1 Reinhard Manns / PL20 1 JUMO CTI-500 PL20 Analytical measurment systems

2 Reinhard Manns / PL20 2 1 Target groups / Application General water technology Monitoring of coolingwater Cleaning baths (e.g. galvanic) Intake to internal waste water plants Monitoring of drinking water Bottle cleaning plants CIP-cleaning (up to max. 100°C; Dairies)

3 Reinhard Manns / PL20 3 2. Concept JUMO CTI - 500 JUMO CTI - 500

4 Reinhard Manns / PL20 4 2. Concept JUMO CTI - 500 JUMO CTI - 500

5 Reinhard Manns / PL20 5 2.1 Gerätekonzept Blockschaltbild Block diagramJUMO CTI - 500 Signal preparation: Using temperature compensation at conductivity signal output conductivity:  temperature compensated conductivity signal  0(4)... 20mA  0(2)... 10V  20... 0(4)mA  10... 0(2)V SETUP-Connection: Configuration/Parametrierung Power supply: DC 24V (DC19..30V) Measuring cell:  inductive cell  Temperature sensor Binary inputs:  2 relay:  2 X Halbleiterschalter  AC/DC 15..50V; 200mA output temperature:  0(4)... 20mA  0(2)... 10V  20... 0(4)mA  10... 0(2)V Operation:  LC-Grafic display  4-Keys  2-LED´s

6 Reinhard Manns / PL20 6 2.2 Concept applications Features: cost effective devise compacte size Devices Head transmitter stand alone version

7 Reinhard Manns / PL20 7 2.2 Concept Features: Operating / Programming Keyboard and Display Setup-program comfortable programming System documentation Power supply DC 19…31V

8 Reinhard Manns / PL20 8 2.3 Measuring areas Features: Measuring area freely configurable Measuring areas 0 … 500µS/cm up to 0 … 2000mS/cm up to 4 areas activated

9 Reinhard Manns / PL20 9 2.4 Measuring units Features: Units µS/cm mS/cm % 5 signs freely editable

10 Reinhard Manns / PL20 10 2.4 Measuring units Features: Concentration measuring Soda lye NaOH Nitric acid HNO 3 1 X free definable graph (entering 20 pairs of variates with setup-programm)

11 Reinhard Manns / PL20 11 2.5 Temperature input Features: Units (Temperature) °C °F Temperature sensor fast response resp.. Protected Fitting inside the measuring cell

12 Reinhard Manns / PL20 12 2.6 Compensation Features: Form of temperature compensation linear natural water (EN 27888) own characteristic curve (Learnfunction) activate up to 4 values

13 Reinhard Manns / PL20 13 2.7 Outputs Features: 2 analog continuous outputs for conductivity / resitance and parallel for temperature (electrical isolated) 0/4..20mA 0..10V 20..4/0mA 10..0V freely programmable in measuring areas

14 Reinhard Manns / PL20 14 2.7 Outputs Features: 2 integrated potential free contacts (semi conductor switches) 0,2 A, AC 14…50V

15 Reinhard Manns / PL20 15 2.8 General concept Features: 2 binary inputs (the function is aktivated by a potential free contact) keyboard locking changing measuring areas changing temperature coefficent desalinisation function

16 Reinhard Manns / PL20 16 2.8 General concept Features: Connections PVC T-part DN32 PVC T-part DN40 Clamp 2 ½” screwed pipe fitting DN50, DIN 11 851(MK DN50, milk cone) SMS 2”

17 Reinhard Manns / PL20 17 2.8 General concept Features: Operating Keyboard Graphic display

18 Reinhard Manns / PL20 18 2.8 General concept Features: Display Output-Modus HAND (Hand-Mode) HOLD (Hold-Mode) K1 K2 E1 E2 ALARM HAND MESSUNG 10.24 18.5 °C mg/l contacts If visible contact is activ Binary inputs If visible input is activ Informations ALARM (e.g.Overrange) KALIB flashing (kalibration timer ended) KALIB (kalibration aktiv) Activity measurement (normal) Temperature and unit value and unit Informations

19 Reinhard Manns / PL20 19 2.8 General concept Features: languages german english french spanish polish

20 Reinhard Manns / PL20 20 3. Application Desalinisation in cooling towers During the process you have to consider the following points: water has to be refilled (by permanent evaporation water is getting lost) change water in sporadic intervals (Desalinisation: by permanent evaporation the liquid is going away, but the ingredients (salt) are increasing. The conductivity is getting higher; deposits in the tubes. You have to prevebt the growth of algae (Addition of biocide; in closed systems is a permanent growth of algae possible; that contaminated the cooling tower and could make problems in process e.g. the degree of efficiency is getting down. Therefor you have to add biocides in periodic lags. A secondary effect is the growing of the conductivity.)

21 Reinhard Manns / PL20 21 3. Application Conductivity measurement on cooling towers (max. conductivity app. 2000…3000µS/cm) JUMO CTI - 500

22 Reinhard Manns / PL20 22 3.1 Possible application General water technology Monitoring of cooling water Air conditions Cleaning baths (e.g. controlling of galvanic bath) betriebliche Endkontrollen

23 Reinhard Manns / PL20 23 3.1 Possible application Monitoring of boilers Drinking water control Reverse Osmosis (RO) Cooling of industrial heaters (e.g. casting houses)