Introductory Lecture Series: The Anesthesia Machine

Slides:



Advertisements
Similar presentations
INHALATION ANAESTHESIA BREATHING SYSTEMS Prepare and monitor anaesthesia in animals INHALATION ANAESTHESIA BREATHING SYSTEMS.
Advertisements

ANESTHETIC MACHINES.
Safety Requirements of the Anesthesia Workstation
Anesthesia Machine Presented by Gil Soto C.R.N.A.
INTERMEDIATE PRESSURE SYSTEM PRESENTED BY: Dr CHITTRA MODERATED BY: Dr JYOTI PATHANIA.
AMIR SALAH MODERN ANAESTHETIC MACHINE MODERN ANAESTHETIC MACHINE 1 of 4.
Presented by: Dr. manoj kumar Moderated by: Dr. Aparna Sharma
Anesthesia Apparatus Checkout Recommendations
Program: Respiratory Therapy Course: Introduction to Respiratory Sciences Lesson: Gas Physics Handle Gas Cylinders With Care.
Inhaled Anesthetic Delivery Systems
Regulators and Flowmeters
Check Valves and Manufacturer-Recommended Leak Test
Scavenging System  To reduce exposure to waste anesthetic gases  Can be active or passive scavenging system Active Scavenging System  Connected to a.
Student Book © 2004 Propane Education & Research CouncilPage Performing Pressure Tests on Gas Distribution Lines Gas personnel must understand.
Presented by : Dr Rashmi Moderator: Dr Gian Chauhan.
Anesthetic Equipment.
Inhaled Anesthetic Delivery Systems
 Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety.
 The annular space. The clearance between the head of the float and the flow tube is known as the annular space.
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U.
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U.
PART 3: Breathing Circuit
Flow Meters Flow: Volume of a fluid passing certain point over a given time Flow Rate. over a given time Flow Rate. Flow meters:sets used to measure flow.
Parts of the Anesthetic Machine
The Anesthesia Gas Machine
HYDRAULICS & PNEUMATICS
Anesthesia Machine VAPORIZERS. Vaporizers Convert liquid anesthetic into a volatile inhalation agent Based on laws of physics You must memorize the chemical.
A). When the carrier gas is quickly switched to 100% nitrous oxide (B). the halothane concentration decreases to 3% within 8 to 10 seconds A new steady-state.
Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab. DCA, Dip. Software statistics PhD(physiology) Mahatma gandhi medical college and research institute,
STEAM HEATING.
Inhaled anesthetic Delivery Systems
Scavenging System A&A Pg 356  To reduce exposure to waste anesthetic gases  Can be active or passive scavenging system Active Scavenging System  Connected.
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U
Lease Automatic Custody Transfer
To Menu.
SERVO-i WITH HELIOX OPTION
Ventilators All you need to know is….
CENTRAL GAS SYSTEMS Without reference, identify principles relating to central gas systems as they pertain to facility and equipment interface with at.
HVAC523 Basic Subsystems.
Introductory Lecture Series: The Anesthesia Machine
Placing Vapor Distribution Systems and Appliances into Operation MODULE 8 System Tests.
Anesthetic Equipment.
Without reference, identify principles about volume/pressure and high frequency ventilators with at least 70 percent accuracy.
Without reference, identify principles about Anesthesia Units with at least 70 percent accuracy.
ANAESTHESIA BREATHING CIRCUITS
PART 3: Breathing Circuit
To Menu.
Anesthesia Machine Circuits
HYDRAULICS & PNEUMATICS
Parts of the Anesthetic Machine A&A Pages
Anesthetic Equipment.
PART 3: Breathing Circuit
Universal Anaesthesia Machine (UAM) – functional testing of a new anaesthesia workstation for use in the developing world De Beer DAH, Walker IA, ♦ Bell.
Mechanical Ventilator
The Anesthesia Machine: Introduction
BREATHING SYSTEMS.
PART 3: Breathing System Parts
Anesthesia Equipment Session 3
Non-Rebreathing System
Parts of the Anesthetic Machine Parts 1 & 2: A&A pages
Mechanical Ventilator 2
Pneumatic schematic of the Dräger Infinity V500 intensive care ventilator. A. Gas-mixture and gas-metering assembly. Gas from the supply lines enters the.
T-6B On-Board Oxygen Generation System (OBOGS)
Non-Rebreathing System
Sophie Young 21st February 2007
The Anaesthesia Machine
Dip. Software based statistics PhD ( physiology), FICA , IDRA
Pressure regulators (reducing valves)
T-6B On-Board Oxygen Generation System (OBOGS)
Pneumatic System Conditioning
Presentation transcript:

Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK

Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout

ventilator Flow meter bellow vaporizer Corrugated tube APL valve Scavenging system Soda lime

The Anesthesia Machine High Intermediate Low Pressure Circuit Mention how the machine is divided into high, intermediate, and low pressure systems

High Pressure System Receives gasses from the high pressure E cylinders attached to the back of the anesthesia machine (2200 psig for O2, 745 psig for N2O) Consists of: Hanger Yolk (reserve gas cylinder holder) Check valve (prevent reverse flow of gas) Cylinder Pressure Indicator (Gauge) Pressure Reducing Device (Regulator) Usually not used, unless pipeline gas supply is off

E Size Compressed Gas Cylinders Cylinder Characteristics Oxygen Nitrous Oxide Carbon Dioxide Air Color White (green) Blue Gray Black/White (yellow) State Gas Liquid and gas Contents (L) 625 1590 Empty Weight (kg) 5.90 Full Weight (kg) 6.76 8.80 8.90 Pressure Full (psig) 2000 750 838 1800

Hanger Yolk Hanger Yolk: orients and supports the cylinder, providing a gas-tight seal and ensuring a unidirectional gas flow into the machine Index pins: Pin Index Safety System (PISS) is gas specificprevents accidental rearrangement of cylinders (e.g.. switching O2 and N2O)

Pressure Reducing Device Reduces the high and variable pressures found in a cylinder to a lower and more constant pressure found in the anesthesia machine (45 psig) Reducing devices are preset so that the machine uses only gas from the pipeline (wall gas), when the pipeline inlet pressure is 50 psig. This prevents gas use from the cylinder even if the cylinder is left open (i.e. saves the cylinder for backup if the wall gas pipeline fails)

Pressure Reducing Device Cylinders should be kept closed routinely. Otherwise, if the wall gas fails, the machine will automatically switch to the cylinder supply without the anesthetist being aware that the wall supply has failed (until the cylinder is empty too).

Intermediate Pressure System Receives gasses from the regulator or the hospital pipeline at pressures of 40-55 psig Consists of: Pipeline inlet connections Pipeline pressure indicators Piping Gas power outlet Master switch Oxygen pressure failure devices Oxygen flush Additional reducing devices Flow control valves

Pipeline Inlet Connections Mandatory N2O and O2, usually have air and suction too Inlets are non-interchangeable due to specific threading as per the Diameter Index Safety System (DISS) Each inlet must contain a check valve to prevent reverse flow (similar to the cylinder yolk) At this point mention that in Sudbury Ontario in the 1970s 23 people died because the N20 and O2 pipelines were crossed over during repairs. Ask them what they would do in this situation? Backup oxygen cylinder turned on Pipeline supply must be disconnected (necessary because the machine will preferentially use the pipeline supply as per the high pressure system regulator design).

Oxygen Pressure Failure Devices Machine standard requires that an anesthesia machine be designed so that whenever the oxygen supply pressure is reduced below normal, the oxygen concentration at the common gas outlet does not fall below 19%

Oxygen Pressure Failure Devices A Fail-Safe valve is present in the gas line supplying each of the flowmeters except O2. This valve is controlled by the O2 supply pressure and shuts off or proportionately decreases the supply pressure of all other gasses as the O2 supply pressure decreases Historically there are 2 kinds of fail-safe valves Pressure sensor shut-off valve (Ohmeda) Oxygen failure protection device (Drager)

Pressure Sensor Shut-Off Valve Oxygen supply pressure opens the valve as long as it is above a pre-set minimum value (e.g.. 20 psig). If the oxygen supply pressure falls below the threshold value the valve closes and the gas in that limb (e.g.. N2O), does not advance to its flow-control valve.

Oxygen Failure Protection Device (OFPD) Based on a proportioning principle rather than a shut-off principle The pressure of all gases controlled by the OFPD will decrease proportionately with the oxygen pressure

Oxygen Supply Failure Alarm The machine standard specifies that whenever the oxygen supply pressure falls below a manufacturer-specified threshold (usually 30 psig) a medium priority alarm shall blow within 5 seconds.

Limitations of Fail-Safe Devices/Alarms Fail-safe valves do not prevent administration of a hypoxic mixture because they depend on pressure and not flow. These devices do not prevent hypoxia from accidents such as pipeline crossovers or a cylinder containing the wrong gas

Limitations of Fail-Safe Devices/Alarms These devices prevent hypoxia from some problems occurring upstream in the machine circuitry (disconnected oxygen hose, low oxygen pressure in the pipeline and depletion of the oxygen cylinder) Equipment problems that occur downstream (for example leaks or partial closure of the oxygen flow control valve) are not prevented by these devices.

Oxygen Flush Valve (O2+) Receives O2 from pipeline inlet or cylinder reducing device and directs high, unmetered flow directly to the common gas outlet (downstream of the vaporizer) Machine standard requires that the flow be between 35 and 75 L/min The ability to provide jet ventilation Hazards May cause barotrauma Dilution of inhaled anesthetic

Second-Stage Reducing Device Located just upstream of the flow control valves Receives gas from the pipeline inlet or the cylinder reducing device and reduces it further to 26 psig for N2O and 14 psig for O2 Purpose is to eliminate fluctuations in pressure supplied to the flow indicators caused by fluctuations in pipeline pressure

Low Pressure System Extends from the flow control valves to the common gas outlet Consists of: Flow meters Vaporizer mounting device Check valve Common gas outlet

Flowmeter assembly When the flow control valve is opened the gas enters at the bottom and flows up the tube elevating the indicator The indicator floats freely at a point where the downward force on it (gravity) equals the upward force caused by gas molecules hitting the bottom of the float

Arrangement of the Flow-Indicator Tubes In the presence of a flowmeter leak (either at the “O” ring or the glass of the flow tube) a hypoxic mixture is less likely to occur if the O2 flowmeter is downstream of all other flowmeters In A and B a hypoxic mixture can result because a substantial portion of oxygen flow passes through the leak, and all nitrous oxide is directed to the common gas outlet * Note that a leak in the oxygen flowmeter tube can cause a hypoxic mixture, even when oxygen is located in the downstream position

Proportioning Systems Mechanical integration of the N2O and O2 flow-control valves Automatically intercedes to maintain a minimum 25% concentration of oxygen with a maximum N2O:O2 ratio of 3:1

Limitations of Proportioning Systems Machines equipped with proportioning systems can still deliver a hypoxic mixture under the following conditions: Wrong supply gas Defective pneumatics or mechanics (e.g.. The Link-25 depends on a properly functioning second stage regulator) Leak downstream (e.g.. Broken oxygen flow tube) Inert gas administration: Proportioning systems generally link only N2O and O2

Vaporizers A vaporizer is an instrument designed to change a liquid anesthetic agent into its vapor and add a controlled amount of this vapor to the fresh gas flow

Classification of Vaporizers Methods of regulating output concentration Concentration calibrated (e.g. variable bypass) Measured flow Method of vaporization Flow-over Bubble through Injection Temperature compensation Thermocompensation Supplied heat

Generic Bypass Vaporizer Flow from the flowmeters enters the inlet of the vaporizer The function of the concentration control valve is to regulate the amount of flow through the bypass and vaporizing chambers Splitting Ratio = flow though vaporizing chamber/flow through bypass chamber The concentration control dial may be located in the bypass chamber of the outlet of the vaporizing chamber Mention that the machine standard requires that all vaporizers on the anesthesia workstation be concentration calibrated (aka variable bypass, direct reading, dial-controlled, automatic plenium, percentage-type, tec-type vaporizers, and vaporizer chamber bypass arrangements

Factors That Influence Vaporizer Output Flow Rate: The output of the vaporizer is generally less than the dial setting at very low (< 200 ml/min) or very high (> 15 L/min) flows Temperature: Automatic temperature compensating mechanisms in bypass chambers maintain a constant vaporizer output with varying temperatures Back Pressure: Intermittent back pressure (e.g. positive pressure ventilation causes a higher vaporizer output than the dial setting) C’ = concentration at the new atmospheric pressure C = the concentration at the old atmospheric pressure (i.e. the concentration dialed into the vaporizer) P’ = the barometric pressure for which c’ is being established P = the barometric pressure for which the vaporizer is calibrated (i.e. at the old atmospheric pressure)

Factors That Influence Vaporizer Output Atmospheric Pressure: Changes in atmospheric pressure affect variable bypass vaporizer output as measured by volume % concentration, but not (or very little) as measured by partial pressure (lowering atmospheric pressure increases volume % concentration and vice versa) Carrier Gas: Vaporizers are calibrated for 100% oxygen. Carrier gases other than this result in decreased vaporizer output.

The Circuit: Circle System Arrangement is variable, but to prevent re-breathing of CO2, the following rules must be followed: Unidirectional valves between the patient and the reservoir bag Fresh-gas-flow cannot enter the circuit between the expiratory valve and the patient Adjustable pressure-limiting valve (APL) cannot be located between the patient and the inspiratory valve

Circle System Advantages: Disadvantages: Relative stability of inspired concentration Conservation of respiratory moisture and heat Prevention of operating room pollution PaCO2 depends only on ventilation, not fresh gas flow Low fresh gas flows can be used Disadvantages: Complex design = potential for malfunction High resistance (multiple one-way valves) = higher work of breathing The final three points of the advantages section are meant to contrast with the Bain circuit

The Adjustable Pressure Limiting (APL) Valve User adjustable valve that releases gases to the scavenging system and is intended to provide control of the pressure in the breathing system Bag-mask Ventilation: Valve is usually left partially open. During inspiration the bag is squeezed pushing gas into the inspiratory limb until the pressure relief is reached, opening the APL valve. Mechanical Ventilation: The APL valve is excluded from the circuit when the selector switch is changed from manual to automatic ventilation

Scavenging Systems Protects the breathing circuit or ventilator from excessive positive or negative pressure. Mention that the ventilator relief valve is the same thing as the overflow valve mentioned in the previous slides

Scavenging Systems

Checking Anesthesia Machines 8 Categories of check: Emergency ventilation equipment High-Pressure system Low-Pressure system Scavenging system Breathing system Manual and automatic ventilation system Monitors Final Position

The end Dr Khezri