Presentation is loading. Please wait.

Presentation is loading. Please wait.

Medical Imaging Systems: MRI Image Formation

Similar presentations


Presentation on theme: "Medical Imaging Systems: MRI Image Formation"— Presentation transcript:

1 Medical Imaging Systems: MRI Image Formation
Instructor: Walter F. Block, PhD 1-3 Notes: Walter Block and Frank R Korosec, PhD 2-3 Tuesday 9 August 1994 at 17h36 S.F. Hilton and Towers, Continental question period Pulse Sequences for Fluid Shear Measurement using Fourier-encoded Velocity Imaging. Session: Flow Quantification Chairs: D. Saloner and L. Frank Departments of Biomedical Engineering 1, Radiology 2 and Medical Physics 3 University of Wisconsin - Madison

2 But so far RF coils only integrate signal
MRI Physics: So far... What we can do so far: 1) Excite spins using RF field at o 2) Record time signal (Known as FID) 3) Mxy decays, Mz grows 4) Repeat. But so far RF coils only integrate signal from entire body. We have no way of forming an image. That brings us to the last of the three magnetic fields in MRI. S?

3 Image Formation Overview
Gradient fundamentals Slice Selection Limit excitation to a slice or slab Can be in any orientation Gradient echo in-plane spatial encoding Radial imaging ( like CT) Frequency encoding Phase encoding

4 3nd Magnetic Field Static High Field Radiofrequency Field (RF)
Termed B0 Creates or polarizes signal 1000 Gauss to 100,000 Gauss Earth’s field is 0.5 G Radiofrequency Field (RF) Termed B1 Excites or perturbs signal into a measurable form O.1 G but in resonant with signal Gradient Fields 1 -4 G/cm Used to determine spatial position of signal MR signal not based directly on geometry

5 Gradient Coils Fig. Nishimura, MRI Principles

6 X Gradient Example: Gx Magnetic field all along z, but magnetic strength can varies spatially with x. Stronger at right, no change in middle, weaker at left.

7 Gradient Coil Fundamentals
Gradient strength directly proportional to current in coil On the order of 100 amps peak Performance Power needed proportional to radius5 Tight bore for patient Strength – G/cm or mT/m 4 G/cm is near peak now for clinical scanners Higher strength with localized gradients (research only) Slew rate Need high voltages to change current quickly T/m/s is high performance Rise to 1 G in .1 ms at 100 T/m/s Limited by peripheral nerve stimulation

8 Magnetic Field Gradient Timing Diagrams

9 B0 (t)(B0 + Gx(t)x) Before, only B0 Now with Gx
Larmor Equation B0 Before, only B0 Precessional Frequency Static Magnetic Field (t)(B0 + Gx(t)x) Now with Gx

10 Gx, Gy, Gz: One for each spatial dimension
Magnetic field all along z, but magnetic strength can vary spatially with x, y, and/or z.

11 Two Object Example of Spatial Encoding
sr(t) Receiver Signal: No gradient Gx On: Beat Frequency Demodulated Signal x m(x) Water

12 Gz Gradient Example L31, slides repeated here: first imaging method, basic procedure, projection reconstruction. The effects of the main magnetic field and the applied slice gradient. In this example, the local magnetic field changes in one-Gauss increments accompanied by a change in the precessional frequency from chin to the top of the head. Image, caption: copyright Proruk & Sawyer, GE Medical Systems Applications Guide, Fig. 11

13 Selective RF Excitation
Recall frequency of RF excitation has to be equal or in resonance with spins Build RF pulse from sum of narrow frequency range

14 Frequency profile of modulated RF pulse
Slice Selection - Consider a pulse B1(t) that is multiplied by cos(ot). This is called modulation . B1(t) is called the RF excitation. o is the carrier frequency = g B0. Mixer B1(t) cos(ot) A(t) cos(ot) Frequency profile of modulated RF pulse o = 2fo wo f

15

16 Frequency Encoding Spin Frequency (x)
Image each voxel along x as a piano key that has a different pitch. MR coil sums the “keys” like your ear. Fig from Illustrator, mag 100% then shrunk in powerpoint

17 Frequency Encoding

18 GRE Pulse Sequence Timing Diagram
rf Slice Select Freq. Encode Signal TE

19 Frequency Encoding & Data Sampling
Generated Signal DAQ Sampled Signal

20 In-plane Encoding MR signal in frequency encoding (x) is Fourier transform of projection of object Line integrals along y Encoding in other direction Vary angle of frequency encoding direction 1D FT along each angle and Reconstruct similar to CT Apply sinusoidal weightings along y direction Spin-warp imaging or phase-encoding By far the most popular

21 2D Projection Reconstruction MRI
ky kx Gx Gy DAQ Reconstruction: convolution back projection or filtered back projection

22 Central Section Theorem in MRI
Object y’ y In MR, echo gives a radial line in spatial frequency space (k-space). x’ ky θ x θ x’ kx F.T. CT Projection MR Signal (t) Interesting - Time signal gives spatial frequency information of m(x,y)

23 k-Space Acquisition (Radial Sampling)
Y readout X readout ky kx kx ky

24 In-plane Encoding MR signal in frequency encoding (x) is Fourier transform of projection of object Line integrals along y Encoding in other direction Vary angle of frequency encoding direction 1D FT along each angle and Reconstruct similar to CT Apply sinusoidal weightings along y direction Apply prior to frequency encoding Repeat several times with different sinusoidal weightings Spin-warp imaging or phase-encoding By far the most popular

25 Phase Encoding: Apply Gy before Freq. encoding
Fig from Illustrator, mag 130%

26 GRE Pulse Sequence Timing Diagram
rf Slice Select Phase Encode Freq. Encode Signal TE

27 k-Space Acquisition ky kx Phase Direction One line of k-space
Encode Sampled Signal DAQ kx ky Phase Direction One line of k-space acquired per TR Frequency Direction

28 k-Space Signal ky kx

29 512 x 512 8 x 8

30 512 x 512 16 x 16

31 512 x 512 32 x 32

32 512 x 512 64 x 64

33 512 x 512 128 x 128

34 512 x 512 256 x 256

35 512 x 512 512 x 32

36 Scan Duration Scan Time = TR  PE  NEX TR = Repetition Time PE = Number of phase encoding values NEX = Number of excitations (averages)

37 GRE Pulse Sequence Timing Diagram
rf Slice Select Phase Encode Freq. Encode Signal TE

38 Images of the Knee -weighted T2-weighted Needs longer TE

39 T2 & T2* Relaxation: Sources of Image Contrast
Mxy T2* Time T2* 1 = T2 + B0

40 Effects of Local Magnetic Inhomogeneity
6GRADECH.AVI The location or phase of spins throughout the x,y plane The first time point shows the spins immediately after the RF pulse. In the left plot, no gradient is played. In the right plot, the gradient waveform has equal positive and negative areas. The marker on top labeled “MAG1” or “MAG2” gives the magnitude of the signal in each case. In the left plot, a local magnetic field distortion is present that causes the vectors to become dephased naturally and the vector sum of the vectors to become smaller over time (T2* decay). This dephasing can be observed over a period of 16 frames in the vector plot on the left. The vector sum of the magnetization vectors during T2* decay is indicated by MAG1. On the right is shown the same distribution of magnetization vectors in the same local field distortion, but with the addition of an applied magnetic field gradient along the x-direction. The vector sum of the magnetization vectors with this additional dephasing is indicated by MAG2. The additional dephasing caused by this gradient can be seen by comparing the left and right vector-field plots, as well as by comparing MAG1 and MAG2. After 8 frames, the field gradient is reversed. This can be seen by observing the behavior of the vectors in the bottom- most row on the right. As well, the value of MAG2 begins to increase. The reversed gradient undoes the dephasing of the originally applied gradient so that, after 16 frames, the vector field plots and their sums are equal. As shown here, a gradient echo does not restore the signal decay caused by field inhomogeneities or other sources of dephasing. It merely corrects its own effects.In the left plot, spins dephase due to a magnetic field inhomogeneity. Gx(t)

41 Perils of Gradient Echo Imaging and T2*
TE = 8 ms TE = 24 ms 0.17 T GE Orthopedic Scanner

42 Image Formation Overview
Gradient fundamentals Slice Selection Limit excitation to a slice or slab Can be in any orientation Gradient echo in-plane spatial encoding Radial imaging Frequency encoding Phase encoding

43 T1-, T2-, and Density-Weighted Images
T1_brain_w_tumor.tif, t2_brain_w_tumor.tif, density_brain_w_tumor.tif From wesolak, janet, series 2, image 16, series 3, image 38, series 3, image 37 T1-weighted T2-weighted r-weighted

44 T2-Weighted Image of the Spine

45 Images of the Knee -weighted T2-weighted

46 T1-, T2-, and Density-Weighted Images
T1_brain_w_tumor.tif, t2_brain_w_tumor.tif, density_brain_w_tumor.tif From wesolak, janet, series 2, image 16, series 3, image 38, series 3, image 37 T1-weighted T2-weighted r-weighted

47 Spin Echo Parameters TR TE T1-weighting short (400 msec)
long (3000 msec) long (100 msec) -weighting

48 Signal vs Weighting T1-weighting long T1, small signal
T1-weighted T2-weighted r-weighted T1-weighting long T1, small signal short T1, large signal T2-weighting long T2, large signal short T2, small signal -weighting high , large signal low , small signal


Download ppt "Medical Imaging Systems: MRI Image Formation"

Similar presentations


Ads by Google