1 생체계측 II Report # 송성진 Medical Instrumentation II

2 Chap 7. Blood pressure and sound
1. Blood pressure : Invasive blood pressure Non-invasive blood pressure 2. Intraocular pressure : Tonometry 3. Stethoscope Medical Instrumentation II

3 P P 1. Tonometry Force per unit Area F = Mg (We Known factor)
Contact Area = A M P P F Ball Ball Force per unit Area Medical Instrumentation II

4 • Measurement structure
LED Intraocular pressure Photo Sensor Ex) Photo diode & Photo Transistor Medical Instrumentation II

5 2. NIBP Cuff pressure Cuff Spuit Stethoscope
Medical Instrumentation II

6 • Heart structure Pulmonary artry Superior Pulmonary Vein Venacava
Coronary artery Coronary vein Aorta Inferior Systemic circulation Pulmonary circulation Coronary circulation Medical Instrumentation II

7 • Blood pressure graph Systolic pressure Diastolic pressure
120mmHg 90mmHg Diastolic pressure NIBP measure the value of systolic & diastolic pressure. Also It isn’t the print graph Medical Instrumentation II

8 • About PCB (Printed Circuit Board)
PCB is Classified MLB(Multi layer Board), Double substrate Spec Circuit design CAD : schematic (OrCAD) CAD : layout(OrCAD) PCB produce Components apply Test & programming Debugging Medical Instrumentation II

9 • Automatic Blood pressure graph & principle
Blue is pressurize Red is decompress P Systolic pressure 120mmHg 80mmHg Cuff pressure decrease -2 ~ -3 mmHg / sec Diastolic pressure Cuff > inner vessel Medical Instrumentation II

10 Osillometric Method • Blood pressure sound graph Noise
Heart sound graph Osillometric Method Medical Instrumentation II

11 3. Invasive BP measurement
• Catheter Diaphragm is flexible Dome Diaphragm Saline Strain gage Pi Po Cable Blood vessel Medical Instrumentation II

12 Catheter have several trouble!!!
Pi(t) Po(t) t t Catheter have several trouble!!! Length Diameter Material Air bubble in Catheter Medical Instrumentation II

13 Equivalent circuit model of catheter – sensor system
Voltage, v, [V]  Pressure, P, [pa] Current, i, [A=C/s]  Moving flow, f, [m³/s] Charge, q, [C]  Volume, v, [m³] EX) m f = mg (Center of Mass ) M Air potential Hight ↑ Pressure ↓ F = Mg A Medical Instrumentation II

14 • Apply measurement Blood flow is applied Current Dome Diaphragm
Saline Strain gage Pi Po Cable Blood vessel Pi & Po is applied voltages Length is applied resistor, Capacitor and Conductor Medical Instrumentation II

15 Pipe represent Resistor.
Resistance A L EX) P1 Pipe represent Resistor. As length & diameter. P2 Medical Instrumentation II

16 Pressure is increasing, Compliance is increasing
2. Capacitance of compliance Pressure is increasing, Compliance is increasing Young’s modulus Medical Instrumentation II

17 3. Inductance / Inertance
Mass is increasing, innersure is increasing, And inductance is increasing As a result, current flow (di/dt) is decreasing. Medical Instrumentation II