Presentation is loading. Please wait.

Presentation is loading. Please wait.

OSU & NBS March Materials Meeting

Similar presentations


Presentation on theme: "OSU & NBS March Materials Meeting"— Presentation transcript:

1 OSU & NBS March Materials Meeting
Materials Working Group Michael Burns, FAA Tech Center March 4th & 5th, 2008

2 Background International research community conducts tests to determine heat release and smoke generating characteristics of airplane cabin interiors Compliance with FAR Ohio State University (OSU) Heat Release Rate and National Bureau of Standards (NBS) Smoke Chamber Testing Principal Investigator – Historically, The Boeing Company Apparatus (OSU & NBS) and method (along with round robin) are industry standard International partners serve as participating laboratories Materials, including finishes or decorative surfaces applied to materials, must comply with CFR including it’s appendix F. This FAR addresses the flammability requirements in compartments occupied by crew or passengers in commercial aircraft. In this presentation the test of interest is the OSU Heat Release Rate and the NBS Smoke Chamber testing For the past ten years or so the Boeing company has been the principal Investigator of the annual round robin tests that were conducted. Thee FAA has assumed this role for the 2006/2007 series of tests. The Round Robin is a tool we employ to identify laboratories that may not be operating their fire tests in compliance with applicable FAA requirements. The basic premise of how a “ROUND ROBIN” test method works is - the Principal Investigator mails out the same samples to each lab Participating labs conduct the smoke chamber test and then return the sample data for analysis. Theoretically, all samples should produce the same results. THE PROBLEM IS….

3 The Problem is… Test results from past years have raised concern
Unexpected/unexplained variability in results Validity and reliable under question Basically… The ambiguity in the data has put these tests on the hot seat

4 Agenda Discuss 2006 OSU / NBS Smoke Chamber Round Robin Tests
FAA examined results and procedures of most recent Int’l study (2006) Present initial findings of 2007 ‘Mini-Study’ FAA conducting independent follow-on work Request for all participating laboratories… Next steps Over the next few minutes, I would like to review the most recent international round robin As follow-on work to this round robin a mini study was organized to look a just a few random labs more closely to try to find any problems that may be present in the field to help explain these errors I will then present conclusions and recommendations as well as a forecast for future plans

5 Review of 2006 Round Robin 27 laboratories participated in the Round Robin Test from North America, Europe, and Asia Some Labs were not specific as to whether they followed the Handbook or the FAA Rule Each lab tested 9 samples (3 samples of 3 materials) Material labeled 5300, 5400, 5500 27 labs participated in all - North America: US and Canada - Europe: Germany, Italy, Spain, France, Austria, Switzerland, United Kingdom - Asia: China and Japan To be in compliance labs can conform to either the FAA Rule or the FAA Handbook however not a combination of the two. The FAA Handbook was written from the rule, however is considered more of an operators manual that is continually updated by way of supplements. There were a total of 243 samples involved made up of 3 samples of 3 different materials identified as sample 5300, 5400 & 5500.

6 Review of 2006 Round Robin Good news… Bad news…
All 27 labs completed the test and reported data Bad news… Data is inconsistent Suspect test equipment and/or procedures were compromised Fully detailed results can be found on the FAA web The Good news… All 27 labs completed the test and reported data The Bad news… Data is unfortunately inconsistent Suspect test equipment and/or procedures may have been compromised. The full report can be found on the FAA fire safety web site

7 2007 Independent Mini-Study
Follow-on ‘Mini-Study’ being conducted to investigate why results were so variable One international and 6 domestic laboratories participating Comprehensive site visits to identify process errors and apparatus deficiencies, review use of procedures and compliance (complete) Conduct sample testing (in progress) Goal is to improve and ensure validity and reliability of future results A Follow-on ‘Mini-Study’ is being conducted to help investigate why results were so variable This mini round robin recently exposed some problems that may be more wide spread than anticipated. This study included one international and 6 domestic laboratories One of the domestic labs are new and requested to be part of this series of tests A comprehensive site visits was completed identifying process errors and apparatus deficiencies. Once these discrepancies are addressed sample testing can be completed The intent of this review is to improve and ensure validity and reliability of future results

8 2007 Mini-Study Initial Findings
Site inspections uncovered several issues that may have affected the outcome of past tests OSU ‘Cold inspection’ issues OSU ‘Hot inspection’ issues NBS ‘Cold inspection’ issues NBS ‘Hot inspection’ issues Furnace defects Equipment altering Next few slides will show a consolidated list (random order) of discrepancies found from the labs that were visited Many problems were found during the visits to the labs in the mini-study A ‘Cold inspection’ of dimensions on the NBS Smoke chamber and OSU were conducted and resulted in numerous issues that I will highlight in the coming slides A ‘Hot inspection’ of operational procedures also revealed some problems And I will discuss two other issues such as furnace defects and equipment altering that may have also been contributing factors.

9 OSU Cold Inspection Issues
Very poor insulation surrounding the OSU as recommended by the FAA handbook. Upper Thermopiles shorted prior to bead (4 out of 5). Upper Thermopiles found out of position (5 out of 5). Not having ability to ensure proper location of upper thermopiles. Missing 1” x 3” baffle plate in chimney section. Found inner cone of chimney section distorted and having a large hole (approx. 2” in diameter). Found cooling manifold of chimney section corroded and having a large hole (approx. ½” wide x 2” in length). Found outer cone of chimney section to be the incorrect thickness (.049” instead of .031”). In the interest of time, I wont read through each of these cold inspection findings, but a quick glance will give you an idea of what we found

10 OSU Cold Inspection Issues…continued
Secondary plate was found covered with debris and having the incorrect diameter holes (120 #28 drill holes ” diam.). Flow straightening tube for orifice plate found installed in reverse position (28” before orifice plate and 11” after). No calibration sheet available for purity of methane Radiation doors found extremely warped, missing insulation and bolts that keep the doors together. Could not verify the proper value is used in the equation to calculate the calibration factor. The handbook states (5.6.6), however the correct value should be Missing radiant door shield washer on injection rod.

11 OSU Cold Inspection Issues…continued
Rear deflector plate (behind glow bars) found to be extremely warped and very close to touching the glow bars. Flow straightening tube for orifice plate found having a crack near the orifice connection. Flow straightening tube for the orifice downstream fitting found to be 1” rather than ¾” as described in the handbook. Difficulties in getting to the thermopile - cause for extremely dirty thermopiles as well as inner cone, chimney and baffle plate caked with soot build up (approaching 1”). [Advised making access hole on lower left side of OSU to facilitate maintenance / cleaning] [Advised fabricating a stool of some sort to assist in cleaning the thermopiles more frequently (after each set of 3 tests as a minimum)]

12 OSU Cold Inspection Issues…continued
Upper pilot found out of position (3/4” above and behind the sample face, holes facing glow bars). Lower pilot tube found incorrectly aligned to sample face (too high) Radiant door mechanism for opening and closing requires adjustment to keep doors closed tightly. Alignment tool for thermopiles does not fit in chimney correctly. [Suggested fabricating a new template which fits correctly and has points instead of flat areas to set thermocouple location] Found exhaust temperature probe (separate from the 5 thermopiles required) located adjacent to center probe. [Suggested relocating to a remote location so as not interfere with the center thermopile reading] Thermopile support bracket (on chimney) found warped and able to rock or move back and forth.

13 OSU Cold Inspection Issues…continued
Operator unaware of proper procedures for filling wet test meter with distilled water (should be filled to pointer tip with no flow etc.) No procedure in place or unable to set sample face the correct distance from the radiation doors once inserted (100mm) No procedure in place or unable to set heat flux gage the correct distance from the radiation doors once inserted (100mm) [Same location as sample face when inserted] Found thermopile alignment tool extremely eroded Excessive play in sample holder injection rod (bad bushings). For calibration, the software does not let equipment stabilize for 2 minutes then capture and average 10 seconds of data.

14 OSU Cold Inspection Issues…continued
Found secondary plate having incorrect number of holes as well as incorrect diameter of holes. There should be 120 #28 drill holes (.140 diam.) and there were 124 #27 drill holes. Found wet test meter filled with sewing machine oil and not distilled water as described in the handbook. Operator claims the manufacturer recommends this, however, documentation must support this. There must also be testing to verify that there is no difference in flow readings between the two fluids. Found mass flow controller used during calibration to achieve set points 1 Liter, 4 Liter, 6 Liter & 8 Liter. Only the 4 Liter set point was used as a calibration reference point and not each flow. Operator states that it is a linear calibration but documentation must prove this. Wet test meter must be installed in line for calibration and each flow verified prior to each calibration.

15 OSU Cold Inspection Issues…continued
Water temperature, barometric pressure and water vapor pressure are used in the calculation for the calibration factor, however, found software having no input for this information. A calibrated millivolt signal was sent to the Data Acquisition System (DAS). Four set points were observed as follows: Calibrated millivolt signal Software reading 28.0 mv mv 36.0 mv mv 42.0 mv mv 50.0 mv mv It is recommended to have the DAS calibrated to obtain correct millivolt readings

16 OSU Cold Inspection Issues…continued
A simulated calibration was conducted using a millivolt generator. The data should have yielded a calibration factor of .187 kW/m2/mv. The software calculated a .194 kW/m2/mv. Since the voltage signals received were lower than the calibration source the calibration factor should have been lower as well, however, it was higher. This leads to the possibility of having incorrect calculations within the software. Malfunctioning mass flow controller caused very large flow spikes and difficulties in trying to conduct a calibration Wet test meter was found to have improper backpressure gage (0”- 6”water) instead a psig gage was installed in its place. This must be approved by the manufacturer as an acceptable change to the equipment. Sewing machine oil was found contaminating the line going to the lower pilot from the wet test meter.

17 OSU Cold Inspection Issues…continued
On the flow straightening tube for the orifice plate a “T” fitting is installed at the orifice fitting upstream location. This “T” fitting provided air to the pilots. This must be relocated so as not to compromise the pressure differential reading across the orifice place.

18 OSU Hot Inspection Issues
Operator not familiar with proper calibration procedures. Plumbing for heat flux gage holder plate found to be incorrect. Center gage was set on an angle and not parallel to sample face. The corner gage was found recessed too much. No water drain or air/oil separator installed in air supply line. Operator unaware of requirements for the number of pilots allowed out at any given time. Center heat flux gage found to be damaged. Operator unaware of which side to burn on Schneller standard core panels (the side with writing should be exposed to the heat). Panels with adhesive film (white film on sample face) should be used only in the NBS chamber. This is only true for Schneller standard core panels. In the interest of time, I wont read through each of these Hot inspection findings, but a quick glance will give you an idea of what we found

19 OSU Hot Inspection Issues…continued
Flame length too long on lower pilot One of the upper pilots found clogged with soot. Problem with rheostat maintaining constant heat flux on glow bars Suggested replacing old rheostat. Found stop on heat flux calibration plate to be set to the incorrect position. This could result in a heat flux setting greater than 3.5 W/cm2. Due to equipment being in such poor condition, unable to complete the inspection or measure the following: Upper pilot location and flame direction Upper and lower pilot flame lengths Sample holders Measure sample position when injected into the OSU (100mm from inner wall)

20 OSU Hot Inspection Issues…continued
Measure heat flux gage position (100mm from inner wall) when setting heat flux Proper procedures for setting heat flux (center and four corners) Proper position of lower pilot Proper procedures for wrapping sample with aluminum foil and installing into sample holder Procedures for cleaning thermopile Procedures for conducting a test Measure diameter of calibration “T” bar holes and location Wet test meter water temperature Wet test meter barometric pressure Wet test meter water vapor pressure Unable to read calibrated millivolt signal on Data Acquisition System

21 OSU Hot Inspection Issues…continued
Items needing to be replaced or repaired Replace Thermopile (upper and lower) Replace outer cone with one of the correct thickness Replace inner cone Replace 1” x 3” baffle plate and wires connecting it to the chimney Replace outer insulation Replace gaskets between inner and outer cones and cooling manifold Replace cooling manifold Fabricate template (100 mm) to ensure sample holder is in correct position relative to radiation doors Replace broken sealing washer on sample injection rod

22 OSU Hot Inspection Issues…continued
Replace sample injection rod guides on holding chamber door Replace upper and lower radiant heat doors including insulation between skins of doors Replace gasket around door seal on holding chamber Replace secondary 18 gage stainless steel plate (120 holes .140 in diameter) Fabricate thermopile alignment tool to fit into chimney Reroute air line to OSU to accommodate an air/oil separator Reverse flow direction of cooling water to heat exchanger on air supply Verify computer program is reading millivolts accurately Replace inspection window on side of OSU

23 NBS Cold Inspection Issues
No means of measuring pressure within the chamber Furnace coil found in 3:00 position and not the 12:00 position as illustrated in the handbook No means of relighting the pilot flames should they go out during a test Pilot tubes found incorrectly aligned to sample face Two center pilots (45 degrees) on burner were found to be oblong and not the correct #54 drill size diameter Found sample holder .020 stainless steel wires made of 2 pieces and not the recommended 1 piece construction Found sample holders with no .020” stainless steel wires installed Upper guides on the sample holders (relative to each other) were found out of alignment Debris/Soot buildup found on inside face of sample holders In the interest of time, I wont read through each of these cold inspection findings, but a quick glance will give you an idea of what we found Pilot tubes found incorrectly aligned to sample face Upper guides on the sample holders (relative to each other) were found out of alignment Poor means of removing chamber contents (smoke) after testing that could be toxic to personnel

24 NBS Cold Inspection Issues…continued
Soot found on lower glass lens (internally) of photomultiplier tube assembly Poor means of removing chamber contents (smoke) after testing that could be toxic to personnel Water filled pressure regulator not vented to a suitable exhaust system Found chambers that failed the leakage rate check of 2” of water in 2 minutes Wall thermocouple found suspended in air and not actually mounted to wall surface Black “eye” observed in the center of the coil (entire coil not heating evenly) Flickering of electronics on control panel Furnace coil found to close to the sample face (internal dimension set to 1 ¼” and not the recommended 1 ½”) Found chambers that failed the leakage rate check of 2” of water in 2 minutes Furnace coil found to close to the sample face (internal dimension set to 1 ¼” and not the recommended 1 ½”) Found the distance between the sample holder and the furnace to be too close (less than 1 ½”) Found misaligned stops (for left/right adjustment) on upper guides used to center the sample in front of and parallel to the furnace

25 NBS Cold Inspection Issues…continued
Found insulation around furnace coil cracked and missing in places. Found the distance between the sample holder and the furnace to be too close (less than 1 ½”) Found misaligned stops (for left/right adjustment) on upper guides used to center the sample in front of and parallel to the furnace Furnace set too low relative to sample holder Debris found inside furnace near coil Found chambers that failed the leakage rate check of 2” of water in 2 minutes Furnace coil found to close to the sample face (internal dimension set to 1 ¼” and not the recommended 1 ½”) Found the distance between the sample holder and the furnace to be too close (less than 1 ½”) Found misaligned stops (for left/right adjustment) on upper guides used to center the sample in front of and parallel to the furnace

26 NBS Hot Inspection Issues
Operator unaware of proper procedures to check chamber for leaks or maximum leakage rate allowable (not more than 2” water in 2 minutes) Operator unaware of proper procedures to set pilot flame length Water level in pressure regulator set to values less than the 4” recommended in the handbook An increase in light signal was observed when the chamber door was closed for testing Operator unaware of requirements for the number of pilot flamelets allowed out at any given time and for how long Improper procedures for inserting sample into chamber (sealed chamber after sample begins to burn) Light source not functioning properly Operator removes the pilot burner prior to setting heat flux These are some of the findings of the Hot inspection Operator unaware of proper procedures to check chamber for leaks or what the maximum leakage rate allowable is Operator unaware of proper procedures to set pilot flame length Operator unaware of requirements for the number of pilot flamelets allowed out at any given time and for how long

27 NBS Hot Inspection Issues…continued
Problem with heat flux controller holding a stable heat flux (range from 4.1 millivolts to 5.0 millivolts). This fluctuation is out of specification for the required heat flux setting of 2.5 +/- .05 W/cm2 Improper aluminum foil wrapping techniques of samples prior to installing in the sample holder Operators not removing entire aluminum foil from sample face prior to testing Operator unaware of pilot flamelets going out during test Unable to close doors tightly during calibration of heat flux due to water lines/calorimeter signal wire passing through the door Poor handling / care of calorimeter Operators not removing entire aluminum foil from sample face prior to testing Operator unaware of pilot flamelets going out during test Poor handling or care of calorimeter – not really aware of how delicate a calorimeter really is and it’s importance in setting the proper heat flux.

28 Furnace Defect A problem with replacement furnaces of the NBS Smoke chamber has recently been observed A dark “eye” in the center of the furnace has been observed on some coils even after being calibrated to the 2.5 W/cm2 value Furnace Defect A problem was observed with some furnace coils having a dark spot in the center even after being set to the correct heat flux.

29 Furnace Defect Drawing 40A (NIST report NISTIR 4917, New Heater and Flux Guage for the NBS Smoke Box) Thought to be the most recent authoritative document, however, original designer did not specify tolerances Original designer/manufacturer of furnace no longer in business The current manufacturer is actively working with the FAA to resolve this issue Apparently the drawing found in report NISTIR 4917 is thought to be the most recent authoritative document, however, has no tolerances specified. This may be the reason for the variability in the manufacturing process. The manufacturer has been contacted and we are actively working together to resolve this issue The following photo’s illustrate the lack of heating in the center of the furnace coil

30 Furnace Defect

31 Furnace Defect

32 defective replacement
Furnace Defect properly functioning furnace defective replacement furnace Here on the left is a properly functioning furnace and on the right is the coil found with the dark eye in the center. The manufacturer states that only ½ of the center loop should not be glowing orange but as you can see in this photo that it is clearly dark more than one complete revolution

33 Alternate Equipment Use of alternate equipment by manufactures in some NBS Smoke Chambers may be reason for erroneous data found in Round Robins Some known alterations include: Changes to Furnace Use of various Heat Flux measuring devices IL 1700 Research Radiometer from International Light (USA) SED033 / WBS465 or SED038 / WBS465 system (Photo multiplier tube replacement) And finally, we suspect that unsanctioned equipment altering may be a part of the problems observed in the data The FAA will be looking into these changes to ensure the test is not being compromised in any way and possibly eliminating some variables.

34 FAA Request For All Participating Labs
Review mini-study findings Do you have similar problems (OSU & NBS)? Conduct thorough internal review of equipment and procedures Review must address the following issues: List discrepancies found during inspection Course of action Estimated completion date for all necessary repairs Notify FAA once review is complete and submit findings no later than 6/1/2008 Request all participating laboratories… Review mini-study findings (do you have similar problems?) Conduct thorough internal review of equipment and procedures before next Round Robin is scheduled Notify FAA once review is complete and submit findings no later than 6/1/2008

35 Conclusions and Recommendations
Mini-Study still in progress. Many equipment and process infractions discovered. Most can be easily resolved now! A check list of items can be provided upon request. The main purpose of this presentation is 1) to enlighten the community of problems we've discovered 2) and improve equipment and techniques so that our future efforts don’t suffer 3) Mike Burns can provide a minimum check list upon request.

36 Next Steps FAA will be the Principal Investigator for next International Round Robin (date on hold) FAA will begin to address maintenance practices / schedules on both the OSU and NBS chamber We are in the process of procuring materials for distribution for the next International Round Robin Study We will be looking into possibly putting together some sort of maintenance recommendations for the OSU and the NBS test apparatus that may be incorporated into the handbook

37 Next Steps FAA is in the process of updating Chapter 6 of the FAA Handbook (NBS) Main focus will address such issues as furnace specification, Heat flux gage/Radiometer use and photometric system Comments on Chapter 6 will be accepted through 7/15/2008 Chapter 5 (OSU) to follow FAA Contact Information: Michael +1 (609) The FAA will begin the process of updating Chapter 6 of the Handbook. Main focus will address furnace specs, heat flux measurement issues and the photometric system Comments will be accepted from now until shortly after the next materials meeting As a reminder please submit all findings of the internal review no later than 6/1/2008

38 QUESTION & ANSWER ANY QUESTIONS, COMMENTS OR SUGGESTIONS?


Download ppt "OSU & NBS March Materials Meeting"

Similar presentations


Ads by Google