FDR Presentation Personal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00
2
Overview Sensors ECG Blood Pressure Temperature Signal Conditioning Filters, Amplifiers,Logic PIC Microcontroller program interface Palm pilot PC
3
Overall Block Diagram
4
Specification Power Supply: 9.0 VDC Battery Operating time 2.8 hours(from Standard 9.0 VDC Duracell battery) Inputs Body Temperature Sensor Detect Temperature in degree Celsius Range between 25 C – 50.5 C Blood Pressure Range between 0 – 5 Psi(259mmHg) Accuracy 0.125 @ 25 C Manual cuff inflation Maximum systolic pressure 250 mmHg ECG Bandwidth 0.3Hz – 150 Hz Detect Heart Rate in bpm One lead ECG
5
Specification Monitor Performance Specification Palm Pilot screen record and plays the patient’s data file on the screen PC screen displays the patient data file Recorded body temperature Systolic and diastolic blood pressure ECG waveform Heart rate Email this data file Interfaces Cables to interface the human body to the device Serial interface from the device to the Palm Serial interface from the palm to PC
6
Specification Indicators LED power status LED blood pressure visual detector Memory Storage 1 MB of Palm is allocated for health monitoring 55.6 minutes of display range
7
Block Diagram
8
Temperature signal conditioner simulation
9
Output voltage of the body Temperature sensor
10
Output voltage of the signal conditioner of the body temperature sensor
11
Blood pressure measurement Blood pressure: is defined as the intra-arterial pressure generated by the pulsatic flow of blood through an artery. The blood pressure is indicated from two numbers. Systolic: represents the pressure while the heart is beating. maximum value for blood pressure Diastolic: represents the pressure when the heart is resting between beats.
12
Flowchart for blood pressure measurement
13
Signal Conditioner of Capacitive Microphone
14
ECG Signal Conditioner
15
Simulation of the frequency domain of ECG
16
Microcontroller circuit
17
Software design block diagram
18
Actual Physical Layout
19
Physical layout of the program interface
20
Demonstration Measure current temperature and blood pressure and display it on the palm. Display temperature,blood pressure and ECG wave in Hyper terminal, then take the data and save it in a file. Finally display it on java applet.
21
Conclusion The Palm pilot is not only an organizer, but also a device to monitor data(Blood pressure, Temperature)
22
Acknowledgement Mr Yongning Prof Zhenyu Guo Mr Faisal Yasin Mr Jose Hernandez Mr Ken Drummond Mr John Petrella Mr Muhammad Afridi