Presentation is loading. Please wait.

Presentation is loading. Please wait.

P. Lecoq CERN February 2011 1 SiPM Workshop, CERN, 16-17 February, 2011 New Approaches in Scintillation Detectors in the Context of HEP Calorimetry and.

Similar presentations


Presentation on theme: "P. Lecoq CERN February 2011 1 SiPM Workshop, CERN, 16-17 February, 2011 New Approaches in Scintillation Detectors in the Context of HEP Calorimetry and."— Presentation transcript:

1 P. Lecoq CERN February 2011 1 SiPM Workshop, CERN, 16-17 February, 2011 New Approaches in Scintillation Detectors in the Context of HEP Calorimetry and Medical Imaging Paul Lecoq CERN, Geneva

2 P. Lecoq CERN February 2011 2 SiPM Workshop, CERN, 16-17 February, 2011 X-  detection Whatever the energy range (GeV for HEP, 100s KeV for medical imaging) X-  detectors record only a global information on the events: – time stamp (limited by the timing resolution of the whole acquisition chain) –Total energy deposit in a crystal (100cc for HEP, 0.05 to 0.1 cc in PET) –Crystal ID for localization (limited by crystal size) Some improvements with COG or DOI capability (several crystal layers or double readout) – 1.5mm depth resolution on 20mm long LYSO crystals for the ClearPEM with APD readout

3 P. Lecoq CERN February 2011 3 SiPM Workshop, CERN, 16-17 February, 2011 Present limitations As a result there is a trade-off between –Energy resolution, which requires the full containment of the conversion chain and therefore « large » detector blocks –Position resolution, for which the detection volume should be restricted to the first interaction point This leads to a trade-off between – Reconstructed spatial resolution –Sensitivity (less events in photopeak for smaller pixels)

4 P. Lecoq CERN February 2011 4 SiPM Workshop, CERN, 16-17 February, 2011  A zoom on the conversion process (HEP) The energy conversion from incoming X or  Rays is a complex process resulting from a cascade of events. Hadronic events are even more complex –Details of the full cascade for HEP with contributions from different conversion mechanisms: scintillation and Cerenkov, would lead to particle identification within the shower

5 P. Lecoq CERN February 2011 5 SiPM Workshop, CERN, 16-17 February, 2011 A zoom on the conversion process (low energy) 

6 P. Lecoq CERN February 2011 6 SiPM Workshop, CERN, 16-17 February, 2011  A zoom on the conversion process (low energy)

7 P. Lecoq CERN February 2011 7 SiPM Workshop, CERN, 16-17 February, 2011 Can we do better? The whole sequence carries a lot more information than what is usually recorded –Position of first impact –Nature of the energy deposition –Timing: the photon flux received from a « photoelectric peak » results from ≤ 50% true photoelectric and ≥ 50% Compton + photoelectric in the same pixel. The photons have different travel time. Difference can be as large as 150ps for the first detected photon. Can we build an imaging calorimeter ?

8 P. Lecoq CERN February 2011 8 SiPM Workshop, CERN, 16-17 February, 2011 Progress towards « intelligent » materials, metamaterials based on –crystalline fibers, –quantum dots, –photonic crystals Progress on photodetectors. Fully integrated and digital mutipixel imaging photodetector Towards a fully digital detector head with photon counting and high accuracy timing measurement capability Towards intellingent, multifunctional metamaterials

9 P. Lecoq CERN February 2011 9 SiPM Workshop, CERN, 16-17 February, 2011 Micro-pulling-down crystal fiber growth BGO YAG:C e LYSO:C e YAP:Ce  =400  m  =1mm  =2mm Courtesy Fibercryst, Lyon

10 P. Lecoq CERN February 2011 10 SiPM Workshop, CERN, 16-17 February, 2011 UV excitation Some crystal fibers YAG and LuAG up to 2m BGO (  : from 0.6 mm to 3 mm ; Length up to 30 cm) LYSO:Ce (  : from 0.6 mm to 3 mm ; Length up to 20 cm)

11 P. Lecoq CERN February 2011 11 SiPM Workshop, CERN, 16-17 February, 2011 ELT  =42m Diffractive Optics

12 P. Lecoq CERN February 2011 12 SiPM Workshop, CERN, 16-17 February, 2011 Diffractives fabrication: IC style optical lithography Diffractive Optics Technologies

13 P. Lecoq CERN February 2011 13 SiPM Workshop, CERN, 16-17 February, 2011 Strong EW symmetry breaking distinguish W and Z in their hadronic decays w/o kinematic constraints ALEPH like resolution ILC goal The calorimetry challenge Possible solution: Dual Readout scintillators Measure on an event to event basis The electromagnetic fraction in a hadronic shower By unfolding the Cerenkov contribution to the total light output

14 P. Lecoq CERN February 2011 14 SiPM Workshop, CERN, 16-17 February, 2011 SiPMTs Concept of a Meta-cable for HEP calorimetry MOEMS diffractive optics light concentrator MOEMS diffractive optics light concentrator SiPMTs Project funded by French ANR and CERN

15 P. Lecoq CERN February 2011 15 SiPM Workshop, CERN, 16-17 February, 2011 Development of new biomarkers First clinical targets: pancreatic/prostatic cancer Tool: dual modality PET-US endoscopic probe –Spatial resolution: 1mm –Timing resolution: 200ps –High sensitivity to detect 1mm tumor in a few mn –Energy resolution: discriminate Compton events EndoTOFPET-US FP7-project

16 P. Lecoq CERN February 2011 16 SiPM Workshop, CERN, 16-17 February, 2011 Coincidence SiPM-SiPM 3 February 2011

17 P. Lecoq CERN February 2011 17 SiPM Workshop, CERN, 16-17 February, 2011 Summary of results

18 P. Lecoq CERN February 2011 18 SiPM Workshop, CERN, 16-17 February, 2011 Internal probe EM tracking sensor PET head active volume 15x15x10mm 3 US Probe with biopsy needle Matrix of 18x18=324 fibers  ≤ 780  m L=10mm Section.cylindrical or square LYSO:Ce, LuAG:Pr,…

19 P. Lecoq CERN February 2011 19 SiPM Workshop, CERN, 16-17 February, 2011 1 chanel of the internal probe A cluster of 26x16 = 416 SPAD/fiber 10 TDC per cluster Diffractive light concentrator between fibers and SPADs Individual SPAD

20 P. Lecoq CERN February 2011 20 SiPM Workshop, CERN, 16-17 February, 2011 Mechanical integration (PET head)

21 P. Lecoq CERN February 2011 21 SiPM Workshop, CERN, 16-17 February, 2011 PET matrix - 1 to 9 coupling 5 mm 1 mm 8  m diameter sensitive spots 50  m apart Microlenses plate with high chromatic aberration 500  m diameter Quantum dot doped Heavy glass fibers = variable

22 P. Lecoq CERN February 2011 22 SiPM Workshop, CERN, 16-17 February, 2011Conclusion Metamaterials provide large design flexibility Combination of –scintillating/Cerenkov heavy crystal fibers –quantum dots in heavy host materials: fluorohafnate glasses –Fully digital SiPM with photon counting capability –Diffractive optics to improve optical coupling between scintillator and photodetector offer fascinating perspectives for the design of complex and highly performant metamaterials

23 P. Lecoq CERN February 2011 23 SiPM Workshop, CERN, 16-17 February, 2011 Meta-X-  Detector Block for Low Energy Applications Heavy scintillating fibers (LuAG, LSO …)  = 500  m Quantum dot loaded polymer with a gradient of QD radius along the fibre Photonic fibers  = 100  m


Download ppt "P. Lecoq CERN February 2011 1 SiPM Workshop, CERN, 16-17 February, 2011 New Approaches in Scintillation Detectors in the Context of HEP Calorimetry and."

Similar presentations


Ads by Google