Presentation is loading. Please wait.

Presentation is loading. Please wait.

Partial Parallel imaging (PPI) in MR for faster imaging IMA Compressed Sensing June, 2007 Acknowledgement: NIH Grants 5RO1CA092004 and 5P41RR008079, Pierre-Francois.

Similar presentations


Presentation on theme: "Partial Parallel imaging (PPI) in MR for faster imaging IMA Compressed Sensing June, 2007 Acknowledgement: NIH Grants 5RO1CA092004 and 5P41RR008079, Pierre-Francois."— Presentation transcript:

1 Partial Parallel imaging (PPI) in MR for faster imaging IMA Compressed Sensing June, 2007 Acknowledgement: NIH Grants 5RO1CA092004 and 5P41RR008079, Pierre-Francois Van de Moortele, Gregor Adriany, Kamil Ugurbil

2 Our coils  Open face coil  16 Channel “closed” coil

3 Intrinsically, surface coils offer a representation of signal as Acquired k-space, How to INTERPOLATE most stably to the Non-acquired data. we will see why it makes completely sense to think about interpolation The sensitivities are complex valued

4 Field of View Courtesy: Douglas Noll, University of Michigan

5 Undersampled images Undersampled individual images

6 The linear system The solution of the linear system gives rice to a spatially varying noise amplification. This is solely dependent on the sensitivities and is referred to as the geometry factor

7 The geometry factor α is the index of an aliased pixel, β n is the index of an unaliased pixel. The reduced FOV is the RSOS of all the channels with a reduced FOV, only for illu. Overall loss when using PPI is SNR red =SNR full /(g sqrt(R))

8 Back to the equation S indv. Channels. E encoding. p un-aliased image S indv. Channels. E encoding. p un-aliased image (all in k- space) K-space Image space Convolution operator

9 E is known, but we can make the formal separation of S, as follows: Two matrix equations, two unknowns E ”acq” includes all of k-space for the sensitivities

10  SENSE/SMASH formalism (get one image) GRAPPA idea, get multiple images. The interpolation is essentially similar to Kriging

11 Courtesy: Yeh, et al, MRM Volume 53, Issue 6, Pages 1383 - 1392

12 GRAPPA Reconstructing the data for EACH coil Courtesy: Griswold et al. MRM, 47(6):1202-1210 (2002)

13 Several reconstruction is found for EACH k-space point- due to the blocks. A weigthed average is used to compute just one

14  GRAPPA formula to reconstruct signal in one channel  ACS (Auto-Calibration Signal) lines (no x) where A represents the acceleration factor. N b is the number of blocks used in the reconstruction, where a block is defined as a single acquired line and A-1 missing lines. 4-8 blocks are needed,l,l

15 Temporal sampling PE Interleaved/segmented (2) Interleaved (2) Works well for imaging of static objects. For dynamic imaging, each image is not only undersampled, but also captures a different part of the “motion”/”change”. The acquisition is assumed faster than the motion time ½ k- space

16 PSF considerations (generally)  Let us start with imaging PE or t Standard PPI used to “unalias” the effects of the psf

17 UNFOLD (does not require multichannels)  Specifically, alternate the sampling by a factor 2, such that Remove aliasing by … ½ k- space Courtesy: Madore. MRM 48:493 (2002).

18 fMRI (UNFOLD) FIG. 16. Results obtained for a single-trial fMRI experiment (4 spiral interleaves, 16 kz phase-encode values, axial images, matrix size 128 3 128, TR 5 250 msec, TE 5 40 msec, 5 mm resolution along z, 24 cm FOV). Bilateral finger tapping was performed while a 2 sec audio cue was on, and then stopped for 12 sec. The acquisition time for a time frame (16 sec) is longer than a paradigm cycle (14 sec). UNFOLD is used to reduce the acquisition time by a factor 8, providing 7 frames per paradigm cycle. a: The acquired frames are corrupted by an 8-fold aliasing in the through slice direction. b: Temporal frequency spectrum for the highlighted image point in a. UNFOLD interleaves 8 spectra into the same temporal bandwidth. Marks are placed on the axis at the locations of the DC, fundamental and harmonic frequencies for the non-aliased material. Selecting only these frequencies, the aliasing seen in a is removed in c. Remove aliased frequency by selective filtering Courtesy: Madore. MRM 48:493 (2002).

19 Extend the concept of aliasing Line in image Unalias the support in x-f space, just like we unalias in x space with SENSE Tsao et al, MRM 50: 1031-1042 (2003)

20 DATA challenge 1. Where is the support in x-f space? Used to define support in x-f space “Equivalent” to a reference scan Interleaved training set Similar concepts hold for radial, where the center is the “prior”. This is used in speech imaging Tsao et al, MRM 50: 1031-1042 (2003)

21 How do the methods compare? k-t SENSE, vs. Sliding Window  Consensus (in cardiac imaging) of: Xu et al. MRM, 57:918- 930 (2007)

22 What does the artifact mean Xu et al. MRM, 57:918- 930 (2007)

23 Looking at the temporal variation (in speech [radial]) K-t SENSE Sliding window t y Comments/Conclusions Michael S. Hansen. Workshop on Non-catesian MRI. 2007

24

25 Formally With localized sensitivities (smooth in image space) The mising information can be determined from the acquired data, if the coeeficients a(i,j,k) are known

26 SMASH  Find weights n k (m) (x) [no x –readout dependence] such that we get a new synthetic sensitivity profile C m comp WE do parallel Imaging by finding ONE combined image (just like SENSE) m is selected depending on how FAR the data must be interpolated. Only one line is used to advance the data

27 Generalised SMASH  Find weights a k (m) (x) [with x – readout dependence] such that (x) Express  We use several phase-encoding lines to generate missing information. For each readout point a new set of weights are comp.

28 Two severe issues  The final image is that of a complex sum image of the individual images. Not optimal for SNR  Total cancellation can occur with such complex sums.  Coils where phase-aligned PRIOR to reconstruction

29 AUTO-SMASH  ACS (Auto-Calibration Signal) lines (no x), not fitting to a harmonic, but a “missing” PE-line


Download ppt "Partial Parallel imaging (PPI) in MR for faster imaging IMA Compressed Sensing June, 2007 Acknowledgement: NIH Grants 5RO1CA092004 and 5P41RR008079, Pierre-Francois."

Similar presentations


Ads by Google