Lecture 6.1 Lecture 6.1 ADVANCED PLASMA DIAGNOSTICTECHNIQUES Fri 23 May 2008, 1 pm LT5 Presented by Dr Ian Falconer Room.

Similar presentations


Presentation on theme: "Lecture 6.1 Lecture 6.1 ADVANCED PLASMA DIAGNOSTICTECHNIQUES Fri 23 May 2008, 1 pm LT5 Presented by Dr Ian Falconer Room."— Presentation transcript:

1

2 Lecture 6.1 Lecture 6.1 ADVANCED PLASMA DIAGNOSTICTECHNIQUES Fri 23 May 2008, 1 pm LT5 Presented by Dr Ian Falconer i.falconer@physics.usyd.edu.au Room 101

3 ITER

4 Langmuir probes

5 Selected ITER diagnostics DiagnosticMeasures Magnetic diagnosticsPlasma current, position, shape, waves.. Spectroscopic & neutral Ion temperature, He & impurity particle analyser systems density,.......... Neutron diagnosticsFusion power, ion temperature profile, …. Microwave diagnosticsPlasma position, shape, electron density, profile, ….. Optical/IR(infra-red) systemsElectron density (Line-average & profile, electron temperature profile, …. Bolometric diagnosticsTotal radiated power, …. Plasma-facing components & Temperature of, and particle flux operational diagnostics to First Wall, ….. Neutral beam diagnosticsVarious parameters

6 Processing plasmas

7 Selected low temperature plasma diagnostics DiagnosticMeasures Langmuir probesPlasma potential, electron temperature & density Magnetic diagnosticsPlasma current, plasma waves, …. Spectroscopic Plasma composition, ion temperature & drift velocity, ……. Microwave diagnosticsPlasma electron density, density profile, …. Laser diagnostics Density etc.of various species in plasma

8 Electrostatic probes (Langmuir probes) Magnetic probes Microwave and optical interferometry Spectroscopic techniques Particle analysis Thomson scattering Nuclear radiation detection Laser diagnostics of processing plasmas PLASMA DIAGNOSTICS

9 General characteristics of a useful plasma diagnostic The diagnostic must not perturb the plasma – i.e. it must not change the conditions within the plasma Plasma diagnostics generally do not give the parameter(s) directly. An understanding of the physics of the processes involved in interpreting diagnostic results is essential

10 Electrostatic probes (Langmuir probes)

11 A short length of fine wire, inserted in a plasma can give valuable information about the plasma properties at a point in the plasma. A Langmuir probe consists of such a short, thin wire inserted into the plasma: the current to/from the probe is measured as its potential is changed.

12 A sheath forms around the probe of thickness ~ Debye length Current to sheath For a Maxwellian velocity distribution But this applies ONLY if the potential of the probe is the same as that of the plasma. How will the current to a Langmuir probe change if we use an external voltage source to change the probe’s potential?

13 A “typical” Langmuir probe characteristic

14 Typical probe characteristic: 1 A.V S is the space or plasma potential (the potential of the plasma in the absence of a probe). There is no E. The current is due mainly to the random motion of electrons (the random motion of the ions is much slower). B.If the probe is more positive than the plasma, electrons are attracted towards the probe and all the ions are repelled. An electron sheath is formed and saturation electron current is reached. X

15 Typical probe characteristic: 2 C. If the probe is more negative than the plasma, electrons are repelled (but the faster ones still reach the probe) and ions are attracted. The shape of this part of the curve depends on the electron velocity distribution. For a Maxwellian distribution with T e > T i, the slope of ln I p plotted against V s is D.The floating potential V f (an insulated electrode would assume this potential) The ion flux = the electron flux so I p = 0.

16 Typical probe characteristic: 3 E. All the electrons are repelled. An ion sheath is formed and saturation ion current is reached.

17 Probe surface

18 Magnetic probes

19 A voltage is induced by the changing magnetic field through this coil Integrating this voltage gives

20 Rogowski coil: measures plasma current Voltage induced in this toroidal coil by the magnetic field passing through area A Integrating

21 Voltage loop: typically used to give the voltage induced in the plasma by the Ohmic heating transformer A voltage is induced between the (open) ends of a (usually) single-turn loop adjacent to the plasma current. This voltage gives the voltage induced in the plasma by the transformer.

22 Measurement of induced voltage in plasma enable calculation of plasma conductivity – and hence temperature

23 Monitoring plasma position. Coils inside and outside the plasma, and voltage loops above and below the plasma, give the position of the plasma within the toroidal vacuum vessel. Signals from these sensors are used for feedback control of the plasma position. But only for toroidal plasmas with a circulating current – tokamaks.

24 Interferometry

25

26

27

28 Thomson scattering

29 Thomson scattering is scattering off free electrons in the plasma. The electrons are set oscillating by the incoming laser beam, and then radiate as dipole radiators. The intensity of the scattered radiation gives the electron density, the double-Doppler broadening of the scattered radiation gives the electron temperature.

30 Layout of a typical Thomson scattering experiment

31 The ITER LIDAR Thomson scattering system

32 An array of non-perturbing diagnostic techniques has been developed to probe both fusion and “processing” plasmas Selection of an appropriate diagnostic depends on the nature of the plasma – and the relative cost of the diagnostics available Effective use of a diagnostic technique depends on a thorough knowledge of the physics of both the plasma and the diagnostic technique adopted Conclusion