Potentiometric Methods A.) Introduction : 1.) Potentiometric Methods: based on measurements of the potential of electrochemical cells in the absence of.

Similar presentations


Presentation on theme: "Potentiometric Methods A.) Introduction : 1.) Potentiometric Methods: based on measurements of the potential of electrochemical cells in the absence of."— Presentation transcript:

1 Potentiometric Methods A.) Introduction : 1.) Potentiometric Methods: based on measurements of the potential of electrochemical cells in the absence of appreciable currents (I →0) 2.) Basic Components: a) reference electrode: gives reference for potential measurement b) indicator electrode: where species of interest is measured c) potential measuring device http:\\asadipour.kmu.ac.ir 24 slides

2 Electrodes and Potentiometry  Potential change only dependent on one ½ cell concentrations  Reference electrode is fixed or saturated  doesn’t change! Reference electrode, [Cl - ] is constant Potential of the cell only depends on [Fe 2+ ] & [Fe 3+ ] Pt wire is indicator electrode whose potential responds to [Fe 2+ ]/[Fe 3+ ] Unknown solution of [Fe 2+ ] & [Fe 3+ ] slides E cell =E cathod -E anod Anod is conventionally reference electode Fe 3+ +e -  Fe 2+ AgCl (s) + e - → Ag + Cl -

3 B.) Reference Electrodes : (Instead of SHE) Need one electrode of system to act as a reference against which potential measurements can be made  relative comparison.  Standard hydrogen electrodes are cumbersome - Requires H 2 gas and freshly prepared Pt surface Desired Characteristics: a) known or fixed potential b) constant response c) insensitive to composition of solution under study d) obeys Nernest Equation e) reversible http:\\asadipour.kmu.ac.ir 24 slides > > >

4 Electrodes and Potentiometry Reference Electrodes 1.)Silver-Silver Chloride Reference Electrode E o = V Activity of Cl - not 1  E (sat,KCl) = V http:\\asadipour.kmu.ac.ir 24 slides

5 Electrodes and Potentiometry Reference Electrodes 2.)Saturated Calomel Reference Electrode (S.C.E)  Saturated KCl maintains constant [Cl - ] even with some evaporation E o = V Activity of Cl - not 1  E (sat,KCl) = V http:\\asadipour.kmu.ac.ir 24 slides

6 Electrodes and Potentiometry Indicator Electrodes 1.)three Broad Classes of Indicator Electrodes  1) Metal Electrodes - Develop an electric potential in response to a redox reaction at the metal surface  2) Ion-selective (Membrane) Electrodes - Selectively bind one type of ion to a membrane to generate an electric potential 3) Molecular Selective Electrode Remember an electric potential is generated by a separation of charge http:\\asadipour.kmu.ac.ir 24 slides

7 1) Metallic Indicator Electrode (3 Main Types) a) Metallic Electrodes of the First Kind b) Metallic Electrodes of the Second Kind c) Metallic Redox Indicators a) Metallic Electrodes of the First Kind i. Involves single reaction ii. Detection of catione derived from the metal used in the electrode iii. Example: use of copper electrode to detect Cu 2+ in solution ½ reaction: Cu e -  Cu (s) E ind gives direct measure of Cu 2+ : E ind = E o Cu – (0.0592/2) log a Cu(s) /a Cu 2+ since a Cu(s) = 1: E ind = E o Cu – (0.0592/2) log 1/a Cu 2+ or using pCu = -log a Cu 2+ : E ind = E o Cu – (0.0592/2) pCu http:\\asadipour.kmu.ac.ir 24 slides

8 b) Metallic Electrodes of the Second Kind i. Detection of anion derived from the interaction with metal ion (M n+ ) from the electrode ii. Anion forms precipitate or stable complex with metal ion (M n+ ) iii. Example: Detection of Cl - with Ag electrode ½ reaction: AgCl(s) + e -  Ag(s) + Cl - E O = V E ind gives direct measure of Cl - : E ind = E o – (0.0592/1) log a Ag(s) a Cl - /a AgCl(s) since a Ag(s) and a AgCl(s) = 1 & E o = V: E ind = – (0.0592/1) log a Cl http:\\asadipour.kmu.ac.ir 24 slides

9 c) Metallic Redox Indicators i. Electrodes made from inert metals (Pt, Au, Pd) ii. Used to detect oxidation/reduction in solution iii. Electrode acts as e - source/sink iv. Example: Detection of Ce 3+ with Pt electrode ½ reaction: Ce 4+ + e -  Ce 3+ E ind responds to Ce 4+ : E ind = E o – (0.0592/1) log a Ce 3+ /a Ce http:\\asadipour.kmu.ac.ir 24 slides

10 2) Membrane Indicator Electrodes a) General i. electrodes based on determination of cations or anions by the selective adsorption of these ions to a membrane surface. ii. Often called Ion Selective Electrodes (ISE) or pIon Electrodes iii. Desired properties of ISE’s 1) minimal solubility – membrane will not dissolve in solution during measurement. – silica, polymers, low solubility inorganic compounds, (AgX) can be used 2) Need some electrical conductivity 3) Selectively binds ion of interest http:\\asadipour.kmu.ac.ir 24 slides

11 C + diffuses across the membrane due to concentration gradient resulting in charge difference across membrane Electrodes and Potentiometry Indicator Electrodes Ion-Selective Electrodes  Responds Selectively to one ion - Contains a thin membrane capable of only binding the desired ion  Does not involve a redox process Membrane contains a ligand (L) that specifically and tightly binds analyte of interest (C + ) A difference in the concentration of C + exists across the outer membrane. The counter-ions (R -,A - ) can’t cross the membrane and/or have low solubility in membrane or analyte solution Remember an electric potential is generated by a separation of charge Potential across outer membrane depends on [C+] in analyte solution http:\\asadipour.kmu.ac.ir 24 slides

12 Electrodes and Potentiometry Indicator Electrodes Ion-Selective Electrodes  Responds Selectively to one ion - Contains a thin membrane capable of only binding the desired ion  Does not involve a redox process Remember an electric potential is generated by a separation of charge Potential across inner membrane depends on [C+] in filling solution, which is a known constant C + diffuses across the membrane due to concentration gradient resulting in charge difference across membrane A difference in the concentration of C + exists across the inner membrane. Electrode potential is determined by the potential difference between the inner and outer membranes: where E inner is a constant and E outer depends on the concentration of C + in analyte solution http:\\asadipour.kmu.ac.ir 24 slides where [C + ] is actually the activity of the analyte and n is the charge of the analyte

13 pH Electrode i. most common example of an ISE based on use of glass membrane that preferentially binds H + ii. Typical pH electrode system is shown pH sensing element is glass tip of Ag/AgCl electrode Two reference electrodes here one SCE outside of membrane //// Combined electrod http:\\asadipour.kmu.ac.ir 24 slides one Ag/AgCl inside membrane

14 Electrodes and Potentiometry pH Electrodes 1.)pH Measurement with a Glass Electrode Ag(s)|AgCl(s)|Cl - (aq) || H + (aq,outside) H + (aq,inside), Cl - (aq)|AgCl(s)|Ag(s) Outer reference electrode [H + ] outside (analyte solution) [H + ] inside Inner reference electrode Glass membrane Selectively binds H + Electric potential is generated by [H + ] difference across glass membrane http:\\asadipour.kmu.ac.ir 24 slides E ref1 E1E1 E2E2 EjEj E ref2 Boundary potential difference (E b ) = E 1 - E 2 E b = c log[H + ] E b = c – pH

15 iii. pH is determined by formation of boundary potential across glass membrane Boundary potential difference (E b ) = E 1 - E 2 where from Nernst Equation: E b = c – 0.059pH -log a H + (on exterior of probe or in analyte solution) constant Selective binding of cation (H + ) to glass membrane http:\\asadipour.kmu.ac.ir 24 slides E b = c log[H + ]

16 Electrodes and Potentiometry pH Electrodes Glass Membrane  Irregular structure of silicate lattice Cations (Na + ) bind oxygen in SiO 4 structure http:\\asadipour.kmu.ac.ir 24 slides

17 Electrodes and Potentiometry pH Electrodes Glass Membrane  Two surfaces of glass “swell” as they absorb water - Surfaces are in contact with [H + ] http:\\asadipour.kmu.ac.ir 24 slides

18 Electrodes and Potentiometry pH Electrodes Glass Membrane  H + diffuse into glass membrane and replace Na + in hydrated gel region - Ion-exchange equilibrium - Selective for H + because H + is only ion that binds significantly to the hydrated gel layer Charge is slowly carried by migration of Na+ across glass membrane Potential is determined by external [H + ] Constant and b are measured when electrode is calibrated with solution of known pH http:\\asadipour.kmu.ac.ir 24 slides

19 iii. pH is determined by formation of boundary potential across glass membrane At each membrane-solvent interface, a small local potential develops due to the preferential adsorption of H + onto the glass surface. Glass Surface http:\\asadipour.kmu.ac.ir 24 slides

20 Electrodes and Potentiometry Junction Potential 1.)Occurs Whenever Dissimilar Electrolyte Solutions are in Contact  Develops at solution interface (salt bridge)  Small potential (few millivolts)  Junction potential puts a fundamental limitation on the accuracy of direct potentiometric measurements - Don’t know contribution to the measured voltage Again, an electric potential is generated by a separation of charge Different ion mobility results in separation in charge http:\\asadipour.kmu.ac.ir 24 slides

21 iv. Alkali Error ‚ H + not only cation that can bind to glass surface - H + generally has the strongest binding ‚ Get weak binding of Na +, K +, etc ‚ Most significant when [H + ] or a H + is low (high pH) - usually pH $ At low a H + (high pH), amount of Na + or K + binding is significant  increases the “apparent” amount of bound H http:\\asadipour.kmu.ac.ir 24 slides

22 v. Acid Error ‚ Errors at low pH (Acid error) can give readings that are too high ‚ Exact cause not known - usually occurs at pH # 0.5 c) Glass Electrodes for Other Cations i. change composition of glass membrane ‚ putting Al 2 O 3 or B 2 O 3 in glass ‚ enhances binding for ions other than H + ii. Used to make ISE’s for Na +, Li +, NH http:\\asadipour.kmu.ac.ir 24 slides

23 Example 18: The following cell was used for the determination of pCrO 4 : SCE||CrO 4 2- (xM), Ag 2 CrO 4 (sat’d)|Ag Calculate pCrO 4 if the cell potential is http:\\asadipour.kmu.ac.ir 24 slides

24 Titration curve.xlsx http:\\asadipour.kmu.ac.ir 24 slides 24


Download ppt "Potentiometric Methods A.) Introduction : 1.) Potentiometric Methods: based on measurements of the potential of electrochemical cells in the absence of."