X-RAY IMAGING INTENSIFIER

Slides:



Advertisements
Similar presentations
X-RAYS.
Advertisements

Radiation Safety Training for Fluoroscopy in Research Radiation Safety Office Indiana University Purdue University Indianapolis and Associated Facilities.
Veterinary Radiation Safety 4 Students Radiation Safety MA Law requires that persons who perform diagnostic radiology procedures on animal patients be.
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
CVT 102.  Time  Distance  Shielding.
Digital FLUORO Summer 2008.
Radiation Occupational exposures and protection A. H. Mehrparvar, MD Occupational Medicine department Yazd University of Medical Sciences.
Veterinary Radiology Safety and Procedures. What is a radiograph? When an xray beam (a form of electromagnetic radiation) penetrates tissue to form a.
Basic quality control tests in medical radiography
Occupational exposure and protective devices
Radiation Exposure and Risk John Hamaty. Evaluation of Radiation Exposure Levels in Cine Cardiac Catheterization Laboratories American Association of.
Digital Radiography Chapter 22. History of Digital Radiography Slower process of conversion because no pressing need to convert to digital radiography.
The X-ray Imaging System
RADIATION SAFETY Phil Facey Lead Superintendent Radiographer
Conventional and Computed Tomography
THE X-RAY UNIT THE BASICS WHAT FOUR COMPONENTS ARE IN TYPICAL RADIOGRAPHIC ROOM? 1. X-RAY TUBE 2. OPERATING CONSOLE 3. HIGH VOLTAGE GENERATOR 4. X-RAY.
Reference Reading Chapter 2: pp  X-rays are produced within the dental x-ray machine  The x-ray machine can be divided into 3 study areas.
Saira Ahmad UOG. CAT Scans CAT Scans ( Computerized axial tomography) Topic:
CREATION OF AN OPTIMAL ANGIO SUITE A. CHATZIOANNOU.
Personal Radiation and Occupational Safety – CT Policy This policy is in both paper form in the lab’s compliance manual, and on a Power Point presentation.
Dental Assisting Radiology
RADIOLOGIC IMAGING EQUIPMENT 1. THE X-RAY ROOM 2.
RAD 350 Chapt. 25 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic motion Invented by Thomas Edison.
Introduction to Fluoroscopy & Radiation Safety
 Veterinary Technicians 101 In cooperation with: The Emma Barnsley Foundation The PEER Program at Texas A&M College of Veterinary Medicine & Biomedical.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Chapter 3 Radiation Safety Vet Tech Institute. Radiation should be respected not Feared! Safety is Always important! Safety is Always important! Stray.
FLUOROSCOPY EQUIPMENT
UNIT 6 Introduction to Radiation Protection
ERCP Q&A.
FLUROSCOPY. What is Fluoroscopy? Fluoroscopy is a method of using low intensity X-ray beams to continuously visualize the area of interest in "real.
Ch. 2 – Anatomy of the X-ray Machine
Northern California SGNA Meeting September 22, 2012.
CHAPTER 3 EQUIPMENT OPERATION AND QUALITY CONTROL
1 DA105 RADIOLOGY RADIATION HEALTH AND SAFETY – Radiation Control for Health and Safety Act – Standardized xray equipment; required filtration,
Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON.
RAD 254 Digital Imaging Basic Elements of Digital Imaging CR/DR.
Angiographic facilities Tortorici Chapter 11. Designed around exam table 500 sq feet min. (25 x 20) 60 sq feet in control room Wide doors, (multiple)
RADIOLOGIC IMAGING EQUIPMENT 1. THE X-RAY ROOM 2.
RADIATION SAFETY Mrs. Brinston. Introduction As a healthcare worker, you know that radiation is an important tool for detecting and treating diseases.
Week 1 Review Don’t forget: You can copy- paste this slide into other presentations, and move or resize the poll.
Radiation Sources in medicine diagnostic Radiology
Designing for Radiation Protection. Design Standards for Radiation Protection – Leakage Radiation.
Week 2: Radiographic Equipment
Radiation Sources in medicine diagnostic Radiology
Week 2 :Radiation Protection
RAD 150 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic (motion)
RAD 354 Chapt. 25 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic motion.
X-Ray Box Imran Awan Experimental High Energy Physics November 12, 2015.
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
Radiation hazards Mamoun Kremli, SA AOT Basic Principles Course.
Tube Exposure Factors Math Technique Contrast and Density
Week 2: Radiographic Equipment
Protection1 Radiation Protection. Protection2 BENEFITS VS. ADVERSE EFFECTS.
IAEA International Atomic Energy Agency Fluoroscopy Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(2)
AJADI ADETOLA DEPARTMENT OF VETERINARY MEDICINE AND SURGERY.
Introduction to and History of Fluoroscopy Radiation Protection
Radiation Badge Training
Pediatric Fluoroscopy
Mobile Fluoroscopy Based on:
Mobile Fluoroscopy Based on:
Introduction to and History of Fluoroscopy Radiation Protection
Pediatric Fluoroscopy
Fluoroscopy Static Image Recording Systems
Pediatric Fluoroscopy
Occupational Radiation Dose Management
Occupational Radiation Dose Management
Fluoroscopy: Static Image Recording Systems
Mobile Fluoroscopy.
Radiation Badge Training
Presentation transcript:

X-RAY IMAGING INTENSIFIER BY RANGANATHAN DURIASAMY www.surgicaltechie.com rengan9095@gmail.com

INTRODUCTION An x-ray image intensifier, is an imaging component which Converts X-rays into a visible image. The term image intensifier refers to a specific component of an x-ray imaging system, which allows low intensity x-rays to be converted to a visible light output. The device contains a low absorbency/scatter input window, typically aluminum, input fluorescent screen, photocathode, electron optics, output fluorescent screen and output window. This device was originally introduced in 1948

A C-arm comprises a generator (X-ray source) and an image intensifier or flat-panel detector. The C-shaped connecting element allows movement horizontally, vertically and around the swivel axes, so that X-ray images of the patient can be produced from almost any angle. The generator emits X-rays that penetrate the patient's body. The image intensifier or detector converts the X-rays into a visible image that is displayed on the C-arm monitor. The doctor can identify and check anatomical details on the image such as blood vessels, bones, kidney stones and the position of implants and instruments at any time.

Clinical applications Modern imaging systems will use the image intensifier as the source of images supplied to a storage system. As a fixed piece of equipment in a dedicated screening room mobile equipment for use in an operation theatre It can be processed and print the on going shot pictures within few minutes Fluoroscopy delivers a dose of approximately 5 Rads per minute in the direct beam  

Components C-arm (encompasses the actual X-ray source and image intensifier) Table Fluoroscopic exposure and program controls Post processing software Viewing monitors Foot control and hand switch Connecting cables and power cord

C-ARM IMAGE INTENSIFIER MONITOR IMAGE INTENSIFIER(CAMERA) C-ARM TABLLE WITH PATIENT X-RAY SOURCE HAND SWITCH CONTROL PANEL C-ARM IMAGE INTENSIFIER MODERN TYPE WITH VIDEO FLUORSOCSOPY

Permanent/Fixed Fluoroscopic Systems Two main configurations in permanently installed fluoroscopic systems 1. One class commonly utilizes a radiolucent patient examination table with an under- table mounted tube and an imaging system mounted over the table 2.Other is commonly referred to as a C-arm system used where greater flexibility in the examination process is needed such as neuro or cardiac imaging

Permanent/Fixed Fluoroscopic Systems Conti., Modern imaging systems on both configurations are limited . All frame rates, storage (local or PACS), image capture devices lower in cost than before. software configurable and based on COTS components .

The non-C-arm - screening rooms. The Types of investigations Endoscopy studies (ERCP) (Some sites will opt for a portable C-arm system for this) Barium studies (swallows, meals, enemas)  Fertility studies (HSG) For diagnosis purpose of the various departments orthopaedic,neuro,gastro)during emergency

C-arm-portable x-ray machine The C-arm systems are commonly used for studies requiring the maximum positional flexibility such as Angiography studies (peripheral, central and cerebral) Therapeutic studies (Line placements .e. Permacath/Hickman, transjugular biopsies, TIPS stent, embolisations) Cardiac studies (PTCA) Orthopedic procedures (ORIF, DHS, MUA, spinal work) - again generally using a portable C-arm maximum flexibility in positional use. Intra-operative Urological surgery Used in paediatric surgery

General configuration A mobile image intensifier has units, 1. X-ray generator 2. imaging system on (C-arm) 3. workstation unit used to store and manipulate the images. The imaging system unit can perform a variety of movements that allow for use in a variety of surgical procedures such as cardiology, orthopedics and urology

X-ray Generator X-ray generator, dose control system and collimator controls are usually housed in the chassis on which the C-arm is mounted. All of the control systems are closed loop systems which are directed by the master controller initial program settings.

Imaging system

IMPORTANT NOTES TO OPERATORS Kvp must be appropriate for the study.For C-arms, typically the kVp and ma are adjusted automatically. Higher kVp, the scatter radiation is more penetrating. Do not override and use kVp much higher than needed Use of wrap around style aprons can help.Lead gloves are required if hands are in the primary beam.

CONTINUE.,, Cover the C-ARM with sterile sheet or draping that readily available in the market. Maintain the sterility during the surgery. Don’t disturb the surgeon and assistants Make sure that door kept closed and all the surgical teams wearing the saftey attire or not

SAFETY MEASURES Lead aprons (shield) Wear a lead apron during fluoroscopy while using 0.5 mm lead equivalent for fluoro aprons - “wrap around” style aprons are best for persons who need to move about in the room.   Thyroid Shield Use of a thyroid shield keeps worker’s radiation Lead glasses For persons who routinely fluoro for long or interventional procedures, lead glasses with side shields can provide additional protection to the lens of the eyes. Note that the operator often has to look sideways from the C-arm x-ray tube to see the image on the monitor. This leaves the lens unprotected if glasses do not have side shields

Safety Measures Conti., Lead Shield Care Hang aprons and shields on racks.Do not bend or fold lead aprons or shields.Folding can cause cracks and tears in the protective material.Periodically inspect shields for evidence of damage. Remove damaged ones from use.  Don’t allow any health workers without wearing safety attires Red light or radiation symbols should be there out the room doors

C-ARM READY FOR USING

Radiation Badges All badges are certified to ensure maximum accuracy of the dose report. Wear the badge flat against your body at your neck. [Unless you are pregnant and issued a waist badge also.] Do not wear badges on a chain, which would allow the badge to turn at various angles as you work. Turn in the badge for processing at the end of the monitoring period. The badges must be sent to the vendor for processing.  

RADIATION

Remember to: Wear only your own badge Wear it whenever working at Tufts Univ. Leave it in a cool, dry place away from radiation when not in use Do not take your badge home. Do not launder the badge or get it wet. Do not expose to heat, such as in a car in summer. Do not open badge. Do not expose the badge to other sources of radiation. Do not wear the badge for personal x-ray or nuclear medicine exams. Turn in your badge for processing in a timely manner. Please cancel badge service when no longer needed.

IMAGES OF X RAY SAFETY SYSTEM

THANK YOU