Presentation is loading. Please wait.

Presentation is loading. Please wait.

Artifacts Ultrasound Physics George David, M.S.

Similar presentations


Presentation on theme: "Artifacts Ultrasound Physics George David, M.S."— Presentation transcript:

1 Artifacts Ultrasound Physics George David, M.S.
Associate Professor of Radiology

2 Nothing to do with Anything

3 Artifacts Assumptions can cause artifacts when assumed conditions are not true sound travels at 1540 m/s sound travels in a straight line All sound attenuation exactly 0.5 dB/cm/MHz

4 Distance from Transducer
Calculation of Distance scanner accurately measures time delay between sound generation & echo reception Distance = Assumed Speed X Measured Delay / 2 Actual Distance to interface 1380 m/s X 58usec / 2 = 4 cm 58 usec 4 cm Calculated Distance to interface V = 1380 m/s 1540 m/s X 58usec / 2 = 4.47 cm (Average speed of sound in soft tissue)

5 Distance from Transducer
Echo positioning on image distance from transducer calculated from assumed speed of sound can place reflector too close to or too far from transducer can alter size or shape of reflector V = 1380 m/s X Actual Object Position X Position of Object on Image V = 1540 m/s

6 Attenuation For all scanning your scanner assumes
soft tissue attenuation .5 dB/cm per MHz Your scanner’s action compensate for assumed attenuation allow operator fine tuning TGC

7 Shadowing Clinical Manifestation Cause Opposite of Enhancement
reduction in imaged reflector amplitude Cause object between this reflector & transducer attenuates ultrasound more than assumed assumed compensation not enough to provide proper signal amplitude intensity under-compensated Opposite of Enhancement Attenuates more than .5 dB/cm/MHz Shadowed Reflector

8 Shadowing Attenuates more than .5 dB/cm/MHz Shadowed Reflector

9 Attenuates less .5 dB/cm/MHz
Enhancement Clinical Manifestation increase in imaged reflector amplitude Cause object between reflector & transducer attenuates ultrasound less than assumed assumed compensation more than needed to provide proper signal amplitude intensity over-compensated Opposite of Shadowing Attenuates less .5 dB/cm/MHz Enhanced reflector

10 Attenuates less .5 dB/cm/MHz
Enhancement Attenuates less .5 dB/cm/MHz Enhanced reflector

11 Shadowing / Enhancing these artifacts not necessarily bad
can help in determining nature of masses “upstream of artifact which caused shadowing / enhancing

12 Scanner Assumptions Echo positioning on image
direction of all sound travel assumed to be direction that sound was transmitted Actual Object Position X Position of Object on Image X Refraction

13 Refraction Artifact refraction alters beam direction direction of sound travel assumed to be direction sound transmitted Actual Object Position X Position of Object on Image X Refraction

14 Refraction Artifact refraction alters beam direction
scanner places dot in wrong location along line of assumed beam direction can alter reflector shape

15 Lobe Artifacts Side Lobes Grating Lobes
beams propagating from a single element transducer in directions different from primary beam reflections from objects here will be placed on main sound transmission line Grating Lobes same as above except for transducer arrays X

16 Range Ambiguity Reflection from 1st pulse reaches transducer after 2nd pulse emitted scanner assumes this is reflection from 2nd pulse places echo too close & in wrong direction 1 2

17 Range Ambiguity To improve any 1 of 3, at least 1 of other 2 must be reduced. many scanners automatically reduce frame rate as depth increases Depth Range Ambiguity Trade-off Lines / Frame Frames / sec (dynamics)

18 Scanner Assumptions Multipath Artifact Actual Object Position X
Position of Object on Image X

19 Multiple Reflection Scenario
reflection from reflector “B” splits at “A” some intensity re-reflected toward “B” Result later false echoes heard scanner places dots behind reflector “B” 1 2 3 A B real 1 2 false 3

20 Artifacts Reverberation (multiple echo) artifact Mirror Image
“comet tail” effect is 1 example can have dozens of multiple reflections between transducer & reflector 2 reflectors Mirror Image common around diaphragm & pleura Real Mirror

21 Artifacts Caused by Shotgun Pellets

22 Multiple Reflection Scenario
Real Mirror

23 Resolution Artifacts Axial and Lateral Resolution Limitations
results in failure to resolve 2 adjacent structures as separate minimum image size equal to resolution in each direction

24 Section Thickness Artifact
anatomy may not be uniform over its thickness universal problem of imaging 3D anatomy in CT & MRI this is known as partial volume effect Thickness

25 Constructive Interference
2 echoes received at same time in phase Result higher intensity + =

26 Destructive Interference
2 echoes received at same time Exactly 180o out of phase Result flat (zero) wave - =

27 Acoustic Speckle texture seen on image may not correspond to tissue texture Results from interference effects between multiple reflectors received simultaneously which can add together constructive interference subtract from one another destructive interference

28 Mirror Image & Doppler Analogous to mirror image artifact discussed previously mirrored structures can include mirrored vessel duplicate image visible on opposite side of strong reflector example: bone Doppler data also duplicated flow & spectrum copied from original vessel

29 Spectral Duplication mirror image of Doppler spectrum appears on opposite side of baseline causes electronic duplication caused by receiver gain set too high overloads receiver True sensing caused by too large Doppler angle beam covers flow in both directions Blood flows toward transducer Blood flows away from transducer

30 Doppler Artifacts Doppler spectrum speckle Cause
same as acoustic speckle random constructive & destructive interference from sound scattered in blood

31 Aliasing Results in detection of improper flow direction
occurs because sampling rate too slow Similar to wagon wheels rotating backwards in movies

32 Insufficient Sampling
Aliasing Sufficient Sampling Insufficient Sampling

33 Aliasing Which way is this shape turning? OR #1 #2 #3

34 Did the shape turn 1/4 turn right or 3/4 turn left?
Aliasing Did the shape turn 1/4 turn right or 3/4 turn left? 1 1/4 turn right? #1 #2 #3

35 Does it help to sample more often?
Aliasing Does it help to sample more often? #2 #1 #1A #2A #3A #3

36 Aliasing Maximum detectable Doppler shift equals half the pulse repetition frequency Sampling rate Same as pulse repetition frequency Must be at least twice highest frequency to be sensed Aliasing occurs when Doppler shift exceeds 0.5 * PRF

37 Aliasing Maximum detectable Doppler Shift not limited for continuous wave Doppler Maximum detectable Doppler Shift is limited for pulsed instruments Maximum Detectable Doppler Shift = half pulse repetition frequency

38 Coping with Aliasing decrease transducer frequency
reduces Doppler shift shift proportional to operating frequency increase pulse repetition frequency decreases maximum imaging depth increases likelihood of range ambiguity for pulsed instruments 77 X fD (kHz) v (cm/s) = fo (MHz) X cosq

39 Coping with Aliasing increase Doppler angle
Reduces relative flow rate between blood & transducer Reduces Doppler shift sensed by scanner q 77 X fD (kHz) v (cm/s) = fo (MHz) X cosq

40 Coping with Aliasing: Baseline Shifting
operator instructs scanner to assume that aliasing is occurring scanner does calculations based on operator’s assumption scanner has no way of determining where in image aliasing occurs Yes No

41 Any Questions?


Download ppt "Artifacts Ultrasound Physics George David, M.S."

Similar presentations


Ads by Google