolution GC-MS/MS: Pesticide analysis in canola oil Evolution GC-MS/MS: Pesticide analysis in canola oil Vivian Watts 1, Ingo Christ 1, Mark Misunis 2.
Forensic Toxicology Use Only Analysis of ETG, ETS using the Thermo Scientific Exactive Mass Spectrometer Kent Johnson Fortes lab, Portland Oregon.
Robert Kettner Determination of the Relevant Impurity Toluene in Formulations by Headspace Gas Chromatography and Flame Ionisation or Mass Spectrometric.
Introduction The pyrolysis products of natural polymers often exhibit very similar neural losses during collision induced dissociation (CID) Some ions.
Chemistry 20  A standard solution is any solution whose concentration is accurately known.  Dilution involves the addition of solvent to a solution.
OVEN Initial temp: 40 'C (On) Maximum temp: 400 'C Initial time: 2.00 min Equilibration time: 0.50 min Ramps: # Rate Final temp Final time
Plasma Free Metanephrines Analysis using LC-MS/MS with Porous Graphitic Carbon Column Xiang He (Kevin) and Marta Kozak Thermo Fisher Scientific.
American Chemical Society, Fall 2012
6850/6890 分流动画 Split/Splitless 进样口 – 部件
James Byrd, Marta Kozak 28 Apr 2011
Optimization of 1,4-Dioxane and Ethanol Detection
Chem. 133 – 5/11 Lecture.
Revathi a/p Rajan Postgraduate Student, School of Health Sciences
Determination of Algae Compounds in Drinking Water
For Forensic Toxicology use Only
LC-MS/MS Analysis of Tetracycline antibiotics in Honey using Advance UHPLC-EVOQ™ SIMA LAB PVT LTD.
A sensitive and repeatable method for characterization of sulfonamides and trimethoprim in honey using QuEChERS extracts with Liquid-Chromatography-Tandem.
EPA Method Equivalency
For Research Use Only. Not for use in diagnostic procedures
Clinical Research Use Only Not For Diagnostic Purposes
Update on Improvements to Dissolved Hydrocarbon Gases in Water Analysis Mark Bruce Ph. D. August 10, 2017.
Ch Concentration Units
Do Now: Define the term concentration as it relates to a solute in a solution. C. Johannesson.
Evaluation and Application of SPME Arrows
Improved Accuracy of Dissolved Hydrocarbon Gas Analysis in Water
Developing a Consensus Test Method for Measuring Volatile Organic Compounds (VOCs) in Water utilizing Headspace Analysis with Gas Chromatography and Mass.
Determining elemental mercury in soils by selective volatilization
The Analysis of Soils and Waters in Accordance with U. S
Fast Analysis of EPA Method 8270 Acid Extractable Compounds Using Gas Chromatography/Tandem Mass Spectrometry (GC/MS/MS) Meredith Clarage1
How to Meet the Analytical Requirements of NJ Low Level TO-15
Deyuan Kong, Brian Morlan, Roopa Kamath, and Sara Mcmillen
Presentation Title 11/28/2018 Confidentiality label
Extraction Procedure (2)
Gas Chromatography.
Darrell Clinton and H. Dorota Inerowicz Purdue University
Presentation transcript:

524.3 Purge Flow Study Anne Jurek – Sr. Applications Chemist Optimize the analytical needs of businesses and industries around the world.

Overview Who is EST analytical? Abstract Experimental parameters examined Observations Conclusions Q&A

Introduction to EST Analytical Purge & Trap Experts Privately held Founded in 1990 Environmental roots Full Support and Service

Purge & Trap Products

Abstract During volatile analysis, samples are purged with an inert gas in order for the Volatile Organic Compounds (VOCs) to be swept out of the sample matrix and onto an analytical trap. For years, the established and recommended purge flow rate and time has been 40ml/min for 11 minutes.

Why do the study? EST analytical is a leader in Purge and Trap research and development. EPA method 524.3 allows for different purge flows and times. Evaluate the ability to shorten cycle times and run more samples. Evaluate how purge times and flows can affect results.

Purge and Trap Steps Standby Purge Dry purge Desorb Preheat Desorb Trap Bake This information is for scientific exchange only. EST analytical does not recommend or suggest methodology outside of published method parameters.

Purge Flow Rate Defined as: The rate in mL/min that the purge gas flows through the sparge vessel during the purge cycle Current flow and time are 40mL/min for 11min

Purge Flow Study Current purge volume is 480mL, how does decreasing the purge volume affect the compound results? Current purge flow is 40mL/min, how does increasing the purge flow affect the compound results?

USEPA Method 524.3 Parameter Recommended Allowable Min. Max. Purge Flow Rate 40mL/min 80mL/min 20mL/min 200mL/min Purge Volume 360mL 520mL 240mL 680mL Purge Volume + Dry Purge Volume 720mL 880mL

Experiment 1 (Baseline) Purge Flow Study Parameter Experiment 1 (Baseline) Experiment 2 Experiment 3 Experiment 4 Purge Flow Rate 40mL/min 65mL/min 100ml/min Purge Time 11min 6.5min 4min Purge Volume 440mL 422.5mL 400mL 260mL Purge Volume + Dry Purge Volume 480mL 462.5mL 300mL

30m x 0.25mm I.D. 1.4µm film thickness (624) GC/MS Parameters GC/MS Inlet Split/Splitless Inlet Temp. 220ºC Inlet Head Pressure 12.153 psi Mode Split Split Ratio 40:1 Column 30m x 0.25mm I.D. 1.4µm film thickness (624) Oven Temp. Program 45ºC hold for 1 min, ramp 15ºC/min to 220ºC, hold for 1.33 min, 14 min run time Column Flow Rate 1mL/min Gas Helium Total Flow 44mL/min Source Temp. 230ºC Quad Temp. 150ºC MS Transfer Line Temp. 180ºC Scan Range m/z 35-300 Scans 5.2 scans/sec Solvent Delay 0.7 min

Evolution Purge and Trap Parameters Purge and Trap Concentrator EST Analytical Evolution Trap Type Vocarb 3000 Valve Oven Temp. 130ºC Transfer Line Temp. Trap Temp. 35ºC Moisture Reduction Trap (MoRT) Temp. 39ºC Purge Time Varied Purge Flow Dry Purge Temp. ambient Dry Purge Flow 40mL/min Dry Purge Time 1.0 min Desorb Pressure Control On Desorb Pressure 5psi Desorb Time 0.5 min Desorb Preheat Delay 10 sec Desorb Temp. 250ºC Moisture Reduction Trap (MoRT) Bake Temp. 210ºC Bake Temp 260ºC Sparge Vessel Bake Temp. 110ºC Bake Time 6 min Bake Flow 85mL/min

Centurion Autosampler Parameters Purge and Trap Auto-Sampler EST Centurion WS Sample Type Water Water Volume 5ml Sample Prime Time 7 sec Loop Equilibration Time 5 sec Sample Transfer Time 10 sec. Syringe Rinse On/10 ml Number of Syringe Rinses 2 Sample Loop Rinse On/10 sec Sample Sweep Time Number of Sparge Rinses On/2 Rinse Volume 5 ml Rinse Transfer Time 10 sec Rinse Drain Time 15 sec Water Heater Temp. 85 sec Internal Standard Vol. 5 µl

524.3 Internal Standard/Surrogate Standard at 5ppm Standard Preparation 524.3 Standard at 200ppm Amount Standard Concentration Final Vol. 200ul 524.3 VOA Mega Mix 2000ug/ml 2ml 524.3 Gas Cal Mix Iodomethane Use 2ml volumetric flask and dilute standards to 2.0ml in purge and trap methanol 524.3 Standard at 20ppm 200µl of 200ppm standard diluted to 2.0ml in purge and trap methanol 524.3 Internal Standard/Surrogate Standard at 5ppm 5µl 524.3 IS/SS Mix 2000 ug/ml 2.0ml Use 2ml volumetric flask and dilute standards to 5.0ml in purge and trap methanol

Calibration Curve Preparation Concentration Standard Standard Amount Final Vol. 0.5ppb 20ppm 2.5µl 100ml 1ppb 5µl 2ppb 10µl 5ppb 25µl 10ppb 200ppm 20ppb 40ppb 20µl Calibration Standards Fill 40ml Vial with final standard leaving no headspace in the vial.

Preserved Water Preparation Preservation of Water Preservative Mass (mg) Final Volume (ml) Ascorbic Acid 25 40 Maleic Acid 200 625 1000 5000

Experimental Baselined the system: Prepared a calibration curve and ran it using 40mL/min for 11 minute purge parameters Ran method detection limit and precision and accuracy samples Compiled all of the data for comparison

Experimental Prepared and ran calibration curves, MDLs and precision and accuracy samples for each of the purge flows and times outlined Compiled all of the data from each purge flow study

Results Purge Flows and Times Avg. Curve Quadratic Regression Avg. Curve RF Avg. Upper PIR (prediction interval of results) (0.5µg/L) Avg. Lower PIR (prediction interval of results (0.5µg/L) Avg. % Recovery (20µg/L) Avg. Precision, %RSD (20µg/L) Baseline 1.000 1.110 110.23 81.65 103.57 4.03 65ml/min for 6.5 min 0.999 1.220 116.53 83.31 95.11 4.66 100ml/min for 4 min 1.273 105.73 68.57 95.85 5.22 65ml/min for 4 min 1.001 112.79 76.89 105.27 2.82

20ppb Chromatograms

Water Abundance Overlay 40mL/min for 11.0 min – black 65mL/min for 6.5 min – blue 65mL/min for 4.0 min – orange 100mL/min for 4.0 min – green

Outlier Prediction Interval Summary Compound 40 ml/min for 11 min 65 ml/min for 6.5 min 100 ml/min for 4 min 65 ml/min for 4 min Upper PIR Lower PIR bromomethane 96.36 63.07 103.14 54.57 111.26 75.60 94.79 41.21 methyl acetate 121.49 77.93 100.19 81.52 113.22 68.50 122.68 93.32 MTBE 120.98 66.45 130.63 112.23 109.33 70.10 103.59 87.26 THF 146.41 53.59 106.41 73.59 208.46 98.40 146.39 82.18 ethylmethacrylate 116.02 101.70 121.26 98.17 89.95 67.76 119.12 87.74 1,2-dibromomethane 100.41 78.45 114.05 85.95 101.47 66.53 133.85 95.29 bromoform 117.55 95.59 132.06 89.09 90.30 75.98 116.46 94.97 1,1,2,2-tetrachloroethane 121.77 103.37 122.39 85.04 99.97 64.60 136.82 108.90 1,2,3-trichloropropane 108.46 86.97 108.41 86.45 124.71 58.15 141.93 87.21 1,2-dibromo-3-chloropropa 129.11 88.60 123.33 84.10 114.15 24.14 131.75 47.68

Response Factor Outlier Summary Compound 40ml/min for 11 min 65 ml/min for 6.5 min 100 ml/min for 4 min 65 ml/min for 4 min methyl acetate 0.360 0.347 0.329 0.203 MTBE 1.327 1.416 1.415 1.090 THF 0.086 0.080 0.075 0.044 ethylmethacrylate 0.749 0.765 0.722 0.513 1,2-dibromomethane 0.496 0.526 0.500 0.357 bromoform 0.304 0.303 0.284 0.197 1,1,2,2-tetrachloroethane 1.348 1.383 1.308 0.870 1,2,3-trichloropropane 0.458 0.459 0.420 0.269 1,2-dibromo-3-chloropropane 0.283 0.266 0.227 0.138

Response Factor Results Bar Graph

Conclusions The effect of purge flow and purge volume can be significant The EPA recommended purge flow of 40ml/min for 11 minutes displayed the least amount of water contamination while the 100ml/min showed the most The 65ml/min flow had similar amounts of water even though the purge times/volumes were different

Conclusions The 65ml/min for 4 minutes produced and overall purge volume of 300ml when taking the dry purge into account. This lower purge volume affected the compound responses of polar compounds and the lower purge response compounds

Conclusions The optimum purge flow and time, when compared with the recommended 40ml/min for 11 minutes, were 65ml/min for 6.5 min The 65ml/min flow was slow enough not to cause water problems and the 6.5min purge time along with the dry purge provided a 462.5 ml purge volume which aided in compound response

Thank You Questions

For more information or application notes visit our website at: www.estanalytical.com Questions: ajurek@estanalytical.com