WU Lei,WANG Xiaolong, LI Chunhua, QU Huamin Mechanical Group, Accelerator Division Institute of High Energy Physics 13 th International Workshops on Accelerator.

Slides:



Advertisements
Similar presentations
BROOKHAVEN SCIENCE ASSOCIATES Abstract Vibrating Wire R&D for Magnet Alignment Animesh Jain, NSLS-II Project The alignment tolerance for a string of magnets.
Advertisements

Alexander Temnykh Cornell University, Ithaca NY, USA
SCU Measurements at LBNL
Multipole Girders - Alignment & Stability (Multipole Girder Alignment technology & R&D) S. Sharma ASD: J. Skaritka, D. Hseuh, V. Ravindranath, G. Miglionico,
1 Long-term Variation of the Magnet Alignment In SPring-8 Storage Ring Main events of magnet alignment Long-term monitoring Variation of magnet alignment.
October 12, 2006 Heinz-Dieter Nuhn, SLAC / LCLS Undulator Good Field Region and Tuning Strategy 1 Undulator Good Field Region and.
Mechanical Vibrations
Wilson Lab Tour Guide Orientation 11 December 2006 CLASSE 1 Focusing and Bending Wilson Lab Tour Guide Orientation M. Forster Mike Forster 11 December.
Beam Loss Analysis Tool for the CTF3 PETS Tank M. Velasco, T. Lefevre, R. Scheidegger, M. Wood, J. Hebden, G. Simpson Northwestern University, Evanston,
Zachary Wolf LCLS Undulator October 30, LCLS Undulator Tuning And Fiducialization Zack Wolf, Yurii Levashov, Achim.
BROOKHAVEN SCIENCE ASSOCIATES Abstract Magnetic Specifications and Tolerances Weiming Guo, NSLS-II Project In this presentation I briefly introduced the.
The 2nd Tevatron Electron Lens and tests of a new electron gun in the framework of Beam-Beam Compensation project Yu. Alexahin, V. Kamerdzhiev, G. Kuznetsov,
Topic 1 different attributes that characterize sensors ETEC 6405.
1 / 19 M. Gateau CERN – Geneva – CH 14th International Magnetic Measurement Workshop September 2005, Geneva, Switzerland.
S. Russenschuck, CLIC-Workshop, WP2-Pacman, R&D projects on rotating coil probe and stretched wire techniques for CLIC / PACMAN Stephan Russenschuck.
ISAT 303-Lab3-1  Measurement of Condition: Lab #3 (2005):  List of parameters of condition: –Linear distance, angular displacement, vibration, displacement,
Chapter 9 Electromagnetic Waves. 9.2 ELECTROMAGNETIC WAVES.
Alignment and Beam Stability
Yichao Jing 11/11/2010. Outline Introduction Linear lattice design and basic parameters Combined function magnets study and feasibility Nonlinear dynamics.
Simulation of direct space charge in Booster by using MAD program Y.Alexahin, N.Kazarinov.
Tilt meter fiducial Guide rail Abstract This poster covers the survey and alignment techniques selected for installation of the SPIRAL2 accelerator devices.
June 14th 2005 Accelerator Division Overview of ALBA D. Einfeld Vacuum Workshop Barcelona, 12 th -13 th September 2005 General 10 th September 2005.
Franz Peters, Georg Gassner WPM/HLS Status and June 17 th, WPM & HLS Status and Schedule Franz Peters, Georg Gassner.
1 BROOKHAVEN SCIENCE ASSOCIATES Abstract Magnet and Girder Integration Lewis Doom, NSLS-II Project National Synchrotron Light Source II (NSLS-II) will.
Orbit Control For Diamond Light Source Ian Martin Joint Accelerator Workshop Rutherford Appleton Laboratory28 th -29 th April 2004.
EKT 451/4 SEM 2 Chapter 6 SENSOR & TRANSDUCER SHAIFUL NIZAM MOHYAR
BEPCII Prealignment Installation Survey and Alignment Accelerator Center of IHEP Xiaolong Wang
Simulation of direct space charge in Booster by using MAD program Y.Alexahin, A.Drozhdin, N.Kazarinov.
CLIC survey and alignment 1 CLIC CES meeting STARTSLIDE CLIC SURVEY AND ALIGNMENT Hélène MAINAUD DURAND.
Engineering Division 1 M321/M331 Mirror Switchyard Design Review Tom Miller
1 Magnetic measurements proposal of the FAIR magnets Magnetic Measurements of Super-FRS dipoles: budgets 1 Budget for dipoles ( ) document FAIR-SFRS-magnetic_measurement_EVM.xls.
Vertical Emittance Tuning at the Australian Synchrotron Light Source Rohan Dowd Presented by Eugene Tan.
CLIC Workshop th -17 th October 2008 Thomas Zickler AT/MCS/MNC 1 CLIC Main Linac Quadrupoles Preliminary design of a quadrupole for the stabilization.
Survey, Alignment and Metrology Fiducials Requirements Fabien Rey ESS Survey, Alignment and Metrology Group ACCELERATOR TECHNICAL BOARD 23/09/2015.
CLIC Stabilisation Day’08 18 th March 2008 Thomas Zickler AT/MCS/MNC/tz 1 CLIC Quadrupoles Th. Zickler CERN.
Magnetic Field Stability Measurements Joe DiMarco 23Oct07.
By Verena Kain CERN BE-OP. In the next three lectures we will have a look at the different components of a synchrotron. Today: Controlling particle trajectories.
An electron/positron energy monitor based on synchrotron radiation. I.Meshkov, T. Mamedov, E. Syresin, An electron/positron energy monitor based on synchrotron.
Lecture 4 - E. Wilson - 23 Oct 2014 –- Slide 1 Lecture 4 - Transverse Optics II ACCELERATOR PHYSICS MT 2014 E. J. N. Wilson.
SL/BI 16/05/1999DIPAC’99 -- JJ Gras -- CERN SL/BI -- Adaptive Optics for the LEP 2 SR Monitors G. Burtin, R.J. Colchester, G. Ferioli, J.J. Gras, R. Jung,
Lecture 4 - E. Wilson –- Slide 1 Lecture 4 - Transverse Optics II ACCELERATOR PHYSICS MT 2009 E. J. N. Wilson.
1 BROOKHAVEN SCIENCE ASSOCIATES 12th International Workshop on Accelerator Alignment September 10-14, 2012 Fermilab, Batavia, Illinois, U.S.A NSLS-II Girder.
1 Research on laser tracker measurement accuracy and data processing Liang Jing IHEP,CHINA
CERN –GSI/CEA MM preparation meeting, Magnetic Measurements WP.
1 BROOKHAVEN SCIENCE ASSOCIATES Beam Stability Overview NSLS-II CFAC Meeting May 8, 2007 S. Krinsky.
Midterm Review 28-29/05/2015 Progress on wire-based accelerating structure alignment Natalia Galindo Munoz RF-structure development meeting 13/04/2016.
EKT 451 CHAPTER 6 Sensor & Transducers.
1 Transducers EKT 451 CHAPTER 6. 2 Definition of a Transducers   Transducer is any device that converts energy in one form to energy in another.  
Yingshun Zhu Design of Small Aperture Quadrupole Magnet for HEPS-TF
Ultra-low Emittance Coupling, method and results from the Australian Synchrotron Light Source Rohan Dowd Accelerator Physicist Australian Synchrotron.
Tutorial On Fiducialization Of Accelerator Magnets And Undulators
Pasquale Arpaia, University of Naples Federico II
Off-axis injection lattice design studies of HEPS storage ring
Mechanical Vibrations
Workshop on Tau-Charm at High Luminosity
For Discussion Possible Beam Dynamics Issues in ILC downstream of Damping Ring LCWS2015 K. Kubo.
Advanced Photon Source Upgrade Project:
Status of Vibration Measurement Work at BNL
Yingshun Zhu Accelerator Center, Magnet Group
Compact and Low Consumption Magnet Design The DESY Experience
R.A.Melikian,YerPhI, , Zeuthen
Multiturn extraction for PS2
SCU Next Phase Meeting July 8, 2014.
Vibration Basics and Shaker Selection
I Alexander Nass for the JEDI collaboration
Vibration Basics and Shaker Selection
Multi-Turn Extraction for PS2 Preliminary considerations
FNAL Superconducting Quadrupole Test
Progress and plans, Mechanics system of CEPC
Breakout Session SC3 – Undulator
Presentation transcript:

WU Lei,WANG Xiaolong, LI Chunhua, QU Huamin Mechanical Group, Accelerator Division Institute of High Energy Physics 13 th International Workshops on Accelerator Alignment,Beijing,Oct13-17,2014

High Energy Photon Source ( HEPS) Energy: 5GeV Circumference: 1296m Lattice: 7BA Emittance:<0.1nm·rad One of 48 typical cells 3.8m quadrupole sextupole

 It is difficult to achieve the required accuracy using magnet fiducialization, coupled with traditional optical survey. so we consider using vibrating wire alignment technique to meet the tolerance  0.03mm between magnet to magnet on one girder. tolerancesMagnet to MagnetGirder to Girder Horizontal  0.03mm  0.05mm Vertical  0.03mm  0.05mm beam direction  0.5mm Roll angle  0.2mrad  0.5mrad HEPS Alignment Tolerances

Cornell University CESR Magnetic center finding of superconducting quadrupoles placed inside the cryostat. Solenoid magnetic center finding. Undulator magnets characterization. SLAC LCLS Fiducialize the quadrupoles between undulator segment. CERN Solenoid magnetic center finding Application examples of vibration wire:

PSI Swiss Free Electron Laser Magnetic axis of quadrupoles finding. BNL NSLS-II Alignment of multipole magnets on one girder. ` The magnet alignment requirements in HEPS is similar to NSLS-II, So the vibrating wire alignment in NSLS-II has been used as our reference.

Based on measurement of the magnetic axis to align the magnets. Further evolution of the pulsed wire method. A single conducting wire is stretched through the magnet aperture and driven by an alternating current. Transversal vibrations are continuously excited by periodic Lorentz Force. By matching current frequency to one of the resonant modes of wire, the vibration amplitude and sensitivity are enhanced and the magnetic induction intensity at the wire position can be calculated. Move the wire across the aperture in horizontal(x) or vertical direction(y), the magnetic induction intensity distribution can be gotten and the magnetic axis can be found.

Derivation of basic theory Design of test bench model Design of sag measurement scheme Calculation of vibrating amplitudes Design of the sensor circuit and sensor calibration Design of data acquisition and control system Finished research work:

Analyze the motion of a section of wire dz. According to Newton’s law, differential equation of the motion in the y-z plane is: homogeneous solution: particular solution due to gravity: particular solution due to Lorentz Force: dz

The difference between fixed end and free end bench is the fixed way of the wire. The wire is fixed and located by V notch at the fixed end bench. At the free end bench, the wire is pulled by weights through a pulley. The quadrupole and sextupole will be borrowed from BEPCII. They are 105Q and 130S. The magnet support girder is 5.4 meters. Two parallel guideways are installed above the girder in order to move magnets conveniently. By move the magnets along the guideways, the magnetic center would be detected in different place along the girder.

For a 7m long wire, sag is unavoidable because of self-weight. To find the vertical magnetic center, accurate sag correction is essential. Establish a coordinate system relative to the wire, measuring sag at different place along the wire(z direction). Using laser tracker and NIVEL200 to adjust the position and orientation of the sensor. Using sensor to measure the vertical position of the wire.

GP1S094HCZ0F is a transmissive photointerrupter with opposing emitter and detector in a molding that provides non-contact sensing. Function: detect the wire vibration amplitudes y d (z,t) change over time due to Lorentz Force. Sensor circuit R D =280Ω Re=18KΩ +13V DC Power Sensor output test bench Calibration:By changing the position of the sensor relative to the wire on the sensor output test bench, a series of points which are the output voltages were gotten. Because the sensor received different optical signals when the wire at different locations, so the output voltages are different.

No shading voltage Left linear region Right linear region Voltage(V) Voltage difference (V) Displacemen t(mm) Displacement difference(mm) voltage/displaceme nt(mV/  m) outputs of one of the four Sensors shading voltage Left linear region Right linear region Voltage(V) Voltage difference (V) Displacemen t(mm) Displacement difference(mm) 0.12 voltage/displaceme nt(mV/  m) There are two linear output parts. In each part, the wire displacement is about 0.12mm. One part will be chosen as the workspace. There was little difference in the output voltages between covering the sensor with light shield and exposing the sensor in the lab light. But the voltages were more stable when the sensor was covered. The sensitivity of the sensor is approximately 30mV/  m.

The output voltages of the sensor were acquired by NI data acquisition devices. In every test position, a series of datum were acquired. The sampling frequency is 500Hz,and the sampling time is 1s. The experimental results show that in each position the difference between the maximum and minimum voltages is approximately 20mV , less than the voltage change caused by 1 micron displacement of the wire.

In one test position, the measures of variation reduced from 20mV to 6mV after filtering. The stability has improved significantly. So adding filters to the circuit and DAQ program is under consideration.

According to the basic theory, the vibration amplitudes is: Vibration amplitudes should be controlled within the voltage linear output part. And it also should be make full use of this region. According to the measurement, the linear region widths about 0.12mm. We designed the wire vibration amplitudes  0.045mm.  Quadrupole magnetic center measurement: Parameter specification: I 0 wire current amplitude GL Q quadrupole integrated gradient L Q quadrupole length  mass per unit length  n harmonic frequency  damping constant l wire length n resonance order j l / z magnet  y distance between magnetic center relative to the wire in y direction

 sextupole magnetic center measurement: According to the distance between the magnetic center and wire, the distance between wire end and magnet, the distance between wire end and sensor,and magnetic field gradient, the AC current can be calculated. We use this method to estimate the AC current in the wire according to the previous magnetic survey datum.  x distance between magnetic center relative to the wire in x direction B”l Q sextupole integrated gradient

Datum will be transported to the IPC by the communication card PXIe- 8360, PCIe-8362 IPC also controls the movement of 2 x-y stages. Using LabVIEW software to do data acquisition and data process. DAQ card NI PXI-6122: 16-bit A/D , input signal ranges (selectable by channel) ±10, ±5, ±2.5, ±1.25V 4 differential analog inputs,  Function: simultaneous sampling the 2 sensors and 1 driving current signals simultaneous sampling, 500KS/s per channel

months draw the mechanical engineering drawings of the test bench, set the manufacturing specifications. purchase outsourcing equipments. write the LabVIEW program of the control system months manufacture the test bench months Set up the whole vibrating wire test bench months Carry out vibrating wire experiment to acquire the test datum, and analyse the experimental results.

The alignment tolerance of multipoles on a long girder in HEPS should be better than 30 microns. Vibrating wire alignment technique is appropriate to achieve the required accuracy. Complete the theoretical research of vibrating wire, and some research schemes were made, and complete the test bench model. Mechanical engineering drawing of test bench is under drawing, the data acquisition program is under writing. A lot of works still need to be done in the following months.