Scanner Validation via the SNM Clinical Trials Network Phantom Program Paul E. Christian Molecular Imaging Program Huntsman Cancer Institute University.

Similar presentations


Presentation on theme: "Scanner Validation via the SNM Clinical Trials Network Phantom Program Paul E. Christian Molecular Imaging Program Huntsman Cancer Institute University."— Presentation transcript:

1 Scanner Validation via the SNM Clinical Trials Network Phantom Program Paul E. Christian Molecular Imaging Program Huntsman Cancer Institute University of Utah

2 Scanner Validation Sub-Committee
Purpose: Ensure scanner meets certain performance criterion needed for multi-center clinical trials via a clinical simulation phantom exercise.

3 Validate scanner performance
PET lesion detectability not consistent across vendors. Imaging/processing protocols not consistent between sites. SUV measurements not consistent across vendors nor sites. PET data submitted to FDA for multi-center drug studies not deemed reliable. Goal: Generate measurement tool to validate reliable multi-center PET data

4 Quality Control Dose Calibrator CT Scanner PET Scanner CT checks
Calibration Inspection PET Scanner Blank scan Constancy PET and CT alignment Record of PMs and calibrations

5 Courtesy: Paul E. Christian
Scanner Constancy Courtesy: Paul E. Christian

6 Commonly Used PET Phantoms
Commercial phantoms IEC NEMA NU-2 ACRIN ACR AAPM QIBA CROs

7 NEMA-IEC NU-2 Phantom 68Ge resin filled phantom 270.8 day T1/2
4.6 mCi Bkgd 0.44 mCi/ml Spheres 1.75 mCi/ml 4:1 sphere/background ratio Atten Coeff mm2/g at 511keV 59+-7 HU

8 Ten Sites, Three vendors (GE 4-PET/CT, 2-PET; Philips 1-PET/CT; Siemens 3-PET/CT)
Kinahan, Doot, U Wash.

9 CTN Oncology Chest Phantom
Clinical simulator to measure: Lesion detectability Lesion quantitation Image noise / texture

10 CTN Oncology Chest Phantom
Fill with F-18 FDG Precision Filling Technique Clinical Concentration Fixed Lesion/Background Ratio

11 Precision 18F Fillable Phantom
Repeat studies using a precision filling technique with F-18 Paul E. Christian, Keith Bigham, unpublished data 2008

12 Imaging Process Fill phantom (precise technique)
Performing imaging according to prescribed protocol Image after specific time delay Reconstruction parameters MD to identify lesions Submit images for review Quantitative measurements (SUVmax and SUVave) At site At core lab

13 Evaluation Criterion:
Clinical Simulation Phantoms Evaluate: Image performance -Accuracy of dose -Acquisition -Reconstruction/filter -Interpretation # Lesions detected Location of detected lesions SUV measurements (site) -Verify SUVs at core lab -Evaluate DICOM images (core lab) Image quality of PET/CT scanner -Uniformity -Resolution -Contrast -Noise -PET/CT alignment -Lesion detectability -Attenuation correction -Quantitative accuracy

14 Scanner Quantitative Validation
Assessment: SUV background accuracy Lesion SUV accuracy DICOM transfer valid Validation of SUV DICOM

15 Scanner Validation Progress
17 scanners validated to date 13 sites 11 more sites have phantoms 20+ additional sites scheduled (worldwide expansion)

16 Scanner Validation Site Problems
DICOM image compatibility Ability to provide quantitative images Ability to provide accurate SUVs Scanner operation/calibration Dose calibrator

17 Scanner Validation Sub-Committee
2010 Goals: Testing brain imaging phantom

18 Scanner Validation Sub-Committee
2010 Goals: Testing myocardial perfusion imaging phantom (3 clinical scenarios)

19 Scanner Validation Sub-Committee
Accomplishments: Development of scanner validation process Implement review and quality assurance testing Provide on-line phantom instructional video Administration of oncology demonstration project Oncology phantom imaging at 40+ sites Development of myocardial perfusion phantom Development of brain phantom

20 Scanner Validation Sub-Committee
2010 Goals: Streamline phantom qualification program Develop central archive for images Develop process for long term maintenance and administration of phantom site qualification Implement multi-site testing of brain and cardiac phantoms


Download ppt "Scanner Validation via the SNM Clinical Trials Network Phantom Program Paul E. Christian Molecular Imaging Program Huntsman Cancer Institute University."