1 QRS Detection Section 6.2 - 6.2.5 18.11.2004 Linda Henriksson BRU/LTL.

Similar presentations


Presentation on theme: "1 QRS Detection Section 6.2 - 6.2.5 18.11.2004 Linda Henriksson BRU/LTL."— Presentation transcript:

1 1 QRS Detection Section 6.2 - 6.2.5 18.11.2004 Linda Henriksson BRU/LTL

2 2 QRS Complex P wave: depolarization of right and left atrium QRS complex: right and left ventricular depolarization ST-T wave: ventricular repolarization

3 3 QRS Detection QRS detection is important in all kinds of ECG signal processing QRS detector must be able to detect a large number of different QRS morphologies QRS detector must not lock onto certain types of rhythms but treat next possible detection as if it could occur almost anywhere

4 4 QRS Detection Bandpass characteristics to preserve essential spectral content (e.g. enhance QRS, suppress P and T wave), typical center frequency 10 - 25 Hz and bandwidth 5 - 10 Hz Enhance QRS complex from background noise, transform each QRS complex into single positive peak Test whether a QRS complex is present or not (e.g. a simple amplitude threshold)

5 5 Signal and Noise Problems 1)Changes in QRS morphology i.of physiological origin ii.due to technical problems 2)Occurrence of noise with i.large P or T waves ii.myopotentials iii.transient artifacts (e.g. electrode problems)

6 6 Signal and Noise Problems http://medstat.med.utah.edu/kw/ecg/image_index/index.html

7 7 Estimation Problem Maximum likelihood (ML) estimation technique to derive detector structure Starting point: same signal model as for derivation of Woody method for alignment of evoked responses with varying latencies

8 8 QRS Detection Unknown time of occurrence 

9 9 QRS Detection

10 10 QRS Detection Unknown time of occurrence and amplitude a

11 11 QRS Detection Unknown time of occurrence, amplitude and width

12 12 QRS Detection

13 13 QRS Detection Peak-and-valley picking strategy Use of local extreme values as basis for QRS detection Base of several QRS detectors Distance between two extreme values must be within certain limits to qualify as a cardiac waveform Also used in data compression of ECG signals

14 14 Linear Filtering To enhance QRS from background noise Examples of linear, time-invariant filters for QRS detection: –Filter that emphasizes segments of signal containing rapid transients (i.e. QRS complexes) Only suitable for resting ECG and good SNR –Filter that emphasizes rapid transients + lowpass filter

15 15 Linear Filtering –Family of filters, which allow large variability in signal and noise properties Suitable for long-term ECG recordings (because no multipliers) Filter matched to a certain waveform not possible in practice  Optimize linear filter parameters (e.g. L 1 and L 2 ) –Filter with impulse response defined from detected QRS complexes

16 16 Nonlinear Transformations To produce a single, positive-valued peak for each QRS complex –Smoothed squarer Only large-amplitude events of sufficient duration (QRS complexes) are preserved in output signal z(n). –Envelope techniques –Several others

17 17 Decision Rule To determine whether or not a QRS complex has occurred Fixed threshold  Adaptive threshold –QRS amplitude and morphology may change drastically during a course of just a few seconds Here only amplitude-related decision rules Noise measurements

18 18 Decision Rule Interval-dependent QRS detection threshold –Threshold updated once for every new detection and is then held fixed during following interval until threshold is exceeded and a new detection is found Time-dependent QRS detection threshold  Improves rejection of large- amplitude T waves  Detects low-amplitude ectopic beats  Eye-closing period

19 19 Performance Evaluation Before a QRS detector can be implemented in a clinical setup –Determine suitable parameter values –Evaluate the performance for the set of chosen parameters Performance evaluation –Calculated theoretically or –Estimated from database of ECG recordings containing large variety of QRS morphologies and noise types

20 20 Performance Evaluation Estimate performance from ECG recordings database

21 21 Performance Evaluation

22 22 Performance Evaluation Receiver operating characteristics (ROC) –Study behaviour of detector for different parameter values –Choose parameter with acceptable trade-off between P D and P F

23 23 Summary QRS detection important in all kinds of ECG signal processing Typical structure of QRS detector algorithm: preprocessing (linear filter, nonlinear transformation) and decision rule For different purposes (e.g. stress testing or intensive care monitoring), different kinds of filtering, transformations and thresholding are needed Multi-lead QRS detectors