The Optimization of Reconstruction Method Reducing Partial Volume Effect in PET/CT 3D Image Acquisition Department of Nuclear Medicine, Samsung Medical.

Slides:



Advertisements
Similar presentations
Change in longest diameter = -19%. Change in sum of perpendicular diameters = -21%
Advertisements

Gamma Camera Quality Control
DATSCAN Quantitation Vs Image Reading Mike Avison Bradford Royal Infirmary.
Evaluation of Reconstruction Techniques
1 Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University.
IMAGE QUALITY.
PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY.
Parameters and Trade-offs
“International Conference on Clinical PET and Molecular Nuclear Medicine” IAEA - IPET 2007 Bangkok, Thailand 10 to 14 November 2007 C. Suarez, R. Pruzzo,
Introduction to PET/CT in Oncology: Practical Aspects Jeffrey T. Yap, PhD Department of Imaging Dana-Farber Cancer Institute.
Computed Tomography III
Comparison of Scintimammography and Dedicated Emission Mammotomography
Kirov A S, MSKCC Overview of Geant4 Use and Issues in Imaging: Emission Tomography (PET and SPECT) Assen S. Kirov Department of Medical Physics Memorial.
Introduction to Medical Imaging
PHYSICS IN NUCLEAR MEDICINE: QUANTITAITVE SPECT AND CLINICAL APPLICATIONS Kathy Willowson Department of Nuclear Medicine, Royal North Shore Hospital University.
Four patients underwent a PET/CT scan with a Philips True Flight Gemini PET/CT scanner. We manually identified a total of 26 central-chest lesions on the.
Coincidence to Image: PET Imaging Jennifer White Marketing Manager SNS Workshop October 13, 2003.
For internal use only / Copyright © Siemens AG All rights reserved. Recent Innovations in PET and hybrid modality technologies Bernard Bendriem,
M Dawood1,2, C Brune2, F Büther1, KP Schäfers1
Jeffrey Yap, PhD Ron Kikinis, MD Wendy Plesniak, PhD -1- CTSA at RSNA 2009 PET/CT Analysis using 3D Slicer Jeffrey Yap PhD Ron Kikinis MD Wendy Plesniak.
EuroMedIm Irène Buvat - May Quantification in emission tomography: challenges, solutions, performance and impact Irène Buvat U678 INSERM,
ASSESSMENT OF INTERNAL BRUISE VOLUME OF SELECTED FRUITS USING MR IMAGING Ta-Te Lin, Yu-Che Cheng, Jen-Fang Yu Department of Bio-Industrial Mechatronics.
Prognostic Value of Tumor Hypoxia,
Prognostic Value of PET Using 18F-FDG in Hodgkin’s Disease for Posttreatment Evaluation J Nucl Med 2003; 44:1225–1231 Intern 魏敬庭.
Scanner Validation via the SNM Clinical Trials Network Phantom Program Paul E. Christian Molecular Imaging Program Huntsman Cancer Institute University.
PET/SPECT Phantom. Side View of Phantom Image Resolution Intrinsic resolution FWHM Intrinsic resolution FWHM Field of view Field of view Measurement:
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) March 18, 2009 Interclinic Comparison of CT Volumetry Quantitative Imaging Biomarker Alliance.
Where do our data come from? Positron Emission Tomography (PET)
Nuclear Medicine: Tomographic Imaging – SPECT, SPECT-CT and PET-CT Katrina Cockburn Nuclear Medicine Physicist.
CT QUALITY MANAGEMENT. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
 Interactive Segmentation For Image Guided Therapy Ohad Shitrit & Tsachi Hershkovich Superviser: Dr. Tammy Riklin Raviv Ben-Gurion University of the Negev.
Part No...., Module No....Lesson No
Conclusions Simulated fMRI phantoms with real motion and realistic susceptibility artifacts have been generated and tested using SPM2. Image distortion.
Impact of Axial Compression for the mMR Simultaneous PET-MR Scanner Martin A Belzunce, Jim O’Doherty and Andrew J Reader King's College London, Division.
Quality Control Bushong Chapter 10.
Improved Hybrid PET Imaging Using Variable Focal Length Fan-Slat Collimators Thomas C. Rust and Dan J. Kadrmas, Ph.D. Medical Imaging Research Laboratory.
Acquisition time6 min1 min 12 s Collimator height25 mm (Anger)12 mm (HiSens) Detector1 layer, 1 pixel / hole3 layers, 1 pixel / hole3 layers, 4 pixels.
Wai-Hoi Wong 1, Hongdi Li 2, Yuxuan Zhang 1, Rocio Ramirez 1, Hossain Baghaei 1, Shaohui An 2, Chao Wang 2, Shitao Liu 2, Yun Dong 2 1 University of Texas.
Nuclear Medicine Physics Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia Image Artifacts in Nuclear Medicine Part B.
Moment of Inertia Comparisons Test CaseParameterUnitsTechnionGeorgia Tech% Error 4K Case Ixxkg.m Iyykg.m
RPC February 2010 SPATIAL RESOLUTION OF HUMAN 3D RPC-PET SYSTEM 1 LIP, Laboratório de Instrumentação e Física Experimental de Partículas, Coimbra,
Journal of Nuclear Cardiology | Official Journal of the American Society of Nuclear Cardiology The high matrix acquisition technique for imaging of atherosclerotic.
PET Scanner Calibration Using a Small Ga68/Ge68 Source
18 F-FDG Colonic Uptake by Oral Anti- Diabetic Drugs in PET/CT Scan Sang Gyu Kim, Hoon Hee Park, Jung Yul Kim, Young Kag Bahn Han Sang Lim, Jae Sam Kim,
The study of new reconstruction method
Segmentation of 3D microPET Images of the Rat Brain by Hybrid GMM and KDE Tai-Been Chen Department of Medical Imaging and Radiological Science,
JESÚS SILVA-RODRÍGUEZ, PABLO AGUIAR, INÉS DOMÍNGUEZ-PRADO, MICHEL HERRANZ, ÁLVARO RUIBAL 18F-Choline: Is shine-through effect an issue for prostate SUV.
Simulations in Medical Physics Y. TOUFIQUE*, R.CHERKAOUI EL MOURSLI*, M.KACI**, G.AMOROS**, *Université Mohammed V –Agdal, Faculté des Sciences de Rabat,
A Study on the Determination of Scan Speed in Whole Body Bone Scan Applying Oncoflash Gwang Gil Yang Department of Nuclear Medicine, Asan Medical Center,
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Characterization of proton-activated implantable markers for proton range verification using PET J. Cho1, G. Ibbott1, M. Kerr1, R. A. Amos2, F. Stingo1,
Prof. Stefaan Vandenberghe Enrico Clementel
Impact of Obesity on Medical Imaging and Image-Guided Intervention
Evaluation of mA Switching Method with Penalized Weighted Least-Square Noise Reduction for Low-dose CT Yunjeong Lee, Hyekyun Chung, and Seungryong Cho.
The Accuracy of the Table Movement During A Whole Body Scan
Proton range verification with PET: 68Zn, Cu and polycarbonate activation SNMMI Young Investigator Symposium, Vancouver, June 9, 2013 Cho J1, Ibbott.
Reconstructions with TOF for in-beam PET
Department of Nuclear Medicine Samsung Medical Center Joon Ho YEON
Journal of Nuclear Cardiology | Official Journal of the American Society of Nuclear Cardiology Reducing the small-heart effect in pediatric gated myocardial.
Journal of Nuclear Cardiology | Official Journal of the American Society of Nuclear Cardiology Standardized Uptake Values in FDG PET/CT for Prosthetic.
2017 Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine Theoretical Study of Benefit of Long Axial Field-of-View PET: Impact.
18F-FDG Colonic Uptake by Oral Anti-Diabetic Drugs in PET/CT Scan
Spatially Varying Frequency Compounding of Ultrasound Images
Chapter 17: Quantitative Nuclear Medicine
U4-14 TITLE Predicting Periodic Properties PURPOSE To predict the density germanium using calculated densities for silicon, tin, and lead. HYPOTHESIS The.
CT images by texture analysis
First demonstration of portable Compton camera to visualize 223-Ra concentration for radionuclide therapy Kazuya Fujieda (Waseda University) J. Kataoka,
Improving Image Accuracy of ROI in CT Using Prior Image
Evaluation of Average CT to reduce the artifact in PET/CT
Fig. 2. Box-plots comparing the percent changes of the quantitative parameters in DCE-MRI (LD, TV, PE), DWI (ADC) and PET/CT (SUV max). The mean percent.
Presentation transcript:

The Optimization of Reconstruction Method Reducing Partial Volume Effect in PET/CT 3D Image Acquisition Department of Nuclear Medicine, Samsung Medical Center Department of Nuclear Medicine, Samsung Medical Center Gun Chul HONG

The underestimate phenomenon due to limit spatial resolution Introduction Low accuracy for quantitative analysisIncreasing error in the small-sized lesion

The evaluation of effect according to sphere size Purpose The variation of SUV the PET image according to the changed reconstruction parameters Selection of parameter value to reduce the PVE effectively

Materials and Methods GE-Discovery STE Scanner Specifications Spatial resolution : 5.0 ㎜ Sensitivity : 8.5 cps / KBq DFOV : 50 ㎝ Cristal : BGO

Materials and Methods NEMA IEC/2001 Body Phantom Phantom Specifications Dimensions : 24.1 x 30.5 x 24.1 ㎝ Spheres Inner Diameter : 10, 13, 17, 22, 28, 37 ㎜ Cylinder Volume : 9.7 ℓ Shipping Weight : 4.9 ㎏

PET Scan 10 min/bed CT scan Sphere 0.06 MBq/ ㎖ Background 0.01 MBq/ ㎖ Materials and Methods

Iteration : 2 ~ 50 Matrix size : 128 Ⅹ 128 Z axis Filter : Standard Image Reconstruction Reconstruction method Diameter : 50 Post Filter : 4.29 FWHM Subset : 1 ~ 56

Measurement of SUV bw set ROI in the image after reconstruction max - true % Difference = max SNR = ( SUV + background SUV ) ( SUV - background SUV ) Reconstruct to measure data 37mm 28mm 22mm 17mm 13mm 10mm Background SUV bw = ( Total activity / Body weight ) ( Tissue activity / tissue mass ) Image Analysis

Reconstruction Image 2 iteration 6 iteration 15 iteration 2 subset8 subset35 subset Results

SUV - Iteration ㎜ ㎜ ㎜ ㎜ ㎜ ㎜ Results Subset : 20 SUV : g/ml sphere number

SUV - Iteration Results 68%

% Difference - Iteration Results

The sum of SNR - Iteration Results # SNR

mm mm mm mm mm mm SUV - Subset Results Iteration : 20 SUV : g/ml sphere number

SUV - Subset Results 170 %

% Difference - Subset Results

# The sum of SNR - Subset Results SNR

We could change the PVE using iteration and subset parameters in the reconstruction method Because of time constrains in performing the data reconstruction, increment of iteration and subset must be considered with time required for reconstruction Summery

As number of iteration and subset is increased, SUV max is increased up 68.1 percent at the 10 ㎜ sphere at the 10 ㎜ sphere The most effective range to decrease the PVE at the small-sized lesion is 6 iteration and 8~20 subset parameters considering reconstruction time the small-sized lesion is 6 iteration and 8~20 subset parameters considering reconstruction time Conclusion

Thank You