DIELECTRIC SYSTEMS Overview INTERNATIONAL SALES MEETING 2009

Similar presentations


Presentation on theme: "DIELECTRIC SYSTEMS Overview INTERNATIONAL SALES MEETING 2009"— Presentation transcript:

1 DIELECTRIC SYSTEMS Overview INTERNATIONAL SALES MEETING 2009
Stefan Zimmermann Intl. Sales Manager PhenixAsia, Taiwan

2 OVERVIEW Phenix Technologies manufactures a wide range of Dielectric Test Sets from smallest kVA ratings in portables to highest power ratings in the world. Dielectric Test Sets are the best sold units of Phenix. Dielectric testing is required, following applicable standards of international institutions, for most electro technical apparatuses and in our transmission/ distribution networks. The most typical applications are: Type testing Routine testing Acceptance testing Preventative maintenance

3 STANDARDS and rule of thumbs in the field
Each facility uses its own test procedures and rules, but all facilities are following certain standards Most commonly known applicable standards are IEC, IEEE, ASTM, AS, etc A standard is a recommendation between the standard committee and the end user Dielectric Testing following IEC 4 and IEEE and standards define that the nominal voltage x2 plus 1000V is to be applied for testing. Un x V

4 What is the definition of a Dielectric, a non-conducting substance or widely known as an ISOLATOR?
The maximum electric field that a dielectric can sustain The voltage which an insulating material can withstand before breakdown occurs, usually expressed as a voltage gradient. The electric voltage gradient at which an insulating material is broken down or "arced through" in volts per millimeter of thickness. The value of a material as an electrical insulator or the resistance to the flow of electric current. What is a Hipot test? We stress the insulation and measure the leakage current in Amperes

5 DIELECTRIC TESTING is required for following products:
Transformers Reactors Coils Cables Cable Joints and Splices Instrument Transformers Power Capacitors Motors Generators Bushings Insulators Switchgears ( Air, Gas, Oil insulated) Lightning Arrestors Circuit Breakers Insulation Materials ( Oil, Paper ) Protective Goods for Live Line Work (Gloves, Hoods, Hoses, Sticks, Blankets, Helmets, Boots, etc)

6 What Dielectric Test Sets does PHENIX offer?
AC (Alternating Current) and DC (Direct Current) or AC and DC Dielectric Test Systems Rule of thumb is: If application is AC, test with AC. If application is DC, test with DC.

7 TEST SYSTEMS Following categories apply to our product range:
Dielectric Test Systems (winding type transformers) Variable Inductance Resonant Test Systems Variable Frequency Resonant Test Systems Variable Frequency Transformer Test Systems (VTTS, VITTS) Variable Frequency Motor Test Systems (VMTS) Modular or Stackable Test Systems for AC and DC Material Test Systems per ASTM D149 and IEC Standard Standard Products (known as Portable Products for field testing) Bench Top Systems / Stand-Up Systems (transportable) AC/ DC Combination Test Systems Custom made AC or DC, AC/DC Test Systems High Performance Instrumentation (PD, Tan-D, RIV)

8 FAQ for STANDARD PRODUCTS
Are they Partial Discharge free? >>Some are…some aren’t > consult with your RSM Can they be transported? >> YES Are longer cables available? Are additional options for controls available? Are additional fixed gap reactors to achieve higher power ratings available? Are custom specifications available? An application questionnaire is required to receive a quote!

9 How to build a custom made AC DIELECTRIC TEST SYSTEM?
Definitions of AC Dielectric Model numbers: Provide a completed questionnaire, ask the right questions i.e. application, standard to be followed, required PD level, location (indoor/outdoor, transportable), intended connection to test specimen, is data report generation required, etc. 6= AC 4= DC 46=AC & DC 8=RTS C= Cylinder T= Tank X= Tap or special feature (Burn,etc) A=Air Type Transformer SA=Standard Analog Controls SD= Standard Digital Controls CE= Panel View 300 OID CD= Panel View 500 OID For example: 6TSA150-20, 6CCE200-40, 6TSD , 4CCE , 46TSA75-150, RTS8TC etc. The model no always reflects a 1hrON/1hrOFF duty!

10 CONTROLS FOR AC/ DC DIELECTRIC
Analog meters (series 600SA) which include: Triple range voltmeter with accuracy of ±2%FS. Triple range ammeter with accuracy of ±3%FS. Standard Digital (series 600SD) which include: Single range digital meter for voltage. 4½ digit with LED display and accuracy of ±(0.8% of reading + 0.2% of range +LSD least significant digit). Single range digital meter for current. 4½ digit with LED display and accuracy of ±(0.8% of reading + 0.2% of range +LSD).

11 CONTROLS FOR AC/ DC DIELECTRIC (continued)
Computerized Economical (series 600CE) which include: Auto-ranging display of voltage. Accuracy of ±(0.8% of reading + 0.2% of range +LSD). Auto-ranging display of current. Failure Memory Voltmeter Function. Auto-Voltage set-point. Stop Watch, Duration, Dwell Timer Functions. Over-Voltage set-point. Auto-Step, Auto-Cycle, and Auto-Sequence Functions. Over-Current set-point. Motorized Regulator. Variable ramp rate (desired rate of rise may be entered). RS-232 Port (Allen-Bradley DF1 Protocol). Graphic Display.

12 CONTROLS FOR AC/DC DIELECTRIC (continued)
Computerized Deluxe (series 600CD) which include: Auto-ranging display of voltage. Accuracy of ±(0.8% of reading + 0.2% of range +LSD). Auto-ranging display of current. Failure Memory Voltmeter Function. Auto-Voltage set-point. Stop Watch, Duration, Dwell Timer Functions. Over-Voltage set-point. Auto-Step, Auto-Cycle, and Auto-Sequence Functions. Over-Current set-point. Motorized Regulator. Variable ramp rate (desired rate of rise may be entered). RS-232 Port (Allen-Bradley DF1 Protocol). Graphic Display. Bar-Graph Display % of output Voltage and % of output Current Peak Memory Voltmeter Function. Peak Memory Current Function.

13 CONTROLS FOR AC/DC DIELECTRIC (continued)
To enbed any type controls following options are available: single, double and triple width pedestals Built into test system/ regulator cabinet Made to fit into existing cabinets Complete temperature controlled control rooms

14 REGULATORS FOR AC DIELECTRIC
Three electro/mechanical models available: TOVT (Toroidal Variable Transformers) used from 7.5kVA-250kVA CTVT (Column Type Variable Transformers used from 250kVA-2500kVA THVT (THOMA Variable Transformers) used from 500kVA-1.5MVA *separate technical memo’s available upon request.

15 FEATURES for COLUMN TYPE REGULATORS
Column type variable transformer (CTVT): The Phenix column type variable transformer, designed for the purpose of controlling output voltage offers following advantages: Linear design (the contacts only move in a straight line) Carbon roller contacts - Rolling friction is less than sliding friction; therefore contact movement does not contribute to heat build-up and does not accelerate wear. Rollers are spring-loaded to ensure positive contact and to provide flexibility for movement. Nickel-plated copper contact surface on the columns leads to less wear and a longer service life. The commutating windings of the CTVT are nickel plated in order to limit the circulating currents developed when the carbon roller rests in contact with more than one winding for a substantial amount of time. The high melting point of the nickel coupled with its ohmic resistance is the key to the heavy-duty operation of the Phenix CTVT. However, the nickel offers negligible voltage drop across the regulated voltage circuit between the secondary winding and the output roller. Therefore, the nickel plating is much better suited for this application than is silver plating.

16 FEATURES for COLUMN TYPE REGULATORS (continued)
Output voltage free is of distortion and harmonics. The low volts per turn is a design advantage of the CTVT. At 0.7 volts per turn (or less) the output resolution is considerably better than other regulation devices. A big advantage of the low volts per turn is that circulating currents are negligible. Choke coils are incorporated into the CTVT to help maintain phase balance. Unique plus-minus design features zero magnetic displacement which reduces overall mechanical stress Drive system - The CTVT uses a variable speed motor, screw and gear drive system to move the roller contacts along the column surface. This drive system insures that the contacts on each column are always in synchronization. If a column should fail you simply take that column out of service and can still operate at a reduced kVA until repairs are made. True sine wave output signal.

17 AC STANDARD PRODUCTS “portables” listed on your pricelist
Single and two piece AC Hipots: Model P is a 40kV, 10mA single piece AC Hipot Model 650-2P is a 50kV, 40mA two piece AC Hipot Model 650/10-2P is a 50kV, 40mA two piece AC Hipot with a tap 10kV, 20mA Model P is a 60kV, 10mA two piece AC Hipot (designed to perform a Vacuum Bottle Test) Refer to Brochures #60105

18 AC STANDARD PRODUCTS “portables” listed on your pricelist
Portable/ Transportable AC Dielectric Test Sets: Model 6CP15-3 is a 15kV, 200mA two piece AC Hipot (designed for Tan-D testing) Model 6CP30-2 is a 30kV, 67mA two piece AC Hipot Model 6CP is a 100kV, 75mA two piece AC Hipot with a tap 50kV, 150mA, mounted on two wheel cart (designed for switchgear testing) Model 6CP is a 100kV, 100mA two piece AC Hipot with a tap 50kV, 200mA mounted on two wheel cart Model 6CP is a 120kV, 83mA two piece AC Hipot with a tap 60kV, 166mA mounted on two wheel cart Model 6CP180-2 is a 180kV, 9mA two piece AC Hipot mounted on two wheel cart Model 6CP is a 200kV, 50mA four piece AC Hipot. Two 100kV,50mA cylinders mounted on 2-wheel carts with one control unit and a separate 200kV divider. Refer to Phenix spec sheets (available on request)

19 AC STANDARD PRODUCTS “portables” listed on your pricelist (continued)
Aerial Lift Test Sets: BK60 is a 60kV, 10mA two piece AC Hipot (designed to test booms of aerial lifts) BK130-1 is a 130kV, 11.7mA two piece AC Hipot (designed to test booms of aerial lifts) BK130 is a 130kV, 50mA two piece AC Hipot (designed to test boom and liner of aerial lifts) 35kV, 171mA Tap for Liner Testing All BK’s are low partial discharge units! Refer to Brochure #60206

20 AC BENCH TOP and STAND UP CABINET PRODUCTS
AC Bench Top and Stand-Up cabinet AC Dielectric Test Systems: Standard Models 3kV-15kV AC Hipot Special Designs up to 25kV 300VA -30KVA Bench Top or Stand-Up Cabinet Designs Air Insulated Transformers with shielded output cables Optional Digital Metering Latching ON/OFF push button or dwell timer Optional Casters, Safety Foot Switch and Flashing Lights available Dry Type Transformer Testing 5 Min ON / 15 Minute OFF Duty Cycle, longer Duty Cycles available Not suitable for Partial Discharge testing Typically used for motor or coil testing but suitable for testing switchgear, fuses, arrestors, bushings, capacitors, cables Optional Variable Reactor to extend testing capability to 15kV, 6A Refer to Brochure #60303

21 AC MODULAR STACKABLE TYPE SYSTEMS
Cylinder type stackable AC Dielectric Test Systems: Standard 100kV, 10.5kVA Cylinder Type Transformers, stackable up to 4 units in series up to 400kV Output Oil Insulated Design External Capacitive Divider Includes base with casters Partial Discharge Free Design Digital Metering Standard Can be expanded up to 400kV anytime Optional caster, Safety Foot Switch and Flashing Lights Typically used for Onsite Testing of Bucket Trucks, Switchgear Insulators or Current Transformers, Calibration of Dividers or Potential Transformers and general purpose High Voltage Lab Applications.

22 AC TANK TYPE TRANSFORMER BASED SYSTEMS
Dead Tank Designs for most applications: Designs available up to 750kV in single tanks. Optional multiple taps to extend testing range Uses bushing tap or internal resistive divider for voltmeter Four different Control Options (SA, SD, CE or CD) Various Duty Cycles (Most suited for continuous duty) Oil Insulated Design Output via Condenser Type Bushings Optional Partial Discharge Free Design High power designs Up to 2MVA Real Power >2MVA Output with compensation Compensation Options (Gapped Core and Primary / Secondary Compensation) Applications: Transformer Testing, Pollution Testing etc…. Refer to Brochure #60402

23 AC CYLINDER TYPE TRANSFORMER BASED SYSTEMS
Insulated Fiberglass Oil Filled Cylinder Designs: Designs available up to 2000kV; (cascade cylinders) Output via electrode on top of system Designs can be expanded (add a cylinder) Designs up to 500kV in single module Uses capacitive divider for voltmeter Various Duty Cycles (Most suited for shorter duty cycles) Optional Partial Discharge Free Designs Four Different Control Options (SA, SD, CE or CD) High power designs Up to 2MVA Real Power >2MVA Output with compensation Compensation Options (Gapped Core and Primary/ Secondary Compensation) Applications: Transformer Testing, Pollution Testing etc…. Refer to Brochure #60402

24 MATERIAL TEST SYSTEMS PER ASTM D149 STANDARD and IEC60243-1
Single Cabinet Complete Testing Solution: Standard Designs available up to 100kV AC Higher voltage designs include low voltage taps Includes Testing Chamber Optional Combination DC Supply or DC only units Optionally Partial Discharge Free Design with integrated PD Detector Test Fixtures per standard are available (to be specified) Hot and Cold Oil baths Only available with CE or CD control options Casters and Writing Shelf are included All units always have 500V/s rate of rise. V 50/60Hz inputs Refer to Brochure #60602

25 TANK TYPE VARIABLE INDUCTANCE RESONANT TEST SYSTEMS
Used for highly capacitive test objects: Dead Tank Designs: Multiple output taps available Uses Capacitive Bushings High Power Tanks available up to 15MVA testing power Most Suitable for continuous duty applications type test and PQ (Pre-Qualification Test) Parallel/ Series Resonance outdoor systems available up to 750kV Low Partial Discharge Options Best Suited for Shielded Rooms (Faraday Cage) Skid Mounted and Trailer Mounted Designs available for onsite testing Available with 800D or 800C control options Most Suited for Cable, Capacitor and Generator Testing Refer to Brochure #80101

26 CYLINDER TYPE VARIABLE INDUCTANCE RESONANT TEST SYSTEMS
Used for the highest voltage capacitive test objects: Up to 2500kVAC Insulating Cylinder Designs (Cascade) Single Cylinders up to 500kV 15MVA Per Cylinder Maximum Power Uses High Voltage Electrode as output Most Suitable for very high voltage applications Series Resonance Mode Only Low Partial Discharge Options Not directly suitable for outdoor use. Upgrades with additional modules possible. Trailer Mounted Designs Onsite testing Available with 800D or 800C control options Most Suited for Cable, Capacitor and Bushing Testing Refer to Brochure #80101

27 TANK TYPE VARIABLE FREQUENCY RESONANT TEST SYSTEMS
Used for onsite tests of cables, GIS or generators: Up to 1000kVAC Dead Tank Designs Multiple output taps available up to 500kV Uses Capacitive Bushings Removable Bushings available up to 265kV Most Suitable for continuous duty applications High Power Tanks available up to 22MVA testing power Parallel Resonance outdoor systems available up to 750kV Low Partial Discharge Options Best Suited for Onsite Testing (3 phase input) Skid Mounted and Trailer Mounted Designs Recommendation per IEC for Frequency is: f=20-300Hz Available with 800C control options Most Suited for Cable, Capacitor and Generator Testing

28 TANK TYPE VARIABLE FREQUENCY RESONANT TEST SYSTEMS (continued)
Used for testing of SF6 Gas Insulated Switchgear: Up to 1000kVAC Dead Tank Designs Uses Oil to SF6 Bushings with Buss flanges for direct connection Various duty cycles applications Low Partial Discharge Options Best Suited for factory testing in production Custom-made designs Available with 800C control options

29 SKID MOUNTED VARIABLE INDUCTANCE RESONANT TEST SYSTEMS
Used for onsite testing of Generators: Multiple output taps available Uses Capacitive Bushings Most Suitable for continuous duty applications Parallel Resonance outdoor systems available Low Partial Discharge Options Best Suited for Onsite Testing Skid Mounted, skid custom-made Available with 800C / D control options

30 CYLINDER TYPE VARIABLE FREQUENCY TEST SYSTEMS
Used for the highest voltage capacitive test objects (GIS): Up to 2400kVAC Insulating Cylinder Designs (Cascade) Single cylinders up to 500kV 20MVA per cylinder maximum power Uses High Voltage Electrode as output Most suitable for very High Voltage applications Series Resonance Mode Only Low Partial Discharge Options Suitable for outdoor/ fair weather testing. Lightweight and highly mobile designs Available with 800C controls 3 Phase input, best for onsite testing Most suited for short cable and GIS testing

31 VARIABLE OR FIXED FREQUENCY RESONANT TEST SYSTEMS
Used for induced testing, long term or research and development: Available in most voltage and power ratings Tank or Cylinder Type Designs Only available with 800C or 600C Control Options Straight Transformer or Resonant Designs Low Partial Discharge Options Suitable for outdoor testing. Standard Systems for accelerated ageing 500Hz or 1000Hz Hybrid systems for induced testing of potential transformers 50 or 60Hz outputs for calibration purposes

32 CUSTOM AC DIELECTRIC TEST SYSTEMS
Automated High Volume Testing Automated Switch Gear (Multiple Tests with one hook up) Units with integrated PD and Tangent Delta Testing Integrated Test Chambers Material Handling Test Fixture Design Custom Software Designs 3 Phase Outputs

33 DC MODULAR STACKABLE DIELECTRIC TEST SYSTEMS
Stackable Designs for low power high voltage testing: Standard 100kV or 60kV 5mA Designs Stackable up to 6 units in series up to 600kV Output Air Insulated Design Internal Resistive Divider Includes Base with Casters Digital Metering Standard Reversible Polarity done manually Can be expanded anytime Optional Safety Foot Switch and Flashing Lights Typically used for Onsite Testing of Bucket Trucks Switchgear Insulators General purpose high voltage lab applications.

34 CYLINDER TYPE DC HIGH VOLTAGE TEST SYSTEMS
Used for the highest voltage capacitive test objects (Cable): Up to 1500kVDC Insulating Cylinder Designs (Cascade) Single Cylinders up to 600kVDC 20mA Per Cylinder Maximum Power Uses High Voltage Electrode as output Most Suitable for very high voltage applications for example DC Transmission lines. Low Partial Discharge Options Suitable for indoor or outdoor testing. Lightweight Highly Mobile Designs Available with 400C controls Very High Voltage DC Testing Automatic reversible Polarity Discharge Resistors / Series Resistors

35 HIGH POWER TANK TYPE DC TEST SYSTEMS
Used for resistive high current applications: Up to 600kVDC Dead Tank Designs Low Ripple 3 Phase Designs Uses High Voltage Bushing as Output Most Suitable for continuous duty applications Suitable for indoor or outdoor testing 3 Phase Rectifier Designs (6 pulse, 12 pulse) Available with 400C controls Very High Voltage DC Testing Reversible Polarity Discharge Resistors / Series Resistors Pollution Testing

36 CUSTOM MADE DC TEST SYSTEMS
High Speed Automated Testing, Low Term Testing: Low Partial Discharge Replacement Windings for Impulse Generators Charging Supplies Capacitive Discharge Units (CDU) Air Reactor Testing High Voltage / High Current Supplies High Power Cable Fault Locators (Thumpers) Multi Conductor Cable Test Systems

37 ACCESSORIES FOR DIELECTRIC TEST SYSTEMS
As part of a Complete Package Phenix can offer the following accessories and options: Supervision of the onsite installation Onsite Training Maintenance Training Operator Training Service and Calibration Contracts Ground and Discharge Sticks (Earthing Rod) 100kV Standard Design Higher Voltages available upon request Sphere Gaps (All Voltages and Sizes) Custom Safety Features Flashing Lights Pressure Mats etc…

38 ACCESSORIES FOR DIELECTRIC TEST SYSTEMS (continued)
Partial Discharge Detectors: PD2U Partial Discharge Detector Optional RIV Measurement Upgrades up to 12 channels Optional Multi Channels for three phase transformer testing Refer to Brochure #90502 Tangent Delta Measuring Systems: Optional add-on to PD detector Extendable up to 500A Analog RIV Systems : SF6 Filled Standard Capacitors 15kV-800kV Outdoor Capacitors available upon request

39 ACCESSORIES FOR DIELECTRIC TEST SYSTEMS (continued)
Cable Terminations available in 3 types: Oil Terminations up to 100kV Open Top Water Terminations Designs up to 200kV Sealed Water Terminations Up to 800KV 60, 90, 120, 240kW Cooling Units PLC Controls standard on all units

40 ACCESSORIES FOR DIELECTRIC TEST SYSTEMS (continued)
Calibration Standards: KVM SERIES 100kV, 200kV, 300kV Higher Voltage Custom Designs Suitable as Working Standard Resistive Standards for DC Hipots Designs up 100kV High Voltage Injection Capacitors: Used for Injection of PD calibration Pulses while High Voltage is applied 100kV, 100pF Standard Design Available up to the highest voltages

41 ACCESSORIES FOR DIELECTRIC TEST SYSTEMS (continued)
Coupling Capacitors: Power Separation Filters In Stock Standard Design 25, 50,100 and 200kV, 1nF Capacitors Higher Voltage Custom Designs Can offered as a Divider High Voltage Filters: Disc or Coil Designs Used to Suppress System Noise Typical attenuation 20dB Made for AC or DC Application Very Low Self Capacitance Available in all current Ratings

42 ACCESSORIES FOR DIELECTRIC TEST SYSTEMS (continued)
Shielded Enclosures for PD testing: PHENIX provides top quality shielded enclosures as part of a package. Generally Phenix uses a European shielded room supplier that builds to our specifications. Installation and erection supervised by a Phenix Engineer Double Shielded Isolation Transformers: Single or 3 Phase Designs Used to suppress mains noise Available in all power ratings Provide input/output voltages and power ratings.

43 Please Contact Phenix with your DIELECTRIC TESTING APPLICATION!
Thank you for your attention!