Burle Leak Check 3/19/2012 Jeff Williams Jun Xie.

Similar presentations


Presentation on theme: "Burle Leak Check 3/19/2012 Jeff Williams Jun Xie."— Presentation transcript:

1 Burle Leak Check 3/19/2012 Jeff Williams Jun Xie

2 System Info Ion gauge not functional due to baratron issues Use RGA to view He background and total system pressure Initial test performed last week 3/15/2012 on bottom side components, oven was on precluding chalice test Chalice tested 3/19/2012, and a 2 nd test of other components also done at that time Pressure in the low 10 -7 Torr range on 3/19 after weekend pumping

3 Observations No change in helium background during testing in static condition There is an increase in helium and other gases (H2, N2) when hi vac valve is actuated (closed) Pressure increase from approx. 2E-7 to 2E-6 in 10 minutes with hi vac closed No discernable difference in level of helium increase whether helium just sprayed in the valve area or not prior to valve actuation

4 Next Step Do one final test with a helium leak detector Close hi vac valve and record pressure vs. time. If outgassing, the rate of rise should decay over time. If a leak is present the rate of rise should remain consistent.


Download ppt "Burle Leak Check 3/19/2012 Jeff Williams Jun Xie."