Download presentation
1
Trigger/Limit/Cycle/Baseline
Mechanical Ventilation
2
Trigger All breaths on a ventilator need to be triggered
3
Trigger All breaths on a ventilator need to be triggered They can be
Time triggered
4
Trigger All breaths on a ventilator need to be triggered They can be
Time triggered Patient triggered
5
Trigger All breaths on a ventilator need to be triggered They can be
Time triggered Patient triggered Flow triggered
6
Trigger All breaths on a ventilator need to be triggered They can be
Time triggered Patient triggered Flow triggered Pressure triggered
7
Time triggering Time triggering means the ventilator starts the breath
8
Time triggering Time triggering means the ventilator starts the breath
This will happen when there is no patient effort
9
Time triggering Time triggering means the ventilator starts the breath
This will happen when there is no patient effort This is based on the set respiratory rate
10
Patient Triggering A breath can be triggered by the patient
11
Patient Triggering A breath can be triggered by the patient
Flow triggering is the most common
12
Patient Triggering A breath can be triggered by the patient
Flow triggering is the most common Pressure triggering is another option
13
Flow Triggering Flow triggering starts with a base flow
This flow is continuously flowing through the circuit
14
Flow Triggering Flow triggering starts with a base flow
This flow is continuously flowing through the circuit The ventilator knows how much flow is going through the circuit The liter flow leaving the ventilator should be the same as the liter flow coming back
15
Flow Triggering As the patient inhales some of the flow goes into the patient lungs This means less flow is returning to the ventilator The ventilator senses that as a patient effort and starts a breath
16
Flow Triggering The difference in the flow coming out of the ventilator and coming back to the ventilator is how sensitivity is determined The less flow needed to be deflected by the patient the more sensitive it is The more flow needed to be deflected by the patient the more difficult it is to trigger, i.e. less sensitive
17
Flow Triggering 10 lpm is leaving the ventilator
10 lpm is returning to the ventilator As a patient inhales and takes in 2 lpm 8 lpm is returning to the ventilator This reduction in returned flow tells the ventilator to start a breath
18
Flow Triggering If you increase the amount of flow needed to be inhaled by the patient the less sensitive the machine is 10 lpm is leaving the ventilator 10 lpm is returning to the ventilator As a patient inhales and takes in 4 lpm 6 lpm is returning to the ventilator
19
Pressure Triggering Negative pressure is generated when a patient inhales
20
Pressure Triggering Negative pressure is generated when a patient inhales This negative pressure is in both the patients lungs and the ventilator circuit
21
Pressure Triggering Negative pressure is generated when a patient inhales This negative pressure is in both the patients lungs and the ventilator circuit The ventilator senses the negative pressure as a patient effort and starts a breath
22
Pressure Triggering So the patient needs to generate -2cmH2O to start a breath If you increase the amount of negative pressure needed the more difficult it is for the patient to trigger, i.e. less sensitive
23
Limiting Limiting is what contols are in effect during the inspiratory phase
24
Limiting Limiting is what contols are in effect during the inspiratory phase After a breath is initiated certain limits will be in effect depending on which type of breath is being delivered
25
Limiting Limiting is what contols are in effect during the inspiratory phase After a breath is initiated certain limits will be in effect depending on which type of breath is being delivered Limiting factors are what define a type of breath
26
Limiting Limiting is what contols are in effect during the inspiratory phase After a breath is initiated certain limits will be in effect depending on which type of breath is being delivered Limiting factors are what define a type of breath The two most common limits are
27
Limiting Limiting is what contols are in effect during the inspiratory phase After a breath is initiated certain limits will be in effect depending on which type of breath is being delivered Limiting factors are what define a type of breath The two most common limits are Pressure
28
Limiting Limiting is what contols are in effect during the inspiratory phase After a breath is initiated certain limits will be in effect depending on which type of breath is being delivered Limiting factors are what define a type of breath The two most common limits are Pressure Flow
29
Pressure Limited Pressure limited breaths have a set peak airway pressure
30
Pressure Limited Pressure limited breaths have a set peak airway pressure The pressure in the patients lungs will not go over the set pressure
31
Pressure Limited Pressure limited breaths have a set peak airway pressure The pressure in the patients lungs will not go over the set pressure The following breaths are pressure limited
32
Pressure Limited Pressure limited breaths have a set peak airway pressure The pressure in the patients lungs will not go over the set pressure The following breaths are pressure limited PCV PSV PRVC
33
Pressure Limited Pressure Control Ventilation (PCV)
Pressure is set with the intention of reaching the peak pressure and holding the pressure for part of the inspiratory time
34
Pressure Limited Pressure Control Ventilation (PCV)
Pressure is set with the intention of reaching the peak pressure and holding the pressure for part of the inspiratory time Pressure Support Ventilation (PSV) Pressure is set to augment a patients own tidal volume
35
Pressure Limited Pressure Control Ventilation (PCV)
Pressure is set with the intention of reaching the peak pressure and holding the pressure for part of the inspiratory time Pressure Support Ventilation (PSV) Pressure is set to augment a patients own tidal volume Pressure Regulated Volume Control (PRVC) Pressure changes with each breath with the intention of reaching a specific volume
36
Flow Limited Flow limiting is used when a specific volume needs to be delivered
37
Flow Limited Flow limiting is used when a specific volume needs to be delivered Each breath has a rigid flow rate and pattern that will deliver the specific volume
38
Flow Limited Flow limiting is used when a specific volume needs to be delivered Each breath has a rigid flow rate and pattern that will deliver the specific volume Therefore the breath is limited to a predetermined flow Volume Control is a type of flow limited breath
39
Pressure vs. Flow Limiting
Pressure limited breaths have a variable flow which allows better gas distribution
40
Pressure vs. Flow Limiting
Pressure limited breaths have a variable flow which allows better gas distribution The pressure is limited therefore if lung conditions change the volume will change
41
Pressure vs. Flow Limiting
Pressure limited breaths have a variable flow which allows better gas distribution The pressure is limited therefore if lung conditions change the volume will change If lungs become stiffer the volume will go down as the pressure remains the same
42
Pressure vs. Flow Limiting
Flow limiting will deliver a volume no matter what the lung conditions
43
Pressure vs. Flow Limiting
Flow limiting will deliver a volume no matter what the lung conditions Volume will be consistent but pressures will change
44
Pressure vs. Flow Limiting
Flow limiting will deliver a volume no matter what the lung conditions Volume will be consistent but pressures will change If the lungs become stiffer the pressures will go up
45
Pressure vs. Flow Limiting
PRVC uses pressure limiting but will automatically adjust the pressure to deliver a volume
46
Pressure vs. Flow Limiting
PRVC uses pressure limiting but will automatically adjust the pressure to deliver a volume So the pressure is not set, the volume is set
47
Pressure vs. Flow Limiting
PRVC uses pressure limiting but will automatically adjust the pressure to deliver a volume So the pressure is not set, the volume is set The ventilator will adjust the pressure to achieve the volume
48
Pressure vs. Flow Limiting
PRVC uses pressure limiting but will automatically adjust the pressure to deliver a volume So the pressure is not set, the volume is set The ventilator will adjust the pressure to achieve the volume This allows for the variable flow which is better for gas distribution
49
Pressure vs. Flow Limiting
PRVC uses pressure limiting but will automatically adjust the pressure to deliver a volume So the pressure is not set, the volume is set The ventilator will adjust the pressure to achieve the volume This allows for the variable flow which is better for gas distribution If the lungs become stiffer the pressures will go up
50
Cycling Cycling is what determines how the breath will stop and allow the patient to exhale
51
Cycling Cycling is what determines how the breath will stop and allow the patient to exhale Most breaths are time cycled
52
Cycling Cycling is what determines how the breath will stop and allow the patient to exhale Most breaths are time cycled Any breath type that requires a set Ti is time cycled These include the following breath types
53
Cycling Cycling is what determines how the breath will stop and allow the patient to exhale Most breaths are time cycled Any breath type that requires a set Ti is time cycled These include the following breath types PRVC
54
Cycling Cycling is what determines how the breath will stop and allow the patient to exhale Most breaths are time cycled Any breath type that requires a set Ti is time cycled These include the following breath types PRVC PCV
55
Cycling Cycling is what determines how the breath will stop and allow the patient to exhale Most breaths are time cycled Any breath type that requires a set Ti is time cycled These include the following breath types PRVC PCV Volume Control
56
Cycling Flow is also used to cycle a breath off
57
Cycling Flow is also used to cycle a breath off
Pressure Support Ventilation (PSV) is flow cycled
58
Cycling Flow is also used to cycle a breath off
Pressure Support Ventilation (PSV) is flow cycled As the inspiratory flow starts to slow down the ventilator will sense that the lungs are getting full and cycle off allowing the patient to exhale
59
Cycling Flow is also used to cycle a breath off
Pressure Support Ventilation (PSV) is flow cycled As the inspiratory flow starts to slow down the ventilator will sense that the lungs are getting full and cycle off allowing the patient to exhale During that same breath it is being pressure limited
60
Cycling Flow is also used to cycle a breath off
Pressure Support Ventilation (PSV) is flow cycled As the inspiratory flow starts to slow down the ventilator will sense that the lungs are getting full and cycle off allowing the patient to exhale During that same breath it is being pressure limited So pressure is being held in the lungs until the flow slows down at which time the ventilator will release the pressure and the patient exhales
61
Baseline Baseline is what limits are in effect during exhalation
Exhalation is passive The only limit will be added pressure, i.e. PEEP The baseline should be assessed via total PEEP to accurately measure the baseline
62
Rise Time or Slope Rise time is part of the inspiration so could be considered a limiting factor
63
Rise Time or Slope Rise time is part of the inspiration so could be considered a limiting factor It is usually not included in the definition of a breath
64
Rise Time or Slope Rise time is part of the inspiration so could be considered a limiting factor It is usually not included in the definition of a breath Rise time adjusts the initial flow rate of a breath
65
Rise Time or Slope Rise time is part of the inspiration so could be considered a limiting factor It is usually not included in the definition of a breath Rise time adjusts the initial flow rate of a breath It determines how much time is allowed for the breath to reach the peak flow needed to deliver the breath in the allotted time (Ti)
66
Rise Time/Slope The less time the Rise time is the faster the breath ramps up to full speed
67
Rise Time/Slope The less time the Rise time is the faster the breath ramps up to full speed The longer the Rise Time is the slower the breath reaches peak flow
68
Rise Time/Slope So if you have an inspiratory time of 1 second and you set your Rise Time at 0.2 seconds the peak flow will be reached in 0.2 seconds creating a blast of air at the beginning of the breath
69
Rise Time/Slope So if you have an inspiratory time of 1 second and you set your Rise Time at 0.2 seconds the peak flow will be reached in 0.2 seconds creating a blast of air at the beginning of the breath If you increase the Rise Time to 0.5 seconds the breath will not reach peak flow until half way through the breath creating a slower overall flow
70
Rise Time/Slope A short Rise Time may be helpful in patient with high inspiratory demands, such as asthma
71
Rise Time/Slope A short Rise Time may be helpful in patient with high inspiratory demands, such as asthma A longer Rise Time is usually more comfortable because there is not the blast of air at the beginning
72
Rise Time/Slope A short Rise Time may be helpful in patient with high inspiratory demands, such as asthma A longer Rise Time is usually more comfortable because there is not the blast of air at the beginning Where it is set is very patient dependent
73
Inspiratory Cycle Off Inspiratory cycle off is part of the flow cycling breath
74
Inspiratory Cycle Off Inspiratory cycle off is part of the flow cycling breath Pressure Support Ventilation (PSV)
75
Inspiratory Cycle Off Inspiratory cycle off is part of the flow cycling breath Pressure Support Ventilation (PSV) Flow cycling is used to end a PSV breath
76
Inspiratory Cycle Off Inspiratory cycle off is part of the flow cycling breath Pressure Support Ventilation (PSV) Flow cycling is used to end a PSV breath Normal inspiration goes to zero flow and then reverses the flow to exhale
77
Inspiratory Cycle Off Inspiratory cycle off is part of the flow cycling breath Pressure Support Ventilation (PSV) Flow cycling is used to end a PSV breath Normal inspiration goes to zero flow and then reverses the flow to exhale Because it is difficult to achieve a tight seal on the circuit and patient, it is impractical to require the patient to get to zero flow before exhaling
78
Inspiratory Cycle Off Inspiratory cycle off is part of the flow cycling breath Pressure Support Ventilation (PSV) Flow cycling is used to end a PSV breath Normal inspiration goes to zero flow and then reverses the flow to exhale Because it is difficult to achieve a tight seal on the circuit and patient it is impractical to require the patient to get to zero flow before exhaling Flow cycling allow the breath to end prior to reaching zero flow
79
Inspiratory Cycle Off Some ventilators have a preset flow rate that will cycle the PSV breath into exhalation
80
Inspiratory Cycle Off Some ventilators have a preset flow rate that will cycle the PSV breath into exhalation Some ventilators allow the flow rate to be adjusted
81
Inspiratory Cycle Off Some ventilators have a preset flow rate that will cycle the PSV breath into exhalation Some ventilators allow the flow rate to be adjusted When setting Inspiratory Cycle Off it is a percentage
82
Inspiratory Cycle Off Some ventilators have a preset flow rate that will cycle the PSV breath into exhalation Some ventilators allow the flow rate to be adjusted When setting Inspiratory Cycle Off it is a percentage The percentage is based on the peak inspiratory flow achieved during the breath
83
Inspiratory Cycle Off Some ventilators have a preset flow rate that will cycle the PSV breath into exhalation Some ventilators allow the flow rate to be adjusted When setting Inspiratory Cycle Off it is a percentage The percentage is based on the peak inspiratory flow achieved during the breath The percent is what determines when the breath will release the pressure and allow exhalation
84
Inspiratory Cycle Off If you set the Inspiratory Cycle Off at 25% the inspiratory breath will end when the flow is 25% of the peak flow during that breath
85
Inspiratory Cycle Off If you set the Inspiratory Cycle Off at 25% the inspiratory breath will end when the flow is 25% of the peak flow during that breath If the peak flow was 60 lpm the breath will end at 25% of that or 15 lpm
86
Inspiratory Cycle Off If you set the Inspiratory Cycle Off at 25% the inspiratory breath will end when the flow is 25% of the peak flow during that breath If the peak flow was 60 lpm the breath will end at 25% of that or 15 lpm The higher the % set the sooner the breath will be cycled off
87
Inspiratory Cycle Off If you set the Inspiratory Cycle Off at 25% the inspiratory breath will end when the flow is 25% of the peak flow during that breath If the peak flow was 60 lpm the breath will end at 25% of that or 15 lpm The higher the % set the sooner the breath will be cycled off This may cut off a patient breath early
Similar presentations
© 2024 SlidePlayer.com Inc.
All rights reserved.