Designing a Microscopy Experiment

Slides:



Advertisements
Similar presentations
Adjusting a Microscope 1Center components on optic axis 2Focus objective 3Focus condenser 4Adjust illumination lamp voltage (intensity) iris diaphragm.
Advertisements

CELL BIOLOGY TECHNIQUES
Microscopy Outline 1.Resolution and Simple Optical Microscope 2.Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3.Newer.
Microscopy Lecture I.
ERT107 MICROBIOLOGY FOR BIOPROCESS ENGINEERING Pn Syazni Zainul Kamal PPK Bioprocess.
Nick Beavers Project Manager Deconvolution from Andy Molnar Software Engineer.
1. Resolution (R) : separation of close objects, light wavelength NA, numerical aperture 2. Contrast : distinction of objects from background “light field”
Blizard Advanced Light Microscopy club Making friends with your microscope.
ABRF meeting 09 Light Microscopy Research Group. Why are there no standards? Imaging was largely an ultrastructure tool Digital imaging only common in.
Fluorophores bound to the specimen surface and those in the surrounding medium exist in an equilibrium state. When these molecules are excited and detected.
Image-Pro Premier Basic Training Course
Lasers and Confocal. Laser Acronym: Light Amplification by Stimulated Emission of Radiation Ordinary light emission: Comes from spontaneous decay of excited.
Comparison and Implications of Three Optical Microscopy Data Acquisition Modalities James Butler Ph.D. Nikon Instruments, Inc. Widefield Fluorescence Confocal.
Biology 177: Principles of Modern Microscopy Lecture 07: Confocal Microscopy Adding the Third Dimension.
1 Confocal Microscopy David Kelly November 2013 Handbook of Biological Confocal Microscopy. Ed. J. Pawley, Plenum Press Fundamentals of Light Microscopy.
Fluorescence microscopy – Principle and practical consideration Hiro Ohkura.
Optics. Thomas Young ( ) Wave Optics Diffraction & Interference Christiaan Huygens ( )
Geometrical Optics and Basic Imaging Light Paths of the Bright Field Microscope E. D. Salmon University of North Carolina at Chapel Hill.
Short pulses in optical microscopy Ivan Scheblykin, Chemical Physics, LU Outline: Introduction to traditional optical microscopy based on single photon.
Laser scanning confocal fluorescence microscopy: an overview Liu Jie B
Introduction to Fluorescence Microscopy. Introduction to fluorescence microscopy Fluorescence Widefield Fluorescence microscopes Filters and Dichroics.
P. Moghe, 125:583 1 Microscopy Techniques for Biomaterials and Cell Based Interfaces Professor Prabhas V. Moghe October 26, :583 Fall 2006.
USE AND CARE OF THE MICROSCOPE LECTURE 1. MICROSCOPY u Light Microscopy: any microscope that uses visible light to observe specimens u Compound Light.
STORM and PALM How to get started. June 2014 Christa Walther.
Media Cybernetics Deconvolution Approaches and Challenges Jonathan Girroir
Study of Protein Association by Fluorescence-based Methods Kristin Michalski UWM RET Intern In association with Professor Vali Raicu.
1 Basics of Digital Imaging Digital Image Capture and Display Kevin L. Lorick, Ph.D. FDA, CDRH, OIVD, DIHD.
Microscope.
A microscope is a Complicated Lens n The microscope can be represented by a single convex lens.
Practical aspects of fluorescence microscopy
Microscopy.
Cameras Course web page: vision.cis.udel.edu/cv March 22, 2003  Lecture 16.
Bioimaging ChemEng 590B: Lecture 15 4/11/13. Rat mammary carcinoma cells 10 min, images every 20 seconds Michele Balsamo, Gertler lab MIT 1. Imaging Cells.
Bioimaging ChemEng 575: Lecture 16 4/15/14. Rat mammary carcinoma cells 10 min, images every 20 seconds Michele Balsamo, Gertler lab MIT 1. Imaging Cells.
TIRF Total Internal Reflection Fluorescence Microscopy specialized fluorescence microscopy technique specifically images a very thin optical section (50-250nm)
Copyright © 2006 Pearson Education, Inc., publishing as Benjamin Cummings Micro 20 Lab 2- Microscopy.
Microscopy. Scale Lenses and the Bending of Light light is refracted (bent) when passing from one medium to another refractive index –a measure of how.
FNI 2B OM 1 Optical Microscopes. FNI 2B OM2 Outline Justification History Components of the Optical Microscope Theory of operation  Basic Microscope.
IPC Friedrich-Schiller-Universität Jena 1 Reduction of out of focus light  Excitation light excites fluorescence more or less within the whole sample.
Agenda How to make the specimen visible – Definition of Contrast
Quantitative Imaging Using imaging to analyze molecular events in living cells Ann Cowan.
BIO 224 Intro to Molecular and Cell Biology. Microscopes are tools frequently used in cell biology Type of microscope used depends on the specimen being.
Basics of Deconvolution
Epi-illumination is form of Kohler Illumination:
Digital Imaging. Digital image - definition Image = “a two-dimensional function, f(x,y), where x and y are spatial coordinates, and the amplitude of f.
Functional cellular imaging by light microscopy MICROSCOPIES.
NANO 230 Micro/Nano Characterization
Molecular Cell Biology Light Microscopy in Cell Biology Cooper Modified from a 2010 lecture by Richard McIntosh, University of Colorado.
Science 10 – Unit C BIOLOGY Chapter 1 – The Microscope.
Use Snell’s law to determine the angle of total internal reflection in the coverslip (without oil). Oil immersion objectives More light (no total internal.
Principles and Practices of Light Microscopy II: Brightfield optics, resolution, and aberrations.
Designing a Microscopy Experiment Kurt Thorn, PhD Director, Image from Susanne Rafelski, Marshall lab.
Designing a Microscopy Experiment Kurt Thorn, PhD Director, Image from Susanne Rafelski, Marshall lab.
Confocal Microscopy Kurt Thorn NIC.
Dr R.Jayaprada CONFOCAL MICROSCOPY.
Principles & Practice of
Microscopes Moisey (2009), Optical Microscope [photograph]. Retrieved from
Microscopy Dr. Bhavesh Patel Principal V.P. and R.P.T.P. Science College Vallabh Vidyanagar –
Microscopy.
Microscopy Group 2 Cabatit, Mendoza, Ramos, Rodriguez, Tan.
Light Microscope Terms and Practices.
Microscope.
Lasers and Confocal.
Oil immersion objectives
NANO 230 Micro/NanoFabrication
LIGHT MICROSCOPY basic
LIGHT MICROSCOPY variations
Harvard Medical School Neurobiology Imaging Facility
Harvard Medical School Neurobiology Imaging Facility
Imaging techniques for next generation plant cell biology.
Presentation transcript:

Designing a Microscopy Experiment Kurt Thorn, PhD Director, NIC@UCSF Image from Susanne Rafelski, Marshall lab

The Chinese-menu view of imaging Imaging Methods Wide-Field TIRF Laser-scanning confocal Multi-point confocal Multi-photon confocal Contrasting techniques Brightfield, phase, DIC Immunofluorescence Physiological dyes Fluorescent proteins FRAP, FLIP, photoactivation FRET, FLIM Experiment: Time Lapse, 3D, multi-point, multi-wavelength, …

Standard microscope capabilities Like all rules, these were made to be broken, but only if you have fancy equipment! Resolution: ~200nm in X and Y, 700 nm in Z Sensitivity: <100 photons Linear detection – quantification is possible Video rate acquisition 4-5 color imaging

By far the most important part: the Objective Lens Obviously, we care about the magnification. What other parameters are important?

Working Distance In general, high NA lenses have short working distances However, extra-long working distance objectives do exist Some examples: 10x/0.3 WD = 15.2mm 20x/0.75 WD = 1.0mm 100x/1.4 WD = 0.13mm

Numerical Aperture = light gathering angle 100X / 0.95 NA  = 71.8° NA = n sin() where 4X / 0.20 NA  = 11.5° = light gathering angle n = refractive index of sample

Resolution of the Microscope limited by the point-spread function Microscope objective collects a limited cone of light from the sample This limits the resolution achievable by the microscope Resolution can be measured by the blurring of a point object → the point-spread function Objective  Sample

Resolution of the Microscope limited by the point-spread function Microscope objective collects a limited cone of light from the sample This limits the resolution achievable by the microscope Resolution can be measured by the blurring of a point object → the point-spread function Y X Z

Resolution of the Microscope limited by the point-spread function Resolution: X-Y, 0.61l / NA; Z, ln / NA2 Resolution for some common objectives, in nm: NA X-Y Z 0.3 1017 16830 0.75 407 2690 0.95 321 1680 1.4 218 770

Light-gathering power Light-gathering power goes as the square of NA All things being equal, a higher NA lens will give a brighter image Increasing magnification generally decreases brightness as light is spread out over more pixels NA Brightness 0.3 0.09 0.75 0.56 0.95 0.90 1.4 1.96

Choosing an objective Questions: What resolution do you need? How bright is your sample? For high resolution, you’ll need high NA. For dim samples, you’ll want high NA, regardless of resolution, to maximize light-gathering. Dim, low-resolution samples (e.g. protein abundance in nucleus): bin camera to trade off resolution for brightness

Choosing an objective Questions: What resolution do you need? How bright is your sample? When to use low NA? Bright samples at low resolution / low magnification If you need long working distance If spherical aberration is a concern If you want large depth of field to get whole structures in focus at once (avoid Z-stacks)

NA, Z-resolution, and Z-stacks Idea – want to record total nuclear fluorescence Depth of field nucleus nucleus High NA Low NA Need multiple Z-sections to capture entire nucleus Only need a single image to capture entire nucleus

Confocal Microscopy Confocal microscopy has the same resolution as widefield, but eliminates out-of-focus light. This improves contrast for thick, heavily stained specimens. However, it usually comes at a cost in sensitivity.

Fluorescence Illumination of a single point Camera Tube lens Excitation light Emission light Objective lens Sample Problem – fluorescence is emitted along entire illuminated cone, not just at focus

The confocal microscope Detector Pinhole Tube lens Excitation light Emission light Objective lens Sample

What do you get?

Confocal vs. Widefield Confocal Widefield 20 mm rat intestine section recorded with 60x / 1.4NA objective

Confocal vs. Widefield Confocal Widefield Tissue culture cell with 60x / 1.4NA objective

The confocal microscope Scan excitation spot point-by-point to build up image Problems: Slow (~1 sec to acquire an image) Low light efficiency (due to use of PMT as detector) Solution: Use multiple pinholes and a camera Detector Pinhole Tube lens Excitation light Emission light Objective lens Sample

Spinning Disk Confocal

Out-of-focus rejection Defocus

Relative Sensitivity Widefield 100 Spinning-Disk Confocal 25 Laser-scanning Confocal 1 See Murray JM et al, J. Microscopy 2007 vol. 228 p390-405

Multi-photon excitation Internal conversion S1 hA hA hE hA S0 2 1 Brad Amos, MRC, Cambridge Multi-photon excitation does not excite out-of-focus light, so you can get rid of pinhole

Total Internal Reflection: TIRF Thinnest optical sections: Images ~100nm section adjacent to coverslip Illuminate through the objective Nabeam: 1.38 Intensity 1.42 Sample on cover glass 1.46 Z

Epifluorescence vs. TIRF Jaiswal et al 2002; cells loaded with FITC-dextran

Which imaging technique should I use? 1-5 mm TIRF (for samples at the coverslip) Wide-field (+deconvolution) 1-20 mm Fast Spinning Disk Confocal Sensitivity Sample Thickness 10-100 mm Line-scanning confocal >20 mm Point scanning Confocal Slow >50 mm 2-photon confocal

Microscope choice Epifluorescence – routine work, low magnification, or thin samples where you don’t need high-resolution 3D reconstruction TIRF – samples at the membrane or otherwise at the coverslip surface; very high signal-to-noise; single molecule imaging Spinning Disk Confocal – Live tissue culture cells, yeast, etc, or thin (<30 mm) tissue sections when you need 3D reconstructions Laser-Scanning Confocal – Thick tissues or specimens

Sample preparation and mounting Mounting media serve several purposes: Stabilizing the sample Preventing photobleaching Clearing the sample Matching refractive index

& Spherical Aberration Index Mismatch & Spherical Aberration objective Immersion fluid n1 Spherical aberration unless n2 = n1 Cover glass n2 Sample Focus at cover slip Focus into sample

& Spherical Aberration Index Mismatch & Spherical Aberration z=0 µm n1=1.515 (oil) z=25 µm n2=1.44 (Vectashield) z=50 µm

What can you do about spherical aberration? Use 0.17 mm coverslips (~ #1.5) Work close to the coverslip Match lenses to the refractive index of your samples, and vice versa For aqueous samples, use water immersion / water dipping lenses For fixed samples and oil immersion lenses, mount your sample in a medium with n = 1.515 Adjust objective correction collar when available Use lower NA lenses

Clearing Clearing media dissolve lipids to make samples more transparent Can be important for thick samples and tissues Commonly used: BABB = 1:2 Benzyl Alchohol : Benzyl Benzoate Methyl Salicylate

Sample Preparation Samples imaged with 20x / 0.75 air objective on spectral confocal Sections acquired ~ 50 mm into tissue Embryonic mouse lungs; samples from Nan Tang, Martin Lab

Sample Preparation Samples imaged with 40x / 1.3 oil objective on spectral confocal Sections acquired ~ 50 mm into tissue Embryonic mouse lungs; samples from Nan Tang, Martin Lab

Dye choices – Fixed samples Common filter set is DAPI / FITC / Rhodamine / Cy5 Dye choices: DAPI / Hoechst / Alexa 350 / Alexa 405 Alexa 488 Rhodamine / Alexa 546 / Alexa 568 Cy5 / Alexa 647 / Atto 647 More than four colors probably requires special filters or spectral imaging.

Dye choices – Live samples Common filter sets: GFP / mCherry, CFP / YFP, CFP / YFP / RFP Two-color choice: GFP / mCherry Three-color: CFP / GFP / mCherry or CFP / YFP / mCherry or BFP / GFP / mCherry Four-color: BFP / CFP / YFP / mCherry or Sapphire / CFP / YFP / mCherry Five-plus colors: possible but tricky, probably requires custom filters or spectral imaging.

Time and noise - tradeoffs The number of photons collected by the camera generally determines the amount of noise in your image Noise = square root (# of photons) Doubling signal to noise ratio requires 4-fold increase in exposure

What does this look like? With 5 e- camera read noise 1000 photons / pixel 100 photons / pixel 10 photons / pixel

Noise and resolution Theoretical perfect data Two spots separated by diffraction limit Slightly oversampled

Noise and resolution With shot noise 1000 ph/pixel at peak

Expected error bars with shot noise Noise and resolution Expected error bars with shot noise 1000 ph/pixel at peak 100 ph/pixel at peak 10 ph/pixel at peak

Noise and resolution High resolution and precise quantitation both require lots of light This means bright samples or long exposures This may cause problems with photobleaching and phototoxicity Be aware of potential tradeoffs between precision, speed, and photobleaching

Colocalization Measures co-occurrence within the resolution limit of the microscope. Does not say anything about molecular interaction Huh et al. 2003

Nothing beats good data Think about what data you need before you take it. Do you need Time resolution? Spatial resolution? Intensity resolution? Day-to-day reproducibility? Spatial uniformity? You can fix a lot of problems with post-processing, but it’s better to fix problems in the data collection!

If you care about it, you should measure it! Spatial uniformity Illumination and detection is not uniform over the field of view of the microscope. Can be measured and corrected with a shading image. Photobleaching may make this hard Temporal uniformity Lamp power and alignment fluctuates from day to day Can measure But best to do experiments same day / same session

Think about data storage Databases are good, but cumbersome Save in manufacturer’s native format so metadata is preserved If not using a database, systematic file names and notes on sample identity are a good idea