POTENTIOMETRY.

Slides:



Advertisements
Similar presentations
Review of Analytical Methods Part 2: Electrochemistry
Advertisements

Potentiometric Methods
Potentiometric Methods A.) Introduction:
15-1 Potentiometry Potential measurements of electrochemical cells Ion selective methods §Reference electrode §Indicator electrode §Potential measuring.
Fig. 22-1a (p.629) A galvanic electrochemical cell at open circuit
Summary of Potentiometry:
Reactions in Aqueous Media
Introduction to Analytical Chemistry
ELECTROCHEMISTRY CHEM 4700 CHAPTER 5 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
MLAB 2401: C LINICAL C HEMISTRY K ERI B ROPHY -M ARTINEZ Designs in Instrumentation Electrochemistry.
Mrs Khadijah Hanim bt Abdul Rahman Sem II 2011/2012 Week 15: 28 & 31 May 2012.
Electroanalytical – Introduction Ch 21, 7th e, WMDS)
Measuring concentration using electrodes Indicator electrodes used with reference electrode to measure potential of unknown solution E cell = E indicator.
POTENTIOMETRY 9th lecture
Electrodes and Potentiometry
Section 11 Potentiometry. Potentiometric Electrodes Potentiometric electrodes measure: Activity not concentration Concepts to review: Activity and affect.
14-6 Cells as chemical probes
Potentiometry. pH is a Unit of Measurement  pH = Power of Hydrogen (H + )  Defined as the Negative Logarithm of Hydrogen Ion Activity  pH = log (1/H.
Chem. 133 – 3/5 Lecture. Announcements Lab –Set 2 Period 2 Labs Set to Finish 3/10 –3/12 will be make up day (for both Period 1 and Period 2 labs) –Set.
Potentiometry and the Detection of Fluoride
Fall 2014, Prof. JB Lee Ion Sensitive FET (ISFET) - What and Why?
Potentiometric Methods
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 15 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Prentice Hall © 2003Chapter 20 Zn added to HCl yields the spontaneous reaction Zn(s) + 2H + (aq)  Zn 2+ (aq) + H 2 (g). The oxidation number of Zn has.
Electrochemistry Chapter and 4.8 Chapter and 19.8.
Fundamentals of Electrochemistry Introduction 1.)Electrical Measurements of Chemical Processes  Redox Reaction involves transfer of electrons from one.
POTENTIOMETRY 8th lecture
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 14
电化学的有关概念: 电池、电极 双电层结构 液接电位、盐桥 传质过程:扩散、迁移、对流 极化:浓差极化、电化学极化 过电位
1 Oxidation Reduction Equilibria and Titrations. 2 Oxidation - Reduction reactions (Redox rxns) involve the transfer of electrons from one species of.
Fundamentals of Electrochemistry Introduction 1.)Electrical Measurements of Chemical Processes  Redox Reaction involves transfer of electrons from one.
A Seminar on Potentiometric and Diazotization Titrations
Electroanalytical methods
Potentiometry and potentiometric measurements. potentiometer A device for measuring the potential of an electrochemical cell without drawing a current.
Potentiometry Dr Hisham E Abdellatef 2010.
Unit 2 Electrochemical methods of Analysis
Electrochemistry Chapter 19. 2Mg (s) + O 2 (g) 2MgO (s) 2Mg 2Mg e - O 2 + 4e - 2O 2- Oxidation half-reaction (lose e - ) Reduction half-reaction.
Announcements Course Evaluations Final is Wednesday Afternoon on May 9 th Homework 14-15, 14-26, 15-6, 16-3, 16-6,
Chapter 15 Electrode Measurements “Chem 7” test in hospital lab. Na +, K +, Cl -, CO 2, glucose, urea, & creatinine ion-selective electrodes.
1 Chapter Eighteen Electrochemistry. 2 Electrochemical reactions are oxidation-reduction reactions. The two parts of the reaction are physically separated.
Electroanalysis measure the variation of an electrical parameter (potential, current, charge, conductivity) and relate this to a chemical parameter (the.
Electrochemical Methods Dr M.Afroz Bakht. Potentiometry Potentiometry is a method of analysis used in the determination of concentration of ions or substances.
John E. McMurry Robert C. Fay C H E M I S T R Y Chapter 17 Electrochemistry.
CHAPTER 11 ELEMENTS OF ELECTROCHEMISTRY Introduction to Analytical Chemistry.
Electrochemistry.
Electrochemistry Chapter 19 Copyright © The McGraw-Hill Companies, Inc. Permission required for reproduction or display.
LECTURE 7: Electrochemistry. Types of electrodes and their using. ass. prof. Yeugenia B. Dmukhalska.
Electrodes and Potentiometry Introduction 1.)Potentiometry  Use of Electrodes to Measure Voltages that Provide Chemical Information - Various electrodes.
Chem. 133 – 3/3 Lecture. Announcements Homework Set 2 (pass out) Grading –Working to get the Electronics labs and exam 1 graded by next Tuesday Lab –today.
Chapter 21 Potentiometry. Potentiometric methods of analysis are based on measuring the potential of electrochemical cells without drawing appreciable.
Home Reading Skoog et al. Fundamental of Analytical Chemistry. Chapters 18, 19, 21.
Ch. 21 Potentiometry 1. General principles
POTENTIALS AND THERMODYNAMICS OF CELLS (1) POTENTIALS AND THERMODYNAMICS OF CELLS (1)
SCI3023 ELECTROCHEMISTRY Chapter 8e: Potentiometry
Chapter 18 Introduction to Electrochemistry. 18A Characterizing oxidation/reduction reactions In an oxidation/reduction reaction or redox reaction, electrons.
9.2 Electrochemical Cells
Inovace bakalářského studijního oboru Aplikovaná chemie Reg. č.: CZ.1.07/2.2.00/
The reversible electrodes
Chapter 7 Electrochemistry § 7.6 Reversible cell.
ELECTRO CHEMISTRY.. References: 1.Engg.Chemistry by Jain and Jain 2.Engg.Chemistry by Dr. R.V.Gadag and Dr. A.Nithyananda Shetty 3.Principles of Physical.
Potentiometry and Ion-Selective Electrodes 1 Lecture 1.
Chem. 133 – 2/28 Lecture.
Potentiometry and potentiometric measurements
Ion Selective Electrode
Potentiometry Potential measurements of electrochemical cells
Chapter 7 Electrochemistry
Electrode Measurements – Ch 15
Chapter 23 Potentiometry.
Presentation transcript:

POTENTIOMETRY

Potentiometric Analysis Based on potential measurement of electrochemical cells without any appreciable current The use of electrodes to measure voltages from chemical reactions

Applications of Potentiometric Analysis

Components of a Potentiometric Cell Reference electrode Salt bridge Analyte Indicator electrode RE SB A IE – Eref + Ej + Eind

Reference electrode Easily assembled Rugged Half-cell with known potential (Eref) Left hand electrode (by convention) Easily assembled Rugged Insensitive to analyte concentration Reversible and obeys Nernst equation Constant potential Returns to original potential Rugged (unaffected by small variations in method parameters)

Indicator electrode Generates a potential (Eind) that depends on analyte concentration Selective Rapid and reproducible response

Salt bridge Prevents mixing up of analyte components Generates potential (Ej) = negligible

                                                                           

Reference Electrodes Standard Hydrogen Electrode Calomel Reference Electrode Silver/Silver Chloride Reference Electrode

SHE Hydrogen Gas Electrode Pt (H2 (1 atm), H+ (1M) SHE is not often used because of fire hazard Inconvenient to use and maintain

Saturated Calomel Electrode Hg2Cl2(s)+2e-2Hg(l)+2Cl-(aq) Aka SCE Easy to prepare Easy to maintain 0.2444 V at 25C Dependent on temp Toxic This electrode has fallen out of favor because of mercury toxicity. It offers no advantages over other electrodes but still lingers on because so many methods used it for decades

SCE Chloride maintains constant ionic strength

Ag/AgCl Ref. Electrode E = 0.199 V AgAgCl (satd),KCl (satd) AgCl(s) + e-Ag(s)+Cl-(aq) E = 0.199 V

Liquid Junction Potential Liquid junction - interface between two solutions containing different electrolytes or different concentrations of the same electrolyte A junction potential occurs at every liquid junction. Caused by unequal mobilities of the + and - ions.

Indicator Electrodes Metallic IE Electrodes of the First Kind Electrodes of the Second Kind Inert Metallic Electrodes (for Redox Systems) Membrane IE Glass pH IE Glass IE for other cations Liquid Membrane IE Crystalline-Membrane IE Gas Sensing Probes

METALLIC INDICATOR ELECTRODES

Electrodes of the First Kind Pure metal electrode in direct equilibrium with its cation Metal is in contact with a solution containing its cation. M+n(aq) + ne-  M(s)

Disadvantages of First Kind Electrodes Not very selective Ag+ interferes with Cu+2 May be pH dependent Zn and Cd dissolve in acidic solutions Easily oxidized (deaeration required) Non-reproducible response

Electrodes of the Second Kind Respond to anions by forming precipitates or stable complex Examples: Ag electrode for Cl- determination Hg electrode for EDTA determination Slope is positive 3rd kind Electrode responds to different cation Competition with ligand complex

Inert Metallic (Redox) Electrodes Inert conductors that respond to redox systems Electron source or sink An inert metal in contact with a solution containing the soluble oxidized and reduced forms of the redox half-reaction. May not be reversible Examples: Pt, Au, Pd, C

MEMBRANE ELECTRODES Aka p-ion electrodes Consist of a thin membrane separating 2 solutions of different ion concentrations Most common: pH Glass electrode

Glass pH Electrode

Properties of Glass pH electrode Potential not affected by the presence of oxidizing or reducing agents Operates over a wide pH range Fast response Functions well in physiological systems Very selective Long lifespan

Theory of the glass membrane potential For the electrode to become operative, it must be soaked in water. During this process, the outer surface of the membrane becomes hydrated. When it is so, the sodium ions are exchanged for protons in the solution: The protons are free to move and exchange with other ions. Charge is slowly carried by migration of Na+ across glass membrane Potential is determined by external [H+]

Alkaline error Exhibited at pH > 9 Electrodes respond to H+ and alkali cations C,D,E and F: measured value is < true value Electrode also responds to other cations Higher pH at lower [Na+]

Acid error Exhibited at pH < 0.5 pH readings are higher (curves A and B) Saturation effect with respect to H+

Selectivity Coefficient No electrode responds exclusively to one kind of ion. The glass pH electrode is among the most selective, but it also responds to high concentration of Na+. When an electrode used to measure ion A, also responds to ion X, the selectivity coefficient gives the relative response of the electrode to the two different species. The smaller the selectivity coefficient, the less interference by X.

Selectivity Coefficient Measure of the response of an ISE to other ions Eb = L’ + 0.0592 log (a1 + kHBb1) kHB = 0 means no interference kHB  1 means there is interference kHB < 1 means negligible interference

LIQUID MEMBRANE ELECTRODES

Liquid Membrane Electrodes Potential develops across the interface between the analyte solution and a liquid ion exchanger (that bonds with analyte) Similar to a pH electrode except that the membrane is an organic polymer saturated with a liquid ion exchanger Used for polyvalent ions as well as some anions Example: Calcium dialkyl phosphate insoluble in water, but binds Ca2+ strongly

                                                                            0.1 M CaCl2 Responsive to Ca2+

                                                                           

Characteristics of Ca+2 ISE Relatively high sensitivity Low LOD Working pH range: 5.5 – 11 Relevant in studying physiological processes

A K+-selective electrode Sensitive membrane consists of valinomycin, an antibiotic

CRYSTALLINE-MEMBRANE ELECTRODES

Crystalline-Membrane Electrodes Solid state electrodes Usually ionic compound Crushed powder, melted and formed Sometimes doped to increase conductivity Operation similar to glass membrane

Crystalline-Membrane Electrodes AgX membrane: Determination of X- Ag2S membrane: Determination of S-2 LaF3 membrane: Determination of F-

F- Selective Electrode A LaF3 is doped with EuF2. Eu2+ has less charge than the La3+, so an anion vacancy occurs for every Eu2+. A neighboring F- can jump into the vacancy, thereby moving the vacancy to another site. Repetition of this process moves F- through the lattice.

Fluoride Electrode

GAS SENSING PROBES

Gas Sensing Probes A galvanic cell whose potential is related to the concentration of a gas in solution Consist of RE, ISE and electrolyte solution A thin gas-permeable membrane (PTFE) serves as a barrier between internal and analyte solutions Allows small gas molecules to pass and dissolve into internal solution O2, NH3/NH4+, and CO2/HCO3-/CO32-

Gas Sensing Probe

DIRECT POTENTIOMETRY A rapid and convenient method of determining the activity of cations/anions

Potentiometric Measurement Ionic composition of standards must be the same as that of analyte to avoid discrepancies Swamp sample and standard with inert electrolyte to keep ionic strength constant TISAB (Total Ionic Strength Adjustment Buffer) = controls ionic strength and pH of samples and standards in ISE measurements

Potentiometric Measurement Calibration Method Standard Addition Method

Special Applications: Potentiometric pH Measurement using Glass electrode One drop of solution Tooth cavity Sweat on skin pH inside a living cell Flowing liquid stream Acidity of stomach

Potentiometric Titration Involves measurement of the potential of a suitable indicator electrode as a function of titrant volume Provides MORE RELIABLE data than the usual titration method Useful with colored/turbid solutions May be automated More time consuming

Potentiometric Titration Curves