Presentation is loading. Please wait.

Presentation is loading. Please wait.

Medical Instrumentation

Similar presentations


Presentation on theme: "Medical Instrumentation"— Presentation transcript:

1 Medical Instrumentation
HW #3 Lim Myeong Jun

2 Contents Resistive sensor Capacitive sensor Inductive sensor LVDT
Problem solution Piezoelectric sensor Electromagnetic spectrum Radiation thermometer

3 Resistive sensor Strain gage : Changes in L & ρ
∵ ρ : Specific resistance Strain gage : Changes in L & ρ Force sensing resistor : Changes in A Thermister : Changes in temperature Potentiometer : Changes in Length

4 Force sensing resistor(FSR)

5 Thermister R Ceramic NTC thermister (Negative temperature coefficient)
PTC thermister (Positive temperature coefficient)

6 Thermister + Vout - Vo, Vs, F, T : Non linear A ) - + B )

7 Thermister C ) Bridge circuit Rs Rp R Linear range D ) Linearization

8 Capacitive & Inductive sensor
Capacitive sensor x A εR Most capacitive sensor is based on changes in x. Inductive sensor Self Inductance Mutual Inductance Difference Inductance

9 LVDT LVDT(Linear Variable Differential Transformer) a c x=D x= δ d f
Primary coil

10 LVDT Magnetic core + Vo - Mutual Inductance Sensor
Line is Vs and dot is Vo at magnetic core

11 LVDT When core is x=0

12 LVDT When core is x=+δ

13 LVDT When core is x=- δ

14 Problem #13 Vertical displacement
Given by Vs(t) = sin(2π*1000t) [V] (f = 1000Hz, T = 1ms) 180º Phase change

15 Problem #13 Method 1: Rectifier + LPF Full x=D wave rectifier x=0 x=-D
D=20mm, 2D=40mm V(t) t

16 Problem #13 Method 2: Phase-Sensitive demodulation Vo(t) x(t)·Vo²(t)
x(t) = sin(2π· 1000t) LPF(Low Pass Filter) V(t) t V(t)

17 Problem #14 In phase-sensitive demodulation Vo(t) Vm(t)
x(t) = sin(2π· 1000t) LPF(Low Pass Filter) V(t)

18 Problem #14 formula

19 Piezoelectric sensor x q : Charge , x : Deflection
Fig. piezoelectric sensor x q : Charge , x : Deflection That means piezoelectric sensor has no DC response.

20 Piezoelectric sensor 1) i(t) Vs(t) = R*i(t) i=0 + IA - Vo(t)

21 Piezoelectric sensor 2) i(t), Current flow - IA + Vo(t) 0V
Change Amplifier

22 Piezoelectric sensor |H| If you plan to use the piezoelectric sensor, for measuring pulse, the lowest frequency of pulse, signal is assumed to be 0.1Hz k/c fc f = 1/2πRC

23 Thanks for your attention!