Presentation is loading. Please wait.

Presentation is loading. Please wait.

Lessons from Retrieval Analysis of Joint Replacements Tim Wright, PhD 3 rd ISOC Meeting - Bologna.

Similar presentations


Presentation on theme: "Lessons from Retrieval Analysis of Joint Replacements Tim Wright, PhD 3 rd ISOC Meeting - Bologna."— Presentation transcript:

1 Lessons from Retrieval Analysis of Joint Replacements Tim Wright, PhD 3 rd ISOC Meeting - Bologna

2 Failure Analysis Explain the unexpected

3 Retrieval Analysis as Failure Analysis

4 Strengths Direct assessment of in vivo “real life” performance Early warning of unanticipated problems Justification for rational design changes Limitations “Failed” devices Clinical data sometimes unknown Retrieval Analysis has Other Purposes

5 Implants: OR  Pathology  Biomechanics Cleaned and cataloged: 77120601 Boxed and labeled (without patient identifiers) Entered into password-protected searchable web- based database (IRB Approved/HIPAA Compliant) Connected to TJR registries (CERT and CORRe) Yr Mo Day # Retrieval Analysis at HSS

6 5 15 25 35 45 55 010 0 Wear Rate (mm 3 /mc) Radiation Dose (Mrad) 5 15..... Courtesy of Harry McKellop, PhD Retrieval Analysis as Postmarket Surveillance Crosslinking of Polyethylene

7 Fatigue Wear and Fracture Cyclic crack driving force Rate of crack growth Bradford et al, CORR 2004

8 Reliance on Preclinical Tests Joint simulators Standardized tests Clinical Relevance?

9 30% to 96% reduction at 2 – 5 yrs THA Clinical Results N = 37 in each group Patient-matched: age, wt, sex Dorr et al, JBJS 2005

10 Retrieval Analysis Examine cross-linked polyethylene acetabular liners for wear damage Compare wear damage to that in a conventional polyethylene liner with identical design and same manufacturer

11 Materials 79 conventional liners (Trilogy, Zimmer) 78 cross-linked liners (Longevity, Zimmer)

12 Materials 79 conventional liners (Trilogy, Zimmer) 78 cross-linked liners (Longevity, Zimmer) 15X

13 Fracture Schroder, et al., 2010 ORS Conventional Poly and Cross-linked Poly 1 Incipient Crack 5 Incipient Cracks, 1 Rim Fx

14 Reliance on Preclinical Tests Joint simulators Standardized tests Clinical Relevance?

15 Preclinical Testing Design Clinical Use Retrieval Analysis Retrieval Analysis as a Design Tool

16 Posterior Stabilized TKA Assure femoral posterior translation (in flexion) Prevent femoral anterior subluxation (at high flexion) Provide large ROM

17 Anterior Impingement Occurs at: –hyperextension –low flexion angles Prevents posterior femoral translation –increases stability –causes wear & deformation Unintended articulation –large contact stresses –low contact area

18 Retrieval analysis of a modern PS design 103 Retrieved Optetrak Tibial Components

19 Examine all inserts for evidence of damage to the tibial post on each face Methods 103 Retrieved Optetrak Tibial Components

20 Demographic Data Radiographic Analysis –Age –Weight –Height –BMI –Length of implantation –Reason for revision –Femoral and tibial varus-valgus angle –Femoral component flexion-extension angle –Tibial posterior slope Methods

21 Rational Design Change Determine stresses associated with box-post impingement in current design Examine design alternatives intended to reduce stresses

22 Model the Anterior Impingement

23 Apply a Load in Hyperextension 445 N

24 Finite Element Computer Model UHMWPE (elastic-plastic) Metal (rigid indenter)

25 Polyethylene Stresses

26 Polyethylene Deformation

27

28

29 Original versus New Design Design Maximum Stress (MPa) Original design 37 New design 24 35% 

30 Reduction to Practice Rounded box to minimize bone resection

31 Direct assessment of in vivo “real life” performance Early warning of unanticipated problems Justification for rational design changes The Importance of Retrieval Analysis

32 Thanks for your attention


Download ppt "Lessons from Retrieval Analysis of Joint Replacements Tim Wright, PhD 3 rd ISOC Meeting - Bologna."

Similar presentations


Ads by Google