Presentation is loading. Please wait.

Presentation is loading. Please wait.

Measurement of the absolute efficiency,

Similar presentations


Presentation on theme: "Measurement of the absolute efficiency,"— Presentation transcript:

1 Measurement of the absolute efficiency,
with a precision better than 2%, of a PMT working in single photoelectron mode Philippe Gorodetzky APC lab, Paris, France VLVT09, Athens, October14, 2009

2 Comparison to a reference
How to calibrate ? Comparison to a reference Calibrated source Calibrated detector (photodiode)

3 1) Calibrated Source GOOD : Only 1 measurement
Time variations of no importance BAD : Control of the spatial variations of the source. ==> IMPOSSIBLE Angle & surface of emission (Liouville)

4 2) Calibrated Detector GOOD : Spatial variations of no importance if comparison made in same conditions BAD : Need 2 measurements => time variations important, but can be controlled through a third detector ==> POSSIBLE

5 Needs NIST photodiodes have a gain of about 0.5. So the light flux has to be reduced ~106 times Avoid geometry problems in illumination => exactly same geometry for PMT and calibrated detector

6 How (is the calibration done)
2 steps Mapping of the photocathodes RELATIVE Comparison to a NIST 1 position ABSOLUTE

7 Mapping of the photocathodes

8 SINGLE PHOTO-ELECTRON
Very few photons We illuminate with an LED of the good wavelength, and pulse at one kHz in order for the ADC to follow. To make a single photoelectron spectrum, while we acquire on the ADC, we lower the quantity of photons sent per pulse until the obtained peak has a stable position*. Then the number of events in that peak lowers while the number of events in the pedestal increases. When do we stop lowering the light? The spectrum is mainly a one pe (the bump), but there is still a « peak » at 2 pe, very weak, which will be troublesome when a discriminator is set between the pedestal and the 1 pe, and we count with a scaler: each 2 pe counts double, and the result will be wrong. scaler scaler One can also use an oscilloscope and watch when the base-line under the pulse begins to fill

9 SINGLE PHOTO-ELECTRON
We use Poisson: and P0, P1, P2 are the respective populations of the pedestal, of the 1 pe and of the 2 pe P0 = (m0 / 0!)e-m = e-m P1 = (m1 / 1!) e-m = m e-m = m P0 P2 = (m2 / 2!) e-m = (m2 / 2) e-m = (m / 2)P1 = (m2 / 2)P0 If one wants that P2 = 1% x P1, (m / 2)P1 = 0.01P1, then m / 2 = 0.01, and m = 0.02 Now, the ratio between P0 and P1: P0 / P1 = P0 / (mP0) = 1 / m will be 1 / 0.02 = 50 In our case, as soon as the pedestal is 50 times more important than the 1 pe, the 2 pe will be less than 1% of the 1 pe Usually, one takes: pedestal = 100 times 1 pe. Then we are sure not to pollute the measurement. Now we can set the discriminator threshold to be in the bottom between the pedestal and the 1 pe (at 0.25 of the 1 pe), and we just have to count in two scalers the pulses sent to the LED and the discriminator output. Exit the ADC: one can pulse until 100 kHz, which allows comfortable statistics in a few minutes. Also, the threshold being in a valley, the measurement will not be very sensitive to a small variation of the threshold, or of the gain due to HV small changes.

10 Mapping of the photocathodes
Reduce the light per pulse & adjust the gain Optical fiber One can also use an oscilloscope and watch when the base-line under the pulse begins to fill

11 Mapping of the photocathodes
In red: Coïncidences between generator & PMT discriminator

12 The photocathode is naked ( = 51 mm)
Mapping : « PMT-JY » The photocathode is naked ( = 51 mm)

13 Mapping of the photocathodes. Here, absolute
Better efficiency if we use only the central part => diaphragm of 20 mm Full pmt (40 mm diameter)

14 Absolute measurement PMT and 1 photodiode at the same time
BUT : very different gains : 1 vs 107 => how to divide a light flux by 107 ?

15 Absolute measurement Use of integrating spheres to reduce light
Measurement of the light flux reduction Measure the PMT efficiency

16 SINGLE PHOTO-ELECTRON

17 Calibration of the system
If one measures 1 nW in the second diode (noise = 1 pW) and mW in the first : Ratio =

18 Calibration of the PMT ANALYSIS 100 kHz: 14.425 nW in NIST
As the ratio = One sends on the PMT: nW / = nW Energy of a 378 nm: E = h = hc/ E = x / = J One knows that 1 J = photons So nW ==> x = photons / sec. In one measurement of 100 sec, we have sent on the PMT: ph We have measured pe ==> efficiency (discri) = / = 17.65% We have to add 8.8% (discri) so efficiency = 19.2 % at 377 nm (PMT center), and 15.8% for full pmt, instead of 22% given by Photonis Discri

19 Absolute measurement Uncertainties :
Flux reduction (ratio) : 3 % (2 NIST diodes) ΔR/R = (ΔI/I + Δα/α)udt + (ΔI/I + Δα/α)o1 Efficiency measurement : 1.7 % (1st NIST cancels out) Δε/ε = ΔR/R - (ΔI/I + Δα/α)udt

20 If one wants a more collimated photon beam
instead of Lambertian distribution 3 Leds NIST Photodiode trans-impedance amp. Integrating sphere   4 cm Amplification of TTL pulses in 40 V pulses with a risetime of 2 ns collimator Another way to look at the set-up: the first sphere is a "perfect" splitter (to the NIST and the first diaphragm) followed by a very stable light reducer.

21 One application: Antares
And why not NESTOR, or km3 ? They calibrate their system with atmospheric muons, but do not know very well (!!!) the efficiency in the back of the tube The light source: - integrating sphere - collimator X,Y,Z, ,  movement in a black box

22 Another application: JEM-EUSO 36 pixel Hamamatsu PMT
We illuminate the center of pixel 22 with a spot of 1 mm size. Assuming a collection efficiency of 70% (Hamamatsu), one gets a quantum efficiency of 40% Less than 1% of the counts are in coincidence in any combination of 2 pixels So, it is not a cross-talk, but a point spread function of 5.8 mm diameter, twice the diameter of the PSF of the lenses, that is 4 times its surface. Hence, Hamamatsu is designing a new PMT, with a better focus (a 64 pixels)


Download ppt "Measurement of the absolute efficiency,"

Similar presentations


Ads by Google