AC and DC generators Global Technologies Company June 2002 A Conventional generators (AC) The mains alternating current (50 Hz) is applied to a circuit.

Similar presentations


Presentation on theme: "AC and DC generators Global Technologies Company June 2002 A Conventional generators (AC) The mains alternating current (50 Hz) is applied to a circuit."— Presentation transcript:

1 AC and DC generators Global Technologies Company June 2002 A Conventional generators (AC) The mains alternating current (50 Hz) is applied to a circuit (basically a step-up transformer) which elevates it from 230V to 70.000V (70kV) the high voltage (still at 50Hz) is applied to a rectifier circuit which: –eliminates the negative portion of the waves (half wave rectifier), or –converts the negative half-waves into positive (full wave rectifier)

2 AC and DC generators Global Technologies Company June 2002 A Conventional generator (AC) Initial waveform Half wave rectifier Full wave rectifier t kV 70 0 t kV 70 0 t kV 70 0

3 AC and DC generators Global Technologies Company June 2002 A Conventional generator(AC) The portion of wave useful for image formation is limited to the “crest”. All the rest doesn’t produce film blackaning, but it is noxious for the patient. t kV 70 0 Useful X-ray portion

4 AC and DC generators Global Technologies Company June 2002 A HF and DC generator The HF generator employs a specific circuitry (INVERTER) which generates a waveform of much higher frequency than the mains. Typical operating frequency of an HF generator is in the order of several tens of kilohertz (kHz) A specific circuits “flattens” the crests of the waves to further increase the amount of useful radiation. Such almost flat waveform is called “constant potential”, or DC.

5 AC and DC generators Global Technologies Company June 2002 A Comparison of waveforms AC generator HF generator t kV t

6 AC and DC generators Global Technologies Company June 2002 A DC Generator Residual oscillation (RIPPLE) t If the ripple is smaller than 4%, the IEC standards allow to give the generator the “constant potential” or DC attribute.

7 AC and DC generators Global Technologies Company June 2002 A HF technology advantages Tubehead more compact and lighter, because the high voltage transformer is smaller Patient dose reduced by 25%, because the “soft” X-ray portion is reduced Complete line voltage fluctuation insensibility Exposure times shorter


Download ppt "AC and DC generators Global Technologies Company June 2002 A Conventional generators (AC) The mains alternating current (50 Hz) is applied to a circuit."