X-Ray Microanalysis – Precision and Sensitivity Recall… wt.fraction I = I SiKα (unknown) / I SiKα (pure std.) K-ratio I = [I SiKα (unknown ) / I SiKα (std.)

Similar presentations


Presentation on theme: "X-Ray Microanalysis – Precision and Sensitivity Recall… wt.fraction I = I SiKα (unknown) / I SiKα (pure std.) K-ratio I = [I SiKα (unknown ) / I SiKα (std.)"— Presentation transcript:

1 X-Ray Microanalysis – Precision and Sensitivity Recall… wt.fraction I = I SiKα (unknown) / I SiKα (pure std.) K-ratio I = [I SiKα (unknown ) / I SiKα (std.) ] x CF CF relates concentration in std to pure element K x 100 = uncorrected wt.%, and … K (ZAF)(100) = corrected wt.%

2 Weight Percent? X-ray intensities are related to mass concentration, not atom concentration Incident electrons penetrate a constant mass of material which will differ as the composition differs Electrons interact with orbital electrons of target atoms lose kinetic energy number of electrons proportional to atomic mass Example: Elements A and B B is heavier than A Pure AMixture of A and B Excited volume

3 If atomic concentration of A = n A the mass concentration is: C A = n A A A / [n A A A +(1-n A )A B ] Where: A A = atomic weight of A A B = atomic weight of B # of excited atoms in pure A = Nm / A A Where:N is Avogadro’s number m is mass penetrated by incident electrons In the compound: # of A atoms excited is = n A Nm / [n A A A +(1-n A )A B ] The X-ray intensity ratio (proportional to the number of excited atoms) is then = {n A Nm / [n A A A +(1-n A )A B ]} / (Nm / A A ) Which is equal to the expression for C A, the mass concentration of A

4 Spatial Resolution D = 0.077 (E 0 1.5 – E C 1.5 ) / ρ ρ = density E 0 = accelerating potential E C = excitation potential Example: Si in fayalite at 15keV ρ = 4.39 E 0 = 15 keV E C = 1.840 keV for SiKα d = 0.98 μm 3σ = 2.9 μm diameter volume containing 99% of X-ray productions X-ray distribution from a point source… X

5 Precision, Accuracy and Sensitivity (detection limits) Precision:Reproducibility Analytical scatter due to nature of X-ray measurement process Accuracy:Is the result correct? Sensitivity:How low a concentration can you expect to see?

6 Accuracy and Precision Wt.% Fe 20253035 Correct value Low precision, but relatively accurate Wt.% Fe 20253035 Correct value High precision, but low accuracy Measured value Standard deviation Ave Std error Ave Std error

7 Accuracy and Precision Wt.% Fe 20253035 Correct value Low precision, but relatively accurate Wt.% Fe 20253035 Correct value High precision, but low accuracy Measured value Standard deviation Ave Std error Ave Std error Ave Std error Precise and accurate

8 Characterizing Error What are the basic components of error? 1)Short-term random error (data set) Counting statistics Instrument noise Surface imperfections Deviations from ideal homogeneity 2)Short-term systematic error (session to session) Background estimation Calibration Variation in coating 3) Long-term systematic error (overall systematic errors that a reproducible session-to-session) Standards Physical constants Matrix correction and Interference algorithms Dead time, current measurement, etc.

9 Frequency of X- ray counts Counts Short-Term Random Error - Basic assessment of counting statistics X-ray production is random in time, and results in a fixed mean value – follows Poisson statistics At high count rates, count distribution follows a normal (Gaussian) distribution

10 68.3% of area95.4% of area 99.7% of area 3σ 2σ 1σ 1σ 2σ 3σ The standard deviation is:

11 Variation in percentage of total counts = (σ C / N)100 So to obtain a result to 1% precision, Must collect at least 10,000 counts

12 Evaluation of count statistics for an analysis must take into account the variation in all acquired intensities Peak (sample and standard) Background (sample and standard) And errors propagated Relative std. deviation Addition and subtraction Multiplication and division

13 Positive and negative offsets for the background measurement, relative to the peak position r+ et r- Total number of measurements on the peak and on the background j pmax, j bmax index of measurements on the peak and on the background j p, j b Intensity (Peak-Bkgd in cps/nA) of the element in the sample e Element concentration in the sample CeCe Intensity (Peak-Bkgd in cps/nA) of the element in the standard s Element concentration in the standard CsCs Background counts B Peak counts P Total counting time tb tb Counting time on the peak tp tp Current from the Faraday cup i

14 For the calibration… And standard deviation…

15 The measured standard deviation can be compared to the theoretical standard deviation … Theo.Dev(%) = 100* S theo / s The larger of the two then represents the useful error on the standard calibration:  ² s = max ((S meas )², or (S theo )²)

16 For the sample, the variance for the intensity can be estimated as… where

17 The intensity on the sample is… Or, in the case of a single measurement… Pk – Bkg cps/nA

18 And the total count statistical error is then (3σ)…

19 An example Calibration X-RayTh Ma Pk-Bg Mean (cps/nA)155.2335 Std.Dev (%)0.372 Theo.Dev (%)0.136 3 Sigma (Wt%)0.563 Pk Mean (cps)3119.686 Bg Mean (cps)34.455 Raw cts Mean (cts)61657 Beam (nA)19.87 S meas0.57746862

20 Sample Th data Wt%currpk cpspk t(sec)bkg cpspk-bk 6.4992200.354098.57800285.08973813.483 This is a very precise number

21 Sensitivity and Detection Limits Ability to distinguish two concentrations that are nearly equal (C and C’) 95% confidence approximated by: N = average counts N B = average background counts n = number of analysis points Actual standard deviation ~ 2σC, so ΔC about 2X above equation If N >> N B, then

22 Sensitivity in % is then… To achieve 1% sensitivity Must accumulate at least 54,290 counts As concentration decreases, must increase count time to maintain precision

23 Example gradient: 0 distance (microns)25 Wt% Ni Take 1 micron steps:Gradient = 0.04 wt.% / step Sensitivity at 95% confidence must be ≤ 0.04 wt.% Must accumulate ≥ 85,000 counts / step If take 2.5 micron steps Gradient = 0.1 wt.% / step Need ≥ 13,600 counts / step So can cut count time by 6X


Download ppt "X-Ray Microanalysis – Precision and Sensitivity Recall… wt.fraction I = I SiKα (unknown) / I SiKα (pure std.) K-ratio I = [I SiKα (unknown ) / I SiKα (std.)"